Independent Verification Program for High-Dose-Rate Brachytherapy Treatment Plans  

Han Youngyih (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine)
Chu Sung Sil (Department of Radiation Oncology, Yonsei University College of Medicine)
Huh Seung Jae (Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine)
Suh Chang-Ok (Department of Radiation Oncology, Yonsei University College of Medicine)
Publication Information
Radiation Oncology Journal / v.21, no.3, 2003 , pp. 238-244 More about this Journal
Abstract
Purpose: The Planning of High-Dose-Rate (HDR) brachytherapy treatments are becoming individualized and more dependent on the treatment planning system. Therefore, computer software has been developed to perform independent point dose calculations with the integration of an isodose distribution curve display into the patient anatomy images. Meterials and Methods: As primary input data, the program takes patients'planning data including the source dwell positions, dwell times and the doses at reference points, computed by an HDR treatment planning system (TPS). Dosimetric calculations were peformed in a $10\times12\times10\;Cm^3$ grid space using the Interstitial Collaborative Working Group (ICWG) formalism and an anisotropy table for the HDR Iridium-192 source. The computed doses at the reference points were automatically compared with the relevant results of the TPS. The MR and simulation film images were then imported and the isodose distributions on the axial, sagittal and coronal planes intersecting the point selected by a user were superimposed on the imported images and then displayed. The accuracy of the software was tested in three benchmark plans peformed by Gamma-Med 12i TPS (MDS Nordion, Germany). Nine patients'plans generated by Plato (Nucletron Corporation, The Netherlands) were verified by the developed software. Results: The absolute doses computed by the developed software agreed with the commercial TPS results within an accuracy of $2.8\%$ in the benchmark plans. The isodose distribution plots showed excellent agreements with the exception of the tip legion of the source's longitudinal axis where a slight deviation was observed. In clinical plans, the secondary dose calculations had, on average, about a $3.4\%$ deviation from the TPS plans. Conclusion: The accurate validation of complicate treatment plans is possible with the developed software and the qualify of the HDR treatment plan can be improved with the isodose display integrated into the patient anatomy information.
Keywords
Chronic myelogenous leukernia; K562 cell; Radiation-induced apoptosis; Hebimycin A; Genistein; Signal transduction;
Citations & Related Records
연도 인용수 순위
  • Reference
1 Fraass B, Doppke K, Hunt M, et al. AAPM TG 53: Quality assurance for clinical radiotherapy treatment planning. Med Phys 1998;25:1773-1829   DOI   ScienceOn
2 Interstitial Collaborative Working Group (ICWG). Interstitial Brachytherapy: Physical, Biological, and Clinical Considerations. Anderson LL , Nath R, Weaver KA, eds.Raven, NY: Weaver, 1990
3 Khan FM. The physics of radiation therapy. 2nd ed. Baltimore, Williams & Wilkins, 1994:418-470
4 Kubo HD, Glasgow GP, Pethel TD, Thomadsen BR,Williamson JF. High-Dose-Rate brachytherapy treatmentdelivery: Report of the AAPM radiation therapy committee task group No.59. Med Phys 1998;25:375-403
5 SawCB,KorbLJ,DarnellB,KrishnaKV,UlewiczD. Independent technique of verifying HDR brachytherapy treatment plans. Int J Radiat Oncol Biol Phys 1998;40:747- 750   DOI   ScienceOn
6 Scala R. Ir-192 High dose rate remote afterloading dose calculationverification. Med Phys 1995;22:902(abst)
7 Miller AV, Davis MG, Horton JL. A method for verifying treatment times for simple high-dose-rate endobronchial brachytherapy procedures. Med Phys 1996;23:1903-1908   DOI   ScienceOn
8 AhmadM,FontenlaDP,CurranJ,Mullokandov E, BeitlerJ,Vikram B. Fastverificationoftreatmenttimesforsingle-catheter HDR brachytherapy. Edocuriether Hypertherm Oncol 1995;11:179-183
9 Meisberger LI,Keller RJ, Shalek RJ. The effectiveattenuation in water of the gamma rays of Gold 198, Iridiium 192, Cesium137,Radium226, and Cobalt60.Radiology1968;90: 953-957   DOI
10 US Nuclear Regulatory Commission. 10 CRF 35.2
11 Van Dyk J, Barnett R,CyglerJ,ShraggeP. Commissioning and quality assurance of treatment planning computers. Int J Radat Oncol Biol Phys 1993; 26:261-273   DOI   ScienceOn
12 CohenGN,AmolsHI,ZaiderM. An independentdose-to-point calucation program for the verification of High- Dose-Rate brachytherapy treatment planning. Int J Radiat Oncol Biol Phys 2000;48:1251-1258   DOI   ScienceOn
13 Kubo HD. Verification of treatment plans by mathematical formulas for single catheter HDR brachytherapy. Med Dosm 1992;17:151-155   DOI
14 Ezzell GA. A manual algorithm for computing dwelltimesfor two-catheter endobronchial treatments using HDR brachytherapy. Med Phys 2000;27:1030-1033   DOI   PUBMED   ScienceOn
15 NathR,AndersonL,LuxtonG,WeaverKA,WilliansonJF,Meigooni AS. Dosimetry of interstitial brachytherapy source: Recommendations of the AAPM Radiation Therahy Committee TaskGroup No. 43. MedPhys1995;22:209-234