Experiment was conducted on the image quality and exposure dose following replacement of $CaWO_4$ system screen BH-III and BX-III which have so far been used for high-voltage hard-radiation quality radiography, with rare earth system screen KO750, combined with high contrast film SRH, while additional filter was altered, Cu 0.8 mm+Al 1.4 mm(HVL : Al 8.8 mm), Cu 1.3 mm+Al 1.0 mm(HVL : Al 10.6 mm) and Cu 1.8 mm+Al 1.5 mm(HVL : Al 11.4 mm). AS a result, visual evaluation did not detect extreme changes in image quality under the respective condition(HVL : $Al\;8.8\;mm{\sim}Al\;11.4\;mm$). It was noted, however, that surface exposure dose declined with an increase in the thickness of the additional filter, as it was $18.9\;{\mu}Gy$ at HVL Al 8.8 mm, $17.5\;{\mu}Gy$ at Al 10.6 mm and $15.7\;{\mu}Gy$ at Al 11.4 mm. Considering the limited rating of X-ray equipment and wear of machinery, however, the range of $Cu\;1.3\;mm{\sim}1.8\;mm+Al\;1.0\;mm{\sim}1.5\;mm(1/16\;VL{\sim}1/32\;VL)$ seemed to be a limit.