• Title/Summary/Keyword: pedicle screws

Search Result 66, Processing Time 0.027 seconds

Short-segment Pedicle Instrumentation of Thoracolumbar Burst-compression Fractures; Short Term Follow-up Results

  • Shin, Tae-Sob;Kim, Hyun-Woo;Park, Keung-Suk;Kim, Jae-Myung;Jung, Chul-Ku
    • Journal of Korean Neurosurgical Society
    • /
    • v.42 no.4
    • /
    • pp.265-270
    • /
    • 2007
  • Objective : The current literature implies that the use of short-segment pedicle screw fixation for spinal fractures is dangerous and inappropriate because of its high failure rate, but favorable results have been reported. The purpose of this study is to report the short term results of thoracolumbar burst and compression fractures treated with short-segment pedicle instrumentation. Methods : A retrospective review of all surgically managed thoracolumbar fractures during six years were performed. The 19 surgically managed patients were instrumented by the short-segment technique. Patients' charts, operation notes, preoperative and postoperative radiographs (sagittal index, sagittal plane kyphosis, anterior body compression, vertebral kyphosis, regional kyphosis), computed tomography scans, neurological findings (Frankel functional classification), and follow-up records up to 12-month follow-up were reviewed. Results : No patients showed an increase in neurological deficit. A statistically significant difference existed between the patients preoperative, postoperative and follow-up sagittal index, sagittal plane kyphosis, anterior body compression, vertebral kyphosis and regional kyphosis. One screw pullout resulted in kyphotic angulation, one screw was misplaced and one patient suffered angulation of the proximal segment on follow-up, but these findings were not related to the radiographic findings. Significant bending of screws or hardware breakage were not encountered. Conclusion : Although long term follow-up evaluation needs to verified, the short term follow-up results suggest a favorable outcome for short-segment instrumentation. When applied to patients with isolated spinal fractures who were cooperative with 3-4 months of spinal bracing, short-segment pedicle screw fixation using the posterior approach seems to provide satisfactory result.

Surgical Treatment of Osteoporotic Vertebral Compression Fractures at Thoraco-Lumbar Levels: Only Pedicle Screw Constructs with Polymethylmethacrylate Augmentation (흉요추부 골다공증성 척추 압박 골절의 수술적 치료: 골시멘트 보강술을 이용한 척추경 나사 고정)

  • Jun, Deuk Soo;Baik, Jong-Min;Park, Ji Hyeon
    • Journal of the Korean Orthopaedic Association
    • /
    • v.54 no.4
    • /
    • pp.327-335
    • /
    • 2019
  • Purpose: To investigate the radiological efficacy of polymethylmethacrylate (PMMA) augmentation of pedicle screw operation in osteoporotic vertebral compression fractures (OVCF) patients. Materials and Methods: Twenty OVCF patients, who underwent only posterior fusion using pedicle screws with PMMA augmentation, were included in the study. The mean follow-up period was 15.6 months. The demographic data, bone mineral density (BMD), fusion segments, number of pedicle screws, and amount of PMMA were reviewed as medical records. To analyze the radiological outcomes, the radiologic parameters were measured as the time serial follow-up (preoperation, immediately postoperation, postoperation 6 weeks, 3, 6 months, and 1 year follow-up). Results: A total of 20 patients were examined (16 females [80.0%]; mean age, 69.1±8.9 years). The average BMD was -2.5±0.9 g/cm2. The average cement volume per vertebral body was 6.3 ml. The mean preoperative Cobb angle of focal kyphosis was 32.7°±7.0° and was improved significantly to 8.7°±6.9° postoperatively (p<0.001), with maintenance of the correction at the serial follow-up, postoperatively. The Cobb angle of instrumented kyphosis, wedge angle, and sagittal index showed similar patterns. In addition, the anterior part of fractured vertebral body height averaged 11.0±5.0 mm and was improved to 18.5±5.7 mm postoperatively (p=0.006), with maintenance of the improvement at the 3-month, 6-month, and 1-year follow-up. Conclusion: The reinforcement of pedicle screws using PMMA augmentation may be a feasible surgical technique for OVCF. Moreover, it appears to be appropriate for improving the focal thoracolumbar/lumbar kyphosis and is maintained well after surgery.

Reduction of Radiation Exposure by Modifying Imaging Manner and Fluoroscopic Settings during Percutaneous Pedicle Screw Insertion

  • Kim, Hyun Jun;Park, Eun Soo;Lee, Sang Ho;Park, Chan Hong;Chung, Seok Won
    • Journal of Korean Neurosurgical Society
    • /
    • v.64 no.6
    • /
    • pp.933-943
    • /
    • 2021
  • Objective : Percutaneous pedicle screw (PPS) fixation is a needle based procedure that requires fluoroscopic image guidance. Consequently, radiation exposure is inevitable for patients, surgeons, and operation room staff. We hypothesize that reducing the production of radiation emission will result in reduced radiation exposure for everyone in the operation room. Research was performed to evaluate reduction of radiation exposure by modifying imaging manner and mode of radiation source. Methods : A total of 170 patients (680 screws) who underwent fusion surgery with PPS fixation from September 2019 to March 2020 were analyzed in this study. Personal dosimeters (Polimaster Ltd.) were worn at the collar outside a lead apron to measure radiation exposure. Patients were assigned to four groups based on imaging manner of fluoroscopy and radiation modification (pulse mode with reduced dose) : continuous use without radiation modification (group 1, n=34), intermittent use without radiation modification (group 2, n=54), continuous use with radiation modification (group 3, n=26), and intermittent use with radiation modification (group 4, n=56). Post hoc Tukey Honest significant difference test was used for individual comparisons of radiation exposure/screw and fluoroscopic time/screw. Results : The average radiation exposure/screw was 71.45±45.75 µSv/screw for group 1, 18.77±11.51 µSv/screw for group 2, 19.58±7.00 µSv/screw for group 3, and 4.26±2.89 µSv/screw for group 4. By changing imaging manner from continuous multiple shot to intermittent single shot, 73.7% radiation reduction was achieved in the no radiation modification groups (groups 1, 2), and 78.2% radiation reduction was achieved in the radiation modification groups (groups 3, 4). Radiation source modification from continuous mode with standard dose to pulse mode with reduced dose resulted in 72.6% radiation reduction in continuous imaging groups (groups 1, 3) and 77.3% radiation reduction in intermittent imaging groups (groups 2, 4). The average radiation exposure/screw was reduced 94.1% by changing imaging manner and modifying radiation source from continuous imaging with standard fluoroscopy setting (group 1) to intermittent imaging with modified fluoroscopy setting (group 4). A total of 680 screws were reviewed postoperatively, and 99.3% (675) were evaluated as pedicle breach grade 0 (<2 mm). Conclusion : The average radiation exposure/screw for a spinal surgeon can be reduced 94.1% by changing imaging manner and modifying radiation source from real-time imaging with standard dose to intermittent imaging with modified dose. These modifications can be instantly applied to any procedure using fluoroscopic guidance and may reduce the overall radiation exposure of spine surgeons.

C7 Posterior Fixation Using Intralaminar Screws : Early Clinical and Radiographic Outcome

  • Jang, Sang-Hoon;Hong, Jae-Taek;Kim, Il-Sup;Yeo, In-Sung;Son, Byung-Chul;Lee, Sang-Won
    • Journal of Korean Neurosurgical Society
    • /
    • v.48 no.2
    • /
    • pp.129-133
    • /
    • 2010
  • Objective : The use of segmental instrumentation technique using pedicle screw has been increasingly popular in recent years owing to its biomechanical stability. Recently, intralaminar screws have been used as a potentially safer alternative to traditional fusion constructs involving fixation of C2 and the cervicothoracic junction including C7. However, to date, there have been few clinical series of C7 laminar screw fixation in the literature. Thus, the purpose of this study is to report our clinical experiences using C7 laminar screw and the early clinical outcome of this rather new fixation technique. Methods : Thirteen patients underwent C7 intralaminar fixation to treat lesions from trauma or degenerative disease. Seventeen intralaminar screws were placed at C7. The patients were assessed both clinically and radiographically with postoperative computed tomographic scans. Results : There was no violation of the screw into the spinal canal during the procedure and no neurological worsening or vascular injury from screw placement. The mean clinical and radiographic follow up was about 19 months, at which time there were no cases of screw pull-out, screw fracture or non-union. Complications included two cases of dorsal breech of intralaminar screw and one case of postoperative infection. Conclusion : Intralaminar screws can be potentially safe alternative technique for C7 fixation. Even though this technique cannot be used in the cases of C7 laminar fracture, large margin of safety and the ease of screw placement create a niche for this technique in the armamentarium of spine surgeons.

A Computed Tomography-Based Anatomic Comparison of Three Different Types of C7 Posterior Fixation Techniques : Pedicle, Intralaminar, and Lateral Mass Screws

  • Jang, Woo-Young;Kim, Il-Sup;Lee, Ho-Jin;Sung, Jae-Hoon;Lee, Sang-Won;Hong, Jae-Taek
    • Journal of Korean Neurosurgical Society
    • /
    • v.50 no.3
    • /
    • pp.166-172
    • /
    • 2011
  • Objective : The intralaminar screw (ILS) fixation technique offers an alternative to pedicle screw (PS) and lateral mass screw (LMS) fixation in the C7 spine. Although cadaveric studies have described the anatomy of the pedicles, laminae, and lateral masses at C7, 3-dimensional computed tomography (CT) imaging is the modality of choice for pre-surgical planning. In this study, the goal was to determine the anatomical parameter and optimal screw trajectory for ILS placement at C7, and to compare this information to PS and LMS placement in the C7 spine as determined by CT evaluation. Methods : A total of 120 patients (60 men and 60 women) with an average age of $51.7{\pm}13.6$ years were selected by retrospective review of a trauma registry database over a 2-year period. Patients were included in the study if they were older than 15 years of age, had standardized axial bone-window CT imaging at C7, and had no evidence of spinal trauma. For each lamina and pedicle, width (outer cortical and inner cancellous), maximal screw length, and optimal screw trajectory were measured, and the maximal screw length of the lateral mass were measured using m-view 5.4 software. Statistical analysis was performed using Student's t-test. Results : At C7, the maximal PS length was significantly greater than the ILS and LMS length (PS, $33.9{\pm}3.1$ mm; ILS, $30.8{\pm}3.1$ mm; LMS, $10.6{\pm}1.3$; p<0.01). When the outer cortical and inner cancellous width was compared between the pedicle and lamina, the mean pedicle outer cortical width at C7 was wider than the lamina by an average of 0.6 mm (pedicle, $6.8{\pm}1.2$ mm; lamina, $6.2{\pm}1.2$ mm; p<0.01). At C7, 95.8% of the laminae measured accepted a 4.0-mm screw with a 1.0 mm of clearance, compared with 99.2% of pedicle. Of the laminae measured, 99.2% accepted a 3.5-mm screw with a 1.0 mm clearance, compared with 100% of the pedicle. When the outer cortical and inner cancellous height was compared between pedicle and lamina, the mean lamina outer cortical height at C7 was wider than the pedicle by an average of 9.9 mm (lamina, $18.6{\pm}2.0$ mm; pedicle, $8.7{\pm}1.3$ mm; p<0.01). The ideal screw trajectory at C7 was also measured ($47.8{\pm}4.8^{\circ}$ for ILS and $35.1{\pm}8.1^{\circ}$ for PS). Conclusion : Although pedicle screw fixation is the most ideal instrumentation method for C7 fixation with respect to length and cortical diameter, anatomical aspect of C7 lamina is affordable to place screw. Therefore, the C7 intralaminar screw could be an alternative fixation technique with few anatomic limitations in the cases when C7 pedicle screw fixation is not favorable. However, anatomical variations in the length and width must be considered when placing an intralaminar or pedicle screw at C7.

The Effect of Postoperative Use of Teriparatide Reducing Screw Loosening in Osteoporotic Patients

  • Kim, Jae Wook;Park, Seung Won;Kim, Young Baeg;Ko, Myeong Jin
    • Journal of Korean Neurosurgical Society
    • /
    • v.61 no.4
    • /
    • pp.494-502
    • /
    • 2018
  • Objective : The loosening of pedicle screws (PS) is one of the frequent problems of spinal surgery in the patients with osteoporosis. Previous studies had revealed that intermittent injection of teriparatide could reduce PS loosening by improving bone mass and quality when their patients took parathyroid hormone for a considerable duration before surgery. However, although the teriparatide is usually used after spine surgery in most clinical situations, there was no report on the efficacy of teriparatide treatment started after spine surgery. The purpose of this retrospective study was to examine the efficacy of teriparatide treatment started immediately after lumbar spinal surgery to prevent pedicle screw loosening in patients with osteoporosis. Methods : We included 84 patients with osteoporosis and degenerative lumbar disease who underwent transforaminal interbody fusion and PS fixation and received parathyroid hormone or bisphosphonate (BP) postoperatively. They were divided into teriparatide group (daily injection of $20{\mu}g$ of teriparatide for 6 months, 33 patients, 172 screws) and BP group (weekly oral administration of 35 mg of risedronate, 51 patients, 262 screws). Both groups received calcium (500 mg/day) and cholecalciferol (1000 IU/day) together. The screw loosening was evaluated with simple radiographic exams at 6 and 12 months after the surgery. We counted the number of patients with PS loosening and the number of loosened PS, and compared them between the two groups. Clinical outcomes were evaluated using visual analog scale (VAS) and Oswestry disability index (ODI) preoperatively, and at 12 months after surgery. Results : There was no significant difference in the age, sex, diabetes, smoking, bone mineral density, body mass index, and the number of fusion levels between the two groups. The number of PS loosening within 6 months after surgery did not show a significant difference between the teriparatide group (6.9%, 12/172) and the BP group (6.8%, 18/272). However, during 6-12 months after surgery, it was significantly lower in the teriparatide group (2.3%, 4/172) than the BP group (9.2%, 24/272) (p<0.05). There was no significant difference in the number of patients showing PS loosening between the teriparatide and BP groups. The teriparatide group showed a significantly higher degree of improvement of the bone mineral density (T-score) than that of BP group (p<0.05). There was no significant difference in the pre- and post-operative VAS and ODI between the groups. Conclusion : Our data suggest that the teriparatide treatment starting immediately after lumbar spinal fusion surgery could reduce PS loosening compared to BP.

The Effect of Distal Hooks in Thoracolumbar Fusion Using a Pedicle Screw in Elderly Patients (척추경 나사못을 이용한 고령 환자의 흉요추부 유합에서 원위부 갈고리의 효과)

  • Lee, Dong-Hyun;Kim, Sung-Soo;Kim, Jung-Hoon;Lim, Dong-Ju;Choi, Byung-Wan;Kim, Jin-Hwan;Kim, Jin-Hyok;Park, Byung-Ook
    • Journal of the Korean Orthopaedic Association
    • /
    • v.52 no.1
    • /
    • pp.83-91
    • /
    • 2017
  • Purpose: To investigate the clinical outcomes of distal hook augmentation using a pedicle screw in thoracolumbar fusion in elderly patients. Materials and Methods: This retrospective multicenter study recruited 20 patients aged 65 years or older, who underwent anterior support and long level posterior fusion in the thoracolumbar junction with a follow-up of one year. To assess the effect of distal hook augmentation, the patients were divided into two groups; the pedicle screw with hook group (PH group, n=10) and the pedicle screw alone group (PA group, n=10). Results: The average age was 72.4 years (65-83 years). The average fusion segment was 4.6 segments (3-6 segments). There were no significant differences in age, sex, causative diseases, bone mineral density of lumbar and proximal femur, number of patients with osteoporosis, and number of fused segments between the two groups (p≥0.05). At 1 year follow-up after surgery, parameters related with distal screw pullout were significantly worse in the PA group. No patients in the PH group had distal screw pullout. However, six patients (60%, 6/10) in the PA group had distal screw pullout. There were no significant differences in the progression of distal junctional kyphosis between the two groups. Conclusion: Distal hook augmentation is an effective procedure in protecting distal pedicle screws against the pullout when long level thoracolumbar fusion was performed in elderly patients aged 65 years or older.

Combination of Anterior and Posterior Subcutaneous Internal Fixation for Unstable Pelvic Ring Injuries: The "Hula Hoop Technique"

  • Balbachevsky, Daniel;Pires, Robinson Esteves;Sabongi, Rodrigo Guerra;Lins, Theophilo Asfora;Carvalho, Geiser de Souza;Fernandes, Helio Jorge Alvachian;Reis, Fernando Baldy dos
    • Journal of Trauma and Injury
    • /
    • v.32 no.1
    • /
    • pp.51-59
    • /
    • 2019
  • Unstable pelvic ring lesions are usually treated with internal fixation. In patients presenting clinical instability or soft tissue complication risk, external fixation is a safe treatment option. However, pin tract infection, insufficient biomechanical properties, difficulty sitting and changing decubitus are important drawbacks related to the treatment. The present study reports the association of anterior and posterior subcutaneous internal fixation by applying spine-designed implants on the pelvic ring disruption: supra-acetabular pedicle screws with an interconnecting rod (Infix), plus posterior transiliac fixation with the same system, which the authors have named the "Hula Hoop Technique".

The Effects of Spinopelvic Parameters and Paraspinal Muscle Degeneration on S1 Screw Loosening

  • Kim, Jin-Bum;Park, Seung-Won;Lee, Young-Seok;Nam, Taek-Kyun;Park, Yong-Sook;Kim, Young-Baeg
    • Journal of Korean Neurosurgical Society
    • /
    • v.58 no.4
    • /
    • pp.357-362
    • /
    • 2015
  • Objective : To investigate risk factors for S1 screw loosening after lumbosacral fusion, including spinopelvic parameters and paraspinal muscles. Methods : We studied with 156 patients with degenerative lumbar disease who underwent lumbosacral interbody fusion and pedicle screw fixation including the level of L5-S1 between 2005 and 2012. The patients were divided into loosening and non-loosening groups. Screw loosening was defined as a halo sign larger than 1 mm around a screw. We checked cross sectional area of paraspinal muscles, mean signal intensity of the muscles on T2 weight MRI as a degree of fatty degeneration, spinopelvic parameters, bone mineral density, number of fusion level, and the characteristic of S1 screw. Results : Twenty seven patients showed S1 screw loosening, which is 24.4% of total. The mean duration for S1 screw loosening was $7.3{\pm}4.1$ months after surgery. Statistically significant risk factors were increased age, poor BMD, 3 or more fusion levels (p<0.05). Among spinopelvic parameters, a high pelvic incidence (p<0.01), a greater difference between pelvic incidence and lumbar lordotic angle preoperatively (p<0.01) and postoperatively (p<0.05). Smaller cross-sectional area and high T2 signal intensity in both multifidus and erector spinae muscles were also significant muscular risk factors (p<0.05). Small converging angle (p<0.001) and short intraosseous length (p<0.05) of S1 screw were significant screw related risk factors (p<0.05). Conclusion : In addition to well known risk factors, spinopelvic parameters and the degeneration of paraspinal muscles also showed significant effects on the S1 screw loosening.

Useful Corrosion - Potential of Magnesium Alloys as Implants

  • Kaya, A. Arslan;Kaya, R. Alper;Witte, Frank;Duygulu, Ozgur
    • Corrosion Science and Technology
    • /
    • v.7 no.3
    • /
    • pp.162-167
    • /
    • 2008
  • Degradable implants have been in use for bone surgery for decades. However, degradable metal implants are one of the new research areas of biomaterials science. Magnesium has good biocompatibility due to its low toxicity, and it is a corroding, i.e. dissolvable, metal. Furthermore, magnesium is needed in human body, and naturally found in bone tissue. There have been some published reports also asserting the potential bone cell activation or bone healing effect of high magnesium ion concentrations. The classic method for achieving intertransverse process fusion involves autogenous iliac crest bone graft. Several investigations have been performed to enhance this type of autograft fusion. However, there is no research which has been undertaken to investigate the efficiency of pure magnesium particles in posterolateral spinal fusion. In this study, corrosion behavior of magnesium metal at the bone interface, the possibility of new bone cell formation and the degree of effectiveness in producing intertransverse process lumbar fusion in a sheep model have been investigated. Cortical bone screws were machined from magnesium alloy AZ31 extruded rod and implanted to hip-bones of sheep via surgery. Three months after surgery, the bone segments carrying these screws were removed from the sacrificed animals. Samples were sectioned to reveal Mg/bone interfaces and investigated using optical microscope, SEM-EDS and radiography. Optical and SEM images showed that there was a significant amount of corrosion on the magnesium screw. The elemental mapping results indicate, due to the presence of calcium and phosphorus elements, that there exists new bone formation at the interface. Furthermore, sixteen sheep were subjected to intertransverse process spinal fusions with pedicle screw fixation at various locations along their spines. Each animal was treated with 5cc autograft bone at one fusion level and 1cc magnesium+5cc autograft bone at the other. Six months after surgery, bone formation was evaluated by gross inspection and palpation, and radiological, histological, scanning electron microscopic and x-ray diffraction analyses. It may be stated that the potential for using useful corrosion of magnesium alloys in medical applications is expected to be significant.