Diffusion-Enhanced Modified Hemodialyzer

  • Lee, Kyung-Soo (Interdisciplinary Program in Biomedical Engineering Major, Seoul National University) ;
  • Lee, Sa-Ram (Interdisciplinary Program in Biomedical Engineering Major, Seoul National University) ;
  • Mun, Cho-Hae (Interdisciplinary Program in Biomedical Engineering Major, Seoul National University) ;
  • Min, Byoung-Goo (Department of Biomedical Engineering, College of Medicine, Seoul National University)
  • Published : 2007.08.30

Abstract

Flow mismatch between blood and dialysate is invariably encountered during conventional hemodialysis, and this deteriorates diffusive mass transfer. A modification of a conventional dialyzer was conceived to prevent this mismatch. The modified dialyzer includes two independent blood flow regions (central and peripheral regions), which were achieved by redesigning the dialyzer cap. Resultantly, the blood stream was divided into two concentric dialyzer regions. Solutes clearances obtained using the modified dialyzers were compared with those of conventional dialyzers. Solutes clearances by conventional dialyzers were uniform, but solutes clearances by modified dialyzers were found to be dependent on the simulated blood split into dialyzer central and peripheral regions. Maximal clearances using the modified dialyzer were improved by up to approximately 7.6% for urea and 7.3% for creatinine, as compared with those of conventional dialyzers. More optimizations are required for clinical applications, but the finding that blood flowrates through central and peripheral fiber bundles can be easily regulated is encouraging.

Keywords

References

  1. J. T. Daugirdas and J. C. Van Stone, Physiologic Principles and Urea Kinetic Modeling. In: Daugirdas JT, Blake PG, Ing TS, eds. Handbook of dialysis, Lippincott Williams & Wilkins, 2000, pp.15-45
  2. C. Ronco, A. Brendolan, C. Crepaldi, M. Rodighiero and M. Scabardi, 'Blood and dialysate flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique,' J. Am. Soc. Nephrol., vol. 13, Suppl 1, pp.S53-6l, 2002
  3. T. Osuga, T. Obata and H. Ikehira, 'Detection of small degree of nonuniformity in dialysate flow in hollow-fiber dialyzer using proton magnetic resonance imaging,' Magn Reson Imaging, vo1.22, no.3, pp.417-420, 2004 https://doi.org/10.1016/j.mri.2003.08.008
  4. C. Ronco, A. Brendolan, C. Crepaldi, M. Rodighiero, P. Everard, M. Ballestri, G. Cappelli, M. Spittle and G. La Greca, 'Dialysate flow distribution in hollow fiber hemodialyzers with different dialysate pathway configurations,' Int. J. Artif. Organs., vol.23, no.9, pp.601-609, 2000 https://doi.org/10.1177/039139880002300902
  5. A. Brendolan, F. Nalesso, A. Fortunato, C. Crepaldi, M. De Cal, S. Cazzavillan, D. Cruz, N. Techawathanawanna and C. Ronco, 'Dialytic performance evaluation of Rexeed: a new polysulfone-based dialyzer with improved flow distributions,' Int. J. Artif. Organs., vo1.28, no. 10, pp.966-975, 2005 https://doi.org/10.1177/039139880502801003
  6. F. Gastaldon, A. Brendolan, C. Crepaldi, P. Frisone, S. Zamboni, V. d'Intini, S. Poulin, R. Hector, A. Granziero, K. Martins, R. Gellert, P. Inguaggiato and C. Ronco, 'Effects of novel manufacturing technology on blood and dialysate flow distribution in a new low flux 'alpha Polysulfone hemodialyzer,' Int. J. Artif. Organs., vo1.26, no.2, pp.105-112, 2003 https://doi.org/10.1177/039139880302600203
  7. C. K. Poh, P. A. Hardy, Z. Liao, Z. Huang, W. R. Clark and D. Gao, 'Effect of spacer yams on the dialysate flow distribution of hemodialyzers: a magnetic resonance imaging study,' Asaio J., vol.49, no.4, pp.440-448, 2003
  8. Z. Liao, E. Klein, C. K. Poh, Z. Huang, P. A. Hardy, S. Morti, W. R. Clark and D. Gao, 'A modified equivalent annulus model for the hollow fiber hemodialyzer,' Int. J. Artif. Organs., vol.27, no.2, pp.110-117, 2004 https://doi.org/10.1177/039139880402700206
  9. C. Ronco, M. Scabardi, M. Goldoni, A. Brendolan, C. Crepaldi and G. La Greca, 'Impact of spacing filaments external to hollow fibers on dialysate flow distribution and dialyzer performance,' Int. J. Artif. Organs., vo1.20, no.5, pp.261-266, 1997 https://doi.org/10.1177/039139889702000505
  10. W. R. Clark and J. H. Shinaberger, 'Effect of dialysate-side mass transfer resistance on small solute removal in hemodialysis,' Blood Purif., vol.18, no.4, pp.260-263, 2000 https://doi.org/10.1159/000014446
  11. Z. Huang, E. Klein, B. Li, C. Poh, Z. Liao, W. R. Clark and D. Gao, 'A new method to evaluate the local clearance at different annular rings inside hemodialyzers,' Asaio J., vol.49, no.6, pp.692-697, 2003 https://doi.org/10.1097/01.MAT.0000093972.49709.08
  12. J. K. Unger, A. J. Lemke and C. Grosse-Siestrup, 'Thermography as potential real-time technique to assess changes in flow distribution in hemofiltration,' Kidney Int., vol.69, no.3, pp. 520-525, 2006 https://doi.org/10.1038/sj.ki.5000143
  13. M. C. Yang and C. C. Lin, 'Influence of design of the hemodialyzer inlet chamber on red blood cell damage during hemodialysis,' Asaio J., vol.47, no.1, pp.92-96, 2001 https://doi.org/10.1097/00002480-200101000-00019