• Title/Summary/Keyword: auto dispenser

Search Result 6, Processing Time 0.028 seconds

A Study on the Usefulness of Auto Dispenser and Optimized Dispensing Method (방사성 의약품 자동 분주장치의 유용성 및 최적화된 분주방법에 관한 고찰)

  • Lee, JeoungEun;Kim, Hosung;Ryu, Jaekwang;Jung, Wooyoung
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.17 no.2
    • /
    • pp.59-66
    • /
    • 2013
  • Purpose: Recently, there is an increase of the number of hospitals using auto dispenser to reduce occupational radiation exposure when drawing up of the $^{18}F-FDG$ dose (5.18 MBq/kg) in a syringe from the dramatic high activity of $^{18}F-FDG$ multidose vial. The aim of this study is to confirm that using auto dispenser actually reduces the radiation exposure for technologists. Also we analyzed the reproducibility of auto dispenser to find optimized dispensing method for the device. Materials and Methods: We conducted three experiments. Comparison of radiation exposure on chest and hands: The chest and hands exposure dose received by technologists during the injection were measured by electronic personal dosimeter (EPD) and ring TLD respectively. Reproducibility of dispensed volume: We draw up the normal saline into 5 and 2 mL syringe using auto dispenser by changing the volume from 1 to 15 mm for 5 times in the same setting of the volume. The weight of 5 normal saline dispensed from the device at same volume was measured using micro balance and calculated standard deviation and coefficient of variation. Reproducibility of dispensed radioactivity: We dispensed 362.6 $MBq{\pm}10%$ of $^{18}F-FDG$ in 5 and 2 mL syringes from the multidose vial of different specific activity. In the same setting of volume, we repeated dispensing for 4 times and compared standard deviation and coefficient of variation of radioactivity between 5 syringes. Results: There was significant difference in the average of chest exposure dose according to the dispensing methods (P<0.05). Also, when dispensing $^{18}F-FDG$ in manual method, exposure dose was 11.5 times higher in right hand and 4.8 times higher in left hand than in auto method. In the result of reproducibility of dispensed volume, standard deviation and coefficient of variation shows decline as the dispensing volume increases. As a result of reproducibility of dispensed radioactivity, standard deviation and coefficient of variation increases as the specific activity increases. Conclusion: We approved that the occupational radiation exposure dose of technologists were reduced when dispensing $^{18}F-FDG$ using auto dose dispenser. Secondly, using small syringes helps to increase reproducibility of auto dose dispense. And also, if you lower the specific activity of $^{18}F-FDG$ in multidose vial below 915-1,020 MBq/mL, you can use auto dispenser more effectively keeping the coefficient of variation lower than 10%.

  • PDF

Effectiveness Evalution of 18F-FDG Auto Dispenser (RIID: Radiopharmaceutical Intelligent Dispenser) (18F-FDG 자동분주기 사용에 따른 유용성 평가)

  • Yoo, Moon-Gon;Moon, Jae-Seung;Kim, Su-Geun;Shin, Min-Yong;Kim, Seung-Chul;Lee, Tea-hun;An, Sung-Hyeun
    • The Korean Journal of Nuclear Medicine Technology
    • /
    • v.22 no.2
    • /
    • pp.79-83
    • /
    • 2018
  • Purpose $^{18}F-FDG$, which is commonly used in PET-CT examinations, is low in capacity and it is difficult to keep the amount of radioactivity busy when the specific activity is high, increasing the amount of space dose and radioactive contamination in the distribution room. Therefore, while evaluating the actual dose administered to patients during the manual dispense process, the medical institution intends to assess the usefulness of the auto dispenser by comparing the differences from the actual dose administered to the patient using the new automatic dispense. Materials and Methods From July 2016 to December 2016, 846 patients were manually administered by workers using $^{18}F-FDG$ and $^{18}F-FDG$ 906 patients were using auto dispenser from July 2017 to December 2017. Results Capacity administered to patients during the manual dispense averaged $35.41{\pm}27.79%$ compared to the recommended dose, and the auto dispenser process showed a small difference of $-2.15{\pm}3.99%$ compared to the recommended dose(p<0.05). Conclusion Working people did not have to touch radioactive medicines directly while they were busy in the auto dispenser, and because of the availability of other tasks far away, the time and distance to receive the exposure were also advantageous. It is believed that future use by many medical institutions will not only reduce the dose to patients but also help reduce the exposure dose to workers.

A Study on the Cash Dispenser Design in Domestic (국내 현금 자동지급기 디자인에 관한 연구)

  • 오성진;이강일
    • Archives of design research
    • /
    • v.5 no.1
    • /
    • pp.21-36
    • /
    • 1992
  • Since 80th, in the monetary markets due to the escalation of creditable promptness dealing of financial goods, Korea monetary business has been confronted with structual contradictions. The first like a men power reduction, inorder to meet and satisfying with cost reduction policy of the holding firms and secondly, escalation and expention of business networks or facilities inorder to meet and satis~'ing with customers continuously seeking for prompt, acurate, and helpfulness service nearly like one similar to a door to door service. Accordingly, the firms eventually poured mere sweaty money into procuring such devices for upgrading. On the contrary, however, regardless the escalation monetary expenditure may be, inorder to promote quality of the same devices are keen necessary because present days customer groups were tended to receives ample service almost like door to door service without moving too$$\mu$h apart from the immediate spot where they belongs tl or veryspot of where actually they spend instead of deep consideration of natures, quality or sttractiveness of service medias what they can received from. Under the present circumstance, the financial business firms, however, would highly suffering and receives pressures from imblalance natures ofincoming and out go of the business budget due to excessive inc1ease of managerial expences. For these reasons, therefor, in order to solves, compensates, improves, and developes most effective financial service for customers can use under most pleasant environment, the establishment of adequate standardization of such service devices based on the foregoing environmental circumstances. This analytical study, therefor, mainly aimed and emphasized based on pre- estimatimation of proper devlopment and use of the unattended servicing devices such as Cash Dispenser and Auto Teller Machine by generalization, automating. and innovation of design to meet compactness and door to door like convenient service for customers which can be achieved by performing constant comparative analysis and continual study toward development of such devices those of which produced by Five (5) major domestic producers around the country.

  • PDF

Study on Optimized Dispensing of F-18 FDG Auto Dispenser (F-18 FDG 자동분주기의 분주 최적화에 관한 연구)

  • Ji, Bong-Geun;Lee, Sang-Hun;Kim, Jong-Eon;Kim, Won-Tae;Ji, Tae-Jeong
    • Journal of radiological science and technology
    • /
    • v.39 no.3
    • /
    • pp.377-384
    • /
    • 2016
  • This is a study on the optimized dispensing of the auto dispenser used for the purpose of reducing the exposure dose and accurate radiation dose of radioisotope with regard to the PET/CT practitioners. The research method was to find the optimized dispensing method through evaluating the results according to the syringe type, dispensing rate, and vial pressure and through the application of corrected values. As a result of this study, 9.38 mCi has been dispensed on average in the case of 5 ml syringe, and the reproducibility close to 10 mCi was shown at the dispense of 9.55 mCi in the case of 3 ml syringe. In the evaluation according to the dispensing rate, the quantity of radioisotope close to 10 mCi was dispensed at the rate of 5 mm/min when the measurement was carried out by increasing the rate by 5 mm/min units in the order of 5, 10, 15 and 20 mm/min. In the evaluation result according to the vial pressure before/after the use of Needle filter, it was measured to be 9.53 mCi before use and 9.84 mCi after use confirming that the dispensing after using Needle filter showed the optimal value. In addition, in the evaluation of radioactivity before/after the application of corrected values according to the increase in dispense frequency, it was measured 9.53 mCi before correction and 10.07 mCi after correction confirming that the value with correction applied was closer to the quantitative value. Thus, a good optimized method was confirmed to use a 3 ml syringe with dispensing rate of 5 mm/min, to use a Needle filter at dispensing, and to set the corrected value of [$y=0.097{\times}x$] according to the dispensing frequency of equipment.

A Study on the Pocket Park as the urban Open Space (도시오픈스페이스로서의 도심지 소공원 디자인에 관한 연구)

  • 이강일
    • Archives of design research
    • /
    • v.5 no.1
    • /
    • pp.7-20
    • /
    • 1992
  • Since 80th, in the monetary markets due to the escalation of creditable promptness dealing of financial goods, Korea monetary business has been confronted with structual contradictions. The first like a men power reduction, inorder to meet and satisfying with cost reduction policy of the holding firms and secondly, escalation and expention of business networks or facilities inorder to meet and satis~'ing with customers continuously seeking for prompt, acurate, and helpfulness service nearly like one similar to a door to door service. Accordingly, the firms eventually poured mere sweaty money into procuring such devices for upgrading. On the contrary, however, regardless the escalation monetary expenditure may be, inorder to promote quality of the same devices are keen necessary because present days customer groups were tended to receives ample service almost like door to door service without moving too$$\mu$h apart from the immediate spot where they belongs tl or veryspot of where actually they spend instead of deep consideration of natures, quality or sttractiveness of service medias what they can received from. Under the present circumstance, the financial business firms, however, would highly suffering and recei\ulcornerves pressures from imblalance natures ofincoming and out go of the business budget due to excessive inc1ease of managerial expences. For these reasons, therefor, in order to solves, compensates, improves, and developes most effective fina\ulcornerncial service for customers can use under most pleasant environment, the establishment of adequate standardization of such service devices based on the foregoing environmental circumstances. This analytical study, therefor, mainly aimed and emphasized based on pre- estimatimation of proper de\ulcornervlopment and use of the unattended servicing devices such as Cash Dispenser and Auto Teller Machine bygeneralization, automating. and innovation of design to meet compactness and door to door like convenient service for customers which can be achieved by performing constant comparative analysis and continual study toward development of such devices those of which produced by Five (5) major domestic producers aro~nd the country.

  • PDF

A Study on Reduction of Radiation Exposure by Nuclear Medicine Radiation Workers (핵의학 방사선 작업종사자 피폭 감소 방안에 대한 연구)

  • Lee, Wanghui;Ahn, Sungmin
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.2
    • /
    • pp.271-281
    • /
    • 2019
  • This study investigated the shielding efficiency of various types of shielding materials and measured the dose by organ using the phantom. Results of Shielding Efficiency Measurement Using Personal Radiation Meter. Among the various shielding materials, 1.1 mm RNS-TX composed of nano tungsten showed the highest shielding efficiency and 0.2 mm lead shielding showed the lowest shielding efficiency. 99mTc 30 mCi was exposed to the phantom for 120 minutes and the result of the measurement of the organs. 20.53 mSv without radiation protective clothing, 8.75 mSv when wearing 0.25 mm Pb protective clothing, 6.03 mSv when wearing 0.5 mm Pb protective clothing. 131I 2 mCi mCi was exposed to the phantom for 120 minutes and the result of the measurement of the organs. 7.71 mSv without radiation protective clothing, 4.88 mSv when wearing 0.25 mm Pb protective clothing, 2.79 mSv when wearing 0.5 mm Pb protective clothing. 18F 5 mCi was exposed to the phantom for 120 minutes and the result of the measurement of the organs. 16.39 mSv without radiation protective clothing, 15.84 mSv when wearing 0.25 mm Pb protective clothing, 12.52 mSv when wearing 0.5 mm Pb protective clothing. None of the radiation workers working in the nuclear medicine department exceeded the dose limit. However, when compared with other workers in the hospital, they showed a relatively high dose. Therefore, it is necessary to prepare measures to reduce and manage the dose of radiation workers in the nuclear medicine department through the wearing of radiation protective clothing made of lightweight, shielding material with good shielding efficiency, circulation task, task sharing, and substitution equipment such as auto dispenser.