This study aims to propose a method for reducing radiation dose in high-voltage radiography using air gap technique while maintaining the same image quality as when using grids. For an experiment, air gaps were set at 10 cm, 15 cm, 20 cm, 25 cm, and 30 cm with a focus-receptor distance of 180 cm; with each air gap distance, tube current was set at 15 mAs, and tube voltage was set at 80 kVp, 85 kVp, 90 kVp, 95 kVp and 100 kVp. Then, radiographs were taken. In a situation of employing a conventional method of using grids, radiographs were taken at 15 mAs and 107 kVp with a focus-receptor distance of 180 cm. According to the results of the experiment, the surface radiation dose from radiography using grids was 0.130 R; the surface radiation dose at a 20cm air gap was 0.124 R; PSNR between these two images was 10.65 [dB]. In conclusion, the air gap distance, which could maintain the image quality similar to that of a case where scattered radiation was removed and grids were used with a small surface radiation dose, was 20 cm. The result of this study is thought to be used as an indicator to remove surface radiation dose in radiography using air gap.
For five diagnostic X-ray generators (DR), four units turned out to be appropriate in tests on the reproducibility of radiation output suggested in the IEC 60601-2-54 standard, but in one unit of the X-ray equipment, an item measured in a combination of 50% of the highest tube voltage of the diagnostic X-ray equipment, the test setting of Group C with authorized output doses between $1{\mu}Gy$ and $5{\mu}Gy$ of mAs turned out to be inappropriate. As a result, the radiation dose to the IEC 60601-2-54 standard for quantification standards proposed by the radiation output from diagnostic X-ray imaging device reproducibility of performance management should be aware that an important evaluation factor.
Purpose: As the demand of a simple and precise method increases to evaluate the performance of the inverter type x-ray unit, we evaluated the usefulness of the recently-introduced X-ray Multi-Function Test Device (moldel : Xi (unfors)-prestige). Method: We compared the performance of X-ray Multi-Function Test Device (XMFTD) which is non-inveasive type device with the performance of Dynalyzer III that has been most widely used inveasive type measure device. Result: X-ray output dose was increased a little in the XMFTD, but both devices were below the performance evaluation standard, 0.002 in the output reproducibility. Linearity of XMFTD were below 0.1 which means that Dynalyzer III showed more excellency in linearity. As the the accuracy of exposure factor, 1.8 and 2 tube voltage, 2.01 and 2.3 tube current were measured. The exposure time was also measured by 0.01 sec ${\pm}10%$. Both devices were within the acceptance of performance evaluatioin standard. Conclusion: We proved the usefulness of X-ray Multi-Function Test Device (model: Xi (unfors)-prestige) to evaluated the performance on reproductibility and linearity of X-ray output and accuracy of exposure factor of inverter type unit.
Recently, the number of interventional procedures has increased dramatically as an alternative of invasive surgical procedure. The need for the quality control program of the angiographic units has also increased, because of concerns about the increased patient dose and the importance of image quality of angiographic units for the successful procedures. The purpose of this study was to propose an optimal guideline for the quality control program of the angiographic units. We reviewed domestic and international standards about medical imaging system and we evaluated the quality of 61 angiographic units in Korea with the use of NEMA 21 phantom. According to the results of our study, we propose a guideline for the quality control program of the angiographic units. Quality control program includes tube voltage test, tube current test, HVL test, image-field geometry test, spatial resolution test, low-contrast iodine detectability test, wire resolution test, phantom entrance dose test. Proposed reference levels are as follows: PAE < $\pm$ 10% in tube voltage test, PAE < $\pm$ 15% in tube current test, minimum 2.3 mmAl at 80 kVp in HVL test, minimum 'acceptable' level at image-field geometry test, 0.8 lp/mm for detector size of 34-40cm, 1.0 lp/mm for detector size of 28-33cm, 1.2 lp/mm for detector size of 22-27cm in spatial resolution test, minimum 200mg/cc in low contrast iodine detectability test, phantom entrance dose should be under 10R/min, 0.012 inch wire should be seen at static wire resolution test, and 0.022 inch wire should be seen at moving wire resolution test.
In dental radiography, panoramic views cause distortion and thus may bring about inaccurate results in the process of quantitative analysis. In this connection, there has recently been an increasing use of cone beam computed tomography (CBCT) that is known to provide high-resolution images and positional information. In this study, a dental computed tomography unit, 'DCT-90-P IMPLAGRAPHY (Vatech, Korea)', was applied to 20 patients for 24 seconds respectively, with a tube voltage of 85kVp and a tube current of 7mA. The data of CBCT were three-dimensionally reconstructed by use of a computer program, and were histomorphometrically analyzed. The results showed that the diameter of mandibular canal is less distorted at a certain inclination of the mandibular body. The image tends to seem more distended in proportion to the distance between the subject and film. Also, the image tends to be affected according as it is out of focus. In conclusion, it requires that the image should be reconstructed in light of anatomic position and structure.
Bubbles are generated by the boiling of the cooling water when an accident occurs in the reactor and then in order to measure the void fraction, the Optical Fiber Probe(OFP) and optical camera are used in thermal hydraulic safety research. However, such an optical method is not suitable for measuring the void fraction in a $17{\times}17$ array of fuel rods due to the geometrical limitations. This study was conducted as a preliminary study using x-ray system and various phantoms before applying to rod bundles. Through radiographic and tomographic experiments, the tube voltage of the x-ray generator was 130 kVp and the tube current was 1 mA. In addition, it is possible to measure the hole of 1mm in size visually through the bubble resolution phantom, and it is confirmed that the contrast is relatively decreased in the inside of the freon in the case of the contrast evaluation using the road phantom. However, we could obtain good image without distortion when reconstructing the image. Bubble generation phantom experiments were used to confirm the flow direction of the bubbles and to acquire tomography images. The image J tool was used to measure the void fraction of 18 % for a single tomography image. This study has carried out previous researches for the measurement of the bubble rate around the nuclear fuel and could be used as a basic research for continuous research.
We measured the radiation exposure for 55 persons (male: 36, female: 19) who was diagnosed with kidney and ureter stones and received ESWL. The absorbed dose was measured at the organ which is expected to absorb relatively much radiation (kidney, bladder, liver). The radiation dose measurement voltage 80kVp, current of 5mA as a fixed model of the human body by using the Rando phantom with Radiophotoluminescent Glass Dosimeter. Absorbed dose was measured for two times (5 minute and 10 minute, each) and converted to effective dose. Mean number of treatment was 1.8 times (1~4) per patient was the mean time of radiation exposure533 seconds (248-2516). For the treatment of right renal stone, the effective dose of right kidney, left kidney, liver and bladder was 2.458mSv, 0.152mSv, 1.404 mSv and 0.019mSv, respectively. For the treatment of left renal stone, the effective dose of right kidney, left kidney, liver and bladder was 2.496mSv, 0.252mSv, 0.178 mSv, and 0.017mSv, respectively. For the treatment of distal ureter stone, the effective dose of right kidney, left kidney and bladder was 0.009mSv, 0.01mSv and 3.742mSv, respectively.
Proceedings of the Korean Vacuum Society Conference
/
2014.02a
/
pp.374-374
/
2014
의료용 X-ray는 과거 analog 방식과, 연구가 진행 중이며 현재 많이 사용되고 있는 digital 방식으로 나누어진다. 최근, 광도전체와 형광체 기반의 flat panel X-ray detector의 발전에 따른 상용화가 이루어지고 있으며, 많은 발전 가능성이 제기되고 있다. flat panel X-ray detector 검출방식은 direct method (직접 방식)와 indirect method (간접 방식)로 나누어진다. 본 연구는 일반적으로 상용화 되어있는 amorphous seleinum (비정질 셀레늄)의 큰 일함수에 의한 저 해상력이라는 단점을 보완하기 위해, 작은 일함수를 가지는 물질을 사용하여, 영상을 얻을 시에 높은 해상력으로 표현할 수 있도록 하고, 원자번호가 높은 물질을 사용하여 X-ray 흡수율을 높일 수 있도록 기존 direct method에 많이 사용되고 있는 amorphous seleinum 기반 digital X-ray detector가 아닌, 이러한 장점을 충족시킬 수 있는 PbI2 물질 층을 사용하여 시편을 제작 하였다. PbI2를 같은 두께로 올린 후, 물질 층 상부에 Au 전극 면적을 다른 size로 제작한 시편으로 X-ray에 노출 시켰다. 이는 상부 전극 size 차이에 따른 신호 차이를 측정하여 전기적 특성을 평가하기 위한 것이다. 전도성을 띠고 있는 ITO (Indium - Tin - Oxide) glass를 이용하여 screen printing 방법으로 제작하였다. PbI2층을 약 160~180 um두께, $3cm{\times}3cm$ size로 5개 제작하였으며, 상부 전극으로는 Au를 진공 증착 시켰다. 상부 전극 size는 각각 시편 5개에 $0.5cm{\times}0.5cm$, $1cm{\times}1cm$, $1.5cm{\times}1.5cm$, $2cm{\times}2cm$, $2.5cm{\times}2.5cm$로 PbI2 물질 층 중앙에 증착 시켰다. 이러한 설정으로 X-ray 노출 시 관찰할 수 있는 PbI2의 전기적인 특성을 평가할 수 있었다. 관전압을 40 kVp, 60 kVp, 80 kVp, 100 kVp, 120 kVp, 140 kVp로 설정하고, 관전류는 100 mA로 설정하였으며, Dark current, Sensitivity를 측정하였다. Dark current와 Sensitivity를 측정한 뒤, 그 값을 이용하여 SNR (신호 대 잡음 비)값을 구해보았다.실험 결과 단위면적당 signal과 SNR을 분석할 수 있었다. 80 kVp로 기준을 잡고 결과 값을 보면 $0.5cm{\times}0.5cm$ 시편에서 2.92 nC/cm2, $2.5cm{\times}2.5cm$ 시편에서 0.84 nC/cm2로 상부 전극 크기가 작을수록 더 좋은 신호를 측정할 수 있었다. 똑같은 기준에서 SNR을 계산 해 보았을 때, $0.5cm{\times}0.5cm$ 시편에서 6.46, $2.5cm{\times}2.5cm$ 시편에서 1.91로 SNR역시 상부 전극 크기가 작을수록 더 큰 값을 확인할 수 있었다. 이러한 결과는 edge-effect의 영향으로 인해 나온 결과라고 할 수 있다. 이러한 실험 결과, detector 제작 시, 같은 물질을 사용하여 더 높은 효율을 내기 위해서는 큰 size의 상부 전극 보다는 작은 size의 상부 전극을 증착 시키는 것이 전기적 특성을 더욱 효율적으로 평가할 수 있을 것이라고 사료된다.
The purpose of this study was to derive the proposals and to suggest the exposure dose reduction scheme on pediatric head CT scan by analyzing and comparing CT dose index (CTDI) and the national diagnostic reference levels. From January 2014 to December, 231 children under 10years who were requested a pediatric head CT scan with head injury were examined. Research methods were to research and analyze the general characteristics kVp, mA test coverage $CTDI_{vol}$ and DLP referring to dose reports and electronic medical record (EMR). As a result, 7.4%(17 patients) of the total subjects in $CTDI_{vol}$ showed a national diagnostic reference levels exceeding. For DLP 41.6%(96 patients) in excess was relatively higher than $CTDI_{vol}$. DLP was exceeded more than about 60% that is higher than the CT dose index presented by Korea Food & Drug Administration. it is cause of high DLP that scan range increased more than about 30% wider than the standard test coverage presented in Health Insurance Review & Assessment Service. In conclusion, it is able to significantly lower the dose if it is complied with checking the baseline scan range of pediatric head CT scan and appropriately adjusting the protocol.
This study proposes effective quality control and maintenance method by developing a new qualitative evaluation method of apron for medical radiation protection. As an experimental material, one of 0.45 mm lead and 100 of 0.45 mm Pb aprons were used and irradiated under the conditions of a tube voltage of 75 kVp and a tube current of 12.5 mAs to obtain an image. and using the Image J program, PSNR values were compared and analyzed. The results showed that there were 40 aprons (less than 11dB), 55 aprons (less than 11dB, less than 30dB), and 5 aprons (30dB or more). In addition, the dose showed a normal distribution for the apron, and 5 aprons with PSNR less than 11dB and 30dB or more were selected and divided into 8 zones, and these groups were statistically significant.
본 웹사이트에 게시된 이메일 주소가 전자우편 수집 프로그램이나
그 밖의 기술적 장치를 이용하여 무단으로 수집되는 것을 거부하며,
이를 위반시 정보통신망법에 의해 형사 처벌됨을 유념하시기 바랍니다.
[게시일 2004년 10월 1일]
이용약관
제 1 장 총칙
제 1 조 (목적)
이 이용약관은 KoreaScience 홈페이지(이하 “당 사이트”)에서 제공하는 인터넷 서비스(이하 '서비스')의 가입조건 및 이용에 관한 제반 사항과 기타 필요한 사항을 구체적으로 규정함을 목적으로 합니다.
제 2 조 (용어의 정의)
① "이용자"라 함은 당 사이트에 접속하여 이 약관에 따라 당 사이트가 제공하는 서비스를 받는 회원 및 비회원을
말합니다.
② "회원"이라 함은 서비스를 이용하기 위하여 당 사이트에 개인정보를 제공하여 아이디(ID)와 비밀번호를 부여
받은 자를 말합니다.
③ "회원 아이디(ID)"라 함은 회원의 식별 및 서비스 이용을 위하여 자신이 선정한 문자 및 숫자의 조합을
말합니다.
④ "비밀번호(패스워드)"라 함은 회원이 자신의 비밀보호를 위하여 선정한 문자 및 숫자의 조합을 말합니다.
제 3 조 (이용약관의 효력 및 변경)
① 이 약관은 당 사이트에 게시하거나 기타의 방법으로 회원에게 공지함으로써 효력이 발생합니다.
② 당 사이트는 이 약관을 개정할 경우에 적용일자 및 개정사유를 명시하여 현행 약관과 함께 당 사이트의
초기화면에 그 적용일자 7일 이전부터 적용일자 전일까지 공지합니다. 다만, 회원에게 불리하게 약관내용을
변경하는 경우에는 최소한 30일 이상의 사전 유예기간을 두고 공지합니다. 이 경우 당 사이트는 개정 전
내용과 개정 후 내용을 명확하게 비교하여 이용자가 알기 쉽도록 표시합니다.
제 4 조(약관 외 준칙)
① 이 약관은 당 사이트가 제공하는 서비스에 관한 이용안내와 함께 적용됩니다.
② 이 약관에 명시되지 아니한 사항은 관계법령의 규정이 적용됩니다.
제 2 장 이용계약의 체결
제 5 조 (이용계약의 성립 등)
① 이용계약은 이용고객이 당 사이트가 정한 약관에 「동의합니다」를 선택하고, 당 사이트가 정한
온라인신청양식을 작성하여 서비스 이용을 신청한 후, 당 사이트가 이를 승낙함으로써 성립합니다.
② 제1항의 승낙은 당 사이트가 제공하는 과학기술정보검색, 맞춤정보, 서지정보 등 다른 서비스의 이용승낙을
포함합니다.
제 6 조 (회원가입)
서비스를 이용하고자 하는 고객은 당 사이트에서 정한 회원가입양식에 개인정보를 기재하여 가입을 하여야 합니다.
제 7 조 (개인정보의 보호 및 사용)
당 사이트는 관계법령이 정하는 바에 따라 회원 등록정보를 포함한 회원의 개인정보를 보호하기 위해 노력합니다. 회원 개인정보의 보호 및 사용에 대해서는 관련법령 및 당 사이트의 개인정보 보호정책이 적용됩니다.
제 8 조 (이용 신청의 승낙과 제한)
① 당 사이트는 제6조의 규정에 의한 이용신청고객에 대하여 서비스 이용을 승낙합니다.
② 당 사이트는 아래사항에 해당하는 경우에 대해서 승낙하지 아니 합니다.
- 이용계약 신청서의 내용을 허위로 기재한 경우
- 기타 규정한 제반사항을 위반하며 신청하는 경우
제 9 조 (회원 ID 부여 및 변경 등)
① 당 사이트는 이용고객에 대하여 약관에 정하는 바에 따라 자신이 선정한 회원 ID를 부여합니다.
② 회원 ID는 원칙적으로 변경이 불가하며 부득이한 사유로 인하여 변경 하고자 하는 경우에는 해당 ID를
해지하고 재가입해야 합니다.
③ 기타 회원 개인정보 관리 및 변경 등에 관한 사항은 서비스별 안내에 정하는 바에 의합니다.
제 3 장 계약 당사자의 의무
제 10 조 (KISTI의 의무)
① 당 사이트는 이용고객이 희망한 서비스 제공 개시일에 특별한 사정이 없는 한 서비스를 이용할 수 있도록
하여야 합니다.
② 당 사이트는 개인정보 보호를 위해 보안시스템을 구축하며 개인정보 보호정책을 공시하고 준수합니다.
③ 당 사이트는 회원으로부터 제기되는 의견이나 불만이 정당하다고 객관적으로 인정될 경우에는 적절한 절차를
거쳐 즉시 처리하여야 합니다. 다만, 즉시 처리가 곤란한 경우는 회원에게 그 사유와 처리일정을 통보하여야
합니다.
제 11 조 (회원의 의무)
① 이용자는 회원가입 신청 또는 회원정보 변경 시 실명으로 모든 사항을 사실에 근거하여 작성하여야 하며,
허위 또는 타인의 정보를 등록할 경우 일체의 권리를 주장할 수 없습니다.
② 당 사이트가 관계법령 및 개인정보 보호정책에 의거하여 그 책임을 지는 경우를 제외하고 회원에게 부여된
ID의 비밀번호 관리소홀, 부정사용에 의하여 발생하는 모든 결과에 대한 책임은 회원에게 있습니다.
③ 회원은 당 사이트 및 제 3자의 지적 재산권을 침해해서는 안 됩니다.
제 4 장 서비스의 이용
제 12 조 (서비스 이용 시간)
① 서비스 이용은 당 사이트의 업무상 또는 기술상 특별한 지장이 없는 한 연중무휴, 1일 24시간 운영을
원칙으로 합니다. 단, 당 사이트는 시스템 정기점검, 증설 및 교체를 위해 당 사이트가 정한 날이나 시간에
서비스를 일시 중단할 수 있으며, 예정되어 있는 작업으로 인한 서비스 일시중단은 당 사이트 홈페이지를
통해 사전에 공지합니다.
② 당 사이트는 서비스를 특정범위로 분할하여 각 범위별로 이용가능시간을 별도로 지정할 수 있습니다. 다만
이 경우 그 내용을 공지합니다.
제 13 조 (홈페이지 저작권)
① NDSL에서 제공하는 모든 저작물의 저작권은 원저작자에게 있으며, KISTI는 복제/배포/전송권을 확보하고
있습니다.
② NDSL에서 제공하는 콘텐츠를 상업적 및 기타 영리목적으로 복제/배포/전송할 경우 사전에 KISTI의 허락을
받아야 합니다.
③ NDSL에서 제공하는 콘텐츠를 보도, 비평, 교육, 연구 등을 위하여 정당한 범위 안에서 공정한 관행에
합치되게 인용할 수 있습니다.
④ NDSL에서 제공하는 콘텐츠를 무단 복제, 전송, 배포 기타 저작권법에 위반되는 방법으로 이용할 경우
저작권법 제136조에 따라 5년 이하의 징역 또는 5천만 원 이하의 벌금에 처해질 수 있습니다.
제 14 조 (유료서비스)
① 당 사이트 및 협력기관이 정한 유료서비스(원문복사 등)는 별도로 정해진 바에 따르며, 변경사항은 시행 전에
당 사이트 홈페이지를 통하여 회원에게 공지합니다.
② 유료서비스를 이용하려는 회원은 정해진 요금체계에 따라 요금을 납부해야 합니다.
제 5 장 계약 해지 및 이용 제한
제 15 조 (계약 해지)
회원이 이용계약을 해지하고자 하는 때에는 [가입해지] 메뉴를 이용해 직접 해지해야 합니다.
제 16 조 (서비스 이용제한)
① 당 사이트는 회원이 서비스 이용내용에 있어서 본 약관 제 11조 내용을 위반하거나, 다음 각 호에 해당하는
경우 서비스 이용을 제한할 수 있습니다.
- 2년 이상 서비스를 이용한 적이 없는 경우
- 기타 정상적인 서비스 운영에 방해가 될 경우
② 상기 이용제한 규정에 따라 서비스를 이용하는 회원에게 서비스 이용에 대하여 별도 공지 없이 서비스 이용의
일시정지, 이용계약 해지 할 수 있습니다.
제 17 조 (전자우편주소 수집 금지)
회원은 전자우편주소 추출기 등을 이용하여 전자우편주소를 수집 또는 제3자에게 제공할 수 없습니다.
제 6 장 손해배상 및 기타사항
제 18 조 (손해배상)
당 사이트는 무료로 제공되는 서비스와 관련하여 회원에게 어떠한 손해가 발생하더라도 당 사이트가 고의 또는 과실로 인한 손해발생을 제외하고는 이에 대하여 책임을 부담하지 아니합니다.
제 19 조 (관할 법원)
서비스 이용으로 발생한 분쟁에 대해 소송이 제기되는 경우 민사 소송법상의 관할 법원에 제기합니다.
[부 칙]
1. (시행일) 이 약관은 2016년 9월 5일부터 적용되며, 종전 약관은 본 약관으로 대체되며, 개정된 약관의 적용일 이전 가입자도 개정된 약관의 적용을 받습니다.