Average Glandular Dose In Mammography

  • Kim, K.H.;Ryu, Y.C.;Oh, C.H.
    • 대한전기학회:학술대회논문집
    • /
    • 대한전기학회 2004년도 학술대회 논문집 정보 및 제어부문
    • /
    • pp.319-321
    • /
    • 2004
  • The average glandular dose (AGD) is determined by the breast entrance skin exposure, x-ray tube target material, beam quality (half-value layer), breast thickness, and breast composition. Almost breast cancer always arises in glandular breast tissue. As a result, the average radiation absorbed dose to glandular tissue is the preferred measure of the radiation risk associated with mammography. If the normalized average glandular dose is known, the average glandular dose can be computed from the product of the normalized average glandular dose and breast entrance skin exposure. In this study, AGD was calculated by the breast thickness and various x-ray energy (HVL) in 50% glandular 50% adipose breast by Mo.-Rh. assembly. AGD is 84 mrad in compressed 5 cm breast. These results show that as increasing the breast thickness, dose also increases. But as increasing the x-ray tube voltage, dose decreases because of high penetrating ratio through the object. But high tube voltage is reducing the subject contrast. From this result, we have to consider the trade-off between subject contrast of image and dose to the patient and choose proper x-ray energy range.

  • PDF

디지털 유방확대촬영술에서 Air gap technique의 유용성 평가 (Evaluation of the Usefulness for Air Gap Technique in Digital Magnification Mammography)

  • 김미영
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제37권2호
    • /
    • pp.101-107
    • /
    • 2014
  • 이 연구는 디지털 유방확대촬영술에서 방사선 피폭을 감소시키며 영상의 질을 유지하기 위한 최적의 촬영조건을 알아보았다. 자동노출방식의 평균유선선량이 수동노출방식에 비해 높은 것으로 나타났다. 자동노출방식은 그리드(grid) 사용 여부에 따라 평균유선선량과 화질평가에 차이가 많았으나, 수동노출방식은 팬텀영상평가 점수에서만 차이가 있었다. 그러나 air gap technique을 사용할 경우 모든 노출방식에서 신호대잡음비의 증가가 뚜렷하게 나타났다. 따라서 유방확대촬영 시 air gap technique을 사용한다면 환자의 평균유선선량을 감소시키고 신호대잡음비를 높여 양질의 화질을 유지할 수 있을 것이다.

유방촬영검사에서 평균유선선량과 장치의 품질관리 실태 (The Average Glandular Dose in Mammography and Quality Control of the Equipment Status)

  • 정홍량;황수련;하본철
    • 한국방사선학회논문지
    • /
    • 제5권3호
    • /
    • pp.111-120
    • /
    • 2011
  • 본 연구는 39개 1차 의료기관을 대상으로 유방촬영 장치의 정도관리 상태에서 실제 방사선 피폭선량을 측정하였다. 유방촬영 시 1차 의료기관의 방사선 피폭선량의 최적화와 진단참고준위값을 제시하고자 연구한 결과는 다음과 같다. 첫째, 영상의학과와 일반의원으로 구분하여 T-검정분석 결과 mAs는 영상의학과 80.16 mAs, 일반의원 77.22 mAs로 평균 78.58 mAs로 측정되었으며, 공기커마율은 영상의학과 8.94 mGy/mR, 일반의원 6.66mGy/mR로 평균 7.71 mGy/mR 측정되었다. 반가층은 영상의학과 0.40 mmAl, 일반의원 0.43 mmAl로 평균 반가층은 평균 0.42mmAl 측정되었고, 평균유선선량은 영상의학과 1.09 mGy 일반의원 1.19 mGy로 1.14 mGy로 측정되었다. 둘째, mAs, 반가층, 현상방법, SID를 두 그룹으로 구분하여 평균치를 T-검정 결과 SID에서 유의한 차이가 있는 것으로 나타났고(P<0.05), 영상의학과의 현상방법에서 현상기 사용 1.00 mGy, CR은 1.17 mGy로 mAs에서 유의한 차이가 있는 것으로 나타났다(P<0.05). 셋째, 항목별 상관분석에서는 평균유선선량에서 영상의학과는 mAs가 일반의원은 SID에서는 상관관계의 유의성이 있는 것으로 나타났다(P<0.05), 넷째, 항목별 회귀분석 결과 평균유선선량에서 mAs가 미치는 영향력은 22.7%, SID가 21.7%로 통계적으로 의미가 있었으며(P<0.05), 영상의학과의 평균유선선량에서 mAs가 미치는 영향력은 29.0%, 일반의원은 평균유선선량에서 SID가 미치는 영향력은 29.1%로 가장 많은 영향력이 있는 것으로 나타났다.

초점/필터에 따른 유방 압박 두께와 평균 유선 선량의 관계 (Relationship of Compressed Breast Thickness and Average Glandular Dose According to Focus/Filter)

  • 이인자
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제32권3호
    • /
    • pp.261-270
    • /
    • 2009
  • 2007년 7월 1일부터 2008년 4월 30일까지 10개월간 대학병원에 내원한 1,969명의 CC촬영건수 3,900건에 대한 유방 압박($13{\sim}15\;daN$)시 두께와 평균 유선 선량(AGD)의 관계를 분석 해본 결과 다음과 같은 결론을 얻었다. 1. 전체 3,900건에 대해 연령 40대와 50대가 전체의 68.69%로 2.679건이었다. 2. 초점/필터 물질별 분포를 보면 41.0%가 Mo/Mo이었고, 34.8%는 Mo/Rh, 24.2%는 Rh/Rh이었다. 3. 초점/필터 물질 별 유방 압박 두께는 Mo/Mo 시 평균 두께 26.9 mm이며, Mo/Rh 시 38.84 mm, Rh/Rh시 48.80 mm로 전체 건수에 대한 평균 두께는 36.27 mm로 나타났다. 4. 초점/필터 물질에 따른 평균 유선 선량(AGD)은 Mo/Mo 시 평균 1.27 mGy, Mo/Rh 시 1.55 mGy, Rh/Rh 시 1.42 mGy로 전체 건수에 대한 평균 유선 선량의 평균은 1.43 mGy로 나타났다. 5. Mo/Mo 시 유방 압박 두께에 따른 평균 유선 선량의 관계는 y = 0.0318x + 0.470이며, Mo/Rh에서는 y = 0.0206x + 0.709, Rh/Rh에서는 y = 0.0248x + 0.335로 유방 압박 두께에 따라서 영향을 많이 받지만, 특히 Mo/Mo 시 유방 두께에 따라서 변화가 컸다.

  • PDF

유방촬영의 화질관리 환경에 대한 고찰 (A Study of Quality Control Environment of Mammography)

  • 황인선;김영근;주형윤
    • 대한디지털의료영상학회논문지
    • /
    • 제12권2호
    • /
    • pp.133-138
    • /
    • 2010
  • The purpose of this study was to survey and test quality control of mammography system. The conclusion of this study is as follows ; First, The rate of pass for phantom image test shows that Film-Screen mammography system(F/S) and computed mammography system(CR) is 80%, Indirect digital mammography system(DR) is 100%. Second, The test of exposure dose shows that F/S is 921 mR. CR is 1,140 mR, DR is 474 mR. The grade of this testament is CR > F/S > DR. Third, The test of average glandular dose shows that F/S is 1,336 mGy, CR is 1,635 mGy, DR is 1,26 mGy. The grade of this testament is CR > F/S > DR. Fourth, The testament of resolution shows as follows F/S is 11~13 Lp/mm, CR is 4~5 Lp/mm, and DR is 5~7 Lp/mm(F/S > DR > CR) Fifth, The survey of projection, cassette, development and reading shows that user are indifference.

  • PDF

가임기 여성의 유방촬영에서 Dose mode 적용의 유용성 분석 (The Usefulness Analysis of Applying Dose Mode in Mammography Women of Childbearing Age)

  • 이소라;손순룡;정재연;이관섭
    • 대한디지털의료영상학회논문지
    • /
    • 제15권2호
    • /
    • pp.19-23
    • /
    • 2013
  • Purpose : Automatic exposure system(AOP mode) in DR Mammography of the STD and DOSE apply to women of childbearing age, the average glandular dose(AGD) and image quality by comparing was to demonstrate the usefulness of DOSE. Materials and Methods : Of the under 40 age 108 patients who visit to our hospital and examined STD and DOSE mammography from January 2008 to July 2013. AGD was obtained by DICOM header information provided by GE Senographe DS. STD and DOSE images were evaluated with obtained patients, Image J program was compared by calculating the SNR. Results : The average AGD of DOSE mode was 0.99mGy, and which decreased by 19% comparing to that of STD mode which was 1.18mGy. The two the average AGDs indicated statistically significant difference(p< .01). The average SNR of STD was 40.26, DOSE displayed, and to 39.68 in, there was no statistical significance. Results : The average AGD using DOSE mode which is one of DR mammography AOP modes decreased by comparing to that of STD mode showing no difference in image quality. The use of DOSE is considered to be useful.

  • PDF

검진 기관에서의 선별 유방촬영술 시행에 따른 연령 제한의 필요성에 대한 연구 (A Study on the Necessity of an Age Limitation in Screening Mammography)

  • 윤하얀;이춘미;안의경;김용환
    • 대한디지털의료영상학회논문지
    • /
    • 제12권1호
    • /
    • pp.33-41
    • /
    • 2010
  • National Cancer Screening Project and Korean Society of Breast Imaging recommend that breast cancer screening should be performed on those aged 40 and above. Nevertheless, this recommendation is usually ignored by a number of medical institutions. The purpose of this study is to emphasize the necessity of an age limitation in screening mammography. Ten institutions were randomly selected and telephone inquiries about patients' age limitation and internal guidelines were set up. The 3,214 women, who underwent screening mammography through 'GE Senography 2000D' in each hospital, were classified into five groups according to age(from 20s to 40s, at intervals of 5). And then, collected data was analyzed by a radiologist in accordance with ACR-BIRADS(American College of Radiology Breast Imaging Reporting and Data System), through which breast parenchymal density and the results of analysis were categorized in order to predict the sensitivity of mammography. Information about craniocaudal-view mammograms was automatically produced by use of GE Senography 2000D, and the average glandular dose was retrospectively analyzed through the program 'Excel 2007.' Two institutions did not set the age limitation. Other seven institutions internally allowed those who wanted to receive mammography regardless of age. Approximately 99% of those aged 20 to 29 were judged as having the dense breast. In those aged 35 to 39, breast parenchymal density tended to be lower, but the fatty breast to increase. In the case of 'category-zero' that does not need additional tests, the rate of 'heterogeneously dense' and 'extremely dense' reached to 83.1% and 15.1% respectively. Regarding dense breasts, there was no sufficient information for image reading. The glandular dose, applied to 3,214, was 1.47mGy on the average. In those aged 20 to 24 who are sensitive to radiation, the average glandular dose indicated 1.59mGy. Those aged 35 and above showed the lowest value, 1.43mGy. In those aged 35 to 39, the breast tended to change from denseness to fattiness. The average glandular dose was lowest in those aged 35 and above, which suggests that screening mammography should be periodically performed on those aged 35 and above in order that breast cancer may be early detected. On the other hand, in those aged less than 35, it is difficult to analyze mammograms due to the high density of breast parenchyma, and also retakes become frequent. In particular, subjects may be exposed to excessive doses. Accordingly, it should be substituted by breast self-examination or clinical breast examination. In case of need, it is advisable to perform ultrasonography.

  • PDF

맘모그래피의 화질과 평균유선조직선량에 관한 검토 (A Study on the Image Quality of Mammography and the Average Glandular Dose)

  • 이인자;김학성;김성수;허준
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제25권2호
    • /
    • pp.47-55
    • /
    • 2002
  • 유방촬영에 대한 화질과 평균유선조직선량에 대하여 실험한 결과 반가층은 장치마다 큰 차이를 나타내지 않았으며, 자동노출기구를 사용할 경우 정확성에 문제가 있으므로 확인이 필요하다. 유방촬영에 사용되는 자동현상기 또한 꼭 관리가 필요하며, 유방촬영 시 사용되는 증감지/필름에 따라서 팬톰 촬영 식별능에 많은 영향을 받고 있으므로 선택 시 신중을 기해야 한다. 유방두께 4.2 cm에 대한 평균유선조직 선량은 관전압에 따라 달라지며, ACR기준인 3.0 mGy보다 훨씬 적은 $0.26{\sim}1.39\;mGy$로 나타났다. 이상에서 알 수 있듯이 유방촬영에서는 촬영장치의 성능뿐만 아니라 HVL, AEC, 자동현상기 관리, 증감지/필름 등 모두 관리가 되어야 하며, 이로 인해 평균유선조직선량은 더 줄일 수 있을 것이다.

  • PDF

유방 촬영 장치의 국제 규격 Beam Quality 기준에 따른 평균 선량 평가에 관한 연구 (A Study on Radiation Beam Quality Set-up of Mammography Equipment and Average Glandular Dose)

  • 박윤희;박지군
    • 한국방사선학회논문지
    • /
    • 제16권3호
    • /
    • pp.303-308
    • /
    • 2022
  • 유방암의 조기 진단에는 X선을 이용한 유방 촬영이 현재 가장 많이 이용되고 있다. X선 장치의 사용 빈도가 증가함에 따라 유방 X선 촬영에서 기인하는 방사선의 위해해 대한 관심이 고조되고 있는 추세이다. 이에 본 연구에서는 고 대조도 및 고해상력을 요구하는 X선 유방촬영에서 유선의 피폭선량을 측정하기 위하여 International Atomic Energy Agency (IAEA)에서 규정한 검사 기준에 대해 IEC 62220-1-2에서 제시하는 국제 권고안의 beam quality 기준을 바탕으로 평균 유선 선량을 평가하는 가이드라인을 제시 하고자 하였다. 그 결과 4.5 cm PMMA 팬톰의 평균 유선선량 값은 30 kV 범위 이내에서 최대 2.3 mGy였으며, 28 kV를 기준으로 1.19 mGy로 평가되었다.

디지털 유방확대촬영술에서 노출방식에 따른 피폭선량 평가 (The Evaluation of Radiation Dose by Exposure Method in Digital Magnification Mammography)

  • 김미영;김화선
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제35권4호
    • /
    • pp.293-298
    • /
    • 2012
  • 디지털 유방촬영술은 유방의 두께 및 유선조직의 밀도에 따라 노출조건이 자동으로 이루어지는 자동노출제어장치를 사용하고 있는데 이는 피폭선량을 증가시키는 원인이 되기도 한다. 이 연구에서는 디지털 유방확대촬영술에서 영상의 질을 유지하며 불필요한 방사선 피폭을 줄이기 위한 최적의 방사선량 촬영조건을 알아보았다. 자동노출방식을 사용할 경우 가장 좋은 영상화질을 얻을 수 있었으나 평균유선선량 역시 가장 높았다. 수동노출방식의 경우 자동노출방식의 55 % 정도의 mAs 값만으로도 기준 팬텀영상 점수와 신호대잡음비를 얻을 수 있으며 평균유선선량은 55 % 이하로 줄일 수 있었다. 위와 같은 결과를 참고로 하면 디지털 유방확대촬영 시 자동노출방식보다는 수동노출방식을 이용함으로써 불필요한 환자의 피폭선량을 줄일 수 있을 것이라 생각된다.