Purpose: Early diagnosis of breast is of the utmost importance to improve prognosis. We have a limitation for mammography and sonography detecting small cancer. Clinical importance of Breast Specific Gamma Imaging (BSGI) has improved for that reason. So We studied performance evaluation test of count rate and resolution with high sensitivity to the low dose of BSGI. Materials and Methods: BSGI of Dilon 6800, point source of $^{99m}Tc$ from 1.85~148 MBq (0.05~4 mCi) at the intervals of 1.85~37 MBq (0.05~1 mCi) was used for the test. Performance evaluation method was performed for measuring deadtime for choosing at the 5 different point in the useful field of view (UFOV), acquired image for 60 seconds. Compared with reference of clinical uptake distribution of breast, activity increased according to the distance change 10, 20, 30, 40, 50 mm in the useful field of view. Results: Counting curve increased according to the activity from 1.85 MBq (0.05 mCi) to the 74 MBq (2 mCi), and it change flat shape over 74 MBq (2 mCi). The variation of the full width of half maximum (FWHM) to the distance is 4.05, 4.73, 5.77, 6.90, 8.00, 9.32 mm in 1.85 MBq (0.05 mCi), 4.30, 4.80, 5.90, 7.00, 8.10, 9.07 mm in 3.7 MBq (0.1 mCi), 4.90, 5.60, 6.20, 7.20, 8.20, 9.10 mm in 5.55 MBq (0.15 mCi), 5.30, 6.10, 6.60, 7.00, 7.90, 8.70 mm in 7.40 MBq (0.2 mCi). Conclusion: Distortions of image would be acquired because of the deadtime in BSGI. We found out the fact that specification of $^{99m}Tc$ reaction under 74 MBq (2 mCi) for BSGI. Second, FWHM distribution change from varied distance from the detector, clearly distinguished the location of the lesion.