• 제목/요약/키워드: AEC (Automatic Exposure Control)

검색결과 42건 처리시간 0.022초

CareDose 4D 사용 시 동일한 스캔조건에서 조직기반설정을 다르게 적용함에 따른 선량 비교: 성인과 소아팬텀 연구 (Radiation Dose Comparison according to Different Organ Characteristics at Same Scan Parameters Using CareDose 4D: An Adult and Pediatric Phantom Evaluation)

  • 공효금;이기백
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제42권4호
    • /
    • pp.271-277
    • /
    • 2019
  • CareDose 4D which is the Siemens's Automatic Exposure Control (AEC) can adjust the level of radiation dose distribution which is based on organ characteristic unlike other manufacturer's AEC. Currently, a wide scan range containing different organs is sometimes examined at once (defined as one scan). The purpose of this study was to figure out which organ characteristic option is suitable when one scan method is utilized. Two types of anthropomorphic phantoms were scanned in the same range which were from frontal bone to carina level according to three different organ characteristics such as Thorax, Abdomen, and Neck. All scans and image reconstruction parameters were equally applied and radiation dose were compared. Radiation dose with Thorax organ characteristic was lower than that with Neck. Also, that with Abdomen oran characteristic was lower than Thorax. There were significant differences in radiation dose according to different organ characteristics at the same parameters (P<0.05). Usage of Neck organ characteristic had a result of the highest radiation dose to all phantom. On the other hand, utilization of Abdomen organ characteristic showed the lowest radiation dose. As a result, it is desirable to set appropriate organ characteristic according to examined body part when you checkup patients. Also, when you implement one scan method, selection of Abdomen-based organ characteristic has reduced more radiation dose compared with two different organ characteristic.

전신 PET/CT 검사에서 환자의 팔 위치에 따른 CT 유효선량의 비교 연구 (A Comparative Study on the CT Effective Dose by the Position of Patient's Arm)

  • 성지혜;박순기;김정선;박승용;정우영
    • 핵의학기술
    • /
    • 제16권1호
    • /
    • pp.44-49
    • /
    • 2012
  • 최근 영상 의학 분야의 발전으로 인한 보편화와 일본의 후쿠시마 원전사태가 일어나면서 방사선 노출에 대한 대중들의 관심이 높아지게 되었다. 이러한 방사선 피폭에 대한 대중들의 관심으로 현재 거의 대부분의 PET/CT검사에서 사용되고 있는 AEC (Automatic Exposure Control) 시스템은 환자의 피폭선량 감소를 위해 피사체의 두께에 따라 최적의 관전류를 조사한다. 또한, 서울아산병원에서는 전신 PET/CT 검사 시선속 경화 인공물(Beam hardening artifact)의 발생을 줄이기 위해 환자의 팔을 머리 위로 올리고 검사를 시행하고 있다. 본 연구에서는 전신 PET/CT검사 시 환자의 팔 위치에 따라 AEC 시스템을 적용한 CT 유효선량의 차이를 비교하고자 한다. 2008년 12월부터 2011년 7월까지 서울아산병원 핵의학과를 내원하여 동일 장비에서 팔을 올렸을 경우와 팔을 내렸을 경우의 전신 18F-FDG PET/CT 검사를 시행한 환자 45명을 연구의 대상으로 하였고, 실험 장비는 Biograph Truepoint 40 (Bio 40), Biograph Sensation 16 (Bio 16), Discovery STe 8 (DSTe 8)을 사용하였다. 각 장비당 15명의 동일 환자를 대상으로, 팔을 올리고 검사하였을 경우와 팔을 내리고 검사하였을 경우의 CT 유효선량을 산출하여 비교 분석하였다. CT 유효선량의 측정 방법은 ImPACT v1.0 프로그램을 사용하였다. 이 연구를 통해 팔을 올리고 검사한 경우에 팔을 내리고 검사한 경우보다 총 15.2%의 환자 유효선량이 감소된 것을 확인할 수 있었다. 전신 PET/CT 검사 시 환자의 팔을 올린 자세는 선속 경화 인공물 발생을 줄이고, 영상의 질을 향상시킬 수 있을 것이라 생각된다.

  • PDF

자동노출제어장치 센서적용을 위한 스크린 프린팅 제작방식의 고흡수율 광도전체 특성평가 (Evaluation of High Absorption Photoconductor for Application to Auto Exposure Control Sensor by Screen Printing Method)

  • 김대국;김교태;박정은;홍주연;김진선;오경민;남상희
    • 한국방사선학회논문지
    • /
    • 제9권2호
    • /
    • pp.67-72
    • /
    • 2015
  • 진단방사선 분야에서는 진단 최적화를 위하여 자동노출제어장치의 활용이 국제적으로 권고되고 있다. 하지만, 기존의 상용화된 광도전체 센서는 제작 공정의 복잡하고 장시간 방사선에 노출될 경우 다양한 성능 저하가 발생하는 문제점이 있다. 이에 본 연구에서는 X-ray 흡수율이 높으면서도 제작이 용이한 장점을 가진 광도전체 기반 센서의 AEC 적용 가능성을 평가하고자 한다. 실험결과, SNR 증가를 통하여 우수한 검출 효율을 가지는 센서의 제작가능성을 확인하였고, 정확한 턴-오프가 가능할 것으로 사료된다. 또한 잠상 영상 및 투과율 실험 결과, 광도전체에 의한 Ghost effect가 나타나지 않음을 확인하였으며, PbO를 제외한 광도전체의 경우 80% - 90%의 우수한 투과율을 확인하였다. 그러므로 상용화된 기존 상품에 대비하여 도핑농도 변화에 따른 성능 저하 및 기계적 안정성이 뛰어나며 제작이 용이한 광도전체 기반의 센서는 AEC 센서로 적용이 가능할 것으로 기대된다.

디지털 흉부영상에서 자동노출제어 및 감도변화를 이용한 영상품질의 정량적인 평가 (Quantitative Evaluation of Image Quality using Automatic Exposure Control & Sensitivity in the Digital Chest Image)

  • 이진수;고성진;강세식;김정훈;김동현;김창수
    • 한국콘텐츠학회논문지
    • /
    • 제13권8호
    • /
    • pp.275-283
    • /
    • 2013
  • 자동노출제어장치를 사용하는 흉부 후전 검사는 이온전리조의 선택에 따라 환자가 받는 피폭선량이 달라진다. 본 논문에서는 자동노출제어장치의 이온전리조의 선택에 따른 최적의 진단 영상을 획득하면서 피폭선량을 최소화하는 방안에 대해 연구하였다. 실험방법은 흉부 후전 검사와 동일한 조건으로 자동노출제어장치의 이온전리조 선택과 감도변화에 따라 실험하였다. 이온전리조의 상단 2개, 하단 1개의 센서를 on/off 선택에 따라 7가지의 경우로 나누어 각각 5회씩 측정하여 평균값을 구하고 피폭선량을 산출하였다. 영상평가는 변조전달함수, 최대신호 대 잡음비, 평균제곱근, 신호 대 잡음비, 대조도 대 잡음비, 평균대표준편차비를 각각 측정하여 평가를 시행하였다. 실험결과 피폭선량 평가에서 이온전리조 상단 2개를 선택한 경우가 다른 조합에 비해 가장 낮은 선량을 나타내었고, 해상력 평가결과로는 감도 625(High)에서 상단2개를 선택한 영상이 두 번째로 높은 공간주파수 1.343 lp/mm를 나타내었다. 상단 2개를 선택한 영상의 평균제곱근 결과값이 두 번째로 낮게 나타났으며, 신호 대 잡음비, 대조도 대 잡음비, 평균 대 표준편차비는 두 번째로 높은 결과값을 나타내었다. 그리고 감도가 증가함에 따라 피폭선량은 감소하였으며, 영상품질 측면에서도 보다 우수한 영상을 얻을 수 있었다. 따라서 피폭선량을 최소화하고, 최적의 의료 영상을 얻기 위해서는 감도 625(High)에서 이온전리조 상단 2개를 선택하는 것이 임상적으로 유용할 것으로 사료된다.

맘모그래피의 화질과 평균유선조직선량에 관한 검토 (A Study on the Image Quality of Mammography and the Average Glandular Dose)

  • 이인자;김학성;김성수;허준
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제25권2호
    • /
    • pp.47-55
    • /
    • 2002
  • 유방촬영에 대한 화질과 평균유선조직선량에 대하여 실험한 결과 반가층은 장치마다 큰 차이를 나타내지 않았으며, 자동노출기구를 사용할 경우 정확성에 문제가 있으므로 확인이 필요하다. 유방촬영에 사용되는 자동현상기 또한 꼭 관리가 필요하며, 유방촬영 시 사용되는 증감지/필름에 따라서 팬톰 촬영 식별능에 많은 영향을 받고 있으므로 선택 시 신중을 기해야 한다. 유방두께 4.2 cm에 대한 평균유선조직 선량은 관전압에 따라 달라지며, ACR기준인 3.0 mGy보다 훨씬 적은 $0.26{\sim}1.39\;mGy$로 나타났다. 이상에서 알 수 있듯이 유방촬영에서는 촬영장치의 성능뿐만 아니라 HVL, AEC, 자동현상기 관리, 증감지/필름 등 모두 관리가 되어야 하며, 이로 인해 평균유선조직선량은 더 줄일 수 있을 것이다.

  • PDF

Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population?

  • Kim, Eun Young;Kim, Tae Jung;Goo, Jin Mo;Kim, Hyae Young;Lee, Ji Won;Lee, Soojung;Lim, Jun-tae;Kim, Yeol
    • Korean Journal of Radiology
    • /
    • 제19권6호
    • /
    • pp.1179-1186
    • /
    • 2018
  • Objective: The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea. Materials and Methods: For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index ($CTDI_{vol}$) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from $CTDI_{vol}$ using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no). Results: Size-specific dose estimate was higher than $CTDI_{vol}$ ($2.22{\pm}0.75mGy$ vs. $1.67{\pm}0.60mGy$, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to $CTDI_{vol}$ was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to $CTDI_{vol}$ was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/$CTDI_{vol}$ in large- vs. standard-sized participants: $1.30{\pm}0.08$ vs. $1.44{\pm}0.08$ (p < 0.001) and without AEC, $1.32{\pm}0.08$ vs. $1.42{\pm}0.06$ (p < 0.001). Conclusion: Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants.

반복적 재구성 알고리즘과 관전류 자동 노출 조정 기법의 CT 영상 화질과 선량에 미치는 영향: 관상동맥 CT 조영 영상 프로토콜 기반의 팬텀 실험 (Effects of Iterative Reconstruction Algorithm, Automatic Exposure Control on Image Quality, and Radiation Dose: Phantom Experiments with Coronary CT Angiography Protocols)

  • 하성민;정성희;장혁재;박은아;심학준
    • 한국의학물리학회지:의학물리
    • /
    • 제26권1호
    • /
    • pp.28-35
    • /
    • 2015
  • 본 논문에서는 반복적 구성 기법과 관전류 노출자동조절 기법이 영상의 화질과 방사선량에 미치는 영향을 관상동맥 전산화단층촬영 혈관조영 영상(coronary computed tomography angiography, CCTA)을 대상으로 팬텀 실험에 기반하여 평가하고자 한다. 이를 위하여 미국 의학물리학회(American Association of Physics in Medicine) 표준의 성능 평가 팬텀을 320 다중검출열 CT로써 촬영하였다. 80 kVp, 100 kVp, 120 kVp의 관전압에 있어서, 관전류 노출자동조절 기법은 저선량 목표 표준편차(SD=44)와 고선량(목표 표준편차=33)의 두 가지 설정으로써 촬영하였다. 재구성 변수로서는 필터보정 역투영(FBP)와 반복적 재구성 방법을 설정하여, 전부 12개의 재구성 영상을 획득하였다(12=3 (80, 100, 120 kVp)${\times}2$ (저선량(목표SD=44), 고선량(목표SD=33))${\times}2$ (필터보정역투영, 반복적 재구성). 영상의 화질은 잡음의 세기(표준편차), 변조전달함수, 대조대잡음비(CNR)에 의하여 평가하였으며, 관전압과 관전류 노출자동조절 기법에서의 목표 선량과 대소 및 재구성 기법의 선택이 화질과 방사선량에 미치는 영향을 관찰하였다. 반복적 재구성 기법은 필터보정역투영 기법보다 영상 잡음을 대폭 감소시켰으며 이는 저선량의 경우 더욱 뚜렷하였다. 즉, 잡음의 세기는 관전류 노출자동조절의 설정이 고선량 (목표SD=33)과 저선량(목표SD=44)인 경우, 각각 평균 38%와 평균 46% 감소하였다. 반복적 재구성 기법에 의하여, 변조전달 함수에 의한 공간적 해상도의 평가에 있어서 미약한 감소를 보였으나, 이로써 잡음 저감과 대조대잡음비(CNR)에 있어서의 현저한 개선을 상쇄할 정도의 영향에는 미치지 못 하였다. 결과적으로, 관상동맥 전산화단층촬영 혈관조영 영상의 획득에서 있어서, 반복적 재구성 기법과 관전류 노출자동조정 기법을 동시에 사용하는 것은 영상의 화질을 개선하면서 공간적 해상도의 저하 등 그 부작용은 최소화함으로써, 합리적으로 획득 가능한 한 최소한의 선량 (ALARA)의 원칙에 충실한 실제 임상적 효과를 의미한다고 기대할 수 있다.

자동노출조절장치를 이용한 흉부 방사선검사 시 갑상샘 장기선량 최소화를 위한 새로운 차폐 방법에 관한 연구 (A Study on New Shielding Method for Minimizing Thyroid Oran Dose on Chest Radiography Used Automatic Exposure Control)

  • 주영철;홍동희;한범희
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제43권5호
    • /
    • pp.323-329
    • /
    • 2020
  • The purpose of this study is to investigate the effect of radiation shielding on the thyroid organ dose and image quality during Chest PA examination using automatic exposure control system. This study was conducted in the patient posture and examination conditions such as Chest PA using human model phantom. An experiment without shielding was set as a control group (non) and the cases of using paper coated with a contrast agent (contrast) and bismuth (bismuth) were used as experimental groups. Compared to non-shielded(non), the dose at bismuth increased about 7% in C(cervical vertebrae)5 and C6 and 14% in C7 and contrast showed dose increases of about 17 to 19% in C5 and C6 and about 21% in C7. As a result of the image quality comparison, when measured in the center of the cervical vertebrae, both SNR and CNR in bismuth increased about 40% higher than non, and contrast showed about 8 to 9% improvement. Compared with soft tissues of the cervix, bismuth reduced SNR by about 15% and CNR by about 13%, in contrast, SNR decreased by 11%, and CNR decreased by about 10%. In the Chest PA using AEC, the method of using the shield in front of the collimator has the advantage to observe the anatomical structure of the neck area well compared to the method using the lead. However, the dose at the neck can be increased by 7-21% depending on shielding materials.

Standing Whole Spine 검사 시 체질량지수 (BMI)에 따른 전방 및 후방장기의 흡수선량 평가 (Estimation of Absorbed Dose for Anterior and Posterior Organs with Body Mass Index in Standing Whole Spine Examination)

  • 심지나;이용구;이영진
    • 전자공학회논문지
    • /
    • 제53권12호
    • /
    • pp.147-151
    • /
    • 2016
  • Standing Whole Spine 검사는 많은 병원에서 자동노출제어장치 (automatic exposure control, AEC)를 사용하고 있어 체질량지수 (body mass index, BMI)에 따라 방사선량이 조절되는 특징이 있으며, 이를 통하여 환자 별 측정 DAP (dose area product) 값을 얻을 수 있다. 하지만, 검사 시 장기의 위치에 따른 흡수선량에 관한 연구는 크게 이루어지지 않고 있으며, 이에 본 연구에서는 Standing Whole Spine 검사 시 환자의 두께정보를 대표하는 BMI와 장기의 위치에 따른 흡수선량의 분포를 평가하고자 한다. 연구의 목적을 위하여 측정된 DAP값을 이용하여 PCXMC에서 환자의 5곳의 장기를 선정 (갑상샘, 유방, 심장, 콩팥, 이자)하여 선량을 계산하였다. 결과적으로, 측정된 DAP값은 BMI에 따라 증가하는 경향을 보였지만 전방 장기인 갑상샘, 유방, 그리고 심장에서는 BMI에 따라 장기선량이 감소하는 경향을 보였다. 또한 후방장기인 콩팥과 이자에서는 BMI와 아무런 상관관계를 가지지 않았다. 결론적으로, 본 연구결과를 통하여 Standing Whole Spine 검사 시 BMI와 장기의 위치에 따라 방사선의 영향이 다르게 나타남을 증명하였다.

일반촬영 자동노출제어장치 사용 시 잔존 CT 조영제와 MRI 조영제에 따른 면적선량의 상관성 연구 (A Study of The Correlation of The Area Dose with Residual CT Contrast Media and MRI Contrast Media during The Use of General Imaging Automatic Exposure Control System)

  • 홍찬우;박진훈;이정민;서영득
    • 한국방사선학회논문지
    • /
    • 제10권8호
    • /
    • pp.619-627
    • /
    • 2016
  • 본 연구에서는 일반촬영에서 자동노출제어장치를 이용하여 검사할 때 체내에 잔존하는 CT 조영제와 MRI 조영제가 면적선량에 미치는 영향에 대해 알아보고자 한다. 복부 두께의 파라핀 팬텀 중앙에 사각형의 홀을 만든 후 CT 조영제, MRI 조영제를 각각 식염수 대 조영제 희석 비율, 10:0, 9:1, 8:2, 7:3, 6:4, 5:5, 4:6, 3:7, 2:8, 1:9, 0:10 %의 비율로 희석하였다. 각 실험은 KUB 촬영의 적정 조건인 78kVp, 320mA로 설정 후 자동노출제어장치를 이용하여 희석 비율 당 총 30회씩의 검사를 하였고, 각 조영제의 희석 비율에 따른 면적선량, 노출지수에 대해 평균 비교와 상관분석을 하였다. 그 결과, CT 조영제와 MRI 조영제는 희석 비율에 따라 면적선량이 다르게 나타났고(p<0.05), 희석 비율이 증가함에 따라 CT 조영제와 MRI 조영제는 면적선량이 증가하는 것으로 나타났다(p<0.05). 각 검사에서 노출지수는 제조사 권고 사항인 200-800 EI값을 나타냈으며, 노출지수와 면적선량은 면적선량이 증가할수록 노출지수도 증가하였다(p<0.05). 결론적으로 CT 조영제와 MRI 조영제는 모두 자동노출제어장치를 사용하는 일반촬영 검사에서 면적선량을 증가시키는 것을 확인하였다. 따라서 CT와 MRI 검사에서 조영제를 사용한 후 일반촬영 검사를 할 때에는 조영제가 충분히 배설된 후 시행하여야 할 것으로 사료된다.