Discrepancies between Calculated and Delivered Dose Distributions of Respiratory Gated IMRT Fields according to the Target Motion Ranges for Lung and Liver Cancer Patients (호흡연동방사선치료시 폐암과 간암환자의 병소 움직임 크기에 따른 선량분포 차이 분석)

  • Kim, Youngkuk;Lim, Sangwook;Choi, Ji Hoon;Ma, Sun Young;Jeung, Tae Sig;Ro, Tae Ik
    • Progress in Medical Physics
    • /
    • v.25 no.4
    • /
    • pp.242-247
    • /
    • 2014
  • To see the discrepancies between the calculated and the delivered dose distribution of IMRT fields for respiratory-induced moving target according to the motion ranges. Four IMRT plans in which there are five fields, for lung and liver patients were selected. The gantry angles were set to $0^{\circ}$ for every field and recalculated using TPS (Eclipse Ver 8.1, Varian Medical Systems, Inc., USA). The ion-chamber array detector (MatriXX, IBA Dosimetry, Germany) was placed on the respiratory simulating platform and made it to move with ranges of 1, 2, and 3 cm, respectively. The IMRT fields were delivered to the detector with 30~70% gating windows. The comparison was performed by gamma index with tolerance of 3 mm and 3%. The average pass rate was 98.63% when there's no motion. When 1.0, 2.0, 3.0 cm motion ranges were simulated, the average pass rate were 98.59%, 97.82%, and 95.84%, respectively. Therefore, ITV margin should be increased or gating windows should be decreased for targets with large motion ranges.

Quantitative Evaluation of Gated Radiation Therapy Using Gamma Index Analysis (감마지표 분석을 통한 호흡연동방사선치료의 정량적 평가)

  • Ma, Sun Young;Choi, Ji Hoon;Jeung, Tae Sig;Lim, Sangwook
    • Progress in Medical Physics
    • /
    • v.24 no.3
    • /
    • pp.198-203
    • /
    • 2013
  • Generally, to evaluate gated radiation therapy, moving phantoms are used to simulate organ motion. Since the target moves in every direction, we need to take into account motion in each direction. This study proposes methods to evaluate gated radiation therapy using gamma index analysis and to visualize adequate gating window sizes according to motion ranges. The moving phantom was fabricated to simulate motion in the craniocaudal direction. This phantom consisted of a moving platform, the I'm MatriXX, and solid water phantoms. A 6 MV photon filed with a field size of $4{\times}4cm^2$ was delivered to the phantom using the gating system, while the phantom moved in the 1-, 2-, 3-, 4-, and 5-cm motion ranges. The gating windows were set at 40~60%, 30~40%, and 0~90%, respectively. The I'm MatriXX acquired the dose distributions for each scenario and the dose distributions were compared with a $4{\times}4cm^2$ static filed. The tolerance of the gamma index was set at 3%/3 mm. The greater the gating window, the lower the pass rate, and the greater the motion range, the lower the pass rate in this study. In case treatment without gated radiation therapy for the target with motion of 2 cm, the pass rate was less than 96%. But it was greater than 99% when gated radiation therapy was used. However gated radiation therapy was used for the target with motion greater than 4 cm, the pass rate could not be greater than 97% when gating window was set as 30~70%. But when the gating window set as 40~60%, the pass rate was greater than 99%.

Evaluation of the Accuracy for Respiratory-gated RapidArc (RapidArc를 이용한 호흡연동 회전세기조절방사선치료 할 때 전달선량의 정확성 평가)

  • Sung, Jiwon;Yoon, Myonggeun;Chung, Weon Kuu;Bae, Sun Hyun;Shin, Dong Oh;Kim, Dong Wook
    • Progress in Medical Physics
    • /
    • v.24 no.2
    • /
    • pp.127-132
    • /
    • 2013
  • The position of the internal organs can change continually and periodically inside the body due to the respiration. To reduce the respiration induced uncertainty of dose localization, one can use a respiratory gated radiotherapy where a radiation beam is exposed during the specific time of period. The main disadvantage of this method is that it usually requests a long treatment time, the massive effort during the treatment and the limitation of the patient selection. In this sense, the combination of the real-time position management (RPM) system and the volumetric intensity modulated radiotherapy (RapidArc) is promising since it provides a short treatment time compared with the conventional respiratory gated treatments. In this study, we evaluated the accuracy of the respiratory gated RapidArc treatment. Total sic patient cases were used for this study and each case was planned by RapidArc technique using varian ECLIPSE v8.6 planning machine. For the Quality Assurance (QA), a MatriXX detector and I'mRT software were used. The results show that more than 97% of area gives the gamma value less than one with 3% dose and 3 mm distance to agreement condition, which indicates the measured dose is well matched with the treatment plan's dose distribution for the gated RapidArc treatment cases.

Development of Respiratory Signal Analysis Program for Accurate Phase Reassignment in 4D CT Reconstruction (4D CT 영상 재구성 시 정확한 위상 변환을 위한 호흡 신호 분석 프로그램 개발)

  • Park, Hae-Jin;Jung, Won-Gyun;Yoon, Jai-Woong;Song, Ju-Young;Suh, Tae-Suk
    • Progress in Medical Physics
    • /
    • v.19 no.4
    • /
    • pp.241-246
    • /
    • 2008
  • Patient's respiration can have an effect on movement of tumor range and peripheral organs. Therefore, the respiratory signal was acquired by relation between external markers and movement of patient's abdomen during radiational therapy in order to minimize the effect of respiration. Based on this technique, many studies of rational therapy to irradiate at particular part of stable respiratory signals have executed and they have been clinically applied. Nevertheless, the phase-based method is preferred to the amplitude-based method for the rational therapy related to respiration. Because stabilization of the respiratory signal are limited. In this study, a in-house respiratory signal analysis program was developed for the phase reassignment and the analysis of the irregular respiratory signals. Various irregular respiratory patterns was obtained from clinical experimental volunteers. After then, the in-house program analyzed the factors affecting to phase assignment which is directly related to irradiated sector. Subsequently, accuracy of phase assignment was improved with removement of irregular signals by self-developed algorithm. This study is considered to be useful for not only image reconstruction and elevation of irradiating accuracy through phase assignment of RPM system but also analysis of respiratory signals. Moreover, development of 4D CT image is planed with phantom researches or clinical experiments based on this program.

  • PDF

Development of Respiratory Monitoring System by Inductive Plethysmography (인덕턴스 호흡감시 시스템의 개발)

  • Kim, Deok-Won;Yeon, Dong-Su;Kim, Su-Chan
    • Journal of Biomedical Engineering Research
    • /
    • v.16 no.3
    • /
    • pp.353-358
    • /
    • 1995
  • The impedance pneumography which is widely used in monitoring respiration is simple to use and noninvasive, but it is sensitive to motion artifacts and insensitive to detect obstructive apnea. A 3-channel respiratory inductive plethysmography (RIP) developed in this study detects inductance change of the inductance band induced by cross-sectional area change of thorax or abdomen as one breathes. It was confirmed that RIP was less sensitive to various motion artifacts but more sensitive to detection of obstructive apnea than impedance pneumography.

  • PDF

Study of the Respiratory Monitoring System by Using the MEMS Acceleration Sensor (MEMS 가속도 센서를 이용한 환자 호흡동작 모니터링 체계 연구)

  • Sung, Jiwon;Yoon, Myonggeun;Chung, Weon Kuu;Kim, Dong Wook;Shin, Dong Oh
    • Progress in Medical Physics
    • /
    • v.24 no.1
    • /
    • pp.61-67
    • /
    • 2013
  • In this study, we developed and evaluated the patient respiration training method which can help to avoid the problems for the limitation of RGRT applicable patient cases. By using the MEMS (micro-electro-mechanical-system) acceleration sensor, we measured movement of motion phantom. We had compared the response of MEMS with commercially introduced real time patient monitoring (RPM) system. We measured the response of the MEMS with 1 dimensional motion phantom movement for 2.5, 3.0, 3.5 second of period and the 2.0, 3.0, 4.0 cm of the amplitudes. The measured period error of the MEMS system was 0.6~6.0% compared with measured period using RPM system. We found that the shape of MEMS signals were similar with RPM system. From this study, we found the possibility of MEMS as patient training system.

Dosimetric Evaluation of Amplitude-based Respiratory Gating for Delivery of Volumetric Modulated Arc Therapy (진폭 기반 호흡연동 체적변조회전방사선치료의 선량학적 평가)

  • Lee, Chang Yeol;Kim, Woo Chul;Kim, Hun Jeong;Park, Jeong Hoon;Min, Chul Kee;Shin, Dong Oh;Choi, Sang Hyoun;Park, Seungwoo;Huh, Hyun Do
    • Progress in Medical Physics
    • /
    • v.26 no.3
    • /
    • pp.127-136
    • /
    • 2015
  • The purpose of this study is to perform a dosimetric evaluation of amplitude-based respiratory gating for the delivery of volumetric modulated arc therapy (VMAT). We selected two types of breathing patterns, subjectively among patients with respiratory-gated treatment log files. For patients that showed consistent breathing patterns (CBP) relative to the 4D CT respiration patterns, the variability of the breath-holding position during treatment was observed within the thresholds. However, patients with inconsistent breathing patterns (IBP) show differences relative to those with CBP. The relative isodose distribution was evaluated using an EBT3 film by comparing gated delivery to static delivery, and an absolute dose measurement was performed with a $0.6cm^3$ Farmer-type ion chamber. The passing rate percentages under the 3%/3 mm gamma analysis for Patients 1, 2 and 3 were respectively 93.18%, 91.16%, and 95.46% for CBP, and 66.77%, 48.79%, and 40.36% for IBP. Under the more stringent criteria of 2%/2 mm, passing rates for Patients 1, 2 and 3 were respectively 73.05%, 67.14%, and 86.85% for CBP, and 46.53%, 32.73%, and 36.51% for IBP. The ion chamber measurements were within 3.5%, on average, of those calculated by the TPS and within 2.0%, on average, when compared to the static-point dose measurements for all cases of CBP. Inconsistent breathing patterns between 4D CT simulation and treatment may cause considerable dosimetric differences. Therefore, patient training is important to maintain consistent breathing amplitude during CT scan acquisition and treatment delivery.

Dose verification for Gated Volumetric Modulated Arc Therapy according to Respiratory period (호흡연동 용적변조 회전방사선치료에서 호흡주기에 따른 선량전달 정확성 검증)

  • Jeon, Soo Dong;Bae, Sun Myung;Yoon, In Ha;Kang, Tae Young;Baek, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.1
    • /
    • pp.137-147
    • /
    • 2014
  • Purpose : The purpose of this study is to verify the accuracy of dose delivery according to the patient's breathing cycle in Gated Volumetric Modulated Arc Therapy Materials and Methods : TrueBeam STxTM(Varian Medical System, Palo Alto, CA) was used in this experiment. The Computed tomography(CT) images that were acquired with RANDO Phantom(Alderson Research Laboratories Inc. Stamford. CT, USA), using Computerized treatment planning system(Eclipse 10.0, Varian, USA), were used to create VMAT plans using 10MV FFF with 1500 cGy/fx (case 1, 2, 3) and 220 cGy/fx(case 4, 5, 6) of doserate of 1200 MU/min. The regular respiratory period of 1.5, 2.5, 3.5 and 4.5 sec and the patients respiratory period of 2.2 and 3.5 sec were reproduced with the $QUASAR^{TM}$ Respiratory Motion Phantom(Modus Medical Devices Inc), and it was set up to deliver radiation at the phase mode between the ranges of 30 to 70%. The results were measured at respective respiratory conditions by a 2-Dimensional ion chamber array detector(I'mRT Matrixx, IBA Dosimetry, Germany) and a MultiCube Phantom(IBA Dosimetry, Germany), and the Gamma pass rate(3 mm, 3%) were compared by the IMRT analysis program(OmniPro I'mRT system software Version 1.7b, IBA Dosimetry, Germany) Results : The gamma pass rates of Case 1, 2, 3, 4, 5 and 6 were the results of 100.0, 97.6, 98.1, 96.3, 93.0, 94.8% at a regular respiratory period of 1.5 sec and 98.8, 99.5, 97.5, 99.5, 98.3, 99.6% at 2.5 sec, 99.6, 96.6, 97.5, 99.2, 97.8, 99.1% at 3.5 sec and 99.4, 96.3, 97.2, 99.0, 98.0, 99.3% at 4.5 sec, respectively. When a patient's respiration was reproduced, 97.7, 95.4, 96.2, 98.9, 96.2, 98.4% at average respiratory period of 2.2 sec, and 97.3, 97.5, 96.8, 100.0, 99.3, 99.8% at 3.5 sec, respectively. Conclusion : The experiment showed clinically reliable results of a Gamma pass rate of 95% or more when 2.5 sec or more of a regular breathing period and the patient's breathing were reproduced. While it showed the results of 93.0% and 94.8% at a regular breathing period of 1.5 sec of Case 5 and 6, it could be confirmed that the accurate dose delivery could be possible on the most respiratory conditions because based on the results of 100 patients's respiratory period analysis as no one sustained a respiration of 1.5 sec. But, pretreatment dose verification should be precede because we can't exclude the possibility of error occurrence due to extremely short respiratory period, also a training at the simulation and careful monitoring are necessary for a patient to maintain stable breathing. Consequently, more reliable and accurate treatments can be administered.

Application of the ExacTrac System in Respiratory Gated Radiotherapy for Lung Cancer Patients (폐암 환자의 호흡연동방사선치료를 위한 ExacTrac 시스템 적용)

  • Ko, Seung Young;Lee, Jung Il
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.3
    • /
    • pp.325-332
    • /
    • 2019
  • This study analyzed the movement of tumors using 4DCT. Appropriate uniform IM were identified using TC, II and CI depending on ITV margins. DVH and NTCP were also compared in each case. Dose analysis on tumors with uniform IM showed that the optimal treatment plan for satisfying all TC, CI, II was evaluated as 2 mm in phase 20 and 3 mm in 40%. That was compared to the dose from the normal tissues of $PTV_{20}$, $PTV_{40}$. In the 20% radiation field, V5, V10, and V20 for the lungs increased 1.49, 1.26, and 0.65%, while 40% increased by 1.9, 2.41 and 1.23%. NTCP had a dose increase of 0.57 to 0.029% from 20% and 40%. There was a dose increase in the spinal cord and heart at uniform IM, but there was no significant difference. These data suggest that the ITV setting of 20%, phase for Respiratory Gated Radiotherapy using Novalis ExacTrac system can be applied with a uniform IM 2 mm and 40% with 3 mm for optimal treatment plan.

Lie detector using impedance and pulse signal for smart phone application (생체신호를 이용한 거짓말 탐지기 안드로이드 어플리케이션 제작)

  • Jung, Su-Min;Yang, Yeong-Joong;Eom, Yeong-Seung;Park, Jin-Ho;Ahn, Chang-Beom
    • Proceedings of the KIEE Conference
    • /
    • 2011.07a
    • /
    • pp.17-18
    • /
    • 2011
  • 본 논문에서는 일반적으로 뇌파, 호흡운동, 표정변화, 심전도, 산소포화도, 인체의 땀 등을 이용하여 거짓말 탐지를 하는 것을 근간으로 하여 일상생활에서 쉽게 거짓말 탐지를 할 수 있는 시스템을 제작하여 제시한다. 이 거짓말 탐지기 시스템 제작을 통해 생체신호를 스마트폰과 연동함으로써 건강기기의 또 다른 방향을 제시할 것으로 기대된다.

  • PDF