Lee, W.H.;Suh, S.S.;In, K.H.;Lee, H.J.;Kim, K.C.;Yoon, C.H.;Auh, Y.H.
49
To inquire its usefulness of the clinical application of intelligent replenishment system of automatic X-ray film processor based on film density, we processed the serial 300 sheets of radiographic film of chest [$14{\times}14"$, HR-C type] and bone [elbow & ankle($8{\times}10"$), skull($10{\times}12"$), hand & foot($11{\times}14"$), pelvis($14{\times}17"$), HR-G type, 68, 70, 77, 85 sheets respectively]. We analyzed the characteristic corves, relative speeds, average gradients and base plus fog densities every twenty five sheets. We also evaluated the developer and fixer replenishment volumes every that time. In the chest and bone radiograph two all, the characteristic curves were little change, and the relative speeds, average gradients and base plus fog densities were within the maximum control limits. The average developer replenishment volumes were about 43m1/sheet and 39m1/sheet respectively. It brings decreased results about 29% in comparison with the conventional replenishment system. In our experiences, we conclude that the intelligent replenishment system of automatic X-ray film processor based on film density maintains image quality consistently, decreases also the replenishment volumes. Therefore, this system will be resulted in economic and environmental effects, and solve problems of over and low replenishment volume.