Clinical Study on Measurement Accuracy of Astigmatic Components

난시성분의 측정 정확도에 관한 임상적 연구

  • Received : 2005.11.25
  • Published : 2006.01.31

Abstract

Clinical evaluation by Astigmatic Components of autorefractors was performed to examine validity and accuracy compared with subjective refraction. The mean cylindrical powers of Astigmatic Components of Subjective Refraction, Closed-view and Open-view were found to be $-0.689{\pm}0.516D$, $-0.691{\pm}0.530D$, $-0.470{\pm}0.507D$, respectively. The Difference of cylindrical Components between Subjective Refraction and Closed, Open-view were found to be $0.002{\pm}0.191D$, $0.015{\pm}0.137D$, respectively. Approximately 85% of Closed-view and Open-view autorefractor measurements were within ${\pm}20^{\circ}$ range of the difference of cylindrical axis with subjective refraction. The mean $J_0$ Components of Astigmatic Components of subjective refraction, Closed-view and Open-view were found to be $-0.155{\pm}0.332D$, $-0.004{\pm}0.178D$, $-0.001{\pm}0.156D$, respectively. The mean $J_{45}$ Components of Astigmatic Components, Closed-view and Open-view were found to be $-0.023{\pm}0.227D$, $-0.026{\pm}0.172D$, $-0.014{\pm}0.182D$, respectively. Both of the Closed-view and Open-view refractor showed available accuracy in measuring astigmatic components.

본 연구는 자각적 굴절검사 값과 비교분석하여 정확도와 신뢰성을 조사하기 위하여 자동굴절검사기기의 난시 구성요소들을 임상적으로 평가하였다. 난시 도수량의 평균은 자각굴절검사, 폐쇄형, 개방형 기기에서 각각 $-0.689{\pm}0.516D$, $-0.691{\pm}0.530D$, -$0.704{\pm}0.507D$로 측정되었으며, 자각검사와의 난시 도수량 차이는 폐쇄형, 개방형 기기 각각에서 $0.002{\pm}0.191D$(p=0.191), $0.015{\pm}0.137D$(p=0.137)이었으며, 난시축은 두 기기 모두 약 85%이상 ${\pm}20$도 이내의 차이를 보였다. $J_0$성분의 평균은 자각식 검사법과 폐쇄형 및 개방형에서 각각 $-0.155{\pm}0.332D$, $-0.004{\pm}0.178D$, $-0.001{\pm}0.156D$로 계산되었다. $J_{45}$ 성분의 평균은 자각식 검사법과 폐쇄형 및 개방형에서 각각 $-0.023{\pm}0.227D$, $-0.026{\pm}0.172D$, $-0.014{\pm}0.182D$로 계산되었다. 두 자동굴절검사기기 모두 난시측정 시 참고할 만한 정확성을 나타내었다.

Keywords