Background: For proper monitoring of the eye lens dose, an appropriate calibration factor of a dosimeter and information about the mean energies of X-rays are indispensable. The scattered X-ray energy spectra should be well characterized in medical practices where eye lenses of medical staffs might be high. Materials and Methods: Scattered X-ray energy spectra were experimentally derived for three different types of X-ray diagnostic and therapeutic equipment, i.e., the computed tomography (CT) scan, the angiography and the fluoroscopy. A commercially available CdZnTe (CZT) spectrometer with a lead collimator was employed for the measurement of scattered X-rays, which was performed in the usual manner. Results and Discussion: From the obtained energy spectra, the mean energies of the scattered X-rays lied between 40 and 60 keV. This also agreed with that obtained by the conventional half value layer method. Conclusion: The scattered X-rays to which medical workers may be exposed in the region around the eyes were characterized by means of spectrometry. The obtained mean energies of the scattered X-rays were found to match the flat region of the dosimeter response.
When X-rays were projected into a patient, there occured the phenomena such as penetration, absorption and scattering etc. The penetrating rays were recorded on films as X-ray image used for diagnosis but scattered rays caused the radiation hazard both to the patient, specialist and technicians. The soft tissue includes many organs which are sensitive to the radiation and in may occupy $40{\sim}50%$ of body weight. Therefore X-rays should be carefully projected to the patient and it is strongly recommended to analyse the distribution of X-rays, when ever the patient is exposed to X-rays. In this study, the distribution of X-ray according to the thickness, the radiation field and the tube voltages (kVp) in soft tissue, the following results were obtained: 1. Total transmitted rays which kept the step with X-ray tube voltage (kVp) increased in proportion to the increasing of X-ray tube voltage. 2. The scattered ray rate in the total transmitted ray was not significantly found with X-ray tube voltage. 3. The affecting factors of the scattered ray rate in total transmitted ray were shown through the radiation field and the thickness. 4. The dose of scattered ray by the angle was observed more in direction of primary ray ($0^{\circ}$) and back scattering ($160^{\circ}$) than in direction of $90^{\circ}$. 5. The more the distance from phantom to the patient should be less distribution of scattered ray.
본 연구에서는 전리함을 이용하여 엑스선 촬영실 내에서 공간산란선량을 측정하고 산란선의 발생원과 조사 조건에 대한 산란율 의존성을 조사하였다. 아크릴 팬톰의 산란선 발생 인자를 조사하기 위해 관전압(40~140 kV)과 조사면($10{\times}10\;cm^2$, $20{\times}20\;cm^2$, $35{\times}35\;cm^2$)에 대한 산란율 변화를 측정하였다. 조사면이 $35{\times}35\;cm^2$일 때 측방산란율은 3.11~4.5%로 측정되었다. 촬영실 내에서 팬톰, 콜리메이터, 엑스선관, 벽면에서 발생하는 산란선량의 영향을 측정하였는데 전체 관전압에 대해 팬톰의 산란선 발생률은 95.4% 이상이었으며 콜리메이터, 엑스선관, 벽면의 산란선 발생률은 각각 2.6%, 1.3%, 0.7% 이하로 나타났다. 관전압 100 kV, 40 mAs 조사 조건에서 촬영실 내 팬톰으로부터 1 m 거리에서 측정한 공간선량은 최대 2 mR 정도로 나타나 조사조건의 최적화 등의 산란선 경감조치가 필요할 것으로 생각된다.
Scattered-rays are produced from objects in x-ray diagnosis and the quality of x-ray image is influenced by presence of the scattered-ray. Therefore, constant ways and means to suppress and to eliminate the scattered-ray were devised. Authors made an experimental study on the scattered-ray content in transmitted-ray using acryl phantoms and obtained the results as follows: 1. The scattered-ray content in transmitted-ray is not influenced by filtration and tube voltage. 2. Irradiated field size and the thickness of object are a potent influence to scattered-ray content.
Purpose: To compare the leakage and scattered radiation from hand-held dental X-ray unit with radiation from fixed dental X-ray unit. Materials and Methods: For evaluation we used one hand-held dental X-ray unit and Oramatic 558 (Trophy Radiologie, France), a fixed dental X-ray unit. Doses were measured with Unfors Multi-O-Meter 512L at the right and left hand levels of X-ray tube head part for the scattered and leakage radiation when human skull DXTTR III was exposed to both dental X-ray units. And for the leakage radiation only, doses were measured at the immediately right, left, superior and posterior side of the tube head part when air was exposed. Exposure parameters of handheld dental X-ray unit were 70 kVp, 3 mA, 0.1 second, and of fixed X-ray unit 70 kVp, 8 mA, 0.45 second. Results: The mean dose at the hand level when human skull DXTTR III was exposed with portable X-ray unit $6.39{\mu}Gy$, and the mean dose with fixed X-ray unit $3.03{\mu}Gy$ (p<0.001). The mean dose at the immediate side of the tube head part when air was exposed with portable X-ray unit was $2.97{\mu}Gy$ and with fixed X-ray unit the mean dose was $0.68{\mu}Gy$ (p<0.01). Conclusions: The leakage and scattered radiation from hand-held dental radiography was greater than from fixed dental radiography.
When X-radiation passes through the human body; some is transmitted some is truly absorbed, and some is scattered. In diagnostic radiography, scattered radiation can reach the film if no protective measures are taken. This scattered ray increased density which not necessary for image formation. We studied about absorbtion, scattered ray and the way of get rid of scatter ray according to the x-ray tube kilovoltage and obtained results as follow; 1. Absorbtion ray increased proportion to KVP. 2. Scattered ray increased at high KVP and thick object. 3. Secondary radiation of the primary increased at high KVP and thick object. 4. Remove .ate of scattered ray decreased at thick object and increase at low KVP make use of 6:1 grid ratio
본 연구에서는 손, 머리, 복부 등에 대한 X선 촬영 시행 시 조사야 크기를 최적화할 경우와 최대화할 경우 검사목적부위로부터 30cm 거리에서의 X선 산란선량이 각각 어느 정도인지를 알아보았다. 그 결과 손, 머리, 복부 등에 대한 X선 산란선량은 첫째, 소인촬영의 경우 조사야 크기를 최적화하였을 때 각각 $0.08{\mu}Sv$, $4.39{\mu}Sv$, $5.56{\mu}Sv$로 나타났고, 조사야 크기를 최대화하였을 때 각각 $0.58{\mu}Sv$, $33.47{\mu}Sv$, $35.93{\mu}Sv$로 나타났으며, 둘째, 성인촬영의 경우 조사야 크기를 최적화하였을 때 각각 $0.40{\mu}Sv$, $14.51{\mu}Sv$, $18.86{\mu}Sv$로 나타났고, 조사야 크기를 최대화하였을 때 각각 $2.78{\mu}Sv$, $107.40{\mu}Sv$, $117.52{\mu}Sv$로 나타났다(P<0.001). 결론적으로, X선 촬영 시 조사야 크기를 필요한 만큼만으로 최대한 줄여주어 최적화시켰을 때에 최대화 시켰을 때보다 피사체 주변의 X선 산란선 발생량은 약 6~7배 정도 감소하였다.
Control of scattered radiation is one of very important factors in the use of medical radiation. In general X-ray exam, the causes, measurement methods, and the kind of detectors of scattered rays within the radiation area are diverse. In this study, the dose of scattered ray was measured by changing the thickness of the polycarbonate phantom and the tube voltage. As a result of measurement of scattered radiation, the results show that the scattered dose significantly(p<.05) increased with growing of thickness of phantom in the tube voltage 40, 50 and 60 kVp(F(p)<.05, R2>64%). As tube voltage increased at all phantom thicknesses, the scattered dose also significantly(p<.05) increased(F(p)<.05, R2>69%). In cases where a significant correlation was shown, the coefficient of determination of more than 60% was shown in regression analysis. The results of this study can be used as data on scattered radiation dose according to the tube voltage and the object thickness in general X-ray imaging exam.
The seattered-ray have a bad effect to the image of x-ray film. Therefore, to obtain the better x-ray image, the most important point is to eliminate the scattered-ray, come from the patient, and the easiest method to eliminate the scattered-ray is to use a proper grid for tube voltages. Authors made an experiment on the scattered-ray contents and the image quality in skull radiography, and obtained the results as follows; 1. The scattered-ray content in skull radiography was 83%, but it could be reduced to 35%-50% by using grid. 2. The image contrast was most increased when the using grid ratio was 8:1. 3. The image quality was not influenced by the scattered-ray contents between 32% and 45%.
본 연구는 진단용 X선 발생장치에서 조사야 외부의 산란선량을 감소시켜 환자의 피폭 선량을 감소시키기 위한 방법이다. 자체 제작한 $150{\times}190mm^2$ 납판을 콜리메이터 하단에 부착함으로써 조사야 이외의 부위에 도달하는 산란선량을 감소시키고자 하였다. 납판을 추가로 삽입 후 X선 관축방향인 X-축의 산란선량을 측정한 결과 26 ~ 36% 감소하는 것으로 나타났으며, 관축의 수직 방향인 Y-축 방향에서는 납판의 사용 유무에 따른 산란선량은 큰 변화를 나타내지 않았다. 이러한 결과는 콜리메이터 Shutter에 의해서 발생하는 산란선 보다는 초점 근방에서 발생하는 초점외 X선에 의한 산란선 영향이 크다는 것을 의미한다. 따라서 기존의 콜리메이터 하단에 추가로 납판을 설치하는 것이 조사야 외부의 X-축 산란선량을 감소시키는 방법으로 판단되어진다.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.