In this study, the influence degree of the scattering of the line depending on the incident energy of the radiation objective was to evaluate a new method of quantitative methods of PSNR is provided.Value of MSE and PSNR of the DR and the Target is placed at the left when there was no change in CR Target, the change appeared to ttaran MSE and PSNR value in tube voltage. Both CR and DR seen that there is a variation of the MSE and PSNR depending on the tube voltage change Computon showed that the scattered radiation effects. If the CR and DR 80 The changes in the MSE and PSNR experience the symptoms suddenly occur in areas kVp was found to affect the optoelectronic Computon E, and at the same time, the scattered radiation detector Computon Computon by scattering due to the photoelectric effect. CR and DR of the imaging device in the future on the basis of the energy bands of the photoelectric effect of 60 kVp 70 kVp, 80 kVp, 90 kVp, compares the value of the PSNR and MSE 100 kVp in accordance with the change of the tube voltage of the CR and device DR proposes jigil scattering study of the degradation of the line quality is achieved.
The purpose of this study is to investigate the effect of posture changes(Anteroposterior projection, Posteroanterior projection) in the plain abdominal examination on breast dose and to examine its clinical usefulness. This study was used a human body phantom and a glass dosimeter. Glass dosimeters were directly inserted from the center and outside of medial and lateral. In this study, the deep dose was measured in the right breast and the surface dose in the left breast. During the abdominal examination, the central X-ray incident point was perpendicularly incident to the image receptor 5 cm above the iliac crest. The exposure parameters were 82 kVp, 320 mA, 50 ms, x-ray field size 14×17 inch The distance between the center X-ray and the detector was fixed at 110 cm, and only the top two AEC chambers were used. As a result of this study, the medial and lateral side doses of the right breast were 535.73±30.68 μGy and 414.46±33.52 μGy for erect AP, and 145.80±18.52 μGy and 148.76±12.92 μGy in erect PA. The superficial breast dose was 754.00±68.36 μGy on the medial side and 674.06±45.58 μGy on the lateral side in the erect AP, 70.66±7.98 μGy on the medial side, and 86.46±15.35 μGy on the lateral side in the erect PA. There was a statistically significant difference in the difference between the mean values of the medial and lateral side doses in the deep and superficial areas of the breast according to the postural change (p <0.01). As a result of this study, If the abdominal radiography was examined in the PA position, the dose reduction effect was 72.78% on the medial side, 64.10% on the lateral side of the deep breast, 90.62% on the medial side, and 87.17% on the lateral side of the superficial breast compared to the AP position.
Mechanical removals in decayed teeth have been performed using drill and sharp hand instruments. These methods have some disadvantages such as pain, local anesthesia and overextended cavities Therefore chemo-mechanical excavation of dentin carious lesions has been introduced. The purpose of this study was to evaluate the efficacy of traditional mechanical methods using burs and chemo-mechanical methods (Carisolv) of caries dentin. Mechanical caries removal was carried with low speed round bur Chemo-mechanical caries excavation was performed with Carisolv (Medi-team), using the Carisolv hand instruments. The mean time to remove caries with two different methods was evaluated and the data analyzed with SPSS software (ver 11.5) by t-test (p < 0.05). For histomorphometry of caries removal were also carried with mechanical or chemo-mechanical (Carisolv) methods from 20 extracted caries permanent molars. Complete caries removal was verified with a $\#$23 sharp explorers, Caries Detector (Kuraray Co. Japan), and standard apical radiography. 1. Chemo-mechanical method was taken more times than mechanical method (1.5 fold) (p < 0.05) 2. Excavation for caries took more time for molar lesion than premolar lesion, and the least time was taken to remove the caries in incisor lesion (p < 0.05). 3. There were no significant differences to remove the caries between the maxilla and mandible (p > 0.05). 4. The remaining carious dentin was detected after the ckemo-mechanical removal of the carious dentin, and no smear layer were seen after the mechanical and chemo-mechanical removal of the carious dentin.
Purpose : The purpose of this study was to reduce Entrance Surface Dose and maintain image quality by changing Source to Image receptor Distance. And we'd like to compare ESD on this study to DRLs in other contries. Materials and Methods : We used indirect DR system(Definium 8000, General Electric, USA)and phantom(ART-200X, Flukebiomedical, USA),glass dosimeters(GD-352M, Asahi Techno Glass, Japan)for this study. The imagies were obtained throuh 80kVp fixed, and different tube currents using AEC mode in $16{\times}16$(inch) field size and changing Source to Image receptor Distance from 100 cm to 130 cm per 10 cm unit. The phantom with attaching 5 glass dosimeters on abdomonal skin was set at supine and erect position as a anterioposterial projection on detector For measuring Entrance Surface Dose. Image analysis was conducted by histograms of Image J(1.46r) which was given from National Institutes of Health(NIH). Results : Due to inverse square law of distance, the tube currents were increasing 42.6 % in supine position and 32.6 % in erect position according to the change of Source to Image receptor Distance. While Entrance Surface Doses were rapidly decreasing 14.2 % in supine position and 29.4 % in erect position according to the change of Source to Image receptor Distance. As the results of histogram using Image J, pixel mean values from 100 cm to 110 cm, 120 cm and 130 cm were decreasing each 1.4%, 2.5%, 2.7%, 4.5%, 2.2 %, 5.8 % in supine, erect position. While standard deviations from 100 cm to 110 cm, 120 cm and 130 cm were increasing each 1.4 %, 2.5 %, 2.5 %, 4.0 %, 2.0 %, 4.9 % Consequently, there are no significant differences in abdomen images taken. Conclusion: As the results described above, we strongly recommend using long Sourceto Image receptor Distance than 100cm that we have been using. So, we should deliver less Entrance Surface Dose to the patients while maintaining image quality in abdomen radiography.
The purpose of this study is to find the optimal method for clinical application by analyzing image quality and radiation output according to parameter combination when using the Automatic Exposure Control (AEC). The experimental method combines 70, 81 kVp with sensitivity S200, S400, S800 and S1000 of the Automatic Exposure Control for Entrance Surface Dose (ESD), current volume, Signal to Noise Ratio (SNR), Contrast to Noise Ratio (CNR), Time-to-Radiation Dose Curve in abdomen and pelvis. And then, image quality and radiation output stability were evaluated. As a results, Entrance Surface Dose, current volume, Signal to Noise Ratio, Contrast to Noise Ratio decreased as the tube voltage and sensitivity were set higher. In addition, the higher tube voltage and sensitivity, the Time-to-Radiation Dose Curve showed a poor output stability. In conclusion, the higher the combination of tube voltage and sensitivity in the use of Automatic Exposure Control, the more problems can be seen in image quality and stability of the radiation output. Therefore, a relatively low combination of tube voltage and sensitivity showed that the image quality and radiation output stability could be optimized by minimizing the error range that would occur when the detector recognized a combination of parameters.
By accessing the current status of FPD system use in the hospitals located in Seoul and Gyeonggi Province as well as the entrance skin dose and the image quality evaluation realized by C-D Phantom, and the image assessment by the medical professionals regarding the radiography for the extremity, the following results were derived. 1. According to the evaluation made in the actual use of FPD system (12 machines), the grid ratio varied from 8:1 to 13:1, and 6 machines used the grid ratio with 12:1, realizing the largest number. Among the machines, there were 8 machines that allowed a removable grid while 3 machines did use a removable grid (25.0%). 2. When it came to the equipments used for the experiment, it showed that the amount of the entrance skin dose increased from 4.13 times up to 4.79 times with the grid use. 3. The difference in the entrance skin dose depending on the changes in the exposure condition(0.5times or 2.0times) was not significantly different regardless of the patients' thickness. 4. In terms of the image quality depending on C-D Phantom, the grid use was distinguished well. However, the images were well distinguishable as the exposure condition got increased. 5. In the clinical assessment, the grid use was less effective for the Hand PA, which was considered to shoot a thin body part. It was evaluated that the grid use was preferred for the Knee AP, which was shooting for a relatively thick body part. Nonetheless, 3 out of 5 people said that they would not use the grid if the entrance skin dose to reduced.
In this study, we investigated the effects of tube voltage (kVp) and tube current (mAs) on an exposure index (EI) applied to a digital radiography. There used an inverter type digital X-ray generator and an image receptor (IR) utilized a portable wireless detector without a grid. Radiographic images were acquired using a cone pyramid phantom produced using a 3D printer. The X-ray tube voltage was increased from 40 kVp to 120 kVp with 10 kVp increment, and the tube current was increased from 1 mAs to 128 mAs with the twice increment. As a result, kVp had logarithmic relationship with the EI as high $R^2$ value, and mAs had a very high linear relationship too. Also, there was a high correlation between the area dose and the EI of the IR, with the $R^2$ value being 0.76 or more. In conclusion, it had showed that mAs affected the EI linearly, and that it could be advantageous to adjust the easy-to-predict mAs to maintain proper image qualities.
A miniature pinscher and two Maltese dogs were present with dyspnea, and radiography and computed tomography (CT) were performed. All dogs were diagnosed with giant bullous emphysema of the massive, hyperexpanded, and hypoattenuating bulla causing contralateral mediastinal shift. Giant bullous emphysema, the rarest form of bullous lung disease, is defined as a bulla that fills more than 30% of the hemithorax, and multi-detector CT scanning could provide the useful information to confirm a diagnosis of giant bullous emphysema in older dogs. The radiographic and computed tomographic features for giant bullous emphysema were described.
Lim, Jong-Nam;Kim, Hyung-Tae;Kim, Min-Hye;Chon, Kwon Su
Journal of the Korean Society of Radiology
/
v.10
no.6
/
pp.451-456
/
2016
Radiation dose to patient is unavoidable when diagnosis is carried out using X-ray. Radiation diagnosis using dual energy X-ray was examined to verify the possibility of medical applications by SNR and image scoring. The dual energy X-ray was realized by combining together two image plates and filter of 0.5 mm thick Cu or Al. Under one X-ray exposure, contrast enhanced image was obtained using two images of image plates. The enhanced image showed higher SNR and image score compared to the first image which was the image recorded with the first image plate. The dual energy X-ray technique would be a very useful method for obtaining higher SNR image and for realizing very low dose, and could be applied to medical applications.
In liquid metal fast breeder reactors, postulated failures of the plant protection system may lead to serious unprotected accidental consequences. Unprotected transients are generically categorized as transient overpower accidents and transient under cooling accidents. In both cases, core meltdown may occur and this can lead to a molten fuel coolant interaction (MFCI). The understanding of MFCI phenomena is essential for study of debris coolability and characteristics during post-accident heat removal. Sodium is used as coolant in liquid metal fast breeder reactors. Viewing inside sodium at elevated temperature is impossible because of its opaqueness. In the present study, a methodology to depict MFCI phenomena using a flat panel detector based imaging system (i.e., real time radiography) is brought out using a woods metal-water experimental facility which simulates the $UO_2-Na$ interaction. The developed imaging system can capture attributes of the MFCI process like jet breakup length, jet front velocity, fragmented particle size, and a profile of the debris bed using digital image processing methods like image filtering, segmentation, and edge detection. This paper describes the MFCI process and developed imaging methodology to capture MFCI attributes which are directly related to the safe aspects of a sodium fast reactor.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.