THE EFFECT OF ABUTMENT HEIGHT ON SCREW LOOSENING IN SINGLE IMPLANT-SUPPORTED PROSTHESES AFTER DYNAMIC CYCLIC LOADING

  • Kim Nam-Gun;Kim Yung-Soo;Kim Chang-Whe;Jang Kyung-Soo;Lim Young-Jun
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.42 no.6
    • /
    • pp.664-670
    • /
    • 2004
  • Statement of problem. One of the common problems of dental implant prosthesis is the loosening of the screw that connects each component, and this problem is more common in single implant-supported prostheses with external connection. Purpose. The purpose of this study was to examine the changes of detorque values of abutment screws with external connection in different abutment heights. Materials and methods. After cyclic loading on three different abutment heights, detorque values were measured. Abutments were retained with titanium abutment screws tightened to 30 Ncm (30.5 kgmm) with digital torque gauge as recommended by the manufacturer. Replacing abutments, implants and titanium abutment screws with new ones at every measurement, initial detorque values were measured six times. In measuring de torque values after cyclic loading, Avana Cemented Abutments of 4.0 mm collar, 7.0 mm height (Osstem Co., Ltd., Seoul, Korea) were used with three different lengths of 5.0, 8.0, 11.0 mm. Shorter abutments were made by milling of 11.0 mm abutment to have the same force-exercised area of 4.5 mm diameter. Sine curve force (20N-320N, 14Hz) was applied, and detorque values were measured after cyclic loading of 2 million times by loading machine. Detorque values of initial and after-loading were measured by digital torque gauge. One-way ANOVA was employed to see if there was any influence from different abutment heights. Results. The results were as follows: 1. The initial detorque value was 27.8$\pm$0.93 kgmm, and the ratio of the initial detorque value to the tightening torque was 0.91(27.8/30.5). 2. Measured detorque values after cyclic loading were declined as the height of the abutment increased, that was, 5.0 mm; 22.3$\pm$0.82 kgmm, 8.0 mm; 21.8$\pm$0.93 kgmm, and 11.0 mm; 21.3$\pm$0.94 kgmm. 3. One-way ANOVA showed no statistically significant differences among these (p>0.05). 4. Noticeable mobility at the implant-abutment interface was not observed in any case after cyclic loading at all.

Comparative Study of the Early Loading of Resorbable Blasting Media and Sandblasting with Large-grit and Acid-etching Surface Implants: A Retrospective Cohort Study

  • Kim, Sung-Beom;Kim, Young-Kyun;Kim, Su-Gwan;Oh, Ji-Su;Kim, Byung-Hoon
    • Maxillofacial Plastic and Reconstructive Surgery
    • /
    • v.36 no.6
    • /
    • pp.247-252
    • /
    • 2014
  • Purpose: This study compares the prognosis (the survival rate and marginal bone loss) of resorbable blasting media (RBM) surface implants and sandblasting with large-grit and acid-etching (SLA) surface implants in the early loading. Methods: This study targeted 123 patients treated by implants installation from January 2008 to March 2010. The loading was initiated in the maxilla within three to four months and in the mandible within one to two months. The types of restoration were single crown and fixed partial prosthesis. Those functioned over one year. The implants were classified by the surface of implants as Group 1: RBM surface (GS III; OSSTEM, Busan, Korea) and, Group 2: SLA surface (Superline; Dentium, Seoul, Korea). The groups were categorized by maxilla and mandible and compared by survival rate, marginal bone loss through clinical records evaluation, and radiographic measurements. Results: The marginal bone loss in the maxilla was $0.14{\pm}0.34mm$ (Group 1) and $0.30{\pm}0.37mm$ (Group 2), a statistically significant difference (P<0.05). In the mandible those were $0.28{\pm}0.54mm$ (Group 1) and $0.20{\pm}0.33mm$ (Group 2), not significant (P>0.05). There was no significant difference of marginal bone loss between maxilla and mandible by groups. During observation there was no implant failure, a survival rate of 100%. Conclusion: Both surfaces showed an excellent survival rate, and the marginal bone loss was not substantial.

Effect of various abutment systems on the removal torque and the abutment settling in the conical connection implant systems (원추형 연결 임플란트에서 지대주 종류에 따른 나사풀림과 침하현상에 관한 연구)

  • Lee, Jin-Seon;Lee, Joon-Seok
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.50 no.2
    • /
    • pp.92-98
    • /
    • 2012
  • Purpose: The aim of this study was to evaluate the effects of different abutment materials on abutment screw loosening and settling-down effect in conical connection type implant system. Materials and methods: Three types of abutment, cementation, gold UCLA, and metal UCLA abutment were used. Two UCLA groups were fabricated in a similar pattern to cementation abutment. Type III gold alloy and Nickel-Chromium alloy was used for casting gold UCLA abutment and metal UCLA abutment, respectively. Fixture and abutment were tightened to 30 Ncm by using digital torque controller and re-tightening was conducted with same force after 10 minutes. Digital torque gauge was used to measure loosening torque and fixture/abutment length was measured by digital micrometer. Dynamic loads between 25 N and 250 N were applied with $0^{\circ}$ angle to the abutment axis. After loading, fixture/abutment length was re-measured and amount of settlement was calculated. Loosening torque value was also measured for comparison Results: All three groups showed significant differences of length when comparing before and after loading, but there was no significant difference of settling amount in all groups. Loosening torque values were significantly decreased when comparing before and after loading in all groups($P$<.05). However, there was no significant difference in loss of loosening torque values when compared to groups. Conclusion: In internal conical connection type implants, dynamic load affected on settlement and loosening torque of implant, but there was no differences between abutments materials. Likewise gold UCLA abutment, metal UCLA abutment might be able to withstand functional load.

A RETROSPECTIVE MULTICENTER CLINICAL STUDY OF INSTALLED US II / SS II IMPLANTS AFTER MAXILLARY SINUS FLOOR ELEVATION (상악동저 거상술 후 Osstem Implant (US II / SS II)의 다기관 후향적 임상연구)

  • Kook, Min-Suk;Park, Hong-Ju;Kim, Su-Gwan;Kim, Young-Kyun;Cho, Yong-Seok;Choi, Gab-Lim;Oh, Young-Hak;Oh, Hee-Kyun
    • Journal of the Korean Association of Oral and Maxillofacial Surgeons
    • /
    • v.34 no.3
    • /
    • pp.341-349
    • /
    • 2008
  • Purpose: The purpose of this study was to evaluate the $Osstem^{(R)}$ implants (US II/SS II implants) through the retrospective study for the clinical success rate during the installation of the $Osstem^{(R)}$ implants (US II/SS II implants) by using of the procedures of maxillary sinus floor elevation. Materials and methods: The current study was researched in the 6 medical institutions: Chonnam National University, Chosun University, Pusan DaeDong Hospital, Bundang Seoul National University Hospital, Ap-Seon Clinic, and All Clinic. Based on the total number of 116 patients whose treatment was the installation of the US II/SS II implants with the procedures of the maxillary sinus floor elevation, they were conferred on the dental records of the patients under the joint consultation of the 6 medical institutions. On the dental recording charts, there were included in as the following; the name of the institutions, gender, age, with or without smoking or drinking, with or without the generalized diseases, the height of the alveolar bone on the operational sites, elapsed edentulous state period, the state of the opposed or adjacent teeth, the methods of the maxillary sinus floor elevation, secondary time period for surgery, the lengths, types, and diameters of implants, with or without bone transplantation or the types of bone, postoperative current bone height, current adjacent soft tissue state of the implants, with or without the success of the installations of the implants. We have done our survey with the clinical and radiolographical examinations and dental questionaries. The success and survival rate of the implants was evaluated. Results 1. Total number of the patients with the installation of the US II implants were 62. The 252 numbers of US II implants were installed on the 89 maxillary sinuses. The patient's mean age was 54.1 years old and there were 36 men and 27 women. 2. Total number of patients with the installation of SS II implant were 57. The 165 numbers of SS II implants were planted on the 80 maxillary sinuses. Their mean age was 48.7 years old and there were 37 men and 20 women. 3. The follow-up period was 30.7 months(21-49 mon) on average. The vertical bone loss of installed implants after the procedures of the maxillary sinus elevation was 1.1 mm on average in SS II and 1.3 mm on average in US II. There existed no statistical significance on each group. The mean enlarged bone height after the maxillary sinus floor elevation was 8.2 mm. 4. For the procedures of the maxillary sinus elevation, the Lateral approach technique occupied 87.1%, which was the most used one. In addition, the most frequently used transplanted bone was autogenous bone only which was 72.7% during the maxillary sinus floor elevation. 5. The complication of maxillary sinus floor elevation were perforation of sinus membrane, disesthesia on doner site, exposure of cover screw and exposure of maxillar bone. 6. The survival rate of US II and SS II after maxillary sinus floor elevation was 99.2% and 95.8%, respectively. And the success rate of US II and SS II after maxillary sinus floor elevation was 97.6% and 89.7%, respectively. Conclusion : On the evaluation of the analysis of our study, both US II and SS II implants showed the excellent clinical results by use of the procedures of maxillary sinus floor elevation.

Comparative analysis of accuracy of how to use the implant torque controllers (사용 방법에 따른 임플란트 토크 조절기의 정확성 비교 분석 연구)

  • Cho, Lee-Ra;Park, Chan-Jin;Huh, Yoon-Hyuk;Lee, Ji-Hye;Kim, Dae-Gon
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.51 no.4
    • /
    • pp.300-306
    • /
    • 2013
  • Purpose: This study was to evaluate the accuracy of how to use the Mechanical Torque-Limiting Devices. Materials and methods: Three different implant manufacturers (Dentium, Osstem, Shinhung) were prepared for each two implant torque controllers. Divided into two groups depending on the method used for removal torque was measured. Repeated measures of ANOVA test (${\alpha}=.05$) was used as statistics to evaluate the effect of repeated loading number on the removal torque. Independent t-test was used to evaluate the difference in removal torque of two groups. Results: The removal torque significantly decreased as the number of loading repetition increased (P<.05). There was significant difference between two groups. Conclusion: Loosen the handle part of the implant torque controllers spring resilience to recover one group compared to the group that did not, showed a little more closed to the reference value.

Influence of Tungsten Carbide/Carbon Coating on the Preload and Tightening Torque of Implant Abutment Screws (임플랜트 지대주 나사의 텅스텐 카바이드/탄소 코팅이 전하중 및 조임회전각에 미치는 영향)

  • Shin, Hyon-Mo;Cho, Wook;Jeong, Chang-Mo
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.25 no.1
    • /
    • pp.53-59
    • /
    • 2009
  • The purpose of this study was to clarify the clinical efficiency of tungsten carbide/carbon coating on obtaining optimal preload of abutment screw compared with non-coated screw in external-hex implant system. In the present work, rotational value and the compressive force between abutment and fixture of abutment screws without coating and with coating tightened to 30Ncm were measured. Mean compressive force of coated screw was 504.6N. Then uncoated screw was tightened to clamping the abutment and the implant to 504.6N, and the tightening torque value and the rotational value was recorded. The following conclusions were drawn within the limitation of this study. Compressive force of coated screw was higher than that of uncoated screw after tightening at 30Ncm. The tightening torque that was nedeed to clamping the uncoated screw equal to coated screw tightened to 30Ncm was 55.6Ncm. In case of equal compressive force, there was no significant difference in rotation value between coated and uncoated screw.

The Effect of Temporary Cement Cleaning Methods on the Retentive Strength of Cementation Type Implant Prostheses (임시 시멘트 제거방법이 시멘트 유지형 임플란트 보철물의 유지력에 미치는 영향)

  • Shin, Hwang-Kyu;Song, Young-Gyun;Shin, Soo-Yeon
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.27 no.2
    • /
    • pp.125-140
    • /
    • 2011
  • The remnant of temporary cement on the intaglio surface of cast restoration may have a negative effect on the retentive strength of permanent cement. This study was to evaluate the effect of temporary cement cleaning methods on the retentive strength of cementation type implant prostheses. Prefabricated implant abutments - height 5.5mm, diameter 4.5mm, 6 degree axial wall taper with chamfer margins were used. Forty copings-abutment specimens were divided into four groups(each n=10) according to the cleaning methods for temporary cement(Temp-$Bond^{(R)}$) as follows : no temporary cementation(the control group), orange solvent, ultrasonic cleaning, air borne-particle abrasion. After the application of temporary cement and the separation, the cleaning procedure was performed according to the protocol of each group. The specimens were cemented with $Premier^{(R)}$ Implant $Cement^{TM}$. After the permanent cementation, the specimens were subjected to thermocycling and pulled out from the specimens with a universal testing machine at a cross-head speed of 0.5mm/min. After the retentive strength test, all the specimens were cleaned using ultrasonic cleaning, abraded with air borne-particles, and steam-cleaned. Likewise, the specimens were temporarily cemented(Temp-$Bond^{(R)}$ NE), cleaned according to the protocol of each group, cemented with $Premier^{(R)}$ Implant $Cement^{TM}$ and subjected to thermocycling and measurement of their retentive strength. The mean of group with orange solvent were significantly lower than those of other groups(p<0.05). There was no significance between group with ultrasonic cleaning and group with air borne-particle abrasion. Group with ultrasonic cleaning and group with air-particle abrasion were no significance at control group. There was no significance between group cemented with Temp-$Bond^{(R)}$ and group cemented with Temp-$Bond^{(R)}$ NE. Within the limitation of this study, it can be concluded that the temporary cement cleaning method with only orange solvent may have a negative effect on the retentive strength of permanent cement. Ultrasonic cleaning and air borne-particle abrasion methods are recommended for the temporary cement cleaning method on cementation type implant prostheses.

Effects of rhBMP-2 with various carriers on bone regeneration in rat calvarial defect (백서 두개골 결손에서 rhBMP-2와 다양한 carrier의 골재생 유도효과)

  • Lee, Seo-Kyoung;Kim, Ji-Sun;Kang, Eun-Jung;Eum, Tae-Kwan;Kim, Chang-Sung;Cho, Kyoo-Sung;Chai, Jung-Kiu;Kim, Chong-Kwan;Choi, Seong-Ho
    • Journal of Periodontal and Implant Science
    • /
    • v.38 no.2
    • /
    • pp.125-134
    • /
    • 2008
  • Purpose: Bone morphogenetic protein (BMP) is a potent differentiating agent for cells of the osteoblastic lineage. It has been used in the oral cavity under a variety of indications and with different carriers. However, the optimal carrier for each indication is not known. This study evaluated the bone regenerative effect of rhBMP-2 delivered with different carrier systems. Materials and Methods: 8 mm critical-sized rat calvarial defects were used in 60 male Sprague-Dawley rats. The animals were divided into 6 groups containing 10 animals each. Two groups were controls that had no treatment and absorbable collagen membrane only. 4 groups were experimentals that contained rhBMP-2 only and applied with absorbable collagen sponge($Collatape^{(R)}$), $MBCP^{(R)}$, Bio-$Oss^{(R)}$ each. The histological and histometric parameters were used to evaluate the defects after 2- or 8-week healing period. The shape and total augmented area were stable in all groups over the healing time. Results: New bone formation was significantly greater in the rhBMP-2 with carrier group than control group. rhBMP-2/ACS was the highest in bone density but gained less new bone area than rhBMP-2/$MBCP^{(R)}$ and rhBMP-2/Bio-$Oss^{(R)}$. The bone density after 8 weeks was greater than that after 2 weeks in all groups. However, rhBMP-2 alone failed to show the statistically significant difference in new bone area and bone density compared to control group. Also $MBCP^{(R)}$ and Bio-$Oss^{(R)}$ particles remained after 8 weeks healing period. Conclusion: These results suggest that rhBMP-2 with carrier system is an excellent inductive agent for bone formation and we can use it as the predictable bone tissue engieering technique. Future study will likely focus on the kinetics of BMP release and development of carriers that is ideal for it.

Influence of Tightening Torque on Implant-Abutment Screw Joint Stability (조임회전력이 임플랜트-지대주 나사 연결부의 안정성에 미치는 영향)

  • Shin, Hyon-Mo;Jeong, Chang-Mo;Jeon, Yonung-Chan;Yun, Mi-Jeong;Yoon, Ji-Hoon
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.46 no.4
    • /
    • pp.396-408
    • /
    • 2008
  • Statement of problem: Within the elastic limit of the screw, the greater the preload, the tighter and more secure the screw joint. However, additional tensile forces can incur plastic deformation of the abutment screw when functional loads are superimposed on preload stresses, and they can elicit the loosening or fracture of the abutment screw. Therefore, it is necessary to find the optimum preload that will maximize fatigue life and simultaneously offer a reasonable degree of protection against loosening. Another critical factor in addition to the applied torque which can affect the amount of preload is the joint connection type between implant and abutment. Purpose: The purpose of this study was to evaluate the influence of tightening torque on the implant-abutment screw joint stability. Material and methods: Respectively, three different amount of tightening torque (20, 30, and 40 Ncm) were applied to implant systems with three different joint connections, one external butt joint and two internal cones. The initial removal torque value and the postload (cyclic loading up to 100,000 cycles) removal torque value of the abutment screw were measured with digital torque gauge. Then rate of the initial and the postload removal torque loss were calculated for the comparison of the effect of tightening torques and joint connection types between implant and abutment on the joint stability. Results and conclusion: 1. Increase in tightening torque value resulted in significant increase in initial and postload removal torque value in all implant systems (P < .05). 2. Initial removal torque loss rates in SS II system were not significantly different when three different tightening torque values were applied (P > .05), however GS II and US II systems exhibited significantly lower loss rates with 40 Ncm torque value than with 20 Ncm (P < .05). 3. In all implant systems, postload removal torque loss rates were lowest when the torque value of 30 Ncm was applied (P < .05). 4. Postload removal torque loss rates tended to increase in order of SS II, GS II and US II system. 5. There was no correlation between initial removal torque value and postload removal torque loss rate (P > .05).

The convergence study on the use of dental impression materials (치과용 인상재의 사용 실태에 대한 융합 연구)

  • Jang, Sun-Ok;Choi, Eun-Mi;Kang, Min-Kyung;Shim, Youn-Soo;Cho, Yoon-Jeong;Oh, Sang-Hwan;Choi, Yu-Ri
    • Journal of Digital Convergence
    • /
    • v.16 no.4
    • /
    • pp.195-203
    • /
    • 2018
  • This study was conducted to investigate the types and methods of dental impression materials used in dental clinics in south Korea, the questionnaire was surveyed by a self-administered method for one dentist in a one dental clinic, the collected data were statistically analyzed using SPSS 20. The most frequently used addition silicone impression materials were used in most dental treatment. It was confirmed that the method of taking impression materials having different flow properties(light body and heavy body) was most preferred at one time, also, the implant impression taking method was the most preferred for two rubber impression materials with different flow properties, both direct and indirect. There were statistically significant differences in age and clinical experience between the methods of impression taking of casting restorations and prosthetic according to general characteristics. These results are expected to be helpful for the development of theoretical and practical work based on the basic data for impression material development and research and the preference of impression materials required for dental professional in clinical practice.