Measurement of Dose outside a 6 MV Field Edge Using Optically Stimulated Luminescent Nano Dot Dosimeters

광자극형광나노닷선량계를 사용한 6 MV 조사야 가장자리 바깥 선량 측정

  • Kim, Jongeon (Department of Radiological Science, Kaya University) ;
  • Kim, Wontae (Department of Radiological Science, Kaya University)
  • 김종언 (가야대학교 방사선학과) ;
  • 김원태 (가야대학교 방사선학과)
  • Received : 2014.11.03
  • Accepted : 2014.12.25
  • Published : 2014.12.30

Abstract

The purpose of this study is(was) to investigate the shielding ratio of 1 mmPb and the off axis ratio outside the field edge at depth of 1 cm from a phantom surface for 6 MV photon beam. A dose of 180 cGy was delivered to a depth of 10 cm for a $10{\times}10cm^2$ and $15{\times}15cm^2$ field in the SAD technique. The off axis ratio was calculated by measuring the dose of optically stimulated luminescent nanoDot dosimeters(OSLnDs) positioned at 2, 4 and 6 cm from the field edge, and the center axis of field. And the shielding ratio of 1 mmPb was calculated by measuring the dose of OSLnDs positioned at 2, 4 and 6 cm from the field edge.. As a result, for a $10{\times}10cm^2$ and $15{\times}15cm^2$ field, the off axis ratios were acquired 0.008-0.023 and 0.011-0.028, respectively. Also the shielding ratios of 1 mmPb were acquired 0.868-0.888 and 0.807-0.842, respectively. These results provide data to protect organs at risk outside the radiation treatment field.

이 연구의 목적은 6 MV 광자 빔에 대하여 팬텀 표면으로부터 1 cm 깊이에서 조사야 가장자리 바깥 축외선량비와 1 mmPb의 차폐비를 조사하는 데 있다. 180 cGy의 선량은 SAD기법에서 $10{\times}10cm^2$$15{\times}15cm^2$ 조사야에 대하여 깊이 10 cm에 전달되었다. 축외선량비는 조사야의 중심축과 가장자리로부터 2, 4, 6 cm에 위치된 광자극형광나노닷선량계(OSLnD)들의 선량을 측정하여 계산하였다. 그리고 1 mmPb의 차폐비는 조사야 가장자리로부터 2, 4, 6 cm에 위치된 OSLnD들의 선량을 측정하여 산출하였다. 결과로서, $10{\times}10cm^2$$15{\times}15cm^2$ 조사야에 대하여, 축외선량비들은 0.008-0.023과 0.011-0.028을 각각 얻었다. 또한 1 mmPb의 차폐비들은 0.868-0.888과 0.807-0.842을 각각 얻었다. 이 결과들은 방사선치료 조사야 바깥에 위치한 위험장기들을 보호하기 위한 자료를 제공한다.

Keywords

References

  1. J.E. Kim, I.C. Im and H.Y. Lee, "Correction factor for the energy dependence of a optically stimulated luminescence dosimeter in diagnostic radiography", Journal of Korean Society of Radiology, Vol. 5, No. 5, pp. 261-265, 2011. https://doi.org/10.7742/jksr.2011.5.5.261
  2. C.S. Reft, "The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovotage photon, megavoltage photon, and electron, proton, and carbon beams", Med. Phys., Vol. 36, No. 5, pp. 1690-1699, 2009. https://doi.org/10.1118/1.3097283
  3. P.A. Jursinic, "Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements", Med. Phys., Vol. 34, No. 12, pp. 4594-4604, 2007. https://doi.org/10.1118/1.2804555
  4. P.H.G. Rosado, M.S. Nogueira, P.L. Squair and P.M.C Oliveira, "Determination of the mean energy for attenuated and unattenuated IEC diagnostic X-ray beams", Inetrnational Nuclear Atlantic Conference, Santos, 2007.
  5. P.H.G. Rosado, M.S. Nogueira, F. genezini and E.C. Vilela, "Measurement of conversion coefficients between free in air kerma and personal dose equivalent for diagnostic X-ray beams", Radiation Measurements, Vol. 43, pp. 968-971, 2008. https://doi.org/10.1016/j.radmeas.2007.11.082
  6. Z. Knezevic, L. Stolarczyk, I. Bessieres, J.M. Bordy, S. Miljanic and P. Olko, "Photon dosimetry methods outside the target volume in radiation therapy: Optically stimulated luminescence (OSL), thermoluminescence (TL) and radiophotoluminescence (RPL) dosimetry", Radiation Measurements, Vol. 57, pp. 9-18, 2013. https://doi.org/10.1016/j.radmeas.2013.03.004
  7. C.R. Edwards and P. J. Mountford, "Near surface photon energy spectra outside a 6 MV field edge", Phys. Med. Biol., Vol. 49, P.293-301, 2004. https://doi.org/10.1088/0031-9155/49/18/N01
  8. E.O. Agyingi, P.N. Mobit and D.A. Sandison, "Energy response of an aluminum oxide detector in kilovoltage and Megavoltage photon beams: an EGSnrc monte carlo simulation study", Radiation Protection Dosimetry, Vol. 118, No. 1, pp. 28-31, 2005. https://doi.org/10.1093/rpd/nci329
  9. P. Mobit, "Comparison of the ebergy-response factor of LiF and $Al_2O_3$ in radiotherapy beams", Radiation Protection Dosimetry, Vol. 119, No. 1-4, pp. 497-499, 2006. https://doi.org/10.1093/rpd/nci676
  10. J.E. Kim, S.H. Kim and H.Y. Lee, "Calibration of Optically Stimulated Luminescent nanoDot Dosimeter for 6 MV Photon Beam" Journal of Korean Society of Radiology, Vol. 7, No. 1, pp.93-98, 2013. https://doi.org/10.7742/jksr.2013.7.1.093
  11. M.F. Chan, Y. Song, L.T. Dauer, J.d. Li, D. Huang and C. Burman, "Estimating dose to implantable cardioverter-defibrillator outside the treatment fields using a skin QED diode, optically stimulated luminescent dosimeters, and LiF thermoluminescent dosimeters", Medical Dosimetry, Vol. 37, No. 98, pp. 334-338, 2012. https://doi.org/10.1016/j.meddos.2011.11.007
  12. A.N. Solan, M.J. Solan, G. Benbarz and M.B. Goodkin, "Treatment of patients with cardiac pacemakers and implantable cardioverter defibrillators during radiotheraphy", Phys. Med. Biol., Vol. 59, No 3, pp. 897-904, 2004.