Determining Ion Collection Efficiency in a Liquid Ionization Chamber in Co-60 Beam

Co-60 빔에서 액체 전리함의 이온 수집 효율 결정 연구

  • Received : 2014.03.03
  • Accepted : 2014.03.16
  • Published : 2014.03.31

Abstract

Liquid ionization chamber is filled with liquid equivalent material unlike air filled ionization chamber. The high density material allow very small-volume chamber to be constructed that still have a sufficiently high sensitivity. However liquid ionization chamber should be considered for both initial recombination and general recombination. We, therefore, studied using the Co-60 beam as the continuous beam and the microLion chamber (PTW) for comparing the ion collection efficiency by Greening theory, two-dose rate method and our experiment method. The measurements were carried out using Theratron 780 as the cobalt machine and water phantom and 0.6 cc Farmer type ionization chamber was used with microLion chamber in same condition for measuring the charge of microLion chamber according to the dose rates. Dose rate was in 0.125~0.746 Gy/min and voltages applied to the microLion chamber were +400, +600 and +800 V. As the result, the collection efficiency by three method was generally less than 1%. In particular, our experimental collection efficiency was in good agreement within 0.3% with Greening theory except the lowest two dose rates. The collection efficiency by two-dose rate method also agreed with Greening theory generally less than 1%, but the difference was about 4% when the difference of two dose rates were lower. The ion recombination correction factors by Greening theory, two-dose rate method and our experiment were 1.0233, 1.0239 and 1.0316, respectively, in SSD 80 cm, depth 5 cm recommended by TRS-398 protocol. Therefore we confirmed that the loss by ion recombination was about 3% in this condition. We think that our experiment method for ion recombination correction will be useful tool for radiation dosimetry in continuous beam.

액체 전리함은 공동 전리함과 달리 감응매질이 물 등가물질로 이루어져 있어서, 감도가 매우 높아서 충분히 작게 만들 수 있기 때문에 기준 조사면 뿐만 아니라 소조사면의 선량 평가에 유용하다는 장점이 있지만, 이온 재결합 손실 계산에 있어 초기 재결합과 일반 재결합을 모두 고려해야 하므로, 사용상에 어려움이 따른다. 본 연구에서는 연속 빔인 코발트 60 빔에서 PTW사의 microLion 액체 전리함을 이용하여 Greening 이론식과 이선량률법 및 다른 실험을 이용하여 수집효율을 구하고, 비교하는 연구를 수행하였다. 이는 코발트 장비인 Theratron 780과 물팬톰을 이용하여 수행하였으며, 선량률에 따른 microLion 전리함의 전하량을 측정하기 위해 0.6 cc 공동 전리함을 같은 조건에서 동시에 측정하였다. 이때 선량률의 범위는 0.125~0.746 Gy/min이였으며, 각 선량률에서 +400, +600 및 +800 전압에 대하여 액체 전리함을 이용하여 전하량을 측정하였다. 측정된 데이터를 이용하여 세 가지 방법에 따라 계산된 수집효율은 대체로 1% 이내로 일치하였다. 특히 본 연구에서 실험을 통해 구한 수집효율은 가장 낮은 두 선량률을 제외하고는 0.3% 이내로 잘 일치함을 보였다. 이선량률법의 경우 Greening 이론식과 비교하여 대체로 1% 이내의 차이를 보였지만, 두 선량률의 차이가 적을 때 대략 4% 가까운 차이를 보임을 확인하였다. TRS-398 물흡수선량 프로토콜에서 권고하는 표면과 선원간의 거리가 80 cm이고, 깊이 5 cm에서 Greening 이론법과 이선량률법, 본 연구에서 실험을 통해 얻은 방법에 의한 이온 재결합 보정계수는 각각 1.0233, 1.0239, 1.0316으로, 이 조건에서 대략 3%의 이온 재결합에 의한 손실이 발생함을 확인하였다. 본 연구에서 실험을 통해 이온 재결합 손실을 계산하는 방법은 다른 두 선량률에서 액체 전리함의 보정된 전하량을 이용하여 손쉽게 결정할 수 있기 때문에 연속 빔에서 선량 평가 시에 매우 유용하게 사용될 수 있으리라 판단된다.

Keywords

References

  1. Wickman G: A liquid ionization chamber with high spatial resolution. Phys Med Biol 19:66-72 (1974) https://doi.org/10.1088/0031-9155/19/1/006
  2. Wickman G, Nyström H: The use of liquids in ionization chambers for high precision radiotherapy dosimetry. Phys Med Biol 37:1789-1812 (1992) https://doi.org/10.1088/0031-9155/37/9/005
  3. Daşu A, Löfroth PO, Wickman G: Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams. Phys Med Biol 43:21-36 (1998) https://doi.org/10.1088/0031-9155/43/1/002
  4. Pardo J, Franco L, Gómez F, et al: Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance. Phys Med Biol 50:1703-1716 (2005) https://doi.org/10.1088/0031-9155/50/8/006
  5. Choi SH, Kim CH, Huh HD, Kim SH, Kim KB: Determination of the beam quality correction factor $\kappa_{Q,Q_{0}}$ for the microLion chamber in a clinical photon beam. J Kor Phys Soc 62(1): 152-158 (2013) https://doi.org/10.3938/jkps.62.152
  6. Eberle K, Engler J, Hartmann G, Hofmann R, Hörandel JR: First tests of a liquid ionization chamber to monitor intensity modulated radiation beams. Phys Med Biol 48:3555-3564 (2003) https://doi.org/10.1088/0031-9155/48/21/007
  7. Chung E, Soisson E, Seuntjens J: Dose homogeneity specification for reference dosimetry of nonstandard fields. Med Phys 39:407-414 (2012) https://doi.org/10.1118/1.3669487
  8. Johansson B, Wickman G, Bahar-Gogani J: General collection efficiency for liquid isooctane and tetramethylsilane in pulsed radiation. Phys Med Biol 42:1929-1938 (1997) https://doi.org/10.1088/0031-9155/42/10/006
  9. Pardo-Montero J, Gómez F: Determining charge collection efficiency in parallel-plate liquid ionization chambers. Phys Med Biol 54:3677-3689 (2009) https://doi.org/10.1088/0031-9155/54/12/005
  10. Tölli H, Sjögren R, Wendelsten M: A two-dose-rate method for general recombination correction for liquid ionization chambers in pulsed beams. Phys Med Biol 55:4247-4260 (2010) https://doi.org/10.1088/0031-9155/55/15/004
  11. Andersson J, Tölli H: Application of the two-dose-rate method for general recombination correction for liquid ionization chambers in continuous beams. Phys Med Biol 56:299-314 (2010)
  12. Greening J: Saturation characteristics of parallel-plate ionization chambers. Phys Med Biol 9:143-154 (1964) https://doi.org/10.1088/0031-9155/9/2/302
  13. PTW, Ionization chamber Type 31018 (microLion), User Manual (2007)
  14. International Atomic Energy Agency (IAEA): Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water IAEA Technical Report Series (TRS) 398 (Vienna: IAEA) (2000)