Comparative Analysis of the Implant Torque Controllers

임플란트 토크 조절기의 비교 분석 연구

  • Kim, Dae-Gon (Department of Prosthodontics and Institute of Oral Science, College of Dentistry, Gangneung-Wonju National University) ;
  • Park, Chan-Jin (Department of Prosthodontics and Institute of Oral Science, College of Dentistry, Gangneung-Wonju National University) ;
  • Cho, Lee-Ra (Department of Prosthodontics and Institute of Oral Science, College of Dentistry, Gangneung-Wonju National University)
  • 김대곤 (강릉원주대학교 치과대학 치과보철학교실 및 구강과학연구소) ;
  • 박찬진 (강릉원주대학교 치과대학 치과보철학교실 및 구강과학연구소) ;
  • 조리라 (강릉원주대학교 치과대학 치과보철학교실 및 구강과학연구소)
  • Received : 2012.01.24
  • Accepted : 2012.03.25
  • Published : 2012.03.30

Abstract

As the popularity of implant treatment using a variety of problems have been reported. The proper torque values recommended by the manufacturer and the actual tightening force applied to the observed differences. Period for each product used in this study using different torque controllers accuracy for each of the comparisons, were analyzed. Error rate varies by the manufacturer, but all the products used in the reference period for increasing the error rate increased. Repeat the initial tightening, but the reference, showed a value close to the reference value for the width of change was larger. However, increasing the number of repeated tightening of the reference value for the error rate increased, but has reduced the width of the observed changes.

임플란트를 이용한 치료가 대중화됨에 따라 다양한 문제가 보고되고 있는데, 이 중에서도 나사풀림현상과 관련된 토크 조절기의 사용에 있어 제조회사에서 권장하는 적절한 토크값과 실제 적용되는 조임력에는 차이가 있음이 여러 연구결과를 통해 보고되고 있다. 본 연구에서는 제품마다 사용기간이 다른 토크 조절기를 이용하여 각각에 따른 정확도를 비교, 분석하고자 하였다. 오차율은 제조회사별로 차이가 있었지만 모든 제품에서 사용기간이 증가할수록 기준값에 대한 오차율은 증가하였다. 그리고 반복 조임 횟수 증가에 따른 측정값의 변화를 살펴보면 반복 조임 초기에는 기준값에 근접한 값을 나타내었지만 기준값에 대한 변화의 폭은 크게 나타났고, 반복 조임 횟수가 증가할수록 기준값에 대한 오차율은 증가하였지만 변화의 폭은 줄어드는 것으로 관찰되었다.

Keywords

References

  1. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Toronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-94. https://doi.org/10.1016/0022-3913(90)90177-E
  2. Wyatt CC, Zarb GA. Treatment outcomes of patients with implantsupported fixed partial prostheses. Int J Oral Maxillofac Implants 1998;13:204-11.
  3. Wennerberg A, Jemt T. Complications in partially edentulous implant patients: a 5-year retrospective follow-up study of 133 patients supplied with unilateral maxillary prostheses. Clin Implant Dent Relat Res 1999;1:49-56. https://doi.org/10.1111/j.1708-8208.1999.tb00091.x
  4. Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Branemark implants in edentulous jaws: a study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants 1991;6:270-6.
  5. Naerti, Quirynen M, van Steenberghe D, Darius P. A six-year prosthodontic study of 509 consecutively inserted implants for the treatment of partial edentulism. J Prosthet Dent 1992;67:236-45. https://doi.org/10.1016/0022-3913(92)90461-I
  6. Hosny M, Duyck J, van Steenberghe D, Naert I. Within-subject comparison between connected and nonconnected tooth-to-implant fixed partial prostheses: up to 14-year follow-up study. Int J Prosthodont 2000;13:340-6.
  7. Priest G. Single-tooth implants and their role in preserving remaining teeth: a 10-year survival study. Int J Oral Maxillofac Implants 1999;14(2):181-188.
  8. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implantsupported single crowns. Clin Oral Implants Res 2008;19(2):119-130. https://doi.org/10.1111/j.1600-0501.2007.01453.x
  9. Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992;7: 26-33.
  10. Khraisat A, Hashimoto A, Nomura S, Miyakawa O. Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system. J Prosthet Dent 2004;91(4):326-334. https://doi.org/10.1016/j.prosdent.2004.01.001
  11. Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single -tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993;6(5):435-439.
  12. Jaarda MJ, Razzoog ME, Gratton DG. Comparison of "look-alike" implant prosthetic retaining screws. J Prosthodont 1995;4(1):23-27. https://doi.org/10.1111/j.1532-849X.1995.tb00310.x
  13. Patterson EA, Johns RB. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants 1992;7(1): 26-33.
  14. Rangert B, Jemt T, Jorneus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989;4(3):241-247.
  15. Weinberg LA, Kruger B. A comparison of implant/prosthesis loading with four clinical variables. Int J Prosthodont 1995;8(5):421-433.
  16. Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent 2002;88(2):183-191. https://doi.org/10.1067/mpr.2002.127604
  17. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10(5):529-536.
  18. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86(1):24-32. https://doi.org/10.1067/mpr.2001.116230
  19. Lang LA, May KB, Wang RF. The effect of the use of a counter-torque device on the abutment-implant complex. J Prosthet Dent 1999;81(4):411-417. https://doi.org/10.1016/S0022-3913(99)80007-9
  20. Jörnéus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7(3):353-359.
  21. Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994;3(1):17-21. https://doi.org/10.1097/00008505-199404000-00002
  22. Dellinges M, Curtis D. Effects of infection control procedures on the accuracy of a new mechanical torque wrench system for implant restorations. J Prosthet Dent 1996;75(1):93-98. https://doi.org/10.1016/S0022-3913(96)90424-2
  23. Goheen KL, Vermilyea SG, Vossoughi J, Agar JR. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants. 1994;9(2):149-155.
  24. Binon PP. Evaluation of the effectiveness of a technique to prevent screw loosening. J Prosthet Dent 1998;79(4):430-432. https://doi.org/10.1016/S0022-3913(98)70157-X
  25. Dellinges MA, Tebrock OC. A measurement of torque values obtained with hand-held drivers in a simulated clinical setting. J Prosthodont 1993;2(4):212-214. https://doi.org/10.1111/j.1532-849X.1993.tb00411.x
  26. Standlee JP, Caputo AA, Chwu MY, Sun TT. Accuracy of mechanical torque-imiting devices for implants. Int J Oral Maxillofac Implants 2002;17(2):220-224.
  27. Standlee JP, Caputo AA. Accuracy of an electric torque-limiting device for implants. Int J Oral Maxillofac Implants 1999;14(2):278-281.
  28. Gutierrez J, Nicholls JI, Libman WJ, Butson TJ. Accuracy of the implant torque wrench following time in clinical service. Int J Prosthodont 1997;10(6):562-567.
  29. Cehreli MC, Akca K, Tonuk E. Accuracy of a manual torque application device for morse-taper implants: a technical note. Int J Oral Maxillofac Implants. 2004;19(5):743-748.
  30. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment Screw Loosening in Single-Implant Restorations: A Systematic Review. Int J Oral Maxillofac Implants 2008;23:681-90.
  31. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-46. https://doi.org/10.1016/j.prosdent.2003.09.012
  32. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-46. https://doi.org/10.1016/j.prosdent.2003.09.012
  33. Kim DG, Cho LR, Park CJ. Comparison of accuracy of implant torque controllers. J Korean Acad Stomatog Func Occ 2008;24:157-68.
  34. Standlee JP, Caputo AA. Accuracy of an electric torque-limiting device for implants. Int J Oral Maxillofac Implants 1999;14:278-81.
  35. McCracken MS, Mitchell L, Hegde R, Mavalli MD. Variability of mechanical torque-limiting devices in clinical service at a US dental school. J Prosthodont 2010;19:20-4. https://doi.org/10.1111/j.1532-849X.2009.00524.x