PET-CT Normalization, Well Counter Correction에 따른 팬텀을 이용한 영상 평가 (Evaluation of Image for Phantom according to Normalization, Well Counter Correction in PET-CT)

  • 이충운;유연욱;문종운;김윤철
    • 핵의학기술
    • /
    • 제27권1호
    • /
    • pp.47-54
    • /
    • 2023
  • Purpose PET-CT imaging require an appropriate quality assurance system to achieve high efficiency and reliability. Quality control is essential for improving the quality of care and patient safety. Currently, there are performance evaluation methods of UN2-1994 and UN2-2001 proposed by NEMA and IEC for PET-CT image evaluation. In this study, we compare phantom images with the same experiments before and after PET-CT 3D normalization and well counter correction and evaluate the usefulness of quality control. Materials and methods Discovery 690 (General Electric Healthcare, USA) PET-CT equiptment was used to perform 3D normalization and well counter correction as recommended by GE Healthcare. Based on the recovery coefficients for the six spheres of the NEMA IEC Body Phantom recommended by the EARL. 20kBq/㎖ of 18F was injected into the sphere of the phantom and 2kBq/㎖ of 18F was injected into the body of phantom. PET-CT scan was performed with a radioacitivity ratio of 10:1. Images were reconstructed by appliying TOF+PSF+TOF, OSEM+PSF, OSEM and Gaussian filter 4.0, 4.5, 5.0, 5.5, 6.0, 6,5 mm with matrix size 128×128, slice thickness 3.75 mm, iteration 2, subset 16 conditions. The PET image was attenuation corrected using the CT images and analyzed using software program AW 4.7 (General Electric Healthcare, USA). The ROI was set to fit 6 spheres in the CT image, RC (Recovery Coefficient) was measured after fusion of PET and CT. Statistical analysis was performed wilcoxon signed rank test using R. Results Overall, after the quality control items were performed, the recovery coefficient of the phantom image increased and measured. Recovery coefficient according to the image reconstruction increased in the order TOF+PSF, TOF, OSEM+PSF, before and after quality control, RCmax increased by OSEM 0.13, OSEM+PSF 0.16, TOF 0.16, TOF+PSF 0.15 and RCmean increased by OSEM 0.09, OSEM+PSF 0.09, TOF 0.106, TOF+PSF 0.10. Both groups showed a statistically significant difference in Wilcoxon signed rank test results (P value<0.001). Conclusion PET-CT system require quality assurance to achieve high efficiency and reliability. Standardized intervals and procedures should be followed for quality control. We hope that this study will be a good opportunity to think about the importance of quality control in PET-CT

  • PDF

201Tl을 이용한 심근관류 SPECT에서 재구성 방법에 따른 작은 용적 심장의 정량 지표 변화 (Quantitative Indices of Small Heart According to Reconstruction Method of Myocardial Perfusion SPECT Using the 201Tl)

  • 김성환;류재광;윤순상;김은혜
    • 핵의학기술
    • /
    • 제17권1호
    • /
    • pp.18-24
    • /
    • 2013
  • $^{201}Tl$을 이용한 심근관류 SPECT 검사는 좌심실의 생존능 및 심장 기능의 정량적 평가를 함에 있어 중요한 방법으로서 현재 영상의 질을 향상시키기 위해 다양한 재구성 방법들이 이용 되고 있다. 하지만 작은 용적 심장에서는 부분용적효과로 인해 재구성 단계에서 정량 지표 값의 오류를 야기 할 수 있으므로 항상 주의 해야 한다. 이에 본 연구는 심근관류 SPECT 검사의 재구성 방법에 따른 좌심실의 정량적 지표 값을 심장 초음파와 서로 비교 함으로써 그 차이의 정도를 확인 한다. 2012년 2월부터 9월까지 본원에 내원하여 심근관류 SPECT 및 심장 초음파 검사를 실시한 278명의 환자(남자 90명, 여자 188명, 평균$65.5{\pm}11.1$세)를 심장 초음파의 ESV 30 mL를 기준으로 삼아 그 이하를 작은 용적 심장, 그 이상을 보통 또는 큰 용적심장으로 구분하였다. 각각 여과 후 FBP 및 OSEM의 방법을 적용하여 EDV, ESV 그리고 LVEF를 산출하였으며, 이를 심장 초음파에서 측정된 지표들과 함께 반복측정 분산분석 방법(Repeated Measures ANOVA)으로 분석하였다. 남녀 간의 EDV는 FBP, OSEM 간 유의한 차이가 없었으나 (p=0.053, p=0.098), 심장 초음파와의 비교에서는 유의한 차이를 보였다(p<0.001). ESV의 변화는 특히 작은 용적 심장을 가진 여성에서 FBP, OSEM, 심장 초음파 모두 유의한 차이(p<0.001)를 보였다. 또한 LVEF에서도 보통 용적 심장을 가진 남녀 모두 FBP, OSEM, 심장 초음파 간 유의한 차이는 보이지 않았으나(p=0.375, p=0.969), 작은 용적 심장을 가진 여성에서 모두 유의한 차이(p<0.001)를 보였다. 핵의학 영상 재구성 방법 간 좌심실의 정량적 지표 값의 변화는 보통 용적 심장을 가진 환자에서는 유의한 차이를 발견 할수 없었으나, ESV를 기준으로 30 mL 이하의 작은 용적 심장, 특히 여성에서는 FBP, IR_OSEM, 심장 초음파 간 유사한 차이를 확인 할 수 있었다. 하지만 이러한 차이는 분석에 사용된 3종류의 모든 감마 카메라에서 OSEM 적용 시FBP 보다 평균적으로 심장 초음파와의 오차가 적은 LVEF값이 산출됨을 확인 할 수 있었다.

  • PDF

심장 CT 검사시 ASIR 적용에 따른 선량 평가 (Radiation dose Assesment according to the Adaptive Statistical Iterative Reconstruction Technique of Cardiac Computed Tomography(CT))

  • 장현철;김현주;조재환
    • 한국콘텐츠학회논문지
    • /
    • 제11권5호
    • /
    • pp.252-259
    • /
    • 2011
  • 심장 CT 검사시 적응식 통계적 반복 재구성법을 이용하여 체질량 지수에 따른 관전압 변화에 대한 방사선 피폭선량에 대하여 알아보고자 하였다. 심장 CT 검사를 시행 받은 환자를 BMI에 따른 관전압에 따라 네 개의 군으로 나누어[A군(n=20), Non-ASIR, BMI < 25, 100 kVp; B군(n=20), Non-ASIR, BMI > 25, 120 kVp; C군(n=20), 40% ASIR BMI < 25, 100 kVp; D군(n=20), 40% ASIR, BMI > 25, 120 kVp] 대동맥 중심부와 우관상동맥, 좌전하행동맥에 관심영역을 설정 한 후 CT값(number)측정하여 평균값과 표준편차를 분석하였다. 영상 잡음은 A군과 C군 사이에는 통계적으로 유의한 차이가 있었으며, A군이 C군보다 노이즈가 유의하게 높았다(group A, 494 ${\pm}$ 32 HU; group C, 482 ${\pm}$ 48 HU: P<0.05). 또한, B군과 D군 사이에는 통계적으로 유의한 차이가 있었으며, B군이 D군보다 노이즈가 유의하게 높았다(group B, 510 ${\pm}$ 45 HU; group D, 480 ${\pm}$ 82 HU: P<0.05). 영상의 정성적 분석에서 관상동맥 분절별로 임상평가 한 결과 평균값은 A군은 4.13${\pm}$0.2, B군은4.18${\pm}$0.1, C군은 4.1${\pm}$0.2, D군은 4.15${\pm}$0.1로 A군, B군, C군, D군 모두 통계적으로 유의한 차이가 없었으며(P>0.05), 모든 군에서 진단에 적절한 영상을 보였다. 피폭선량은 A군은 8.6${\pm}$0.9, B군은 14.9${\pm}$0.4, C군은 5.8${\pm}$0.5, D군은 10.1${\pm}$0.6 mSv 로 나타났다.

PET/CT 장비와 영상 재구성 차이에 따른 EQ PET을 이용한 표준섭취계수의 평가 (Evaluation of Standardized Uptake Value applying EQ PET across different PET/CT scanners and reconstruction)

  • 윤석환;김병진;문일상;이홍재
    • 핵의학기술
    • /
    • 제22권1호
    • /
    • pp.35-42
    • /
    • 2018
  • 양전자 방출 단층촬영에서 정량적 평가에 통상적으로 쓰이는 표준섭취계수(SUV : Standardized Uptake Value)는 종양의 병소와 병기의 진단 그리고 치료 성적 평가에 있어서 사용되고 있는 유용한 지표이다. 하지만 SUV는 환자의 체격, 검사까지의 시간, 부분용적 효과, 관심영역의 설정, 검사장비, 영상재구성 차이 등에 따라 변화 되므로 다양한 인자의 영향을 고려하여야만 한다. 그 중에서도 PET/CT 장비와 영상의 재구성 방법의 차이에 의한 SUV의 차이를 SIEMENS 사의 EQ PET을 이용하여 방사능의 차이를 보정 할 수 있게 되었다. 그러므로 본 연구에서는 phantom 실험과 FDG PET 임상영상의 SUVmax를 비교 하여 EQ PET을 적용함으로써, SUV의 변화를 평가 하고 EQ PET의 유용성에 대해서 검증해보고자 하였다. 본원의 3대의 PET/CT 장비인 Biograph true point 40, Biograph mCT 40, Biograph mCT 64 장비를 이용하여 $^{18}F-FDG$를 주입 한 NEMA IEC body phantom 영상을 획득 한 후 OSEM3D+PSF, OSEM3D+TOF, OSEM3D+PSF+TOF의 알고리즘을 이용하여 영상을 재구성 하였다. 각각 재구성된 영상에서 관심영역의 방사능 농도를 측정한 후, EARL에서 권고하는 NEMA IEC body phantom의 방사능 농도의 회복계수 값을 비교하여 서로 다른 PET system 간에 방사능 농도의 차이를 줄일 수 있는 EQ filter 값을 산출 하였다. 산출한 EQ Filter 값을 장비와 영상 재구성에 따라 적용하여 팬텀의 회복계수와 61명의 폐암환자의 FDG PET 영상에서 종양의 SUVmax를 비교 분석 하였다. 3대의 PET/CT 장비에서 영상의 재구성 알고리즘을 달리하여 획득한 phantom의 영상의 6개구의 평균 변동계수는 EQ PET 적용 전 후의 값은 OSEM3D 재구성 영상에서 각각 0.05, 0.04 OSEM3D+TOF 재구성 영상에서는 각각 0.04, 0.03, OSEM3D+PSF 재구성 영상에서는 각각 0.04, 0.03, OSEM3D+PSF+TOF 재구성 영상에서는 각각 0.03, 0.02 값을 보여 EQ PET 적용 후 팬텀의 6개 구의 평균 변동계수는 감소하였다. 임상 영상 비교 에서는 폐암환자 종양의 SUVmax는 OSEM3D, OSEM3D+TOF, OSEM3D+PSF, OSEM3D+PSF+TOF 순으로 증가하였으며, EQ PET 적용 후 재구성 영상에 대한 SUVmax는 통계적으로 유의한 차이가 없었다(p=1.000). PET/CT 영상에서 서로 다른 PET/CT 장비와 영상의 재구성 방법의 차이에 의해 발생하는 방사능 농도의 차이를 EQ PET을 이용하여 측정함으로서 장비와 영상 재구성별 방사능 농도의 편차가 감소되었다. PET 영상에서 정량적 평가에 이용되는 SUV의 편차를 줄일 수 있어 종양의 병소와 병기의 진단, 그리고 치료 성적평가의 정량적 평가 신뢰도가 향상 될 것으로 사료된다.

ASIR를 이용한 두부 CT의 영상 잡음 평가 및 피폭선량 분석 (Evaluation of Image Noise and Radiation Dose Analysis In Brain CT Using ASIR(Adaptive Statistical Iterative Reconstruction))

  • 장현철;김경근;조재환;서정민;이행기
    • 한국방사선학회논문지
    • /
    • 제6권5호
    • /
    • pp.357-363
    • /
    • 2012
  • 본 연구는 두부 컴퓨터 단층 촬영 검사 시 적응식 통계적 반복 재구성법인 알고리즘을 적용하여 노이즈 및 화질평가, 피폭선량의 감소에 대하여 알아보고자 하였다. 두부 CT 검사 시 ASIR를 적용하지 않은 군[A군], ASIR 50 % 적용한군 [B군]으로 나누어 검사하였다. 팬텀연구에서 측정된 CT 노이즈 평균값의 측정결과는 B군이 A군보다 중심부(A)와 주변부(B, C, D)에서 각 각 46.9 %, 48.2 %, 43.2 %, 47.9 %가 감소되었다. 영상화질 평가에서 정량적 분석방법으로 CT 값(number)을 측정하여 잡음(noise) 정도를 분석하였다. 영상 잡음은 A군과 B군 사이에는 통계적으로 유의한 차이가 있었으며, A군이 B군보다 영상 잡음이 유의하게 높았다(group A ;우엽에서 31.87 HU, 좌엽에서 31.78 HU, group B ; 우엽에서 26.6 HU, 좌엽에서 30.42 HU : P<0.05). 영상의 정성적 평가방법으로 두부 임상 영상 평가표에 의해 평가한 결과 80점 만점에 A군의 관찰자 1의 점수는 73.17 점, 관찰자 2의 점수는 74.2 점으로 평가하였으며, B군의 관찰자 1의 점수는 71.77 점, 관찰자 2의 점수는 72.47 점으로 평가하였다. 통계적으로 유의한 차이가 없었으며(P>0.05), 진단에 적절한 영상을 보였다. 피폭선량은 ASIR 50 % 적용하여 검사함으로써 영상의 질적 저하 없이 방사선 피폭선량을 47.6 % 감소 시킬수 있었다. 결론적으로 임상 부위에 ASIR가 적용이 된다면 훨씬 더 적은 선량으로도 검사가 가능할 것으로 사료되며, 검사 시에 검사자가 판단하는데 있어 긍정적인 요인이 될 것으로 사료된다.

Evaluation of the Accuracy of Distance Measurements on 3D Volume-rendered Image of Human Skull Using Multi-detector CT: Effects of Acquisition Section Thickness and Reconstruction Section Thickness

  • Haijo Jung;Kim, Hee-Joung;Lee, Sang-Ho;Kim, Dong-Wook;Soonil Hong;Kim, Dong-Hyeon;Son, Hye-Kyung;Wonsuk Kang;Kim, Kee-Deog
    • 한국의학물리학회:학술대회논문집
    • /
    • 한국의학물리학회 2002년도 Proceedings
    • /
    • pp.457-460
    • /
    • 2002
  • The image quality of three-dimensional (3D) images has been widely investigated by the qualitative analysis method. A need remains for an objective and quantitative method to assess the image quality of 3D volume-rendered images. The purpose of this study was to evaluate the quantitative accuracy of distance measurements on 3D volume-rendered images of a dry human skull by using multi-detector computed tomography (MDCT). A radiologist measured five times the twenty-one direct measurement line items composed among twelve reference points on the skull surface with a digital vernier caliper. The water filled skull specimen was scanned with a MDCT according to the section thicknesses of 1.25, 2.50, 3.75, and 5.00 mm for helical (high quality; pitch 3:1) scan mode. MDCT data were reconstructed with its acquisition section thickness and with 1.25 mm section thickness for all scans. An observer also measured seven times the corresponding items on 3D volume-rendered images with measuring tools provided by volumetric analysis software. The quantitative accuracy of distance measurements on the 3D volume-rendered images was statistically evaluated (p-value < 0.05) by comparatively analyzing these measurements with the direct distance measurements. The accuracy of distance measurements on the 3D volume-rendered MDCT images acquired with 1.25, 2.50, 3,75 and 5.00 mm section thickness and reconstructed with its section thickness were 48%, 33%, 23%, and 14%, respectively. Meanwhile, there were insignificant statistical differences in accuracy of distance measurements among 3D volume-rendered images reconstructed with 1.25 mm section thickness for the each acquisition section thickness. MDCT images acquired with thick section thickness and reconstructed with thin section thickness in helical scan mode should be effectively used in medical planning of 3D volume-rendered images. The quantitative analysis of distance measurement may be a useful tool for evaluating the quantitative accuracy and the defining optimal parameters of 3D volume-rendered CT images.

  • PDF

PET/CT 검사에서 Q.Clear 기법의 유용성에 대한 고찰 (The Usefulness of Q.Clear Technique in PET / CT)

  • 최용훈;김정열;최영숙;임한상;김재삼
    • 핵의학기술
    • /
    • 제21권2호
    • /
    • pp.31-36
    • /
    • 2017
  • 최근 PET/CT 장비의 성능의 발전과 다양한 기법의 개발로 민감도와 해상도등 영상 품질을 개선할 수 있게 되었다. 본 논문에서는 GE사의 Discovery IQ 장비의 Q.Clear (a fully convergent iterative reconstruction) 기법을 이용하여 영상의 질 향상에 유용성이 있는지 알아보고자 한다. 장비는 Discovery IQ (GE Healthcare, MI, USA)를 사용하였다. NEMA IEC Body Phantom의 배후방사능과 열소 체적(10 mm, 13 mm, 17 mm, 22 mm)의 비를 1:4로 하고 3분간 촬영하여 VPHDs (VUE Point High-Definition SharpIR)와 Q.Clear의 대조도를 비교 분석하였다. PET/SPECT Performance Phantom에 $^{18}F-FDG$를 187 MBq을 주입 후 4분간 촬영하여 해상도와 균일도를 비교 분석하였다. 그리고 100명의 임상 환자에서 질환의 종류와 상관없이 2 cm 미만의 작은 병소의 SUVmax를 측정하여 t-test 통계분석하였다. NEMA IEC Body Phantom에서 VPHDs와 Q.Clear의 대조도가 $63.6{\pm}5.7%$, $75{\pm}4.8%$로 나왔고 PET/SPECT Performance Phantom에서 해상도는 VPHDs가 9.2 mm, Q.Clear가 7.3 mm로 나왔다. 균일도는 Q.Clear가 10.8% 더 우수하였다. 임상 환자의 t-test 통계 결과 p-value가 0.021로 유의한 차이가 있었다. 임상환자에서 SUVmax는 Q.Clear에서 높게 측정 되었으며, 신호대 잡음 비도 우수하였다. 이는 부분체적효과의 영향을 줄였기 때문으로 볼 수 있다. Phantom test와 임상 환자의 결과 모두 Q.Clear를 적용 하였을 때 영상품질이 향상된 것을 확인하였다. 이러한 영상 품질 향상은 병소를 더욱 정확하게 발견할 수 있고 나아가 선량저감과 환자평가 그리고 영상 분석 등 다양한 방면에서 활용할 수 있을 것으로 사료된다.

  • PDF

국내 라이선스 패션잡지 표지에 나타난 패션특성의 비교분석 - ELLE, VOGUE, W를 중심으로 - (Comparative Analysis of Fashion Characteristics on the Cover of Domestic Licensed Fashion Magazines - Focused on ELLE, VOGUE, W -)

  • 이현지;이경희
    • 한국의류산업학회지
    • /
    • 제21권1호
    • /
    • pp.1-12
    • /
    • 2019
  • The purpose of this study is to examine the fashion characteristics of fashion magazine cover by comparing and analyzing the formative characteristics of fashion, visual design characteristics and illustration vocabulary on the cover of 3 fashion magazines. The data analysis criteria consisted of the formative elements of fashion (fashion design element, fashion coordination element) and visual design element (color, illustration lexical layout, model photograph type). Data analysis methods were statistical analysis, stepwise lexical analysis, and content analysis. The results of the study are as follows. First, the formative characteristics of fashion on the cover of fashion magazines show that ELLE is a feminine and elegant characteristics, VOGUE is a modern, chic and mannish characteristics, and W is avant-garde and neutral characteristics. Second, visual design characteristics on the cover of fashion magazines, ELLE and VOGUE use modern and simple modern sensibility by using monotonous background color and background color number, and W showed original image characteristic by using various colors. Third, as a result of the illustration lexical analysis on the cover of fashion magazines, 4 core keywords of trend, star, event, and life appeared in 3 magazines in common. Elle differentiates by innovation, Vogue by discrimination, W by reconstruction.

CT 및 초음파 스캔 데이터를 이용한 체적 측정의 정확도 및 유용성 (Accuracy and Usefulness of Volume Measurement using CT and Ultrasound Scan Data)

  • 김현주;이후민;윤준
    • 한국방사선학회논문지
    • /
    • 제16권3호
    • /
    • pp.289-294
    • /
    • 2022
  • 본 연구는 CT와 USG 스캔 데이터를 이용하여 팬텀 실험과 환자 스캔 데이터를 활용하는 임상실험으로 체적 측정의 정확도 및 유용성을 알아보았다. 그 결과 CT와 USG 스캔 데이터 모두 다양한 체적의 실제 원형 팬텀의 체적과 유의한 차이가 없었으며(p>0.05), 동일 환자의 방광 체적 측정 결과 CT 데이터는 236.9±2.1, USG 영상은 236.9±1로 측정되어 통계분석 결과 유의한 차이가 없는 것으로 분석되었다(p>0.05). 본 결과의 임상 적용은 추가 임상실험이 필요하지만, 환자 피폭, 검사자의 스캔 기술, 그리고 CT 재구성 경험 등을 고려하여 CT 또는 초음파 장치를 선택 적용한다면 의료영상을 활용한 체적 측정의 유용성 측면에서 기초자료로 적용 가치가 있을 것으로 사료된다.

L-spine Bone SPECT/CT에서 획득된 저선량 CT 영상을 이용한 용적 골밀도 결과의 유용성 (Usefulness of volumetric BMD measurement by using low dose CT image acquired on L-spine Bone SPECT/CT)

  • 고현수;박순기;김은혜;최종숙;정우영;이동윤
    • 핵의학기술
    • /
    • 제27권2호
    • /
    • pp.99-109
    • /
    • 2023
  • Purpose: CT scan makes up for the weak point of the nuclear medicine image having a low resolution and also were used for attenuation correction on image reconstruction. Recently, many studies try to make use of CT images additionally, one of them is to measure the bone mineral density(BMD) using Quantitative CT(QCT) software. BMD exams are performed to scan lumbar and femur with DXA(Dual-Energy X-Ray Absorptiometry) in order to diagnose bone disease such as osteopenia, osteoporosis. The purpose of this study is to identify the usefulness of QCT_BMD analyzed with low dose CT images on L-spine Bone SPECT/CT comparing with DXA_BMD. Materials and Methods: Fifty five women over 50 years old (mean 66.4 ± 9.1) who took the both examinations(L-spine Bone SPECT/CT with SIEMENS Intevo 16 and DXA scan with GE Lunar prodigy advance) within 90 days from April 2017 to July 2022, BMD, T-score and disease classification were analyzed. Three-dimensional BMD was analyzed with low dose CT images acquired on L-spine Bone SPECT/CT scan on Mindways QCT PROTM software and two-dimensional BMD was analyzed on DXA scan. Basically, Lumbar 1-4 were analyzed and the patients who has lesion or spine implants on L-spine were excluded for this study. Pearson's correlation analysis was performed in BMD and T-score, chi-square test was performed in disease classification between QCT and DXA. Results: On 55 patients, the minimum of QCT_BMD was 18.10, maximum was 166.50, average was 82.71 ± 31.5 mg/cm3. And the minimum of DXA-BMD was 0.540, maximum was 1.302, average was 0.902 ± 0.201 g/cm2, respectively. The result shows a strong statistical correlation between QCT_BMD and DXA_BMD(p<0.001, r=0.76). The minimum of QCT_T-score was -5.7, maximum was -0.1, average was -3.2 ± 1.3 and the minimum of DXA_T-score was -5.0, maximum was 1.7, average was -2.0 ± 1.3, respectively. The result shows a statistical correlation between QCT T-score and DXA T-score (p<0.001, r=0.66). On the disease classification, normal was 5, osteopenia was 25, osteoporosis was 25 in QCT and normal was 10, osteopenia was 25, osteoporosis was 20 in DXA. There was under-estimation of bone decrease relatively on DXA than QCT, but there was no significant differences statistically by chi-square test between QCT and DXA. Conclusion: Through this study, we could identify that the QCT measurement with low dose CT images QCT from L-Spine Bone SPECT/CT was reliable because of a strong statistical correlation between QCT_BMD and DXA_BMD. Bone SPECT/CT scan can provide three-dimensional information also BMD measurement with CT images. In the future, rather than various exams such as CT, BMD, Bone scan are performed, it will be possible to provide multipurpose information via only SPECT/CT scan. In addition, it will be very helpful clinically in the sense that we can provide a diagnosis of potential osteoporosis, especially in middle-aged patients.