• 제목/요약/키워드: Zirconia ceramic implant

검색결과 29건 처리시간 0.024초

사출성형 및 열간가압 소결법으로 제작된 지르코니아 세라믹 임플란트의 소결물성 및 미세구조적 결함 (Sintered Properties and Microstructural Defects of Zirconia Ceramic Implant Fabricated by Injection Molding and Hot Isostatic Pressing (HIP))

  • 박현정;박정식;이종국
    • 열처리공학회지
    • /
    • 제36권4호
    • /
    • pp.215-222
    • /
    • 2023
  • 3Y-TZP (3 mol% yttria-stabilized tetragonal zirconia polycrystals, 3Y-TZP) ceramics are emerging as dental implant materials due to their superior optical and mechanical properties as well as excellent biophysical properties, in spite of low bioactivity. In this study, we investigated to sintered properties and microstructural defects of dental zirconia implants fabricated by ceramic injection molding and post-HIP (Hot isostatic pressing) processing and analyzed the processing parameters related with the obtainment of its high sinterd density. Sintered and microstructural parameters, i.e, apparent density, grain size and phase composition of zirconia implants fabricated by injection molding were dependent on the fixtute size and implant type. Maximum sintered density of 99.2% and minimum grain size of 0.3-0.4 ㎛ were obtained from large-scaled 2-body sample. In 1-body ceramic implant, high sintered density of 99.5% was obtained, but it had a little monoclinic phase and wide grain size distribution.

최근 치과용 지르코니아 CAD/CAM 수복물의 임상적 고찰 (A clinical consideration of current dental zirconia CAD/CAM restorations)

  • 이해형
    • 대한치과의사협회지
    • /
    • 제49권5호
    • /
    • pp.279-285
    • /
    • 2011
  • Currently there is no dental ceramic material can be used in all dental situations need to be restored. However, in view of recent clinical reports, the most viable alternative is zirconia ceramic. Clinical success of dental zirconia restorations strongly depends on proper selection of materials, accurate laboratory procedure and final cementation, which can be achievable with the correct understanding of zirconia. As dental materials, zirconia ceramics have a very bright future, because they are being used increasingly in the anterior region as implant fixtures, as well as crown and bridge restorations and implant abutments. Many dental ceramics showing poor clinical performance have been gone from the dental market. However, in terms of outstanding mechanical properties and esthetic nature, new dental materials can replace zirconia ceramics will not be available in the foreseeable future.

임상가를 위한 특집 1 - 치과심미수복용 세라믹의 최신 특성평가 (Recent characteristics of dental esthetic restorative ceramics)

  • 오승한
    • 대한치과의사협회지
    • /
    • 제51권1호
    • /
    • pp.6-11
    • /
    • 2013
  • Dental ceramics is well known to have excellent esthetics, biocompatibility as well as high compressive strength. However, the fragility of ceramics against tensile and shear loads leading to the delayed fracture of micro crack on ceramic surface and the backwardness of ceramic fabrication technique limit the usage of ceramic materials in dentistry. Among all ceramic materials, zirconia has been introduced to overcome the drawback of conventional dental ceramics in the field of dentistry due to the nature of zirconia featuring proper opalescence and high fracture toughness. Also, novel manufacturing techniques enable ceramic materials to prepare high esthetic anterior and posterior all ceramic system. In this paper, it is introduced and discussed that novel techniques characterizing the bond strength between zirconia core and veneering ceramics and analyzing the fluorescence of dental ceramics in order to overcome the gap between the results of basic research and the feasibility of the results in the field of dental clinics.

Monolithic zirconia framework으로 제작된 fixed detachable prostheses를 이용한 심미적인 임플란트 전악 수복 증례 (Esthetic Full Zirconia Fixed Detachable Implant-Retained Restorations Manufactured from Monolithic Zirconia : Clinical Report)

  • 홍준태;최유성;한세진;조인호
    • 구강회복응용과학지
    • /
    • 제28권3호
    • /
    • pp.253-268
    • /
    • 2012
  • 수직적, 수평적 골흡수가 심한 환자에서 임플란트를 이용한 전악수복의 경우 경조직과 연조직 이식을 통해서 임플란트를 원하는 위치에 식립할 수도 있지만, 치은과 치아의 기능과 심미를 회복할 수 있는 fixed detachable prostheses를 대체 술식으로 사용할 수도 있다. 이러한 증례에서 다양한 재료가 수복물의 제작에 사용 가능하지만, metal/acrylic 보철물에서는 레진치아의 파절 및 탈락이 일어날 수 있고, metal/ceramic 혹은 zirconia/ceramic 보철물에서는 도재의 chipping이나 파절과 같은 문제가 발생할 수 있다. 이에 최근에 심미적이면서도 기능적인 보철수복을 위해 zirconia에 도재를 축성하지 않고 임상적으로 적용가능한 monolithic zirconia framework이 출시되어 사용되고 있다. 본 임상 증례는 심미적인 요구도가 높은 완전 무치악 환자에서 임플란트를 식립하고 chipping이나 파절의 위험을 감소시키기 위해 monolithic zirconia framework을 이용해 만들어진 complete fixed detachable 보철물을 이용한 수복에 대해 보고하고자 한다. 이번 증례에서 보철물은 심미적, 기능적으로 만족스러웠으며, 2년 간의 정기검사에서 임상적인 합병증은 보고되지 않았다.

임상가를 위한 특집 3 - 심하게 흡수된 치조제를 가진 환자에서 CAD/CAM을 이용한 임플란트 고정성 보철치료 (Implant fixed prosthetic treatment using CAD/CAM system in a patient with severe alveolar resorption)

  • 최유성
    • 대한치과의사협회지
    • /
    • 제50권3호
    • /
    • pp.126-139
    • /
    • 2012
  • Loss of dentition can lead to not only compromised esthetics and functions of the patient, but also alveolar bone resorption. Bone grafting with prosthetic reconstruction of the gingiva can be selected for the treatment, and it provides many benefits as prosthetic gingival reconstruction does not require a complicated surgical process and is available within a short period of time, with stable clinical results. However, conventional porcelain fused to metal prosthesis has certain limits due to its size, and deformation after several firing procedures. In this clinical report, the author would like to introduce a patient with severe alveolar resorption who was treated with gingiva-shaped zirconia/titanium CAD/CAM implant fixed prosthesis for esthetic and functional rehabilitation. Clinical reports Clinical report 1, 2 : A case of loss of anterior dentition with atrophied alveolar bone. Implant retained zirconia bridge applied with Procera implant bridge system to simulate the gingiva. Upper structure was fabricated with zirconia all ceramic crown. Clinical report 3, 4 : A case of atrophied maxillary alveolus was reconstructed with fixed implant prosthesis, a CAD/CAM designed titanium structure covered wi th resin on its surface. Anterior dentition was reconstructed with zirconia crown. Conclusion and clinical uses. All patients were satisfied with the outcome, and maintained good oral hygiene. Zirconia/titanium implant fixed prosthesis fabricated by CAD/CAM system was highly accurate and showed adequate histological response. No critical failure was seen on the implant fixture and abutment overall. Sites of severe alveolar bone loss can be rehabilitated by implant fixed prosthesis with CAD/CAM system. This type of prosthesis can offer artificial gingival structure and can give more satisfying esthetics and functions, and as a result the patients were able to accept the outcome more fondly, which makes us less than hard to think that it can be a more convenient treatment for the practitioners.

인공 고관절 골두용 세라믹 복합재료에 대한 세라믹-세라믹 접촉 마멸 특성 분석 (Ceramic-Ceramic Wear Zirconia/Alumina Composites For The Application Of Total Hip Joint Implant)

  • 김환;이권용;김대준;이명현;서원선
    • 한국윤활학회:학술대회논문집
    • /
    • 한국윤활학회 2004년도 학술대회지
    • /
    • pp.355-361
    • /
    • 2004
  • Ceramic femoral heads in the total hip replacement have been developed to reduce the polyethylene liner wear, Alumina and zirconia (3Y-TZP) having the excellent tribological properties are coupled against acetabular cups of polyethylene and are used in clinical application worldwide. However, alumina has a risk of catastrophic failure, and zirconia has the low temperature degradation in spite of enhanced fracture toughness. Recently, novel zirconia/alumina composite is very attractive due to the low temperature degradation (LTD)-free character and high fracture toughness. In the present study, we focus on the wear of ceramic on ceramic, which are able to be used as femoral heads and acetabular cups. Therefore, LTD-free zirconia/alumina composites with three compositions are made in a form of disk and cylinder, and the wear of the composites is performed on pin-on-disk type wear tester. The wear is conducted with or without lubricant. All the composites fabricated with the different composition show the good wear resistance.

  • PDF

Fracture resistance of implant- supported monolithic crowns cemented to zirconia hybrid-abutments: zirconia-based crowns vs. lithium disilicate crowns

  • Elshiyab, Shareen H;Nawafleh, Noor;Ochsner, Andreas;George, Roy
    • The Journal of Advanced Prosthodontics
    • /
    • 제10권1호
    • /
    • pp.65-72
    • /
    • 2018
  • PURPOSE. The aim of this in vitro study was to investigate the fracture resistance under chewing simulation of implant-supported posterior restorations (crowns cemented to hybrid-abutments) made of different all-ceramic materials. MATERIALS AND METHODS. Monolithic zirconia (MZr) and monolithic lithium disilicate (MLD) crowns for mandibular first molar were fabricated using computer-aided design/computer-aided manufacturing technology and then cemented to zirconia hybrid-abutments (Ti-based). Each group was divided into two subgroups (n=10): (A) control group, crowns were subjected to single load to fracture; (B) test group, crowns underwent chewing simulation using multiple loads for 1.2 million cycles at 1.2 Hz with simultaneous thermocycling between $5^{\circ}C$ and $55^{\circ}C$. Data was statistically analyzed with one-way ANOVA and a Post-Hoc test. RESULTS. All tested crowns survived chewing simulation resulting in 100% survival rate. However, wear facets were observed on all the crowns at the occlusal contact point. Fracture load of monolithic lithium disilicate crowns was statistically significantly lower than that of monolithic zirconia crowns. Also, fracture load was significantly reduced in both of the all-ceramic materials after exposure to chewing simulation and thermocycling. Crowns of all test groups exhibited cohesive fracture within the monolithic crown structure only, and no abutment fractures or screw loosening were observed. CONCLUSION. When supported by implants, monolithic zirconia restorations cemented to hybrid abutments withstand masticatory forces. Also, fatigue loading accompanied by simultaneous thermocycling significantly reduces the strength of both of the all-ceramic materials. Moreover, further research is needed to define potentials, limits, and long-term serviceability of the materials and hybrid abutments.

Biomechanical behavior of CAD/CAM cobalt-chromium and zirconia full-arch fixed prostheses

  • Barbin, Thais;Silva, Leticia Del Rio;Veloso, Daniele Valente;Borges, Guilherme Almeida;Presotto, Anna Gabriella Camacho;Barao, Valentim Adelino Ricardo;Groppo, Francisco Carlos;Mesquita, Marcelo Ferraz
    • The Journal of Advanced Prosthodontics
    • /
    • 제12권6호
    • /
    • pp.329-337
    • /
    • 2020
  • PURPOSE. To verify the influence of computer-aided design/computer-aided manufacturing (CAD/CAM) implant-supported prostheses manufactured with cobalt-chromium (Co-Cr) and zirconia (Zr), and whether ceramic application, spark erosion, and simulation of masticatory cycles modify biomechanical parameters (marginal fit, screw-loosening torque, and strain) on the implant-supported system. MATERIALS AND METHODS. Ten full-arch fixed frameworks were manufactured by a CAD/CAM milling system with Co-Cr and Zr (n=5/group). The marginal fit between the abutment and frameworks was measured as stated by single-screw test. Screw-loosening torque evaluated screw stability, and strain analysis was explored on the implant-supported system. All analyses were performed at 3 distinct times: after framework manufacturing; after ceramic application in both materials' frameworks; and after the spark erosion in Co-Cr frameworks. Afterward, stability analysis was re-evaluated after 106 mechanical cycles (2 Hz/150-N) for both materials. Statistical analyses were performed by Kruskal-Wallis and Dunn tests (α=.05). RESULTS. No difference between the two materials was found for marginal fit, screwloosening torque, and strain after framework manufacturing (P>.05). Ceramic application did not affect the variables (P>.05). Spark erosion optimized marginal fit and strain medians for Co-Cr frameworks (P<.05). Screw-loosening torque was significantly reduced by masticatory simulation (P<.05) regardless of the framework materials. CONCLUSION. Co-Cr and Zr frameworks presented similar biomechanical behavior. Ceramic application had no effect on the biomechanical behavior of either material. Spark erosion was an effective technique to improve Co-Cr biomechanical behavior on the implant-supported system. Screw-loosening torque was reduced for both materials after masticatory simulation.

방사선 조사된 상악골에서 all-on-4 임플란트에 의해 지지되는 지르코니아 고정성 보철물 수복 증례 (Zirconia ceramic fixed dental prosthesis with all-on-4 concept implants for irradiated maxilla: A case report)

  • 최은주;조혜원
    • 대한치과보철학회지
    • /
    • 제55권2호
    • /
    • pp.218-224
    • /
    • 2017
  • 방사선치료를 받은 상악에 임플란트 지지 고정성 보철물을 이용하여 수복하는 것은 골치유능력이 낮기 때문에 상세한 치료계획이 필요하다. All-on-4 개념에 의한 임플란트 식립은 골이식을 피하면서 임플란트를 매식할 수 있어 유리하다. 일반적으로 경사된 임플란트에는 기성 경사형 지대주를 사용해왔다. 본 임상증례에서는 computer-aided design and computer-aided manufacturing (CAD/CAM)으로 제작된 지대주를 사용하였다. 본 증례는 all-on-4 개념에 의해 임플란트를 매식하고 CAD/CAM titanium 지대주를 제작한 다음 CAD/CAM zirconia 고정성 보철물로 수복하여 좋은 결과를 얻었기에 보고하고자 한다.

All-ceramic versus titanium-based implant supported restorations: Preliminary 12-months results from a randomized controlled trial

  • Weigl, Paul;Trimpou, Georgia;Grizas, Eleftherios;Hess, Pablo;Nentwig, Georg-Hubertus;Lauer, Hans-Christoph;Lorenz, Jonas
    • The Journal of Advanced Prosthodontics
    • /
    • 제11권1호
    • /
    • pp.48-54
    • /
    • 2019
  • PURPOSE. The aim of the present randomized controlled study was to compare prefabricated all-ceramic, anatomically shaped healing abutments followed by all-ceramic abutments and all-ceramic crowns and prefabricated standard-shaped (round-diameter) titanium healing abutments followed by final titanium abutments restored with porcelain-fused-to-metal (PFM) implant crowns in the premolar and molar regions. MATERIALS AND METHODS. Forty-two patients received single implants restored either by all-ceramic restorations (test group, healing abutment, final abutment, and crown all made of zirconia) or conventional titanium-based restorations. Immediately after prosthetic incorporation and after 12 months of loading, implant survival, technical complications, bone loss, sulcus fluid flow rate (SFFR) as well as plaque index (PI) and implant stability (Periotest) were analyzed clinically and radiologically. RESULTS. After 12 months of loading, an implant and prosthetic survival rate of 100% was observed. Minor prosthetic complications such as chipping of ceramic veneering occurred in both groups. No statistical significant differences were observed between both groups with only a minimum of bone loss, SFFR, and PI. CONCLUSION. All-ceramic implant prostheses including a prefabricated anatomically shaped healing abutment achieved comparable results to titanium-based restorations in the posterior region. However, observational results indicate a benefit as shaping the peri-implant soft-tissue with successive provisional devices and subsequent compression of the soft tissue can be avoided.