DOI QR코드

DOI QR Code

Development of High Resolution Micro-CT System for In Vivo Small Animal Imaging

소형 동물의 생체 촬영을 위한 고해상도 Micro-CT 시스템의 개발

  • 박정진 (경희대학교 동서의료공학과) ;
  • 이수열 (경희대학교 동서의료공학과) ;
  • 조민형 (경희대학교 동서의료공학과)
  • Published : 2007.02.28

Abstract

Recently, small-animal imaging technology has been rapidly developed for longitudinal screening of laboratory animals such as mice and rats. One of newly developed imaging modalities for small animals is an x-ray micro-CT (computed tomography). We have developed two types of x-ray micro-CT systems for small animal imaging. Both systems use flat-panel x-ray detectors and micro-focus x-ray sources to obtain high spatial resolution of $10{\mu}m$. In spite of the relatively large field-of-view (FOV) of flat-panel detectors, the spatial resolution in the whole-body imaging of rats should be sacrificed down to the order of $100{\mu}m$ due to the limited number of x-ray detector pixels. Though the spatial resolution of cone-beam CTs can be improved by moving an object toward an x-ray source, the FOV should be reduced and the object size is also limited. To overcome the limitation of the object size and resolution, we introduce zoom-in micro-tomography for high-resolution imaging of a local region-of-interest (ROI) inside a large object. For zoom-in imaging, we use two kinds of projection data in combination, one from a full FOV scan of the whole object and the other from a limited FOV scan of the ROI. Both of our micro-CT systems have zoom-in micro-tomography capability. One of both is a micro-CT system with a fixed gantry mounted with an x-ray source and a detector. An imaged object is laid on a rotating table between a source and a detector. The other micro-CT system has a rotating gantry with a fixed object table, which makes whole scans without rotating an object. In this paper, we report the results of in vivo small animal study using the developed micro-CTs.

Keywords

References

  1. T. F. Massoud and S. S. Gambhir, 'Molecular imaging living subject: seeing fundamental biological processes in a new light,' Genes & Development, vol. 17, pp. 545-580, 2003 https://doi.org/10.1101/gad.1047403
  2. M. G. Pomper, 'Molecular imaging: an overview,' Acad. Radiol., vol. 8, pp. 1141-1153, 2001 https://doi.org/10.1016/S1076-6332(03)80728-6
  3. E. L. Ritman, 'Molecular imaging in small animals-roles for micro-CT,' J. Cell. Biochem. Supp., vol. 36, pp. 116-224, 2002
  4. T. Hildebrand and P. Ruegsegger, 'A new method for the modelindependent assessment of thickness in three-dimensional images,' J. Micros., vol. 185, Pt. 1, pp. 67-75, 1997 https://doi.org/10.1046/j.1365-2818.1997.1340694.x
  5. J. S. Thomsen, A. Laib, B. Koller, S. Prohaska, LI. Mosekilde and W. Gowin, 'Stereological measures of trabecular bone structure: comparison of 3D micro computed tomography with 2D histological sections in human proximal tibial bone biopsies,' J. Micros., vol. 218, Pt. 2, pp. 171-179, 2005 https://doi.org/10.1111/j.1365-2818.2005.01469.x
  6. A. Odgaard, 'Three-dimensional methods for quantification of cancellous bone architecture,' Bone, vol. 20, no. 4 pp. 315-328, 1997 https://doi.org/10.1016/S8756-3282(97)00007-0
  7. A. Barbier, C. Martel, M. C. De Vernejoulm F. Tirode, M. Nys, G. Mocaer, C. Morieux, H. Murakami and F. Lacheretz, 'The visualization and evaluation of bone architecture in the rat using three-dimensional x-ray microcomputed tomography,' J. Bone and Min. Research, vol. 17, pp 37-44, 1999 https://doi.org/10.1007/s007740050061
  8. P. Ruegsegger, B. Koller and R. Muller, 'A microtomographic system for the nondestructive evaluation of bone architecture,' Calcified Tissue Int., vol. 58, pp. 24-29, 1996 https://doi.org/10.1007/BF02509542
  9. J.H. Kinney, N.E. Lane, D.L. Haupt, 'In vivo, three-dimensional microscopy of trabecular bone,' J. Bone and Min. Research, vol.10, pp. 264-270, 1995
  10. D. A. Jaffray and J. H. Siewerdsen, 'Cone-beam computed tomography with a flat-panel imager: Initial performance characterization,' Med. Phys., vol. 27, no. 6, pp. 1311-1323, 2000 https://doi.org/10.1118/1.599009
  11. R. Ning, 'Flat panel detector-based cone-beam volume CT angiography imaging: System evaluation,' IEEE Trans. Med. Imaging, vol. 19, pp. 949-963, 2000 https://doi.org/10.1109/42.887842
  12. S. C. Lee, H. K. Kim, I. K. Chun, S. Y. Lee and M. H. Cho, 'A flat panel detector based micro-CT system: performance evaluation for small-animal imaging,' Phys. Med. Biol., vol. 48, pp. 4173-4185, 2003 https://doi.org/10.1088/0031-9155/48/24/014
  13. I. K. Chun, M. H. Cho, S. C. Lee, M. H. Cho and S. Y. Lee, 'X-ray micro-tomography system for small-animal imaging with zoom-in imaging capability,' Phys. Med. Biol., vol. 49, pp. 3889-3902, 2004 https://doi.org/10.1088/0031-9155/49/17/005
  14. S. Y. Lee, M. H. Cho, S. C. Lee, I. K. Chun, J. J. Park, 'Small animal x-ray micro-CT with zoom-in imaging capability,' in Proc. 15th IEEE Trans. Med. Imag., San Diego, Oct. 2006
  15. L. A. Feldkamp, L. C. Davis and J. W. Kress, 'Practical conebeam algorithm,' J. Opt. Soc. Am. A, vol. 1 no. 6, pp.612-619, 1984 https://doi.org/10.1364/JOSAA.1.000612
  16. R. M. Lewitt and R. H. T. Bates, 'Image reconstruction from projection-III : Projection completion method(theory) ,' Optik, vol. 50, pp.180-204, 1978