DOI QR코드

DOI QR Code

Economical Gas Chamber for In-situ Gas Measurement and Analysis of Gas Response Characteristics according to Sensor Voltage

인시투 가스 측정이 가능한 경제적 가스 챔버 구현 및 센서 전압에 따른 가스 응답 특성 분석

  • Choi, Yun-Suk (Department of Computer Information Engineering, Hoseo University) ;
  • Lee, In Hwan (School of Mechanical Engineering, Chungbuk National University)
  • 최연석 (호서대학교 컴퓨터정보공학부) ;
  • 이인환 (충북대학교 기계공학부)
  • Received : 2019.04.03
  • Accepted : 2019.04.26
  • Published : 2019.05.31

Abstract

Breath analysis using a portable device is better than the classical breath analysis system in terms of installation and operation. There is an increasing need to develop cost-effective equipment for testing gas sensors from the viewpoint of functionalities that can be applied applicable to portable devices. In the present study, an economical gas chamber for in-situ gas measurement is implemented with a single gas chamber without using expensive gas storage and control equipment; the gas response characteristics are analyzed using the above-described chamber. The main features of the implemented gas chamber are simple injection procedure, improved gas diffusion, easy measurement and cleaning, support for low-power mode measurement function for portable devices, and open source platform. Moreover, an analysis of gas response characteristics based on changes in sensor voltage show that the sensitivity and 90% response time are affected by the sensor voltage. Furthermore, the sensitivity graph has an inflection point in a specific range. The gas sensor applied in this study showed fast response speed and high sensitivity for sensor voltages of 3.0-3.5 V, regardless of the concentration of acetone gas, the target gas used in this study.

Keywords

References

  1. Manolis, A., "The diagnostic potential of breath analysis", Clin. Chem., Vol. 29, No. 1, pp. 5-15, 1983. https://doi.org/10.1093/clinchem/29.1.5
  2. Cao, W. and Duan, Y., "Breath analysis: potential for clinical diagnosis and exposure assessment", Clin. Chem., Vol. 52, No. 5, pp. 800-811, 2006. https://doi.org/10.1373/clinchem.2005.063545
  3. Miekisch, W., Schubert, J. K. and Noeldge Schomburg, G. F. E., "Diagnostic potential of breath analysis - focus on volatile organic compounds.", Clin. Chim. Acta, Vol. 347, No. 1-2, pp. 25-39, 2004. https://doi.org/10.1016/j.cccn.2004.04.023
  4. Amann, A., Spanel, P. and Smith, D., "Breath analysis : the approach towards clinical applications", Mini-Rev. Med. Chem., Vol. 7, No. 2, pp. 115-129, 2007. https://doi.org/10.2174/138955707779802606
  5. Samar, K. K., James, A. B., Radhakrishnan, N. and Anna, M. J., "Breath Acetone Analyzer : Diagnostic Tool to Monitor Dietary Fat Loss", Clin. Chim., Vol. 39, No. 1, pp. 87-92, 1993. https://doi.org/10.1093/clinchem/39.1.87
  6. Righettoni, M., Tricoli, A., Gass, S., Schmid, A., Amann, A. and Pratsinis, S. E., "Breath acetone monitoring by portable Si:WO3 gas sensors", Analytica Chimica Acta, Vol. 738, No. 1, pp. 69-75, 2012. https://doi.org/10.1016/j.aca.2012.06.002
  7. FIGARO, TGS822 Product information, Figaro Engineering Inc., pp. 1-10, 1987.
  8. FIGARO, Technical information for TGS8100, Figaro Engineering Inc., pp. 1-10, 2014.