This study examined the effectiveness degree of a protective apron that is taken not to be exposed to the first ray or scattered rays, for X-ray of thick subject like lateral lumbar, and the results are as follows; First, spatial dose by scattered rays is shielded by 3 mmPb protective apron, 86.8% at a distance of 50 cm, 92.7% at 100 cm, and 95.6% at 200 cm, when minimizing the field size, while 89% at a distance of 50 cm, 92.3% at 100 cm, and 95.2% at 200 cm, when maximizing the field size. Second, 1st exposure dose is shielded by 3 mmPb protective apron, 93.7% at a distance of 50 cm, 94.4% at 100 cm, and 93.6% at 200 cm, when minimizing the field size, while 93.7% at a distance of 50 cm, 93.6% at 100 cm, and 94.2% at 200 cm, when maximizing the field size.