Influence of Tungsten Carbide/Carbon Coating on the Preload and Tightening Torque of Implant Abutment Screws

임플랜트 지대주 나사의 텅스텐 카바이드/탄소 코팅이 전하중 및 조임회전각에 미치는 영향

  • Shin, Hyon-Mo (Department of Dentistry, Graduate School, Pusan National University) ;
  • Cho, Wook (Department of Dentistry, Graduate School, Pusan National University) ;
  • Jeong, Chang-Mo (Department of Dentistry, Graduate School, Pusan National University)
  • 신현모 (부산대학교 치과대학 치과보철학교실) ;
  • 조욱 (부산대학교 치과대학 치과보철학교실) ;
  • 정창모 (부산대학교 치과대학 치과보철학교실)
  • Received : 2009.01.03
  • Accepted : 2009.03.25
  • Published : 2009.03.31

Abstract

The purpose of this study was to clarify the clinical efficiency of tungsten carbide/carbon coating on obtaining optimal preload of abutment screw compared with non-coated screw in external-hex implant system. In the present work, rotational value and the compressive force between abutment and fixture of abutment screws without coating and with coating tightened to 30Ncm were measured. Mean compressive force of coated screw was 504.6N. Then uncoated screw was tightened to clamping the abutment and the implant to 504.6N, and the tightening torque value and the rotational value was recorded. The following conclusions were drawn within the limitation of this study. Compressive force of coated screw was higher than that of uncoated screw after tightening at 30Ncm. The tightening torque that was nedeed to clamping the uncoated screw equal to coated screw tightened to 30Ncm was 55.6Ncm. In case of equal compressive force, there was no significant difference in rotation value between coated and uncoated screw.

본 연구는 external butt joint 연결 형태를 가진 임플랜트 시스템에서 티타늄 합금 나사와 텅스텐 카바이드와 탄소로 코팅한 나사의 압축력 및 조임 회전각 측정을 통해 동일한 전하중을 얻는 데 있어 코팅나사 사용이 갖는 임상적 효율성을 간접적으로 비교해 보고자 하였다. 본 연구에서는 Osstem Implant의 US II 시스템 고정체(${\phi}=4mm$)와 Cemented abutment를 사용하여 티타늄 합금(Ta) 지대주 나사 및 텅스텐 카바이드와 탄소로 표면 처리한(WC/CTa) 지대주 나사에 30Ncm의 조임 회전력을 적용한 후 압축력 및 조임 회전각을 측정하였다. 다음 WC/CTa 지대주 나사에서 측정된 압축력의 평균값을 계산하고 이를 Ta 지대주 나사에 적용하여 동일한 압축력이 가해질 때의 조임 회전력 값을 측정하였다. 그리고 실험결과 측정된 조임 회전력의 평균값을 Ta 나사에 적용한 후 조임 회전각을 측정하여 비교해 본 결과 다음과 같은 결론을 얻었다. 30Ncm의 조임 회전력에서 텅스텐 카바이드/탄소 코팅 나사가 티타늄 합금 나사보다 압축력과 회전각이 더 높게 나타났다. 텅스텐 카바이드/탄소 코팅 나사에 30Ncm의 조임 회전력을 적용하여 발생한 압축력을 얻기 위해서는 티타늄 합금 나사의 경우 55.6Ncm의 조임 회전력이 필요하였으며, 회전각에는 두 나사 간에 차이가 없었다.

Keywords

Acknowledgement

Supported by : 부산대학교병원

References

  1. Branemark P-I, Hansson BO, Adell R, Breine U, Lindstrom J, Hallen O, Ohman A. Osseointegrated implants in the treatment of the edentulous jaw: Experience from a ten year period. Scand J Plastic Reconst Surg 1977;11(suppl)
  2. Adell R, Lekholm U, Rockler B, Branemark P-I. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg 1981;10:387-416 https://doi.org/10.1016/S0300-9785(81)80077-4
  3. Albrektsson T. A multicenter report on osseointegrated oral implants. J Prosthet Dent 1988;60:75-84 https://doi.org/10.1016/0022-3913(88)90355-1
  4. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Tronto study. Part I: Surgical results. J Prosthet Dent 1990;63:451-457 https://doi.org/10.1016/0022-3913(90)90237-7
  5. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Tronto study. Part II: The prosthetic results. J Prosthet Dent 1989;64:53-61
  6. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Tronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-194 https://doi.org/10.1016/0022-3913(90)90177-E
  7. Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994;3:17-21 https://doi.org/10.1097/00008505-199404000-00002
  8. Jemt T. Modified single and short-span restorations supported by osseointegrated fixtures in the partially edentulous jaw. J Prosthet Dent 1986;55:243-7 https://doi.org/10.1016/0022-3913(86)90352-5
  9. Jemt T, Lekholm U, Grondahl K. 3-year followup study of early single implantb restorations ad modum Branemark. Int J Periodontics Restorative Dent 1990;10:340-9
  10. Rangert B, Jemt T, Jrnus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989:4:241-247
  11. Rangert B, Sullivan R. Learning from history The transition from full arch to posterior partial restorations. Nobelpharma News 1995;9(2):67
  12. Misch CE, Wang HL, Misch CM, Sharawy M, Lemons J, Judy KW. Rationale for the application of immediate load in implant dentistry: Part I. Implant Dent 2004;13:207-17 https://doi.org/10.1097/01.id.0000140461.25451.31
  13. Shigley J. Mechanical Engineering Design,ed 3. New York: McGraw Hill, 1987:227-52
  14. Robb TT, Porter SS. Increasing abutment screw rotation by applying a thingold coating [abstract 1641]. J Dent Res 1998;77(special issue):837
  15. Robb TT, Porter SS. Increasing implantabutment preload by thin gold coating abutment screws [abstract 1642]. J Dent Res 1998;77(special issue):837
  16. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32 https://doi.org/10.1067/mpr.2001.116230
  17. Vigolo P. Cemented versus screw-retained implant- supported single tooth crowns: A 4-year prospective clinical study. Int. J Oral Maxillofac Implants 2004; 19:260-5
  18. Liu Y, Gubisch M, Hild W, Scherge M, Spiess L, Knedlik Ch, Schaefer JA. Nanoscale multilayer WC/C coatings developed for nanopositioning. part II: Friction and wear. Thin Solid Films 2005;488: 140-8 https://doi.org/10.1016/j.tsf.2005.03.020
  19. Choi JU, Jeong CM, Jeon YC, Lim JS, Jeong HC, Eom TG. Influence of tungsten carbide/carbon http://searchc.naver.com/ntk/coating on the preload of implant abutment screws. J Korean Acad Prosthodon 2006;44:229-42
  20. Park JK, Jeong CM, Jeon YC, Yoon JH. Influence of tungsten carbide/carbon coating of implant-abutment screw on screw loosening. J Korean Acad Prosthodon 2008;46:137-47
  21. McGlumphy EA, Mendel DA, Holloway JA. Implant Screw Mechanics. Dent Clinics North Am 1998; 42:71-89
  22. Schwarz MS. Mechanical complications of dental implants. Clin Oral Impl Res 2000;11:156-8 https://doi.org/10.1034/j.1600-0501.2000.011S1156.x
  23. Motosh N. Development of design charts for bolts preloaded up to the plastic range. J Eng Ind 1976;98:849-51