• Title/Summary/Keyword: Image Receptor

Search Result 76, Processing Time 0.023 seconds

Evaluation of Relationship between Radiation Dose and Image Quality according to Source to Image Receptor Distance in Rib Series Radiography (늑골 방사선검사 시 X선관 초점-영상수신체간 거리에 따른 환자선량과 화질의 연관성 평가)

  • Joo, Young-Cheol;Jung, Young-Jin
    • Journal of radiological science and technology
    • /
    • v.41 no.5
    • /
    • pp.391-396
    • /
    • 2018
  • The purpose of this study was to compare and analyze the patient dose according to the distance between the X-ray tube focus and the image receptor, and to propose a new method for quantitatively evaluating the image quality. Using this quantitative evaluation method, the optimal distance for increasing x-ray image quality with low radiation dose was estimated between source and image receptor in Rib series radiography. Phantom images were obtained by changing the distance between focus and image receptor (100 cm and 180 cm). The patient radiation dose was estimated using entrance surface dose and dose area product. In order to evaluate image quality objectively, a non - reference image evaluation method was employed with paper and salt noise and Gaussian filter. As a result of this study, when the SID was changed from 100 cm to 180 cm, the entrance surface dose decreased by 4 ~ 5 times and the dose area product decreased by 3 times. In addition, there is no significant difference in image quality between of SID 180 cm and SID 100 cm. In conclusion, it was demonstrated that performing the rib series radiography at SID 180 cm is an optimal method to reduce the exposure dose and improve the image quality.

A Study on Patients Dose and Image Quality according to Source to Image receptor Distance in Abdomen Radiography: comparison of ESD measured and DRLs in other countries (복부일반촬영시 선원과 검출기간의 거리변화에 따른 영상 화질 및 피폭선량에 관한 연구)

  • Jang, Ji-Sung;Choi, Weon-Keun;Jung, Jae-Yon;Lee, Kwan-Sub;Ha, Dong-Yoon
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.14 no.2
    • /
    • pp.39-46
    • /
    • 2012
  • Purpose : The purpose of this study was to reduce Entrance Surface Dose and maintain image quality by changing Source to Image receptor Distance. And we'd like to compare ESD on this study to DRLs in other contries. Materials and Methods : We used indirect DR system(Definium 8000, General Electric, USA)and phantom(ART-200X, Flukebiomedical, USA),glass dosimeters(GD-352M, Asahi Techno Glass, Japan)for this study. The imagies were obtained throuh 80kVp fixed, and different tube currents using AEC mode in $16{\times}16$(inch) field size and changing Source to Image receptor Distance from 100 cm to 130 cm per 10 cm unit. The phantom with attaching 5 glass dosimeters on abdomonal skin was set at supine and erect position as a anterioposterial projection on detector For measuring Entrance Surface Dose. Image analysis was conducted by histograms of Image J(1.46r) which was given from National Institutes of Health(NIH). Results : Due to inverse square law of distance, the tube currents were increasing 42.6 % in supine position and 32.6 % in erect position according to the change of Source to Image receptor Distance. While Entrance Surface Doses were rapidly decreasing 14.2 % in supine position and 29.4 % in erect position according to the change of Source to Image receptor Distance. As the results of histogram using Image J, pixel mean values from 100 cm to 110 cm, 120 cm and 130 cm were decreasing each 1.4%, 2.5%, 2.7%, 4.5%, 2.2 %, 5.8 % in supine, erect position. While standard deviations from 100 cm to 110 cm, 120 cm and 130 cm were increasing each 1.4 %, 2.5 %, 2.5 %, 4.0 %, 2.0 %, 4.9 % Consequently, there are no significant differences in abdomen images taken. Conclusion: As the results described above, we strongly recommend using long Sourceto Image receptor Distance than 100cm that we have been using. So, we should deliver less Entrance Surface Dose to the patients while maintaining image quality in abdomen radiography.

  • PDF

Assessment of Dose and Image Quality according to the Change of Distance from Source to Image Receptor and the Examination Posture during the Skull Lateral Radiography (두부 측 방향 방사선검사 시 선원 영상수용체간 거리와 검사 자세 변화가 선량과 영상품질에 미치는 영향)

  • Eun-Hye, Kim;Young-Cheol, Joo;Han-Yong, Kim;Dong-Hwan, Kim
    • Journal of radiological science and technology
    • /
    • v.45 no.6
    • /
    • pp.483-489
    • /
    • 2022
  • This study proposes a new skull lateral examination, and provides an improved examination environment for patients and radiologists. The study was divided into three groups. One group was divided into the SID (source to image receptor distance) 110 ㎝ and 180 ㎝ in the skull lateral posture, the other group The other group was divided into an position in contact with the detector and an position without contact with the detector, and the other group was divided into male and female groups, considering that the difference in shoulder width between adult males and females would affect the dose and image quality. For dose evaluation, the ESD (entrance surface dose) was measured at the EAM (external auditory meatus), and the conditions were applied equally at 70 ㎸p, 200 ㎃, and 10 ㎃s. For image quality evaluation, SNR (signal to noise ratio) and CNR (contrast to noise ratio) were measured in frontal sinus, EAM, and sella turcica. As a result of ESD comparison, when sid 110 ㎝ to sid 180 ㎝ was changed among the three groups, ESD values decreased the most to 729.18±4.62 μ㏉ and 224.18±0.74 μ㏉ at 180 ㎝ (p<0.01). The values of SNR and CNR were statistically significant (p<0.01), but there was no qualitative difference. This shows that when the SID is 180 ㎝, it is possible to reduce the dose without lowering the image quality. So, It is suggested that the SID 180 ㎝ is used without contacting the detector when examining the skull lateral.

Evaluation of Image Receptor Characteristics in Computed Radiography System Using Exposure Index in International Electrotechnical Commission (I) (IEC 규정 노출지수를 활용한 디지털 방사선 영상시스템에서의 영상 수용체간 특성평가 (I))

  • Park, Hyemin;Yoon, Yongsu;Roh, Younghoon;Kim, Sungjun;Na, Chanyoung;Han, Taeho;Kim, Jungsu;Jeong, hoiwoun;Kim, Jungmin
    • Journal of radiological science and technology
    • /
    • v.42 no.4
    • /
    • pp.291-299
    • /
    • 2019
  • The International Electrotechnical Commission (IEC) has regulated the definition and requirements of the exposure index (EI). In this study, we calculated the EI of several image receptors in digital radiography system of two different manufacturers according to the method as per IEC, and evaluated the relationship with incident air kerma. To calculate the EI, w e obtained the characteristics curve of each image receptor by increasing the incident air kerma at RQA 3, 5, 7 and 9, respectively. As a result, there was no significant difference in the EI values between different image receptors of the same manufacturer, but EI values of different manufacturer was different despite the same air kerma was incident. Therefore, understanding the characteristics of the digital radiography systems is important in order to use EI as a tool for measuring and managing the radiation dose.

The Effect of Source to Image-Receptor Distance(SID) on Radiation Dose for Digital Chest Radiography (Digital Chest Radiography에서 방사선량에 대한 Source to Image-Receptor Distance (SID)의 영향)

  • Kwon, Soonmu;Park, Changhee;Park, Jeongkyu;Son, Woonheung;Jung, Jaeeun
    • Journal of the Korean Society of Radiology
    • /
    • v.8 no.4
    • /
    • pp.203-210
    • /
    • 2014
  • Chest radiography has been typically performed at SID of 180 cm. Image quality and patient dose were investigated between 180 cm and 340 cm by 20 cm intervals at 120 kVp and 320 mAs with the AEC. VGA was performed for qualitative assessment and SNR was analysed for quantitative assessment on the image of the chest phantom. Patients dose was measured by ESAK and PCXMC was used for effective dose. As a result, when using the standard of SID of 180 cm which is typically used in the clinical practice, in the case of ESAK, 240 cm, 280 cm, and 320 cm were 8.7%, 11.47%, and 13.56% respectively therefore significant reduction was confirmed. In the case of effective dose, 2.89%, 4.67%, and 6.41% in the body and 5.08%, 6.09%, and 9.6% in lung were reduced. In the case of SNR, 9.04%, 8.24%, and 11.46% were respectively decreased especially, by 8.03% between SID of 260 cm and 300 cm, but SNR was 5.24 up to 340 cm. There were no significant differences in VGA thus the image is valuable in diagnosis. It is predicted that increasing SID up to 300 cm in digital chest radiography can reduce patient dose without decreasing image quality.

Radiographic examination protocol and patient dose in lateral cephalometric radiography in Korea (국내 의료기관에서 측방두부규격방사선촬영시 임상에서의 촬영조건 및 환자 선량)

  • Choi, Jin-Woo
    • Imaging Science in Dentistry
    • /
    • v.40 no.4
    • /
    • pp.165-169
    • /
    • 2010
  • Purpose : To survey the radiographic examination protocol for lateral cephalometric radiographic examinations and to measure their patient doses in Korea and to compare the dose according to the size of hospital, the type of image receptor system, and the installation duration. Materials and Methods : The radiographic examination protocols (kVp, mA, and exposure time) for lateral cephalometric radiography were surveyed with 61 cephalometric radiographic equipments and their patient dose-area product (DAP) measured with a DAP meter (DIAMENTOR M4-KDK, PTW, Freiburg, Germany) for 51 cephalometric radiographic equipments. The radiographic examination protocols and patient doses were compared according to the size of hospital (university dental hospital, dental hospital, and dental clinic), the type of image receptor system (film-based, DR and CR type) and the installation duration, respectively. SPSS 12.0.1 for Windows (SPSS Inc., Chicago, USA) was used for independent t-test and ANOVA test. Results : The average protocols were 77.0 kVp, 12.7 mA, 6.2 second for cephalometric radiography. The average patient dose (DAP) was $128.0mGy\;cm^2$ and 3rd quartile dose (DAP) $161.1mGy\;cm^2$ for cephalometric radiography for adult male. There was no statistically significant difference at average patient DAP according to the size of hospital, the type of image receptor system, and the installation duration, repectively. Conclusion : The average patient dose was $128.0mGy\;cm^2$ and the third quartile patient dose $161.1mGy\;cm^2$ for lateral cephalometric radiography for adult male in Korea.

PET Radiopharmaceuticals for Tumor Imaging (종양 영상을 위한 PET 방사성의약품)

  • Choe, Yearn-Seong
    • The Korean Journal of Nuclear Medicine
    • /
    • v.36 no.1
    • /
    • pp.8-18
    • /
    • 2002
  • Early and accurate diagnosis of tumors using positron omission tomography (PET) has been the focus of considerable interest due to its high metastasis and mortality rates at late detection. PET radiopharmaceuticals-which exhibit a high tumor-to-background uptake ratio, and appropriate metabolic characteristics, and pharmacokinetics-are attractive tools for tumor imaging. Tumor imaging by these radiopharmaceuticals are based on metabolic and receptor imaging. The former is based on accelerated metabolism in tumor tissue compared to normal tissue and the rate roughly corresponding to the rate of growth of tumors. Radiopharmaceuticals for this purpose include radiolabeled sugars, amino acids, and nucleosides which detect increased glucose utilization, protein synthesis, and DNA synthesis, respectively. Tumor receptor imaging is based on the proliferation of tumor cells regulated by many hormones and growth factors, which bind to the corresponding receptors and exhibit the biological responses Radiopharmaceuticals used to image the tumor receptor systems may be ligands for the specific receptors and antibodies for the growth factor receptors. Some antitumor agents have been labeled with radionuclides and used to study in vivo biodistribution and pharmacokinetics in humans. This overview describes typical PET radiopharmaceuticals used for tumor imaging based on their uptake mechanisms.

Mast Cell Increase and Stem Cell Factor Receptor (c-kit) Expression in Helicobacter pylori-infected Gastritis (Helicobacter pylori 감염 위염에서의 비만세포 증가와 Stem Cell Factor Receptor (c-kit)의 발현)

  • Jekal, Seung-Joo
    • Korean Journal of Clinical Laboratory Science
    • /
    • v.37 no.1
    • /
    • pp.41-46
    • /
    • 2005
  • It is known that mast cells (MCs) are increased in H. pylori-infected gastritis and its increase is mediated by stem cell factor (c-kit ligand). To determine the mechanism of mast cell recruitment and activation by stem cell factor, weinvestigated the expression of stem cell factor receptor (c-kit) in H. pylori-positive and -negative gastric mucosa. Biopsy specimens from 16 H. pylori-negative and 20 positive subjects were examined. H. pylori infection in gastric mucosa was examined by the Warthin-Starry method. MC and c-kit were identified by immunohistochemisty, using a monoclonal antihuman MC tryptase antibody and a polyclonal anti-human c-kit antibody. Densities of MC and c-kit positive cell were measured by a computerized image analysis system. MCs were detected in the lamina propria of both H. pylori-positive and -negative gastric mucosa. Densities of MC and c-kit positive cell were significantly greater in H. pylori-positive than -negative subjects. c-kit was located on the surface of MCs. These results indicate that stem cell factors may be one of the factors involved in mast cell increase and that they activate mast cells by binding with c-kit.

  • PDF

Growth Rates of Transgenic Mice Containing Growth Hormone Receptor Gene

  • Kim, H. J;K. Naruse;S. M. Chang;K. S. Im;Lee, S. H.;Park, C. S.;D. I. Jin
    • Korean Journal of Animal Reproduction
    • /
    • v.27 no.4
    • /
    • pp.333-338
    • /
    • 2003
  • Transgenic mice containing GH Receptor (GHR) gene fused to metallothionein promoter were analyzed to evaluate effect of GHR expression on growth in vivo. Three founder mice lines contained copies of GHR transgene and transmitted these genes into F$_1$ and F$_2$ progenies. The mRNA expression of transgene was identified using RT-PCR with GHR genes in tissues. To analyze the effects of transgenes on growth performance, body weights of pups were measured at 4, 10 and 14 weeks after birth. The body weight of transgenic mice was higher compared with that of non-transgenic control mice regardless of sex (P<0.05). Body weights between transgenic and non-transgenic mice were increased with aging. Overall, GHR transgenic mice tended to grow about 10 to 15 % faster than non-transgenic mice without any pathological defects.