Dose verification for Gated Volumetric Modulated Arc Therapy according to Respiratory period  

Jeon, Soo Dong (Department of Radiation Oncology, ASAN Medical Center)
Bae, Sun Myung (Department of Radiation Oncology, ASAN Medical Center)
Yoon, In Ha (Department of Radiation Oncology, ASAN Medical Center)
Kang, Tae Young (Department of Radiation Oncology, ASAN Medical Center)
Baek, Geum Mun (Department of Radiation Oncology, ASAN Medical Center)
Publication Information
The Journal of Korean Society for Radiation Therapy / v.26, no.1, 2014 , pp. 137-147 More about this Journal
Abstract
Purpose : The purpose of this study is to verify the accuracy of dose delivery according to the patient's breathing cycle in Gated Volumetric Modulated Arc Therapy Materials and Methods : TrueBeam STxTM(Varian Medical System, Palo Alto, CA) was used in this experiment. The Computed tomography(CT) images that were acquired with RANDO Phantom(Alderson Research Laboratories Inc. Stamford. CT, USA), using Computerized treatment planning system(Eclipse 10.0, Varian, USA), were used to create VMAT plans using 10MV FFF with 1500 cGy/fx (case 1, 2, 3) and 220 cGy/fx(case 4, 5, 6) of doserate of 1200 MU/min. The regular respiratory period of 1.5, 2.5, 3.5 and 4.5 sec and the patients respiratory period of 2.2 and 3.5 sec were reproduced with the $QUASAR^{TM}$ Respiratory Motion Phantom(Modus Medical Devices Inc), and it was set up to deliver radiation at the phase mode between the ranges of 30 to 70%. The results were measured at respective respiratory conditions by a 2-Dimensional ion chamber array detector(I'mRT Matrixx, IBA Dosimetry, Germany) and a MultiCube Phantom(IBA Dosimetry, Germany), and the Gamma pass rate(3 mm, 3%) were compared by the IMRT analysis program(OmniPro I'mRT system software Version 1.7b, IBA Dosimetry, Germany) Results : The gamma pass rates of Case 1, 2, 3, 4, 5 and 6 were the results of 100.0, 97.6, 98.1, 96.3, 93.0, 94.8% at a regular respiratory period of 1.5 sec and 98.8, 99.5, 97.5, 99.5, 98.3, 99.6% at 2.5 sec, 99.6, 96.6, 97.5, 99.2, 97.8, 99.1% at 3.5 sec and 99.4, 96.3, 97.2, 99.0, 98.0, 99.3% at 4.5 sec, respectively. When a patient's respiration was reproduced, 97.7, 95.4, 96.2, 98.9, 96.2, 98.4% at average respiratory period of 2.2 sec, and 97.3, 97.5, 96.8, 100.0, 99.3, 99.8% at 3.5 sec, respectively. Conclusion : The experiment showed clinically reliable results of a Gamma pass rate of 95% or more when 2.5 sec or more of a regular breathing period and the patient's breathing were reproduced. While it showed the results of 93.0% and 94.8% at a regular breathing period of 1.5 sec of Case 5 and 6, it could be confirmed that the accurate dose delivery could be possible on the most respiratory conditions because based on the results of 100 patients's respiratory period analysis as no one sustained a respiration of 1.5 sec. But, pretreatment dose verification should be precede because we can't exclude the possibility of error occurrence due to extremely short respiratory period, also a training at the simulation and careful monitoring are necessary for a patient to maintain stable breathing. Consequently, more reliable and accurate treatments can be administered.
Keywords
Gated VMAT; Respiratory period; Gamma index;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Yen-Cho Huang, M.Sc, Chien-Yi Yeh at al.: Clinical practice and evaluation of electronic portal imaging device for VMAT. Medical Dosimetry 2013;38:35-41   DOI   ScienceOn
2 Anton Mans, Peter Remeijer, Igor Olaciregui-Ruiz at al.: 3D Dosimetric Verification of volumetric modulated arc therapy by portal dosimetry. Radiotherapy and Oncology 2010;94:181-187   DOI   ScienceOn
3 Stine Korreman, Joakim Medin, Flemming Kjær- Kristoffersen: Dosimetric verification of RapidArc treatment delivery. Acta Oncologica 2009;48:185-191   DOI   ScienceOn
4 Bedford JL, Warrington AP: Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat Oncol Biol Phys 2009 Feb 1;73(2):537-45
5 Jiwon Sung, Myonggeun Yoon, Weon Kuu Chung at al.: Evaluation of the Accuracy for Respiratory-gated RapidArc. PROGRESS in MEDICAL PHYSICS 2013;June;2;24
6 Ruijiang Li, Edward Mok, Bin Han at al.: Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images. Medical Physics 2012;39(5):2686-2693   DOI
7 Jiango Qian, Lei Xing, Wu Liu, et al.: Dose verification for respiratory-gated volumetric modulated arc therapy. Physics In Medicine And Biology 2011;56:4827-4838   DOI   ScienceOn
8 Giorgia Nicolini, Eugenio Vanetti, Alessandro Clivio at al.: Pre-clinical evaluation of respiratory-gated delivery of volumetric modulated arc therapy with RapidArc. Physics In Medicine And Biology 2010;55:347-357   DOI   ScienceOn
9 Ramesh Boggula, Mattias Birkner, Frank Lohr, et al.: Evaluation of a 2D detector array for patient-specific VMAT QA with different setups. Physics In Medicine And Biology 2011;56: 7163-7177   DOI   ScienceOn
10 Varatharaj Chandraraj, Sotririos Stathakis, Ravikumar Manickam et al.: Comparison of four commercial devices for RapidArc and sliding window IMRT QA. Journal of Applied Clinical Medical Physics 2011;12:338-349
11 M.Stasi, S. Bresciani, A. Miranti at al.: Pretreatment patientspecific IMRT quality assurance: a correlation study between gamma index and patient clinical dose volume histogram. Medical Physics 2012;Dec;39(12):7626-34   DOI   ScienceOn
12 Francisco CH, Silvia VC: A Probability Approach to the Study on Uncertainty Effects on Gamma Index Evaluations in Radiation Therapy. Computational and Mathematical Methods in Medicine 2011; 861869:10
13 Krishna Murthy K: Patient-specific quality assurance of RapidArc treatments: Portal prediction dosimetry compared with phantom studies. Biomedical Imaging and Intervention Journal 2012;8(4):e28
14 Eduard Schreibmann, Anees Dhabaan, Eric Elder at al.: Patient-specific quality assurance method for VMAT treatment delivery. Medical Physics 2009;Oct;36:4530   DOI   ScienceOn
15 Giorgia Nicolini, Eugenio Vanetti, Alessandro Clivio at al.: The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiation Oncology 2008;3:24   DOI   ScienceOn