• Title/Summary/Keyword: Dose verification program

Search Result 27, Processing Time 0.024 seconds

IMRT and IMRS Checking the Dose Distribution in the Small Field Evaluation of Measurement by Changes in SAD (IMRT 및 IMRS에서 Small Field의 선량분포 확인시 SAD 변화에 따른 측정의 유용성 평가)

  • Ko, Seung-Young;Kim, Sung-Joon;Park, Gir-Yong;Son, Mi-Suk;Lee, Nam-Ki;Kim, Jin-Soo
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.22 no.1
    • /
    • pp.33-39
    • /
    • 2010
  • Purpose: It is very important to confirm conformance of dose distribution that is formed with treatment planning from IMRS or IMRT. It has been a problem dropped accuracy and conformance when the field size is getting smaller because of character of the 2D ion chamber. Verification of MatriXX Phantom dose distribution with a change in the SAD. Dose distribution measurement and analysis to improve the accuracy and should be useful to evaluate the award. Materials and Methods: A use of Novalis linear accelerator 6 MV photon beams. In general, IMRS were 25 patients with small field size. The selected patients were divided into three groups on the basis of the field size. SAD was changed from 80 to 130 cm and field size to determine the dose distribution to the change, each dose was measured using MatriXX Phantom. Analysis of measured values obtained from the program for each patient through the treatment planning system comparison and analysis of the dose distribution and gamma values were expressed. Result: SAD 80, 100, and 120 cm in size in the gamma value to the investigation of patients less than $3\;cm^2$ average 0.939, 0.969, and 0.979, respectively. Patients with more than $5\;cm^2$ 0.962, 0.983, and 0.988, respectively. $5\;cm^2$ or more patients 0.982, 0.990, and 0.992, respectively. Conclusion: The error rate of less than $3\;cm^2$ field size is increased rapidly. If the field size is increased, resolution is increased by 2D ion chambers. It has been approved that it can be credible if it is around $3\;cm^2$ when measuring dose distribution using MatriXX. Adjusting geometric field size by changing SAD is likely to be very useful when you measure dose distribution using MatriXX.

  • PDF

Comparison Study of Conventional Film-based and CT-reconstruction method in HDR Brachytherapy (고선량률 근접 방사선 치료에서 기존의 필름 방법과 CT 재구성 방법의 비교 연구)

  • 장지나;이형구;윤세철;서태석
    • Progress in Medical Physics
    • /
    • v.15 no.2
    • /
    • pp.63-69
    • /
    • 2004
  • HDR brachytherapy administers a large dose of radiation in a short time compare with LDR, and its optimization for treatment is related to several complex factors, such as physical, radiation and optimization algorithms, so there is a need for these to be verified for accurate dose delivery. In our approach, a previous study concerning the phantom for dose verification has been modified, and a new pelvic phantom fabricated for the purpose of localization, including a structure enabling the use of a CT or MRI system. In addition, a comparison study was performed to verify an orthogonal method that is commonly used for brachytherapy localization by comparing target coordinates from a CT system. Since the developed phantom was designed to simulate the clinical setups of cervix cancer, it included an air-filled bladder and a rectum structure shaped sphere and cylinder An N-shaped localizer was used to obtain precision coordinates from both CT and films. Moreover, the IDL 5.5 software program for Windows was used to perform coordinates analysis based on an orthogonal algorithm. The film results showed differences within 1.0 mm of the selected target points compare with the CT coordinates. For these results, a Plato planning system (Nucletron, Netherlands) could be independently verified using this phantom and software. Furthermore, the new phantom and software will be efficient and powerful qualify assurance (QA) tools in the field of brachytherapy QA.

  • PDF

Independent Verification Program for High-Dose-Rate Brachytherapy Treatment Plans (고선량률 근접치료계획의 정도보증 프로그램)

  • Han Youngyih;Chu Sung Sil;Huh Seung Jae;Suh Chang-Ok
    • Radiation Oncology Journal
    • /
    • v.21 no.3
    • /
    • pp.238-244
    • /
    • 2003
  • Purpose: The Planning of High-Dose-Rate (HDR) brachytherapy treatments are becoming individualized and more dependent on the treatment planning system. Therefore, computer software has been developed to perform independent point dose calculations with the integration of an isodose distribution curve display into the patient anatomy images. Meterials and Methods: As primary input data, the program takes patients'planning data including the source dwell positions, dwell times and the doses at reference points, computed by an HDR treatment planning system (TPS). Dosimetric calculations were peformed in a $10\times12\times10\;Cm^3$ grid space using the Interstitial Collaborative Working Group (ICWG) formalism and an anisotropy table for the HDR Iridium-192 source. The computed doses at the reference points were automatically compared with the relevant results of the TPS. The MR and simulation film images were then imported and the isodose distributions on the axial, sagittal and coronal planes intersecting the point selected by a user were superimposed on the imported images and then displayed. The accuracy of the software was tested in three benchmark plans peformed by Gamma-Med 12i TPS (MDS Nordion, Germany). Nine patients'plans generated by Plato (Nucletron Corporation, The Netherlands) were verified by the developed software. Results: The absolute doses computed by the developed software agreed with the commercial TPS results within an accuracy of $2.8\%$ in the benchmark plans. The isodose distribution plots showed excellent agreements with the exception of the tip legion of the source's longitudinal axis where a slight deviation was observed. In clinical plans, the secondary dose calculations had, on average, about a $3.4\%$ deviation from the TPS plans. Conclusion: The accurate validation of complicate treatment plans is possible with the developed software and the qualify of the HDR treatment plan can be improved with the isodose display integrated into the patient anatomy information.

A Study on the Verification and Improvement to Locate and Determine the Radioactive Contamination Using a Whole Body Counter (전신계측기를 이용한 원전종사자 방사성오염 위치확인과 내부방사능 측정개선에 관한 연구)

  • Kim, Hee-Geun;Kong, Tae-Young
    • Journal of Radiation Protection and Research
    • /
    • v.34 no.1
    • /
    • pp.37-42
    • /
    • 2009
  • Whole body counters (WBCs) are used to monitor radiation workers for internal contamination of radionuclides at domestic nuclear power plants (NPPs). A WBC is a scintillation detector using sodium iodide (NaI) and provides the identification of inhaled radionuclide and the measurement of its internal radioactivity in a short time. However, it is often possible to estimate external contamination as internal contamination due to radionuclides attached to the skin of radiation workers and this leads to an excessively conservative estimation of radioactive contamination. In this study, several experiments using a WBC and the Korean humanoid phantom were performed to suggest the more systematic method of discrimination between external and internal contamination. Furthermore, a WBC geometry experiment was conducted to suggest the optimal WBC geometry in consideration of deposited areas inside the body for dominant radionuclides at NPPs. The procedure of measurement and estimation of internal radioactivity for radiation workers at NPPs was improved on the basis of experimental results. Thus, it is expected to prevent from estimating internal exposure dose conservatively owing to the application of accurate whole body counting program to NPPs.

Understanding the Response Characteristics of X-ray Verification Film (X-선 Verification 필름의 반응 특성에 관한 연구)

  • Yeo Inhwan;Seong Jinsil;Chu Sung Sil;Kim Gwi Eon;Suh Chang Ok;Burch Sandra E.;Wang Chris K.
    • Radiation Oncology Journal
    • /
    • v.16 no.4
    • /
    • pp.505-515
    • /
    • 1998
  • Purpose : This study is intended to understand the sensitometric characteristics and the emulsion properties of the commercially available CEA TVS film in comparison with the Kodak X-Omat V film. Materials and Methods : For this purpose, we have formulated an analytic expression of the characteristic curves for x-ray film exposed to mixed radiation of electrons, photons, and visible light. This mathematical expression was developed based on reaction-rate and target-hit theories. Unlike previous expressions. it relates optical density to emulsion properties such as grain size and silver bromide content We have also developed a quantity which characterizes the film response to visible light relative to that to photons and electrons. This quantity could be expressed as a function of grain area. Thus, we have developed mathematical expressions and quantities with which the emulsion properties of the films can be revealed based on the sensitometric characteristics. Demonstrating the use of this analytical study, we exposed CEA and Kodak verification films to the mixed radiation of electrons, photons, and visible light, and interpreted the experimental results accordingly. Results : We have demonstrated that: (1) the saturation density increases as the silver bromide content increases, (2) the time required to reach the threshold dose (to which the film begins to respond) when films are exposed to visible light decreases as the grain size increases, and (3) the CEA film contains more silver bromide. whereas the Kodak film contains larger grains. These findings were supported by the data provided by the manufacturers afterward. Conclusion : This study presented an analytical and experimental basis for understanding the response of X-ray film with respect to the emulsion properties.

  • PDF

Dose verification for Gated Volumetric Modulated Arc Therapy according to Respiratory period (호흡연동 용적변조 회전방사선치료에서 호흡주기에 따른 선량전달 정확성 검증)

  • Jeon, Soo Dong;Bae, Sun Myung;Yoon, In Ha;Kang, Tae Young;Baek, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.1
    • /
    • pp.137-147
    • /
    • 2014
  • Purpose : The purpose of this study is to verify the accuracy of dose delivery according to the patient's breathing cycle in Gated Volumetric Modulated Arc Therapy Materials and Methods : TrueBeam STxTM(Varian Medical System, Palo Alto, CA) was used in this experiment. The Computed tomography(CT) images that were acquired with RANDO Phantom(Alderson Research Laboratories Inc. Stamford. CT, USA), using Computerized treatment planning system(Eclipse 10.0, Varian, USA), were used to create VMAT plans using 10MV FFF with 1500 cGy/fx (case 1, 2, 3) and 220 cGy/fx(case 4, 5, 6) of doserate of 1200 MU/min. The regular respiratory period of 1.5, 2.5, 3.5 and 4.5 sec and the patients respiratory period of 2.2 and 3.5 sec were reproduced with the $QUASAR^{TM}$ Respiratory Motion Phantom(Modus Medical Devices Inc), and it was set up to deliver radiation at the phase mode between the ranges of 30 to 70%. The results were measured at respective respiratory conditions by a 2-Dimensional ion chamber array detector(I'mRT Matrixx, IBA Dosimetry, Germany) and a MultiCube Phantom(IBA Dosimetry, Germany), and the Gamma pass rate(3 mm, 3%) were compared by the IMRT analysis program(OmniPro I'mRT system software Version 1.7b, IBA Dosimetry, Germany) Results : The gamma pass rates of Case 1, 2, 3, 4, 5 and 6 were the results of 100.0, 97.6, 98.1, 96.3, 93.0, 94.8% at a regular respiratory period of 1.5 sec and 98.8, 99.5, 97.5, 99.5, 98.3, 99.6% at 2.5 sec, 99.6, 96.6, 97.5, 99.2, 97.8, 99.1% at 3.5 sec and 99.4, 96.3, 97.2, 99.0, 98.0, 99.3% at 4.5 sec, respectively. When a patient's respiration was reproduced, 97.7, 95.4, 96.2, 98.9, 96.2, 98.4% at average respiratory period of 2.2 sec, and 97.3, 97.5, 96.8, 100.0, 99.3, 99.8% at 3.5 sec, respectively. Conclusion : The experiment showed clinically reliable results of a Gamma pass rate of 95% or more when 2.5 sec or more of a regular breathing period and the patient's breathing were reproduced. While it showed the results of 93.0% and 94.8% at a regular breathing period of 1.5 sec of Case 5 and 6, it could be confirmed that the accurate dose delivery could be possible on the most respiratory conditions because based on the results of 100 patients's respiratory period analysis as no one sustained a respiration of 1.5 sec. But, pretreatment dose verification should be precede because we can't exclude the possibility of error occurrence due to extremely short respiratory period, also a training at the simulation and careful monitoring are necessary for a patient to maintain stable breathing. Consequently, more reliable and accurate treatments can be administered.

Examinations on Applications of Manual Calculation Programs on Lung Cancer Radiation Therapy Using Analytical Anisotropic Algorithm (Analytical Anisotropic Algorithm을 사용한 폐암 치료 시 MU 검증 프로그램 적용에 관한 고찰)

  • Kim, Jong-Min;Kim, Dae-Sup;Hong, Dong-Ki;Back, Geum-Mun;Kwak, Jung-Won
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.24 no.1
    • /
    • pp.23-30
    • /
    • 2012
  • Purpose: There was a problem with using MU verification programs for the reasons that there were errors of MU when using MU verification programs based on Pencil Beam Convolution (PBC) Algorithm with radiation treatment plans around lung using Analytical Anisotropic Algorithm (AAA). On this study, we studied the methods that can verify the calculated treatment plans using AAA. Materials and Methods: Using Eclipse treatment planning system (Version 8.9, Varian, USA), for each 57 fields of 7 cases of Lung Stereotactic Body Radiation Therapy (SBRT), we have calculated using PBC and AAA with dose calculation algorithm. By developing MU of established plans, we compared and analyzed with MU of manual calculation programs. We have analyzed relationship between errors and 4 variables such as field size, lung path distance of radiation, Tumor path distance of radiation, effective depth that can affect on errors created from PBC algorithm and AAA using commonly used programs. Results: Errors of PBC algorithm have showned $0.2{\pm}1.0%$ and errors of AAA have showned $3.5{\pm}2.8%$. Moreover, as a result of analyzing 4 variables that can affect on errors, relationship in errors between lung path distance and MU, connection coefficient 0.648 (P=0.000) has been increased and we could calculate MU correction factor that is A.E=L.P 0.00903+0.02048 and as a result of replying for manual calculation program, errors of $3.5{\pm}2.8%$ before the application has been decreased within $0.4{\pm}2.0%$. Conclusion: On this study, we have learned that errors from manual calculation program have been increased as lung path distance of radiation increases and we could verified MU of AAA with a simple method that is called MU correction factor.

  • PDF