Disposable Strip-Type Biosensors for Amperometric Determination of Galactose

  • Gwon, Kihak;Lee, Seonhwa;Nam, Hakhyun;Shin, Jae Ho
    • Journal of Electrochemical Science and Technology
    • /
    • v.11 no.3
    • /
    • pp.310-317
    • /
    • 2020
  • A development of disposable strip-type galactose sensor for point-of-care testing (POCT) was studied, which was constructed using screen-printed carbon electrodes. Galactose levels were determined by the redox reaction of galactose oxidase in the presence of potassium ferricyanide as an electron transfer mediator in a small sample volume (i.e., less than 1 µL). The optimal performance of biosensor was systematically designated by varying applied potential, operating pH, mediator concentration, and amount of enzyme on the electrode. The sensor system was identified as a highly active for the galactose measurement in terms of the sensitivity (slope = 4.76 ± 0.05 nA/µM) with high sensor-to-sensor reproducibility, the linearity (R2 = 0.9915 in galactose concentration range from 0 to 400 µM), and response time (t95% = <17 s). A lower applied potential (i.e., 0.25 V vs. Ag/AgCl) allowed to minimize interference from readily oxidizable metabolites such as ascorbic acid, acetaminophen, uric acid, and acetoacetic acid. The proposed galactose sensor represents a promising system with advantage for use in POCT.

Electrochemical Determination of Immobilization Technique for Glucose Sensor Fabrication (포도당 센서의 제작을 위한 고정화 방법의 전기화학적 결정)

  • 정태훈;홍석인;노봉수;정용섭;윤정원;김태진
    • KSBB Journal
    • /
    • v.13 no.1
    • /
    • pp.52-57
    • /
    • 1998
  • The present work proposes a simple electrochemical method applicable to any immobilization processes of oxidase using a Clark type oxygen electrode as a base transducer. The present work suggests an optimal immobilization technique among three different methods of glucose oxidase(GOD) onto one side of $37[\mu}$mthick blend membranes, composed o 80% of cellulose triacetate and 20% of polycaprolactone, on the basis of the maximum Michaelis-Menten parameter(Vm) determined by either steady state or transient analyses. The electrode system was made of disk type gold cathode(4mm diameter) and Ag/AgCl anode. One side of the blend membrane was in contact with the cathode surface while the other side was immobilized with GOD either in covalent-bond or cross-linked forms, the latter being covered by $25{\mu}$m thick dialysis membrane of cellulose acetate. The resultant current density was on-line monitored by a potentiostat while glucose level was varied from 1 to 20 mM. The present study shows that direct cross-linking of GOD with glutaraldehyde was mostly preferred for fabrication of glucose sensor, on the basis of resultant kinetic parameters from either steady state or transient analyses.

  • PDF

Application of Acacia as an Alternative to Horseradish Peroxidase for the Determination of Hydrogen Peroxide (과산화수소 정량을 위한 서양고추냉이 과산화효소 대용 아카시아의 활용)

  • Yoon, Kil Joong
    • Applied Chemistry for Engineering
    • /
    • v.28 no.3
    • /
    • pp.369-374
    • /
    • 2017
  • The curtailment of production cost is important for the mass production of biosensors. Since horseradish peroxidase, which is a key material of enzyme electrodes for hydrogen peroxide analysis is rather expensive, this has been a limiting factor for fabricating carbon paste based enzyme electrodes. In this paper, the acacia leaf tissue as a zymogen easily obtainable in our living environment was used as an alternative to horseradish peroxidase for developing a hydrogen peroxide sensor and the electrochemical properties were evaluated. Ten or more electrochemical parameters alongside the other experimental results acquired by the potentiostatic method demonstrated that our enzyme electrodes can be used for the quantitative analysis of hydrogen peroxide. This also indicates that acacia leaves can take the place of the marketed peroxidase.

Disposable Electrochemical Immunosensors for the Detection of Herbicide (제초제 검출을 위한 전기화학적 일회용 면역센서)

  • Chang, Seung-Cheol
    • Journal of Sensor Science and Technology
    • /
    • v.20 no.1
    • /
    • pp.35-39
    • /
    • 2011
  • A disposable electrochemical immunosensor system has been developed for the detection of herbicide in aqueous samples. Disposable screen printed carbon electrodes(SPCE) were used as basic electrodes and an enzyme, horseradish peroxidase (HRP), and anti-herbicide antibodies was immobilised on to the working electrode of SPCE by using avidin-biotin coupling reactions. An herbicide-glucose oxidase conjugates have been used for the competitive immunoreaction with sample herbicides. The enzymatic reaction between the conjugated glucose oxidase and glucose added generates hydrogen peroxide, which was reduced by the peroxidase immobilised. The latter process caused an electrical current change, due to direct re-reduction of peroxidase by a direct electron transfer mechanism, which was measured to determine the herbicides in the sample. The optimal operational condition was found to be: $20\;{\mu}gl-1$ deglycosylated avidin loading to the working electrode and working potential +50 mV vs. Ag/AgCl. The total assay time was 15 min after sample addition. The detection limits for herbicides, atrazine and simazine, were found to be 3 ppb and 10 ppb, respectively.

Sensitive and Selective Electrochemical Glucose Biosensor Based on a Carbon Nanotube Electronic Film (탄소나노튜브 전자 필름을 이용한 고감도-고선택성 전기화학 글루코스 센서)

  • Lee, Seung-Woo;Lee, Dongwook;Seo, Byeong-Gwuan
    • Applied Chemistry for Engineering
    • /
    • v.33 no.2
    • /
    • pp.188-194
    • /
    • 2022
  • This work presents a non-destructive and straightforward approach to assemble a large-scale conductive electronic film made of a pre-treated single-walled carbon nanotube (SWCNT) solution. For effective electron transfer between the immobilized enzyme and SWCNT electronic film, we optimized the pre-treatment step of SWCNT with p-terphenyl-4,4"-dithiol and dithiothreitol. Glucose oxidase (GOx, a model enzyme in this study) was immobilized on the SWCNT electronic film following the positively charged polyelectrolyte layer deposition. The glucose detection was realized through effective electron transfer between the immobilized GOx and SWCNT electronic film at the negative potential value (-0.45 V vs. Ag/AgCl). The SWCNT electronic film-based glucose biosensor exhibited a sensitivity of 98 ㎂/mM·cm2. In addition, the SWCNT electronic film biosensor showed the excellent selectivity (less than 4 % change) against a variety of redox-active interfering substances, such as ascorbic acid, uric acid, dopamine, and acetaminophen, by avoiding co-oxidation of the interfering substances at the negative potential value.

Electrochemical Detection of Pesticide in Living Plant and Fish Brain Cell

  • Lee, Chang-Hyun;Ly, Suw-Young
    • Journal of Environmental Science International
    • /
    • v.19 no.8
    • /
    • pp.941-949
    • /
    • 2010
  • The three electrode system was used to detect the pesticide fenitrothion ($C_9H_{12}NO_5PS$. MW=277.24) using cyclic voltammetry (CV) and square wave anodic stripping voltammetry (SWASV). The working electrode was mercury immobilized on a carbon nanotube paste electrode (Hg-CNTPE). At the optimized condition, the limit of detection (LoD) was 0.6 ppt ($2.16{\times}10^{-12}\;M$), and the relative standard deviation was 0.035% (n=15). And there is more sensitive in detecting fenitrothion than common type carbon nanotube paste electrode. When it was implanted into the brain of live fish (carp), the existence of fenitrothion was measured without any destruction or damage of tissue.

Structural and Electrical Properties of Nanotube as Various Second Anodizing Time for Biosensor (바이오 센서로의 응용을 위한 2차 양극산화 시간에 따른 나노튜브의 구조적, 전기적 특성)

  • Kim, Yong-Jun;Lee, Tae-Ho;Jung, Hye-Rin;Lee, Sung-Gap
    • Journal of the Korean Institute of Electrical and Electronic Material Engineers
    • /
    • v.26 no.10
    • /
    • pp.741-744
    • /
    • 2013
  • In this study, we fabricated anodic aluminum oxide (AAO) membrane by two step anodizing process for pH detection. The structural properties were observed by X-ray diffraction (XRD) and field emission scanning electron microscope (FE-SEM). Electrochemical measurements of the pH sensor have been performed in capacitance-voltage (C-V) and drift rates. The characterization of AAO membrane exhibited high sensitivity (99.1 mV/pH) at second anodizing time of 4 min.

Electrocatalytic Properties of Metal-dispersed Carbon Paste Electrodes for Reagentless L-lactate Biosensors (금속이 첨가된 탄소전극의 전기화학적 특성과 이를 이용한 L-lactate 바이오센서의 개발)

  • 윤현철;김학성
    • KSBB Journal
    • /
    • v.11 no.4
    • /
    • pp.489-496
    • /
    • 1996
  • Metal dispersed carbon paste electrodes were fabricated, and their electrochemical properties were investigated. Among various metal dispersed carbons, platinum-dispersed carbon paste electrode showed most efficient electrocatalytic characteristics. The overpotential for the oxidation of NADH was significantly lowered in the platinum-dispersed carbon paste electrode, and catalytic current was also enhanced. Based on these electrocatalytic observations, L-lactate biosensor using L-lactate dehydrogenase was constructed to evaluate its performance in terms of sensitivity and stability.

  • PDF

Voltammetric Assay of Silver Ions in Frog's Tissue

  • Ly, Suw-Young;Lee, Jin-Hui;Lee, Chang-Hyun
    • Journal of the Korean Applied Science and Technology
    • /
    • v.30 no.1
    • /
    • pp.139-145
    • /
    • 2013
  • The electrochemical analysis of silver ion was performed using cyclic voltammetry (CV) and square-wave (SW) stripping voltammetry, and electrode cell systems were fabricated with graphite pencil electrode (GE) of working, reference and counter electrodes. Also electrolyte was the use of sea water as electrolyte solutions instead of ionic controlled solutions. The optimum analytical conditions for the cyclic and stripping parameters were determined using GE. The results approached the microgram working ranges of SW(ug/L) and CV(ug/L) Ag, and the optimum conditions were applied to frog's tissue and the food samples.

Development of Three-dimensional Chamber-type Glucose Sensor Using Micromachining Technology (마이크로머시닝 기술을 이용한 3차원 마이크로 챔버형 글루코스 센서의 개발)

  • Kim Sung Ho;Kim Chang Kyo
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.6 no.1
    • /
    • pp.24-28
    • /
    • 2005
  • A micromachined biochip with a three dimensional silicon chamber was developed for the construction of biosensors. Anisotropic etching was used fur the formation of the chamber on the p-type silicon wafer(100) and then was glued to the Pyrex glass bottom-substrate with pre-deposited platinum electrode. The electrochemical characterization of its Pt electrode and Ag/AgCl reference electrode was investigated.

  • PDF