Radiation Dose Comparison according to Different Organ Characteristics at Same Scan Parameters Using CareDose 4D: An Adult and Pediatric Phantom Evaluation  

Kong, Hyo-Geum (Department of Radiology, Kangwon National University Hospital)
Lee, Ki-Baek (Department of Radiology, Seoul Asan Medical Center)
Publication Information
Journal of radiological science and technology / v.42, no.4, 2019 , pp. 271-277 More about this Journal
Abstract
CareDose 4D which is the Siemens's Automatic Exposure Control (AEC) can adjust the level of radiation dose distribution which is based on organ characteristic unlike other manufacturer's AEC. Currently, a wide scan range containing different organs is sometimes examined at once (defined as one scan). The purpose of this study was to figure out which organ characteristic option is suitable when one scan method is utilized. Two types of anthropomorphic phantoms were scanned in the same range which were from frontal bone to carina level according to three different organ characteristics such as Thorax, Abdomen, and Neck. All scans and image reconstruction parameters were equally applied and radiation dose were compared. Radiation dose with Thorax organ characteristic was lower than that with Neck. Also, that with Abdomen oran characteristic was lower than Thorax. There were significant differences in radiation dose according to different organ characteristics at the same parameters (P<0.05). Usage of Neck organ characteristic had a result of the highest radiation dose to all phantom. On the other hand, utilization of Abdomen organ characteristic showed the lowest radiation dose. As a result, it is desirable to set appropriate organ characteristic according to examined body part when you checkup patients. Also, when you implement one scan method, selection of Abdomen-based organ characteristic has reduced more radiation dose compared with two different organ characteristic.
Keywords
Radiation Dose; CareDose 4D; Automatic Exposure Control; One Scan Method; Organ Characteristic; Computed Tomography;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 McCollough CH. Automatic exposure control in CT: are we done yet? Radiology. 2005;237(3):755-6.   DOI
2 Lee KB, Goo HW. Quantitative Image Quality and Histogram-Based Evaluations of an Iterative Reconstruction Algorithm at Low-to-Ultralow Radiation Dose Levels: A Phantom Study in Chest CT. Korean J Radiol. 2018;19(1):119-29.   DOI
3 Leipsic J, Nguyen G, Brown J, Sin D, Mayo JR. A prospective evaluation of dose reduction and image quality in chest CT using adaptive statistical iterative reconstruction. Am J Roentgenol. 2010;195(5):1095-9.   DOI
4 Goo HW. CT radiation dose optimization and estimation:an update for radiologists. Korean J Radiol. 2012;13(1):1-11.   DOI
5 Baker ME, Dong F, Primak A, Obuchowski NA, Einstein D, Gandhi N, et al. Contrast-to-noise ratio and low-contrast object resolution on full- and low-dose MDCT: SAFIRE versus filtered back projection in a low-contrast object phantom and in the liver. Am J Roentgenol. 2012;199(1):8-18.   DOI
6 Singh S, Kalra MK, Hsieh J, Licato PE, Do S, Pien HH, et al. Abdominal CT: comparison of adaptive statistical iterative and filtered back projection reconstruction techniques. Radiology. 2010;257(2):373-83.   DOI
7 Wang L, Gong S, Yang J, Zhou J, Xiao J, Gu JH, et al. CARE Dose 4D combined with sinogram-affirmed iterative reconstruction improved the image quality and reduced the radiation dose in low dose CT of the small intestine. J Appl Clin Med Phys. 2019;20(1):293-307.   DOI
8 Lee MK, Lim CH. Survey of CT Practice and Collective Effective Dose Estimation. Journal of Radiological Science and Technology. 2010;33(3):231-8.
9 Choi MH, Jang JS, Lee KB. A Study on the Indirect Radiation Exposure of the Medical Personnel Who is Responsible for Patient Safety in CT Examination. Journal of Radiological Science and Technology. 2019;42(2):105-11.   DOI
10 Lee KB, Lee WH, Lee JH, Lee BR, Oh GB. Dose reduction and image quality assessment in MDCT using AEC (D-DOM & Z-DOM) and in-plane bismuth shielding. Radiat Prot Dosimetry. 2010;141(2):162-7.   DOI
11 Soderberg M, Gunnarsson M. Automatic exposure control in computed tomography--an evaluation of systems from different manufacturers. Acta Radiol. 2010;51(6):625-34.   DOI
12 Philipp MO, Kubin K, Hormann M, Metz VM. Radiological emergency room management with emphasis on multidetector-row CT. Eur J Radiol. 2003;48(1):2-4.   DOI
13 Park GJ, Kim YO, Yeo SG, Dae CM, Min KH. Automatic Exposure Control Systems with Inappropriate Topogram : A Phantom study in Chest CT on Radiation dose and Image quality between different manufactures. Journal of Korean Society of Computed Tomographic Technology. 2011;13(2):227-38.
14 Nguyen D, Platon A, Shanmuganathan K, Mirvis SE, Becker CD, Poletti PA. Evaluation of a Single-Pass Continuous Whole-Body 16-MDCT Protocol for Patients with Polytrauma. Am J Roentgenol. 2009; 192(1):3-10.   DOI
15 Hauser H, Bohndorf K. Radiologic emergency management in multiple trauma cases. Radiologe. 1998; 38(8):637-44.   DOI
16 Mazumdar A, Siegel MJ, Narra V, Luchtman-Jones L. Whole-Body Fast Inversion Recovery MR Imaging of Small Cell Neoplasms in Pediatric Patients: A Pilot Study. Am J Roentgenol. 2002;179(5):1261-6.   DOI
17 Marin D, Nelson RC, Rubin GD, Schindera ST. Body CT: technical advances for improving safety. Am J Roentgenol. 2011;197(1):33-41.   DOI
18 Lee KB, Kim JH. The Effect of Radiation Dose and Image Quality using AEC(Automatic Exposure Control) with Inappropriate Scout images : A Chest Phantom Experiment with Two Different AEC Modes. Journal of Korean Society of Computed Tomographic Technology. 2016;18(1):47-56.
19 Wood TJ, Moore CS, Stephens A, Saunderson JR, Beavis AW. A practical method to standardise and optimise the Philips DoseRight 2.0 CT automatic exposure control system. J Radiol Prot. 2015;35(3):495-506.   DOI
20 Russell MT, Fink JR, Rebeles F, Kanal K, Ramos M, Anzai Y. Balancing radiation dose and image quality: clinical applications of neck volume CT. Am J Neuroradiol. 2008;29(4):727-31.   DOI