• Title/Summary/Keyword: 광자극형광나노닷선량계

Search Result 4, Processing Time 0.019 seconds

Calibration of Optically Stimulated Luminescent nanoDot Dosimeter for 6 MV Photon Beam (6 MV 광자 빔에 대한 광자극형광나노닷선량계의 교정)

  • Kim, Jongeon;Kim, Seonghu;Lee, Hyoyeong
    • Journal of the Korean Society of Radiology
    • /
    • v.7 no.1
    • /
    • pp.93-98
    • /
    • 2013
  • The purpose of this study is to investigate the calibration of an optically stimulated luminescent nanoDot dosimeter(OSLnD) to 6 MV photon beam. Dose ranges of the calibration of linear and non-linear from the analysis of dose response of the OSLnD were decided. To evaluate the accuracy of calibration equation and the calibration, the sets of the calibration and quality control dosimeter were used to make. The calibrations were performed by the linear and the non-linear in the dose range of 0~300 cGy and 20~1300 cGy, respectively. The errors of the calibration were acquired less than 0.1% respectively from the measurement of the quality control dosimeters for the calibration of linear and the non-linear. This study provides the calibration equation of the OSLnD to the 6 MV photon beam.

Determination of the Effective Energy of X-Ray Beam Using Optically Stimulated Luminescent nanoDot Dosimeters (광자극형광나노닷선량계를 사용한 X선 빔의 유효에너지 결정)

  • Kim, Jongeon;Lee, Sanghun
    • Journal of the Korean Society of Radiology
    • /
    • v.9 no.6
    • /
    • pp.375-379
    • /
    • 2015
  • The purpose of this study is to determine the effective energy of a polyenegetic X-ray beam. The half value layer(HVL) of aluminum for 80 kVp X-ray beam was measured by using optically stimulated luminescent nanoDot dosimeters(OSLnDs). The linear attenuation coefficient(${\mu}$) was calculated using the measured HVL. And the mass attenuation coefficient(${\mu}/{\rho}$) was obtained by dividing the linear attenuation coefficient by the density(${\rho}$) of aluminum. The effective energy($E_{eff}$) of the obtained mass attenuation coefficient was determined using data of the X-ray mass attenuation coefficients for photon energies of aluminum given by National Institute of Standards and Technology(NIST). As a result, the HVL value is 2.262 mmAl. The ${\mu}$ value is $3.06cm^{-1}$. The ${\mu}/{\rho}$ value is $1.114cm^2/g$. And the $E_{eff}$ value was determined at 29.79 keV.

Measurement of Dose outside a 6 MV Field Edge Using Optically Stimulated Luminescent Nano Dot Dosimeters (광자극형광나노닷선량계를 사용한 6 MV 조사야 가장자리 바깥 선량 측정)

  • Kim, Jongeon;Kim, Wontae
    • Journal of the Korean Society of Radiology
    • /
    • v.8 no.7
    • /
    • pp.449-454
    • /
    • 2014
  • The purpose of this study is(was) to investigate the shielding ratio of 1 mmPb and the off axis ratio outside the field edge at depth of 1 cm from a phantom surface for 6 MV photon beam. A dose of 180 cGy was delivered to a depth of 10 cm for a $10{\times}10cm^2$ and $15{\times}15cm^2$ field in the SAD technique. The off axis ratio was calculated by measuring the dose of optically stimulated luminescent nanoDot dosimeters(OSLnDs) positioned at 2, 4 and 6 cm from the field edge, and the center axis of field. And the shielding ratio of 1 mmPb was calculated by measuring the dose of OSLnDs positioned at 2, 4 and 6 cm from the field edge.. As a result, for a $10{\times}10cm^2$ and $15{\times}15cm^2$ field, the off axis ratios were acquired 0.008-0.023 and 0.011-0.028, respectively. Also the shielding ratios of 1 mmPb were acquired 0.868-0.888 and 0.807-0.842, respectively. These results provide data to protect organs at risk outside the radiation treatment field.

A Study on the Probability of Secondary Carcinogenesis during Gamma Knife Radiosurgery (감마나이프 방사선 수술시 2차 발암 확률에 관한 연구)

  • Joo-Ah, Lee;Gi-Hong, Kim
    • Journal of the Korean Society of Radiology
    • /
    • v.16 no.7
    • /
    • pp.843-849
    • /
    • 2022
  • In this study, the probability of secondary carcinogenesis was analyzed by measuring the exposure dose of surrounding normal organs during radiosurgery using a gamma knife. A pediatric phantom (Model 706-G, CIRS, USA) composed of human tissue-equivalent material was set to four tumor volumes of 0.25 cm3, 0.51 cm3, 1.01 cm3, and 2.03 cm3, and the average dose was 18.4 ± 3.4 Gy. After installing the Rando phantom on the table of the gamma knife surgical equipment, the OSLD nanoDot dosimeters were placed in the right eye, left eye, thyroid, thymus gland, right lung, and left lung to measure each exposure dose. The probability of cancer occurrence due to radiation exposure of surrounding normal organs during gamma knife radiosurgery for acoustic schwannoma disease was 4.08 cancers per 100,000 at a tumor volume of 2.03 cm3. This study is expected to be used as useful data in relation to stochastic effects in the future by studying the risk of secondary radiation exposure that can occur during stereotactic radiosurgery.