Calibration of Optically Stimulated Luminescent nanoDot Dosimeter for 6 MV Photon Beam  

Kim, Jongeon (Department of Radiological Science, Kaya University)
Kim, Seonghu (Department of Radiology, Gyeongsang National University Hospital)
Lee, Hyoyeong (Department of Nuclear Medicine, Pusan National University Yangsan Hospital)
Publication Information
Journal of the Korean Society of Radiology / v.7, no.1, 2013 , pp. 93-98 More about this Journal
Abstract
The purpose of this study is to investigate the calibration of an optically stimulated luminescent nanoDot dosimeter(OSLnD) to 6 MV photon beam. Dose ranges of the calibration of linear and non-linear from the analysis of dose response of the OSLnD were decided. To evaluate the accuracy of calibration equation and the calibration, the sets of the calibration and quality control dosimeter were used to make. The calibrations were performed by the linear and the non-linear in the dose range of 0~300 cGy and 20~1300 cGy, respectively. The errors of the calibration were acquired less than 0.1% respectively from the measurement of the quality control dosimeters for the calibration of linear and the non-linear. This study provides the calibration equation of the OSLnD to the 6 MV photon beam.
Keywords
Optically stimulated luminescent nanoDot dosimeter; Dose response; Calibration;
Citations & Related Records
연도 인용수 순위
  • Reference
1 P.A. Jursinic, "Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements", Med. Phys, Vol. 34, No. 12, pp.4594-4604, 2007.   DOI   ScienceOn
2 D.W. Kim, W.K. Chung, D.O. Shin, M.G. Yoon and U.J. Hwang,"Dose response of commercially available optically stimulated luminescent detector, Al2O3:C for megavoltage photons and electrons", Radiation Protection Dosimetry, pp.1-8, 2011.
3 M.S. Akselrod, A.C. Lucas, J.C. Polf and S.W. Mckeever, "Optically stimulated luminescence of Al2O3:C", Radiat. Meas, Vol. 29, pp.391-399, 1998.   DOI   ScienceOn
4 microStar user manual, Landauer Inc., 2009.
5 Frequently asked questions about landaue's optically stimulated luminescence(OSL) detectors, Landauer Inc
6 A. Viamonte, L.A.R. Darosa, L.A. Buckley, A. Cherpak and J.E. Cygler, "Radiotherapy dosimetry using a commercial OSL system", Med. Phys, Vol. 35, No. 4, pp.1261-1266, 2008.   DOI   ScienceOn
7 V. Schembri and B.J.M. Heijmen, "Optically stimulated luminescence(OSL) of carbon-doped aluminum oxide ( Al2O3:C) for film dosimetry in radiotherapy", Med. Phys, Vol. 34, No. 6, pp.2113-2118, 2007.   DOI   ScienceOn
8 M.S. Yukihara, V.H. Whitley, S.W.S. Mckeever, A.E. Akselrod and M.S. Akselrod, "Effects of high-dose irradiation on the optically stimulaetd luminescence of Al2O3:C", Radiat. Meas, Vol. 38, pp.317-330, 2004.   DOI   ScienceOn
9 E.G.Yukihara, G. Mardirossian, M. Mirzasadeghi, S. Guduru and S. Ahmad, Evaluation of Al2O3:C optically stimulated luminescence(OSL) dosimeters for passive dosimetry of high-energy photon and electron beams in radiotherapy, Med. Phys, Vol. 35, No. 1, pp.260-269, 2008.   DOI   ScienceOn
10 P.A. Jursinic, "Changes in optically stimulated luminescent dosimeter(OSLD) dosimetric characteristics with accumulated dose", Med. Phys, Vol. 27, No. 1, pp.132-140, 2010.