• Title/Summary/Keyword: Detorque

Search Result 22, Processing Time 0.014 seconds

Comparative Analysis of the Implant Torque Controllers (임플란트 토크 조절기의 비교 분석 연구)

  • Kim, Dae-Gon;Park, Chan-Jin;Cho, Lee-Ra
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.28 no.1
    • /
    • pp.27-36
    • /
    • 2012
  • As the popularity of implant treatment using a variety of problems have been reported. The proper torque values recommended by the manufacturer and the actual tightening force applied to the observed differences. Period for each product used in this study using different torque controllers accuracy for each of the comparisons, were analyzed. Error rate varies by the manufacturer, but all the products used in the reference period for increasing the error rate increased. Repeat the initial tightening, but the reference, showed a value close to the reference value for the width of change was larger. However, increasing the number of repeated tightening of the reference value for the error rate increased, but has reduced the width of the observed changes.

Comparative analysis of accuracy of how to use the implant torque controllers (사용 방법에 따른 임플란트 토크 조절기의 정확성 비교 분석 연구)

  • Cho, Lee-Ra;Park, Chan-Jin;Huh, Yoon-Hyuk;Lee, Ji-Hye;Kim, Dae-Gon
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.51 no.4
    • /
    • pp.300-306
    • /
    • 2013
  • Purpose: This study was to evaluate the accuracy of how to use the Mechanical Torque-Limiting Devices. Materials and methods: Three different implant manufacturers (Dentium, Osstem, Shinhung) were prepared for each two implant torque controllers. Divided into two groups depending on the method used for removal torque was measured. Repeated measures of ANOVA test (${\alpha}=.05$) was used as statistics to evaluate the effect of repeated loading number on the removal torque. Independent t-test was used to evaluate the difference in removal torque of two groups. Results: The removal torque significantly decreased as the number of loading repetition increased (P<.05). There was significant difference between two groups. Conclusion: Loosen the handle part of the implant torque controllers spring resilience to recover one group compared to the group that did not, showed a little more closed to the reference value.