DOI QR코드

DOI QR Code

Fabrication and Performance Evaluation of a Micro Separation Chip of Magnetic Beads Using Magnetophoretic Flow

자기영동을 이용한 자성입자 분리 마이크로 칩 제작 및 성능평가

  • 고정상 (부산대학교 기계공학부, 부산대학교 기계기술연구소(RIMT))
  • Published : 2007.04.01

Abstract

We developed a microfluidic platform able to control the trap and release of magnetic beads used for separation of a specific biomolecules. The magnetic beads can be trapped and released conditionally by controlling the difference between the Stokes force induced by the fluid flow and magnetic force resulting from a permanent magnet. The permanent magnet of CoNiP alloy is electroplated. It is characterized to have the 1369 Oe of coercivity, 1762 Gauss of remanence, and 0.603MGOe of (BH)max. Through the experimental and numerical investigation, the magnetic beads are trapped under the flow velocity of 17 ${\mu}m/s$ and are released perfectly above the velocity of 174 ${\mu}m/s$.

Keywords

References

  1. Schutte, H., Kroner, K.H., Hummel W. and Kula, M.-R, 1983, 'Recent Development in Separation and Purification of Biomolecules,' Annals of the New York Academy of Sciences, Vol. 413, pp.270 https://doi.org/10.1111/j.1749-6632.1983.tb47899.x
  2. Albright, C. D., Frost, J. K., Pressman, N. J., 2005, 'Centrifugal Separation of Cells in Sputum Specimens from Patients with Undifferentiated Carcinoma,' Cytometry, Vol. 7, Issue 6, pp.536-543 https://doi.org/10.1002/cyto.990070607
  3. Patonay, G., Salon, J., Sowell, J. and Strekowski, L., 2004, 'Noncovalent Labeling of Biomolecules with Red and Near-Infrared Dyes,' Molecules, Vol. 9, pp. 40-49 https://doi.org/10.3390/90300040
  4. Lee, S. Y., Seo, K. S. and Cho, Y.-H., 2004, 'A DNA Microextractor Using Crossed Field Electrophoresis,' Trans. of the KSME (A), Vol. 28, No. 8, pp. 1135-1139 https://doi.org/10.3795/KSME-A.2004.28.8.1135
  5. Cunico, R. L., Gooding, K. M. and Richmond, T.W., 1998, Basic HPLC and CE of Biomolecules, Bay Bioanalytical Laboratory
  6. Kemshead. J. T. and Ugelstad, J., 1985, 'Magnetic Separation Techniques: Their Application to Medicine,' Molecular and Cellular Biochemistry, Vol. 67, No. 1, pp. 11-18
  7. Choi, J. W., Ahn, C. H., Bhansali, S. and Henderson, H. T., 2000, 'A New Magnetic Bead-based, Filterless Bio-separator with Planer Electromagnet Surfaces for Integrated Bio-detection Systems,' Sensors and actuators, Physical B, Vol. 68, pp. 34-39 https://doi.org/10.1016/S0925-4005(00)00458-5
  8. Rong, R., Choi, J.-W. and Ahn, C.H., 2006, 'An Onchip Magnetic Bead Separator for Biocell Sorting,' J. Micromech. Microeng., Vol. 16, 2783-2790 https://doi.org/10.1088/0960-1317/16/12/035
  9. Nguyen, N.-T. and Wereley, S. T., 2006, Fundamentals and Applications of Microfluidics, Artech House
  10. Cho, C.-H., Cho, W., Ahn, Y. and Hwang, S.-Y., 2006, 'PDMS/Glass Serpentine Microchannel Chip for PCR with Bubble Suppression in Sample Injection,' Trans. of the KSME (A), Vol. 30, No. 10, pp. 1261-1268 https://doi.org/10.3795/KSME-A.2006.30.10.1261