Analyzed the Computed Tomography Dose Index (CTDI) to the Pediatric Brain CT by Reason of the Observation for the Exposure Dose: Base on a Hospital  

Lee, Jae-Seung (동의대학교 방사선학과)
Kim, Hyun-Jin (동의대학교 대학원 보건의과학과)
Im, In-Chul (동의대학교 방사선학과)
Publication Information
Abstract
The purpose of this study was to derive the proposals and to suggest the exposure dose reduction scheme on pediatric head CT scan by analyzing and comparing CT dose index (CTDI) and the national diagnostic reference levels. From January 2014 to December, 231 children under 10years who were requested a pediatric head CT scan with head injury were examined. Research methods were to research and analyze the general characteristics kVp, mA test coverage $CTDI_{vol}$ and DLP referring to dose reports and electronic medical record (EMR). As a result, 7.4%(17 patients) of the total subjects in $CTDI_{vol}$ showed a national diagnostic reference levels exceeding. For DLP 41.6%(96 patients) in excess was relatively higher than $CTDI_{vol}$. DLP was exceeded more than about 60% that is higher than the CT dose index presented by Korea Food & Drug Administration. it is cause of high DLP that scan range increased more than about 30% wider than the standard test coverage presented in Health Insurance Review & Assessment Service. In conclusion, it is able to significantly lower the dose if it is complied with checking the baseline scan range of pediatric head CT scan and appropriately adjusting the protocol.
Keywords
Computed Tomography Dose Index; Diagnostic Reference Level; Dose Report; Electronic Medical Record; Dose Length Product;
Citations & Related Records
Times Cited By KSCI : 1  (Citation Analysis)
연도 인용수 순위
1 J. Sherck, C. Shatney, K. Sensaki, and V. Selivanov, "The accuracy of computed tomography in the diagnosis of blunt small-bowel perforation," The American journal of surgery, Vol.168, No.6, pp.670-675, 1994.   DOI   ScienceOn
2 Neil. Bhattacharyya and Marvin P. Fried, "The accuracy of computed tomography in the diagnosis of chronic rhinosinusitis," The Laryngoscope, Vol.113, No.1, pp.125-129, 2003.   DOI   ScienceOn
3 D. J. Brenner and E. J. Hall, "Computed tomography an increasing source of radiation exposure," N Engl J Med, Vol.357, No.22, pp.2277-2284, 2007.   DOI   ScienceOn
4 E. Dougeni, K. Faulkner, and G. Panayiotakis, "A review of patient dose and optimisation methods in adult and pediatric CT scanning," Eur J Radiol, Vol.81, No.44, pp.e665-683, 2011.
5 D. H. Yang and H. W. Goo, "Pediatric 16-slice CT protocol: radiation dose and image quality," J. Korean Raiol. Soc., Vol.59, No.1, pp.333-347, 2008.   DOI
6 이정근, 장성주, 장영일, "소아 CT 촬영 시 방사선피폭과 저감화 방법", 한국콘텐츠학회논문지 Vol.14, No.1, pp.356-363, 2014.   DOI
7 D. H. Yang and H. W. Goo, "Pediatric 16-slice CT protocol: radiation dose and image quality," J. Korean Raiol. Soc., Vol.59, pp.289-296, 2001.
8 D. J. Brenner and E. J. Hall, "Computed tomography: an incerasing source of radiation exposuer," N. Engl. J. Med., Vol.357, pp.2277-2284, 2007.   DOI   ScienceOn
9 ICRP, "2007 recommendations of the International Commission on Radiological Protection," ICRP publication No.103, 2007.
10 한국식품의약품안전처, "일반 영상의학검사의 환자선량권고량 가이드라인", 방사선안전관리 시리즈, No.30, 2012.
11 한국식품의약품안전처, "어린이 CT 영상의학검사의 환자선량권고량 가이드라인", 방사선안전관리 시리즈, No.31, 2012.
12 ICRP, "Managing patient dose in computed tomography," ICRP publication, No.87, 2000.
13 IEC, "Evaluation and routine testing in medical imaging departments, part 2-6: constancy tests X-ray equipment for computed tomography," International Electrotechnical Commission (Geneva, Switzerland), IEC 1223-2-6, 1994.
14 S. Edyvean, "Type testing of CT scanners: methods and methodology for assessing imaging performance and dosimetry," Medical Devices Agency (London, UK), Report No.mA/98/25, 1998.
15 Health Insurance Review & Assessment Service, "CT Examination and Repeat CT Examination Guidelines," p.39, 2014.
16 UNSCEAR 2000 Report Vol.1. "Sources and Effects of Ionization Radiation," Annex D Medical radiation exposure, UNSCER, 2000.