Analysis of dosimetric leaf gap variation on dose rate variation for dynamic IMRT (동적 세기조절방사선 치료 시 선량률 변화에 따른 선량학적엽간격 변화 분석)

  • Yang, Myung Sic;Park, Ju Kyeong;Lee, Seung Hun;Kim, Yang Su;Lee, Sun Young;Cha, Seok Yong
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.28 no.1
    • /
    • pp.47-55
    • /
    • 2016
  • To evaluate the position accuracy of the MLC. This study analyzed the variations of the dosimetric leaf gap(DLG) and MLC transmission factor to reflect the location of the MLC leaves according to the dose rate variation for dynamic IMRT. We used the 6 MV and 10 MV X-ray beams from linear accelerator with a Millennium 120 MLC system. We measured the variation of DLG and MLC transmission factor at depth of 10 cm for the water phantom by varying the dose rate to 200, 300, 400, 500 and 600 MU/min using the CC13 and FC-65G chambers. For 6 MV X-ray beam, a result of measuring based on a dose rate 400 MU/min by varying the dose rate to 200, 300, 400, 500 and 600 MU/min of the difference rate was respectively -2.59, -1.89, 0.00, -0.58, -2.89%. For 10 MV X-ray beam, the difference rate was respectively ?2.52, -1.69, 0.00, +1.28, -1.98%. The difference rate of MLC transmission factor was in the range of about ${\pm}1%$ of the measured values at the two types of energy and all of the dose rates. This study evaluated the variation of DLG and MLC transmission factor for the dose rate variation for dynamic IMRT. The difference of the MLC transmission factor according to the dose rate variation is negligible, but, the difference of the DLG was found to be large. Therefore, when randomly changing the dose rate dynamic IMRT, it may significantly affect the dose delivered to the tumor. Unless you change the dose rate during dynamic IMRT, it is thought that is to be the more accurate radiation therapy.

  • PDF

Assessment for the Utility of Treatment Plan QA System according to Dosimetric Leaf Gap in Multileaf Collimator (다엽콜리메이터의 선량학적엽간격에 따른 치료계획 정도관리시스템의 효용성 평가)

  • Lee, Soon Sung;Choi, Sang Hyoun;Min, Chul Kee;Kim, Woo Chul;Ji, Young Hoon;Park, Seungwoo;Jung, Haijo;Kim, Mi-Sook;Yoo, Hyung Jun;Kim, Kum Bae
    • Progress in Medical Physics
    • /
    • v.26 no.3
    • /
    • pp.168-177
    • /
    • 2015
  • For evaluating the treatment planning accurately, the quality assurance for treatment planning is recommended when patients were treated with IMRT which is complex and delicate. To realize this purpose, treatment plan quality assurance software can be used to verify the delivered dose accurately before and after of treatment. The purpose of this study is to evaluate the accuracy of treatment plan quality assurance software for each IMRT plan according to MLC DLG (dosimetric leaf gap). Novalis Tx with a built-in HD120 MLC was used in this study to acquire the MLC dynalog file be imported in MobiusFx. To establish IMRT plan, Eclipse RTP system was used and target and organ structures (multi-target, mock prostate, mock head/neck, C-shape case) were contoured in I'mRT phantom. To verify the difference of dose distribution according to DLG, MLC dynalog files were imported to MobiusFx software and changed the DLG (0.5, 0.7, 1.0, 1.3, 1.6 mm) values in MobiusFx. For evaluation dose, dose distribution was evaluated by using 3D gamma index for the gamma criteria 3% and distance to agreement 3 mm, and the point dose was acquired by using the CC13 ionization chamber in isocenter of I'mRT phantom. In the result for point dose, the mock head/neck and multi-target had difference about 4% and 3% in DLG 0.5 and 0.7 mm respectively, and the other DLGs had difference less than 3%. The gamma index passing-rate of mock head/neck were below 81% for PTV and cord, and multi-target were below 30% for center and superior target in DLGs 0.5, 0.7 mm, however, inferior target of multi-target case and parotid of mock head/neck case had 100.0% passing rate in all DLGs. The point dose of mock prostate showed difference below 3.0% in all DLGs, however, the passing rate of PTV were below 95% in 0.5, 0.7 mm DLGs, and the other DLGs were above 98%. The rectum and bladder had 100.0% passing rate in all DLGs. As the difference of point dose in C-shape were 3~9% except for 1.3 mm DLG, the passing rate of PTV in 1.0 1.3 mm were 96.7, 93.0% respectively. However, passing rate of the other DLGs were below 86% and core was 100.0% passing rate in all DLGs. In this study, we verified that the accuracy of treatment planning QA system can be affected by DLG values. For precise quality assurance for treatment technique using the MLC motion like IMRT and VMAT, we should use appropriate DLG value in linear accelerator and RTP system.

Evaluation of Dosimetric Leaf Gap (DLG) at Different Depths for Dynamic IMRT (동적 세기조절방사선치료에서 깊이에 따른 DLG변화 분석)

  • Chang, Kyung Hwan;Kwak, Jungwon;Cho, Byungchul;Jeong, Chiyoung;Bae, Jae Beom;Yoon, Sang Min;Lee, Sang-wook
    • Progress in Medical Physics
    • /
    • v.26 no.3
    • /
    • pp.153-159
    • /
    • 2015
  • This study is to evaluate thedosiemtric leaf gap (DLG) at different depths for dynamic intensity-modulated radiation therapy (IMRT) in order to evaluate the absolute dose and dose distribution according to the different positions of tumors and compare the measured and planned the multileaf collimator (MLC) transmission factor (T.F.) and DLG values. We used the 6 MV and 15 MV photon beam from linear accelerator with a Millenium 120 MLC system. After the import the DICOM RT files, we measured the absolute dose at different depths (2 cm, 5 cm, 10 cm, and 15 cm) to calculate the MLC T. F. and DLG. For 6 MV photon beam, the measured both MLC T. F. and DLG were increased with the increase the measured depths. When applying to treatment planning systemas fixed transmission factor with its value measured under the reference condition at depth of 5 cm, although the difference fixed and varied transmission factor is not significant, the dosiemtric effect could be presented according to the depth that the tumor is placed. Therefore, we are planning to investigate the treatment planning system whichthe T. F. and DLG factor according to at the different depths can be applied in the patient-specific treatment plan.