• 제목/요약/키워드: Implant-abutment

검색결과 505건 처리시간 0.025초

The reliability of Anycheck device related to healing abutment diameter

  • Lee, Dong-Hoon;Shin, Yo-Han;Park, Jin-Hong;Shim, Ji-Suk;Shin, Sang-Wan;Lee, Jeong-Yol
    • The Journal of Advanced Prosthodontics
    • /
    • 제12권2호
    • /
    • pp.83-88
    • /
    • 2020
  • PURPOSE. The purpose of this in vitro study was to examine the reliability of the Anycheck device and the effect of the healing abutment diameter on the Anycheck values (implant stability test, IST). MATERIALS AND METHODS. Thirty implants were placed into three artificial bone blocks with 10 Ncm, 15 Ncm, and 35 Ncm insertion torque value (ITV), respectively (n = 10). (1) The implant stability was measured with three different kinds of devices (Periotest M, Osstell ISQ Mentor, and Anycheck). (2) Five different diameters (4.0, 4.5, 4.8, 5.5, and 6.0 mm) of healing abutments of the same height were connected to the implants and the implant stability was measured four times in different directions with Anycheck. The measured mean values were statistically analyzed. RESULTS. The correlation coefficient between the mean implant stability quotient (ISQ) and IST value was 0.981 (P<.01) and the correlation coefficient between the meant periotest value (PTV) and IST value was -0.931 (P<.01). There were no statistically significant differences among the IST values with different healing abutment diameters. CONCLUSION. There was a strong correlation between the Periotest M and Anycheck values and between the ISQ and IST. The diameter of the healing abutment had no effect on the Anycheck values.

수종 임플랜트 지대주나사의 반복하중 후 나사풀림에 관한 연구 (A STUDY OF SCREW LOOSENING AFTER DYNAMIC CONTINOUS FATIGUE TEST OF SEVERAL ABUTMENT SCREW)

  • 김진만;한중석;이선형;양재호;이재봉;김영수
    • 대한치과보철학회지
    • /
    • 제41권4호
    • /
    • pp.519-531
    • /
    • 2003
  • Statement of problem : Chronic implant screw loosening remains a problem in restorative practices. Some implant manufactureres have introduced abutment screws with treated material, surfaces and macrostructures in an effort to reduce potential loosening. Purpose : This study evaluated the materials and loading cycles on detorque value after dynamic continous fatigue test in the sinulated conditions of posterior single restoration. Material and method : Fourteen of each of the following abutment screws - titanium alloy, gold alloy, gold-tite, and titanium alloy modified - were used in test. SEM is used to verify macrostructures of each screws. $ZrO_2/Al_2O_3$ composite abutment was tightened on $4{\times}10.0mm$ titanium external implant at 30 Ncm. Cyclic loading machine delivered dynamic loading forces between 20 and 320N for 100,000, 200,000, 300,000, 500,000, and 1,000,000 cycles at frequencies 14Hz. Torque and detorque value after loading was measured. Results : All measued screws had different screw length and thread form. Titanium modified screw had greater detorque value than others before and after cyclic loadings(p<0.05). All abutment screws had no significant change in mean percentage of detorque value after loading to initial value after less than 500.000 cyclic loadings, but significant lower value after 1,000,000 cycles(p<0.05). Conclusion : Within limintations of this study all abutment screws may be loosend after about 1 year use. Annual check-up is nessasary to prevent screw loosening.

Implant에 사용되는 CAD/CAM 전용 zirconia abutment의 표면적과 디자인에 따른 유지력 차이 (Differences in retention of the reduction direction depending on dimensions and design CAD/CAM zirconia abutment for implant)

  • 남택모;김한곤;김병식;임시덕
    • 대한치과기공학회지
    • /
    • 제32권4호
    • /
    • pp.317-325
    • /
    • 2010
  • Purpose: The aim of this study is to make some basic materials to find retention force difference based on the total height of CAD/CAM zirconia abutment used for implant, retention force difference based on how to regulate the height of the abutment, retention force difference based on the size and retention force difference based on the design group. Methods: The retention force was measured by being pulled at the speed of 1mm/min after being combined with zirconia block and abutment using Temp-BOND of Kerr. The experiment was done at the research lab of OSTEM in Busan by useing UNIVERSAL TESTING MACHINE on March 3rd, 2010. Results: After analysing the total height and the retention force, p-value had minor difference by 0.01 statistically. Namely, 3mm, 4mm, 5mm had the retention force difference and we could see retention force difference on 3mm and 5mm at the post test. After analyzing how to regulate the height and retention force, p-value had minor difference by 0.000 statistically. Namely, 1mm and 2mm had the retention difference and we could see that 1mm and 2mm with the total height had retention difference. After analyzing the retention force based on the size, p-value had minor retention force difference by 0.000 statistically. Namely, 7 different size had retention force difference and we could see the size 21.9mm, 32.9mm, 32.9mm, 38.4mm, 48.9mm and 54.9mm had retention force difference. Conclusion: Namely 9 different design group had retention difference and we could see that 9 design group with 5.6.7.8 design group and 9 design group with 1.2.3.4. design group had retention force.

A two-short-implant-supported molar restoration in atrophic posterior maxilla: A finite element analysis

  • Song, Ho-Yong;Huh, Yoon-Hyuk;Park, Chan-Jin;Cho, Lee-Ra
    • The Journal of Advanced Prosthodontics
    • /
    • 제8권4호
    • /
    • pp.304-312
    • /
    • 2016
  • PURPOSE. The aim of this study was to investigate the stress distribution of 2-short implants (2SIs) installed in a severely atrophic maxillary molar site. MATERIALS AND METHODS. Three different diameters of internal connection implants were modeled: narrow platform (NP), regular platform (RP), and wide platform (WP). The maxillary first molars were restored with one implant or two short implants. Three 2SI models (NP-oblique, NP-vertical, and NP-horizontal) and four single implant models (RP and WP in a centered or cantilevered position) were used. Axial and oblique loadings were applied on the occlusal surface of the crown. The von Mises stress values were measured at the bone-implant, peri-implant bone, and implant/abutment complex. RESULTS. The highest stress distribution at the bone-implant interface and the peri-implant bone was noticed in the RP group, and the lowest stress distribution was observed in the 2SI groups. Cantilevered position showed unfavorable stress distribution with axial loading. 2SI types did not affect the stress distribution in oblique loading. The number and installation positions of the implant, rather than the bone level, influenced the stress distribution of 2SIs. The implant/abutment complex of WP presented the highest stress concentration while that of 2SIs showed the lowest stress concentration. CONCLUSION. 2SIs may be useful for achieving stable stress distribution on the surrounding bone and implant-abutment complex in the atrophic posterior maxilla.

나사풀림과 나사파절이 동반된 실패한 임플란트 고정성 보철물의 회복증례 (Reconstruction of failed fixed implant prosthesis accompanied by abutment screw loosening and fracture : A case report)

  • 배은빈;신영근;전영찬;정창모;윤미정;이소현;허중보
    • 대한치과의사협회지
    • /
    • 제55권7호
    • /
    • pp.457-466
    • /
    • 2017
  • In the case of failed fixed implant prosthesis accompanied by abutment screw fracture, fractured screw fragment must be removed to use the existing implant fixtures. A 61-year-old male patient, who had a failed maxillary fixed implant prosthesis accompanied by three abutment screw fracture, hoped to reconstruct the maxillary implant prosthesis, while maintaining the existing implant fixtures. To use the existing implant fixtures, fractured screw fragments were removed. A maxillary implant overdenture using available existing implants was planned. Bar-attachment with Locator was used for implant splinting, denture stability, and retention. Final impression was taken after treatment of peri-implantitis. Jaw relation registration was taken to evaluate available interarch space for bar-attachment. After fabricating bar-attachment, centric relation was taken. Implant overdenture using bar-attachment with Locator was delivered after wax-denture evaluation. This case report showed that a satisfactory clinical result was achieved by implant overdenture using existing implant fixtures in a maxillary edentulous patient.

  • PDF

지르코늄 임플란트 지대주의 나사결합부 안정성에 관한 연구 (Screw Joint Stability under Cyclic Loading of Zirconia Implant Abutments)

  • 이미순;서규원;류재준
    • 대한치과보철학회지
    • /
    • 제47권2호
    • /
    • pp.164-173
    • /
    • 2009
  • 연구 목적: 심미적 장점을 지닌 지르코니아 지대주와 금속 지대주 사이에 반복 하중 전, 후의 풀림 회전력을 비교하여 지대주 재질에 따른 나사 안정성에 차이가 있는지를 비교 분석해 보고자 하는 것이다. 연구 재료 및 방법: 실험군은 크게 외측 연결 구조(US II, Osstem Implant, Korea)와 내측 연결 구조(GS II, Osstem Implant, Korea)의 임플란트 시스템을 두 군으로 나누었다. 각각의 군에서 지르코니아 지대주군와 금속 지대주군으로 나누어 실험하였다. 각군당5개씩의 시편을 제작하였다. 임플란트 고정체를 별도 제작된 지그에 고정시킨다. 디지털 토크게이지를 이용하여 30 Ncm의 조임 회전력을 가하고 10분 후 동일 조임 회전력을 한번 더 적용하였다. 다음 각 지대주 나사의 초기 풀림 회전력을 디지털 토크게이지로 측정하였다. 동일 시편을 다시 30 Ncm의 조임 회전력을 가하여 체결한 후, 유압식 동적 재료시험기(Instron, USA)에 임플란트 매식체를 치과용 임플란트 피로시험에 관한 ISO/FPIS 14801:2003(E)규정을 참고하여 고정시켰다. 최소 하중은 10 N, 최대 하중은 250 N의 sine형 반복 하중을 $30^{\circ}$의 경사각도로, 하중 주기는 14 Hz로, 100만 회의 반복 하중을 적용한 후 풀림 회전력을 측정하였다. 결과: 1. 모든 시스템에서 지대주 나사의 풀림 회전력이 조임 회전력 보다 감소하였으나, 나사 풀림 현상은 나타나지 않았다. 2. 반복 하중 전과 후의 나사 풀림 회전력은 두 임플란트 시스템 모두에서 지르코니아 지대주가 금속 지대주보다 더 컸다(P<.05). 3. 반복 하중에 따른 풀림 회전력 상실률은 두 임플란트 시스템 모두에서 지르코니아 지대주와 금속 지대주가 차이가 없는 것으로 나타났다(P>.05). 4. 금속 지대주에서 반복 하중에 따른 풀림 회전력 상실률은 내측 연결 형태의 GS II 시스템이 외측 연결형태의 US II 시스템보다 작았다(P<.05). 5. 지르코니아 지대주에서는 반복 하중에 따른 풀림 회전력 상실률은 두 임플란트 시스템 간에 차이가 없는 것으로 나타났다(P>.05). 결론: 본 실험을 통해 100만회 반복 하중 하에서 지르코니아 지대주가 금속 지대주에 비해 나사 결합부 안정성이 더 크다는 결론을 내릴 수 있다.

지대주 나사 풀림 방지를 위한 새로운 Lock Screw 시스템의 효과에 대한 유한요소해석적 연구 (Finite element analysis of the effect of novel Lock Screw system preventing abutment screw loosening)

  • 임은섭;김종은;김지환;박영범
    • 구강회복응용과학지
    • /
    • 제35권3호
    • /
    • pp.132-142
    • /
    • 2019
  • 목적: 본 유한요소해석적 연구는 새로운 Lock screw 시스템을 소개하고 그 역학적 특성을 분석하여 Lock screw가 지대주 나사 풀림을 방지할 수 있는지 알아보려고 한다. 연구 재료 및 방법: Lock screw는 임플란트 지대주 내부에 체결하여 지대주 나사 두부에 압축력을 가하는 장치이다. 그 효과를 알아보기 위해 CAD 프로그램을 이용하여 모델링을 하였고, 이를 다양한 하중조건 하에서 유한요소해석법을 이용해 분석하였다. 우선 지대주 나사의 조임회전력에 따른 전하중을 측정하고 이론 값과 비교하여 해석모델을 검증하였다. 검증된 해석 모델을 외부 하중이 없는 것과 178 N을 부여한 것으로 나누어 Lock screw의 조임회전력을 각각 10, 20, 30 Ncm으로 변화시켰고, 이때 임플란트 구성요소의 응력 분포가 어떠한 특성을 보이는지 살펴보았다. 결과: 여러 하중조건 하에서 Lock screw의 사용은 임플란트 구성요소의 항복강도를 넘어서는 등가응력을 발생시키지 않았다. 또한 지대주-지대주 나사 계면에서 축방향 하중의 증가를 보였다. 결론: Lock screw의 사용은 임플란트 구성요소에 과도한 응력을 가하지 않으며, 지대주-지대주 나사 계면의 마찰력을 증가시켜 지대주 나사 풀림을 방지할 수 있을 것으로 판단된다.

External vs internal implant : 각각의 장점과 고려사항 (External vs internal connection implant system)

  • 성동재;홍성진;하승룡
    • 대한치과의사협회지
    • /
    • 제54권3호
    • /
    • pp.184-190
    • /
    • 2016
  • The osseointegration of titanium implants has been examined over the past 50 years. Many implant systems have been introduced and have become popular to the implant dentistry. The designs of the connection between implant fixture and abutment are divided into external vs internal connection. From beginning, the $Br{\aa}nemark$ system was characterized by an external hexagon. Internal connection has been developed to reduce stress transferred to the bone. These differences may have impact on the clinical procedures and protocols, laboratory and components costs, and incidence of complications. Therefore, the clinician has to know the different biomechanical features and understand their implications to produce successful implant-supported prosthesis with an external or an internal connection system.

  • PDF

Clinical study on screw loosening in dental implant prostheses: a 6-year retrospective study

  • Lee, Ki-Young;Shin, Kyung Su;Jung, Ji-Hye;Cho, Hye-Won;Kwon, Kyung-Hwan;Kim, Yu-Lee
    • Journal of the Korean Association of Oral and Maxillofacial Surgeons
    • /
    • 제46권2호
    • /
    • pp.133-142
    • /
    • 2020
  • Objectives: In this study, we determined the incidence and pattern of screw loosening in patients who received dental implants. Materials and Methods: Patients who received implants between January 2008 and October 2013 and completed their prosthetic rehabilitation were evaluated for the incidence, frequency, and onset of screw loosening using dental charts and radiographs. The association between each factor and screw loosening was analyzed using the chi-square test and a multivariate analysis with binary logistic regression models (P<0.05). Results: Total 1,928 implants were placed in 837 patients (448 males, 389 females), whose follow-up period after loading varied from 0.25 to 70 months (mean period, 31.5 months). Screw loosening occurred in 7.2% of implants. Most cases occurred less than six months after loading. Among those, 22.3% experienced recurrent screw loosening. Screw loosening was most common in the molar region (8.5%) and frequently associated with an implant diameter of ≥5 mm (14.2%). External implant-abutment connections (8.9%) and screw-retained implant prostheses (10.1%) showed higher incidence of problems than internal implant-abutment connections and cement-retained implants, respectively. Screw loosening was most common in implant prostheses with single crowns (14.0%). Conclusion: Within the limits of the current study, we conclude that the incidence of screw loosening differs significantly according to the position of implant placement, the type of implant and manufacturer, implant diameter, the type of implant-abutment connection, the type of retention in the implant prosthesis, and the type of implant prosthesis.

단일치 임플란트에서 고정체와 지대주 직경의 차이에 따른 광탄성 응력 분석 (Photoelastic Stress Analysis of Single Implant Restoration According to Implant Fixture Size and Abutment Diameter)

  • 이진한;조혜원
    • 구강회복응용과학지
    • /
    • 제24권3호
    • /
    • pp.253-267
    • /
    • 2008
  • 이 연구의 목적은 임플란트 고정체와 지대주간의 직경 차이가 임플란트 지지조직에 발생하는 응력에 미치는 영향을 평가 하는 것이다. 본 연구에는 세 가지 직경(4.0, 5.0, 6.0㎜)의 3i 임플란트 고정체에 지대주의 직경을 달리하여 수복하고, 하중 조건(15, 30 lb)에 따라 임플란트 지지조직에 발생한 응력의 정도와 분포를 광탄성 응력 분석법을 이용하여 비교, 분석하였다. 연구결과 고정체의 직경이 증가할수록, 상부 보철물에 가해지는 하중에 대하여 고정체 변연부의 응력집중이 높게 나타났으며, 상부 보철물의 근원심 폭경을 증가시키는 것이나, 임플란트 고정체의 직경에 비해 작은 지대주를 사용하는 것은 고정체 주위의 응력양상에 영향을 주지 않았다.