Screw Joint Stability under Cyclic Loading of Zirconia Implant Abutments  

Lee, Mi-Soon (Esthetic Restorative Dentistry, Korea University Graduate School of Clinical Dentistry, Korea University Institute for Clinical Dental Research)
Suh, Kyu-Won (Esthetic Restorative Dentistry, Korea University Graduate School of Clinical Dentistry, Korea University Institute for Clinical Dental Research)
Ryu, Jae-Jun (Esthetic Restorative Dentistry, Korea University Graduate School of Clinical Dentistry, Korea University Institute for Clinical Dental Research)
Publication Information
The Journal of Korean Academy of Prosthodontics / v.47, no.2, 2009 , pp. 164-173 More about this Journal
Abstract
Purpose: The purpose of this study was to evaluate the effect of abutment material on screw-loosening before and after cyclic loading. Among the different materials of abutments, zirconia and metal abutment were used. Material and methods: Two types of implant systems: external butt joint(US II, Osstem Implant, Korea) and internal conical joint(GS II, Osstem Implant, Korea) were used. In each type, specimens were divided into two different kinds of abutments: zirconia and metal(n=5). The implant was rigidly held in a special holding to device ensure fixation. Abutment was connected to 30 Ncm with digital torque gauge, and was retightened in 30 Ncm after 10 minutes. The initial removal torque values were measured. The same specimens were tightened in 30 Ncm again and held in the cycling loading simulator(Instron, USA) according to ISO/FPIS 1480. Cycling loading tests were performed at loads 10 to 250 N, for 1 million cycles, at 14 Hz,(by subjecting sinusoidal wave from 10 to 250 N at a frequency of 14 Hz for 1 million cycles,) and then postload removal torque values were evaluated. Results: 1. In all samples, the removal values of abutment screw were lower than tightening torque values(30 Ncm), but the phenomenon of the screw loosening was not observed. 2. In both of the implant systems, initial and postload removal torque of zirconia abutment were significantly higher than those of metal abutment(P<.05). 3. In both of the implant systems, the difference in removal torque ratio between zirconia abutment and metal abutment was not significant(P>.05). 4. In metal abutments, the removal torque ratio of GS II system(internal conical joint system) was lower than that of US II system(external butt joint system)(P<.05). 5. In zirconia abutments, the difference in removal torque ratio between the two implant systems was not significant(P>.05). Conclusion: Zirconia abutment had a good screw joint stability in the condition of one million cycling loading.
Keywords
zirconia abutment; screw joint stability; cyclic loading; removal torque value; removal torque ratio;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Rangert B, Jemt T, Jorneus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989;4:241-7   PUBMED
2 Hemmings KW, Schmitt A, Zarb GA. Complications and maintenance requirements for fixed prostheses and overdentures in the edentulous mandible: a 5-year report. Int J Oral Maxillofac Implants 1994;9:191-6   PUBMED
3 Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75   DOI   ScienceOn
4 Shin HM, Jeong CM, Jeon YC, Jeong HC, Eom TG. Influence of tightening torque on implant-abutment screw joint stability. [MS dissertation.] Korea: Pusan National University; 2007
5 Prestipino V, Ingber A. Esthetic high-strength implant abutments. Part I. J Esthet Dent 1993;5:29-36   DOI   PUBMED
6 Prestipino V, Ingber A. Esthetic high-strength implant abutments. Part II. J Esthet Dent 1993;5:63-8   DOI   PUBMED
7 Gehrke P, Dhom G, Brunner J, Wolf D, Degidi M, Piattelli A. Zirconium implant abutments: fracture strength and influence of cyclic loading on retaining-screw loosening. Quintessence Int 2006;37:19-26   PUBMED
8 Butz F, Heydecke G, Okutan M, Strub JR. Survival rate, fracture strength and failure mode of ceramic implant abutments after chewing simulation. J Oral Rehabil 2005;32:838-43   DOI   ScienceOn
9 Att W, Kurun S, Gerds T, Strub JR. Fracture resistance of single-tooth implant-supported all-ceramic restorations: an in vitro study. J Prosthet Dent 2006;95:111-6   DOI   ScienceOn
10 Mcglumphy EA. Keeping implant screws tight : the solution. J Dent Symp. 1993 Aug;1:20-3   PUBMED
11 Kirkwood WF, Feng WW, Scott RG, Streit RD, Goldberg A. Mechanical properties and science of engineering materials. In: Blake A[ed]. Handbook of Mechanics, Materials, and Structures. London: Wiley, 1985:320-8
12 Leempoel PJ, Van't Hof MA, de Haan AF. Survival studies of dental restorations: criteria, methods and analyses. J Oral Rehabil 1989;16:387-94   DOI   PUBMED
13 Carlsson GE, Haraldson T. Functional response in tissueintegrated prosthesis, 4th ed. Quintessence Publ, co, 1986:74-78
14 Richter EJ. In vivo vertical forces on implants. Int J Oral Maxillofac Implants 1995;10:99-108   PUBMED
15 Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66   DOI   PUBMED   ScienceOn
16 Sutter F, Weber HP, Sorensen J, Belser U. The new restorative concept of the ITI dental implant system: design and engineering. Int J Periodont Rest Dent 1993;13:409-31
17 Bickford JH. An introduction to the design and behavior of bolted joints. New York: Marcel Deckker.; 1981. p.247-64
18 Jemt T, Lekholm U. Implant treatment in edentulous maxillae: a 5-year follow-up report on patients with different degrees of jaw resorption. Int J Oral Maxillofac Implants 1995;10:303-11   PUBMED
19 Kallus T, Bessing C. Loose gold screws frequently occur in full-arch fixed prostheses supported by osseointegrated implants after 5 years. Int J Oral Maxillofac Implants 1994;9:169-78   PUBMED
20 Jemt T. Fixed implant-supported prostheses in the edentulous maxilla. A five-year follow-up report. Clin Oral Implants Res 1994;5:142-7   DOI   PUBMED   ScienceOn
21 Heydecke G, Sierraalta M, Razzoog ME. Evolution and use of aluminum oxide single-tooth implant abutments: a short review and presentation of two cases. Int J Prosthodont 2002;15:488-93   PUBMED   ScienceOn
22 Andersson B, Taylor A, Lang BR, Scheller H, Sch?rer P, Sorensen JA, Tarnow D. Alumina ceramic implant abutments used for single-tooth replacement: a prospective 1-to 3-year multicenter study. Int J Prosthodont 2001;14:432-8   PUBMED   ScienceOn
23 Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995;73:36-43   DOI   ScienceOn
24 Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Toronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-94   DOI   ScienceOn
25 Korioth TW, Cardoso AC, Versluis A. Effect of washers on reverse torque displacement of dental implant gold retaining screws. J Prosthet Dent 1999;82:312-6   DOI   ScienceOn
26 Binon PP. Implants and components: entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94   PUBMED
27 Glauser R, Sailer I, Wohlwend A, Studer S, Schibli M, Sche'rer P. Experimental zirconia abutments for implantsupported single-tooth restorations in esthetically demanding regions: 4-year results of a prospective clinical study. Int J Prosthodont 2004;17:285-90   PUBMED   ScienceOn
28 Piconi C, Maccauro G. Zirconia as a ceramic biomaterial. Biomaterials 1999;20:1-25   DOI   ScienceOn
29 ISO/FDIS 14801 Dentistry - Fatigue test for endosseous dental implants, International Organization for Standardization, 2003(E)
30 Wiskott HW, Nicholls JI, Belser UC. Stress fatigue: basic principles and prosthodontic implications. Int J Prosthodont 1995;8:105-16   PUBMED   ScienceOn
31 Jemt T, Laney WR, Harris D, Henry PJ, Krogh PH Jr, Polizzi G, Zarb GA, Herrmann I. Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991;6:29-36   PUBMED
32 Naert I, Quirynen M, van Steenberghe D, Darius P. A sixyear prosthodontic study of 509 consecutively inserted implants for the treatment of partial edentulism. J Prosthet Dent 1992;67:236-45   DOI   ScienceOn
33 Mericske-Stern R, Zarb GA. In vivo measurements of some functional aspects with mandibular fixed prostheses supported by implants. Clin Oral Implants Res 1996;7:153-61   DOI   ScienceOn
34 Binon PP. Evaluation of machining accuracy and consistency of selected implants, standard abutments, and laboratory analogs. Int J Prosthodont 1995;8:162-78   PUBMED   ScienceOn
35 Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73   DOI   ScienceOn
36 Adrian ED, Krantz WA, Ivanhoe JR, Turner KA. A silicone obturator for the access canal in an implant-retained fixed prosthesis. J Prosthet Dent 1991;65:597   DOI   ScienceOn
37 Gratton DG, Aquilino SA, Stanford CM. Micromotion and dynamic fatigue properties of the dental implant-abutment interface. J Prosthet Dent 2001;85:47-52   DOI   ScienceOn
38 Yildirim M, Fischer H, Marx R, Edelhoff D. In vivo fracture resistance of implant-supported all-ceramic restorations. J Prosthet Dent 2003;90:325-31   DOI   ScienceOn
39 Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-9   PUBMED
40 Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36   PUBMED
41 Scherrer SS, Kelly JR, Quinn GD, Xu K. Fracture toughness (KIc) of a dental porcelain determined by fractographic analysis. Dent Mater 1999;15:342-8   DOI   ScienceOn
42 Merz BR, Hunenbart S, Belser UC. Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519-26   PUBMED
43 Branemark PI. Osseointegration and its experimental background. J Prosthet Dent 1983;50:399-410   DOI   ScienceOn
44 Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194-9   DOI   ScienceOn
45 Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res 2000;11:156-8   DOI   ScienceOn
46 Leempoel PJ, Van't Hof MA, de Haan AF. Survival studies of dental restorations: criteria, methods and analyses. J Oral Rehabil 1989;16:387-94   DOI   PUBMED
47 Faulkner MG, Wolfaardt JF, Chan A. Measuring abutment/implant joint integrity with the Periotest instrument. Int J Oral Maxillofac Implants 1999;14:681-8   PUBMED
48 Prestipino V, Ingber A. All-ceramic implant abutments: esthetic indications. J Esthet Dent 1996;8:255-62   DOI   PUBMED
49 Andresson B. Implants for single-tooth replacement. A clinical and experimental study on the Branemark CeraOne system. Swed Dent J Suppl. 1995;108:1-41   PUBMED
50 Yildirim M, Edelhoff D, Hanisch O, Spiekermann H. Ceramic abutments--a new era in achieving optimal esthetics in implant dentistry. Int J Periodontics Restorative Dent 2000;20:81-91   PUBMED
51 Richter EJ. In vivo vertical forces on implants. Int J Oral Maxillofac Implants 1995;10:99-108   PUBMED
52 Bates JF, Stafford GD, Harrison A. Masticatory function - a review of the literature. III. Masticatory performance and efficiency. J Oral Rehabil 1976;3:57-67   DOI   PUBMED
53 Binon P, Sutter F, Beaty K, Brunski J, Gulbransen H, Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants 1994;9:48-63
54 Cho HW, Dong JK, Jin TH, Oh SC, Lee HH, Lee JW. A study on the fracture strength of implant-supported restorations using milled ceramic abutments and all-ceramic crowns. Int J Prosthodont 2002;15:9-13   PUBMED   ScienceOn
55 Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75   DOI   ScienceOn
56 Gross M, Abramovich I, Weiss EI. Microleakage at the abutment-implant interface of osseointegrated implants: a comparative study. Int J Oral Maxillofac Implants 1999;14:94-100   PUBMED