The Korean Journal of Nuclear Medicine Technology (핵의학기술)
- Volume 14 Issue 1
- /
- Pages.24-27
- /
- 2010
- /
- 1229-9901(pISSN)
- /
- 2982-8406(eISSN)
Performance Evaluation of Breast Specific Gamma Imaging
Breast Specific Gamma Imaging 장비의 성능평가
- Cha, Eun-Sun (Department of Nuclear Medicine, Samsung Medical Center) ;
- Kwak, In-Suk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Noh, Ik-Sang (Department of Nuclear Medicine, Samsung Medical Center) ;
- Yeon, Joon-Ho (Department of Nuclear Medicine, Samsung Medical Center) ;
- Kim, Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
- Choi, Choon-Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
- Seok, Jae-Dong (Department of Nuclear Medicine, Samsung Medical Center)
- 차은선 (삼성서울병원 핵의학과) ;
- 곽인석 (삼성서울병원 핵의학과) ;
- 노익상 (삼성서울병원 핵의학과) ;
- 연준호 (삼성서울병원 핵의학과) ;
- 김기 (삼성서울병원 핵의학과) ;
- 최춘기 (삼성서울병원 핵의학과) ;
- 석재동 (삼성서울병원 핵의학과)
- Received : 2010.01.06
- Accepted : 2010.01.26
- Published : 2010.06.05
Abstract
Purpose: Early diagnosis of breast is of the utmost importance to improve prognosis. We have a limitation for mammography and sonography detecting small cancer. Clinical importance of Breast Specific Gamma Imaging (BSGI) has improved for that reason. So We studied performance evaluation test of count rate and resolution with high sensitivity to the low dose of BSGI. Materials and Methods: BSGI of Dilon 6800, point source of
Breast cancer는 조기 병변의 발견이 매우 중요하나 크기가 작은 암세포를 구별해 내는 데 한계가 있어 예민도 96.4% 이상의 성능을 얻을 수 있는 BSGI가 도입되어 임상적인 중요성이 강조되고 있다. 따라서 예민한 감도를 보이는 BSGI 장비를 이용하여 계수율, 해상력 등의 성능 평가를 하고자 한다. Dilon사의 BSGI 장비와