• Title/Summary/Keyword: Implant fracture

Search Result 307, Processing Time 0.027 seconds

Effect of force during stumbling of the femur fracture with a different ce-mented total hip prosthesis

  • El Sallah, Zagane Mohammed;Ali, Benouis;Abderahmen, Sahli
    • Biomaterials and Biomechanics in Bioengineering
    • /
    • v.5 no.1
    • /
    • pp.11-23
    • /
    • 2020
  • Total hip prosthesis is used for the patients who have hip fracture and are unable to recover naturally. To de-sign highly durable prostheses one has to take into account the natural processes occurring in the bone. Finite element analysis is a computer based numerical analysis method which can be used to calculate the response of a model to a set of well-defined boundary conditions. In this paper, the static load analysis is based, by se-lecting the peak load during the stumbling activity. Two different implant materials have been selected to study appropriate material. The results showed the difference of maximum von Misses stress and detected the frac-ture of the femur shaft for different model (Charnley and Osteal) implant with the extended finite element method (XFEM), and after the results of the numerical simulation of XFEM for different was used in deter-mining the stress intensity factors (SIF) to identify the crack behavior implant materials for different crack length. It has been shown that the maximum stress intensity factors were observed in the model of Charnley.

Comparison of Absorbable Mesh Plate versus Titanium-Dynamic Mesh Plate in Reconstruction of Blow-Out Fracture: An Analysis of Long-Term Outcomes

  • Baek, Woon Il;Kim, Han Koo;Kim, Woo Seob;Bae, Tae Hui
    • Archives of Plastic Surgery
    • /
    • v.41 no.4
    • /
    • pp.355-361
    • /
    • 2014
  • Background A blow-out fracture is one of the most common facial injuries in midface trauma. Orbital wall reconstruction is extremely important because it can cause various functional and aesthetic sequelae. Although many materials are available, there are no uniformly accepted guidelines regarding material selection for orbital wall reconstruction. Methods From January 2007 to August 2012, a total of 78 patients with blow-out fractures were analyzed. 36 patients received absorbable mesh plates, and 42 patients received titanium-dynamic mesh plates. Both groups were retrospectively evaluated for therapeutic efficacy and safety according to the incidence of three different complications: enophthalmos, extraocular movement impairment, and diplopia. Results For all groups (inferior wall fracture group, medial wall fractrue group, and combined inferomedial wall fracture group), there were improvements in the incidence of each complication regardless of implant types. Moreover, a significant improvement of enophthalmos occurred for both types of implants in group 1 (inferior wall fracture group). However, we found no statistically significant differences of efficacy or complication rate in every groups between both implant types. Conclusions Both types of implants showed good results without significant differences in long-term follow up, even though we expected the higher recurrent enophthalmos rate in patients with absorbable plate. In conclusion, both types seem to be equally effective and safe for orbital wall reconstruction. In particular, both implant types significantly improve the incidence of enophthalmos in cases of inferior orbital wall fractures.

THE TRANSMANDIBULAR IMPLANT SYSTEM (The Transmandibular Implant System)

  • Moon, Hong-Seok
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.35 no.3
    • /
    • pp.435-444
    • /
    • 1997
  • Many of the problems which are faced to the edentulous patients are related to a minimal amount of available mandibular bone volume and height. Most of the patients with mild atrophy of the mandible are treated using endosseous implant prosthodontics. TMI(Trans-mandibular Implant) can be used in case of severe mandibular atrophy, poor bone quality, atrophy of the mandible with exposed of inferior alveolar nerve, osteoporosis and the fracture of the atrophic mandible. Also it can eliminate the need for bone grafting and vestibuloplasty. The TMI is a rigid box frame structure which controls and distribute the masticatory force over the severely resorbed mandible. The box frame structure consist of a superstructure, baseplate, 4 transosseous posts, and 5 cortical screws. This is a case report that also describes about the transmandibular implant reconstruction system.

  • PDF

Removal of fractured implant screws: case report (파절된 임플란트 나사의 제거: 증례보고)

  • Kim, Tae-Su;Lee, Jae-Hyun;Lee, Won-Sup;Lee, Su-Young
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.31 no.1
    • /
    • pp.60-66
    • /
    • 2015
  • Screw loosening and screw fracture of abutment is one of most frequent mechanical complications in implant restoration. Fractured fragments in implant restoration like abutment and screw should be completely removed and the procedure needs minimal damage to the fixture of implant. In some cases, it could fail to remove fractured fragments and cause a lot of damage to the fixture of implant. These situations could render implant unusable at the worst. This article describes three different situations and simple techniques for successful removal of fractured fragments without damage of implants. The procedures used are described in this clinical report.

Fatigue Strength of Dental Implant in Simulated Body Environments and Suggestion for Enhancing Fatigue Life (생체유사환경 하의 치과용 임플란트의 피로강도 평가 및 수명 향상법)

  • Kim, Min Gun
    • Transactions of the Korean Society of Mechanical Engineers A
    • /
    • v.38 no.3
    • /
    • pp.259-267
    • /
    • 2014
  • Fatigue tests were performed in various simulated body environments reflecting various factors (such as body fluids, artificial saliva) relevant within a living body. First, the fatigue limit under a simulated body environment (artificial saliva) was evaluated and the governing factors of implant fatigue strength were looked into by observing the fracture mode. The fatigue life of an implant decreased in the artificial saliva environment compared with that in the ringer environment. Furthermore, in the artificial saliva environment, the implant fracture mode was fatigue failure of fixture as opposed to the abutment screw mode in the ringer environment. In the fatigue test, corrosion products were observed on the implant in the simulated body environment. A larger amount of corrosion products were generated on the artificial saliva specimen than on the ringer specimen. It is thought that the stronger corrosion activity on the artificial saliva specimen as compared with that on the ringer specimen led to an overall decrease of fatigue life of the former specimen. In the case of the implant with a nitrided abutment screw eliminated hardened layer (TixN), a several times increase in fatigue life is achieved in comparison with tungsten carbide-coated implants.

Three-Dimensional Pre-Bent Titanium Implant for Concomitant Orbital Floor and Medial Wall Fractures in an East Asian Population

  • Lee, Kyung Min;Park, Ji Ung;Kwon, Sung Tack;Kim, Suk Wha;Jeong, Eui Cheol
    • Archives of Plastic Surgery
    • /
    • v.41 no.5
    • /
    • pp.480-485
    • /
    • 2014
  • Background The objective of this article is to evaluate clinical outcomes of combined orbital floor and medial wall fracture repair using a three-dimensional pre-bent titanium implant in an East Asian population. Methods Clinical and radiologic data were analyzed for 11 patients with concomitant orbital floor and medial wall fractures. A combined transcaruncular and inferior fornix approach with lateral canthotomy was used for the exposure of fractures. An appropriate three-dimensional preformed titanium implant was selected and inserted according to the characteristics of a given defect. Results Follow-up time ranged from 2 to 6 months (median, 4.07 months). All patients had a successful treatment outcome without any complications. Clinically significant enophthalmos was not observed after treatment. Conclusions Three-dimensional pre-bent titanium implants are appropriate for use in the East Asian population, with a high success rate of anatomic restoration of the orbital volume and prevention of enophthalmos in combined orbital floor and medial wall fracture cases.

Implant Therapy and Temporomandibular Disorder

  • Kim, Young-Kyun;Yun, Pil-Young
    • Journal of Korean Dental Science
    • /
    • v.2 no.1
    • /
    • pp.4-10
    • /
    • 2009
  • Though implant treatment is considered as a common treatment option for edentulous patients, there have been few studies on the temporomandibular disorder (TMD) related with implant treatment. The purposes of this study were to evaluate the relevance of TMD to the implant patients and to evaluate the risk factors of TMD in relation with implant treatment. For the evaluation of various risk factors of TMD in relation with implant therapy, clinical evaluation focused on patient factors and implant factors. From a group of 694 patients, 25 patients (3.6 %) were included in this study. The majority of the patients were included in the asymptomatic 'adaptive' group. Parafunction was detected in 11 patients, 8 patients were male. Four patients having parafunction showed complications such as implant failure or fracture of the suprastructure. From the results, it is possible that TMJ related symptoms are developed or aggravated after implant therapy, which requires relatively more chair time; so TMJ examination should be included in the pre-operative evaluation for dental implant patients. Also, it is important to treat patients after they sign an informed consent that includes a detailed explanation on the possibility of TMD during treatment.

  • PDF

Influence of preparation depths on the fracture load of customized zirconia abutments with titanium insert

  • Joo, Han-Sung;Yang, Hong-So;Park, Sang-Won;Kim, Hyun-Seung;Yun, Kwi-Dug;Ji, Min-Kyung;Lim, Hyun-Pil
    • The Journal of Advanced Prosthodontics
    • /
    • v.7 no.3
    • /
    • pp.183-190
    • /
    • 2015
  • PURPOSE. This study evaluated the fracture load of customized zirconia abutments with titanium insert according to preparation depths, with or without 5-year artificial aging. MATERIALS AND METHODS. Thirty-six identical lithium disilicate crowns (IPS e.max press) were fabricated to replace a maxillary right central incisor and cemented to the customized zirconia abutment with titanium insert on a $4.5{\times}10$ mm titanium fixture. Abutments were fabricated with 3 preparation depths (0.5 mm, 0.7 mm, and 0.9 mm). Half of the samples were then processed using thermocycling (temperature: $5-55^{\circ}C$, dwelling time: 120s) and chewing simulation (1,200,000 cycles, 49 N load). All specimens were classified into 6 groups depending on the preparation depth and artificial aging (non-artificial aging groups: N5, N7, N9; artificial aging groups: A5, A7, A9). Static load was applied at 135 degrees to the implant axis in a universal testing machine. Statistical analyses of the results were performed using 1-way ANOVA, 2-way ANOVA, independent t-test and multiple linear regression. RESULTS. The fracture loads were $539.28{\pm}63.11$ N (N5), $406.56{\pm}28.94$ N (N7), $366.66{\pm}30.19$ N (N9), $392.61{\pm}50.57$ N (A5), $317.94{\pm}30.05$ N (A7), and $292.74{\pm}37.15$ N (A9). The fracture load of group N5 was significantly higher than those of group N7 and N9 (P<.017). Consequently, the fracture load of group A5 was also significantly higher than those of group A7 and A9 (P<.05). After artificial aging, the fracture load was significantly decreased in all groups with various preparation depths (P<.05). CONCLUSION. The fracture load of a single anterior implant restored with lithium disilicate crown on zirconia abutment with titanium insert differed depending on the preparation depths. After 5-year artificial aging, the fracture loads of all preparation groups decreased significantly.

Bone Cement-Augmented Short Segment Fixation with Percutaneous Screws for Thoracolumbar Burst Fractures Accompanied by Severe Osteoporosis

  • Jung, Hyun Jin;Kim, Seok Won;Ju, Chang Il;Kim, Sung Hoon;Kim, Hyen Sung
    • Journal of Korean Neurosurgical Society
    • /
    • v.52 no.4
    • /
    • pp.353-358
    • /
    • 2012
  • Objective : The purpose of this study was to determine the efficacy of bone cement-augmented short segment fixation using percutaneous screws for thoracolumbar burst fractures in a background of severe osteoporosis. Methods : Sixteen patients with a single-level thoracolumbar burst fracture (T11-L2) accompanying severe osteoporosis treated from January 2008 to November 2009 were prospectively analyzed. Surgical procedures included postural reduction for 3 days and bone cement augmented percutaneous screw fixation at the fracture level and at adjacent levels without bone fusion. Due to the possibility of implant failure, patients underwent implant removal 12 months after screw fixation. Imaging and clinical findings, including involved vertebral levels, local kyphosis, canal encroachment, and complications were analyzed. Results : Prior to surgery, mean pain score (visual analogue scale) was 8.2 and this decreased to a mean of 2.2 at 12 months after screw fixation. None of the patients complained of pain worsening during the 6 months following implant removal. The percentage of canal compromise at the fractured level improved from a mean of 41.0% to 18.4% at 12 months after surgery. Mean kyphotic angle was improved significantly from $19.8^{\circ}$ before surgery to 7.8 at 12 months after screw fixation. Canal compromise and kyphotic angle improvements were maintained at 6 months after implant removal. No significant neurological deterioration or complications occurred after screw removal in any patient. Conclusion : Bone cement augmented short segment fixation using a percutaneous system can be an alternative to the traditional open technique for the management of selected thoracolumbar burst fractures accompanied by severe osteoporosis.

Characteristics of Abutment Screw Structure for Dental Implant (치과용 임플란트 지대주 나사 구조에 관한 연구)

  • Song, Jong-Beop;Choi, Il-kyung;Jung, Hyo-kyung;Kwon, Soon-Hong;Kwon, Soon-Gu;Park, Jong-Min;Kim, Jong-Soon;Jung, Sung-Won;Choi, Won-Sik
    • Journal of the Korean Society of Industry Convergence
    • /
    • v.20 no.2
    • /
    • pp.169-176
    • /
    • 2017
  • Dental implants are required to have biomechanical functions and biostability in order to perform authoring, pronunciation, and aesthetic functions in the oral cavity. In terms of biostability, pure titanium for medical have good biostability and no rejection in the alveolar bone. with appropriate strength in terms of strength as well as biocompatibility. In recent years, various surgical methods and devices have been developed to improve the convenience and safety of the procedure. However, as the number of procedures increases, the screw loosening of the abutment screw connecting the artificial root and the abutment There are many reports of artificial root and abutment fracture. Fig. 1 is an example of a case where the upper part of the abutment screw is arbitrarily modified to remove the abutment by the abutment fracture due to the loosening of the abutment screw. The fundamental cause of abduction of the abutment screw is caused by the slight movement due to the lowering of the retention force of the abutment screw. It is necessary to minimize loosening of the abutment screw to avoid problems such as fracture during the period of using the implant. The purpose of this study is to investigate the structure of the abutment screw to prevent the loosening of the abutment screw by forming 0.5mm slot.