A Study on the Image Quality and the Scattered Ray content in the Skull Radiography

두부(頭部) X선촬영시(線撮影時) 산란선(散亂線)과 화질(畵質)에 관(關)한 검토(檢討)

  • Huh, Joon (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
  • Kim, Chang-Kyun (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
  • Kang, Hong-Seok (Dept. of Radiotechnology, Incheon Junior Nursing College and Allied Health Sciences) ;
  • Lee, Sun-Sook (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University) ;
  • Lee, In-Ja (Dept. of Radiotechnology, Junior College of Public Health and Medical Technology, Korea University)
  • 허준 (고려대학교 보건전문대학 방사선과) ;
  • 김창균 (고려대학교 보건전문대학 방사선과) ;
  • 강홍석 (인천간호보건전문대학 방사선과) ;
  • 이선숙 (고려대학교 보건전문대학 방사선과) ;
  • 이인자 (고려대학교 보건전문대학 방사선과)
  • Published : 1985.12.30

Abstract

The seattered-ray have a bad effect to the image of x-ray film. Therefore, to obtain the better x-ray image, the most important point is to eliminate the scattered-ray, come from the patient, and the easiest method to eliminate the scattered-ray is to use a proper grid for tube voltages. Authors made an experiment on the scattered-ray contents and the image quality in skull radiography, and obtained the results as follows; 1. The scattered-ray content in skull radiography was 83%, but it could be reduced to 35%-50% by using grid. 2. The image contrast was most increased when the using grid ratio was 8:1. 3. The image quality was not influenced by the scattered-ray contents between 32% and 45%.