Impact on Retrievability by Cement Variety for Implant Restorations Equipped with a Lingual Slot

  • Lee, Ji-Hong;Lee, Kyu-Bok
    • Journal of Korean Dental Science
    • /
    • v.11 no.1
    • /
    • pp.14-20
    • /
    • 2018
  • Purpose: The purpose of this study is to measure and compare the removal torques of different cements applied in attachments of zirconia restorations on titanium (Ti) abutments fitted with retrievable cement-type slot (RCS) on the lingual side for the better retrievablity by use of a slot driver. Materials and Methods: Three types of cements were used in the experiment: two permanent cements in $RelyX^{TM}$ U200 (RU) (3M ESPE) which is a resin cement and $FujiCem^{TM}$ (FC) (GC) which is a resin-modified glass ionomer cement, and a temporary cement in $Freegenol^{TM}$ temporary cement (TC) (GC). Measurements of removal torques were conducted as follows; an attached sample was fixed on the equipment customized for the experiment; a slot driver was connected to a MGT12 (Mark-10 Corp.), a torque measurement instrument; the sample had the driver fitted to its RCS and then was rotated until the it was removed; and finally, the maximum torque value was recorded. Result: As for the removal torque measurement results, the average values were $47.9{\pm}2.6Ncm$ for RU, $43.4{\pm}1.5Ncm$ for FC, and $20.9{\pm}1.0Ncm$ for TC. The statistical analysis using Kruskal-Wallis test yielded the significance probability of P<0.05 (P=0.002), which confirmed the presence of significant differences between the three groups. Conclusion: All three cements exhibit clinically acceptable levels of removal torque when applied to an upper zirconia implant restoration fitted with a lingual slot, with RU and FC, the two permanent cements, having the significantly higher values than that of TC, the temporary cement.

Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants

  • Mohammed, Hnd Hadi;Lee, Jin-Han;Bae, Ji-Myung;Cho, Hye-Won
    • The Journal of Advanced Prosthodontics
    • /
    • v.8 no.1
    • /
    • pp.62-69
    • /
    • 2016
  • PURPOSE. The purpose of this study was to evaluate the effects of abutment screw length and cyclic loading on the removal torque (RTV) in external hex (EH) and internal hex (IH) implants. MATERIALS AND METHODS. Forty screw-retained single crowns were connected to external and internal hex implants. The prepared titanium abutment screws were classified into 8 groups based on the number of threads (n = 5 per group): EH 12.5, 6.5, 3.5, 2.5 and IH 6.5, 5, 3.5, 2.5 threads. The abutment screws were tightened with 20 Ncm torque twice with 10-minute intervals. After 5 minutes, the initial RTVs of the abutment screws were measured with a digital torque gauge (MGT12). A customized jig was constructed to apply a load along the implant long axis at the central fossa of the maxillary first molar. The post-loading RTVs were measured after 16,000 cycles of mechanical loading with 50 N at a 1-Hz frequency. Statistical analysis included one-way analysis of variance and paired t-tests. RESULTS. The post-loading RTVs were significantly lower than the initial RTVs in the EH 2.5 thread and IH 2.5 thread groups (P<.05). The initial RTVs exhibited no significant differences among the 8 groups, whereas the post-loading RTVs of the EH 6.5 and EH 3.5 thread groups were higher than those of the IH 3.5 thread group (P<.05). CONCLUSION. Within the limitations of this study, the external hex implants with short screw lengths were more advantageous than internal hex implants with short screw lengths in torque maintenance after cyclic loading.

EFFECT OF DIFFERENT SURFACE TREATMENTS TO INCREASE BIOCOMPATIBILITY OF DENTAL IMPLANT (임플랜트 표면처리가 생체활성에 미치는 영향)

  • Lee, Ho-Jin;Song, Kwang-Yeob;Yoon, Tae-Ho
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.44 no.5
    • /
    • pp.594-605
    • /
    • 2006
  • Statement of problem: Modification of titanium implant surface has potential to ensure clinically favorable performance that several surface modification technologies have been introduced. Among the methods. anodizing method and sol-gel hydroxyapatite coating method have gained much interest due to its roughness and chemical composition of the coating layer, but more of its biocompatibility result is required. Purpose : The purpose of this study was to compare bone-implant interface shear strength of four different surface treated implants as time elapsed. Resonance frequency analysis(RFA) and removal torque measurement methods were employed to measure implant stability at one week and six week after implantation. Material and method: A total of 80 screw-shaped implant [20 machined, 20 resorbable media blasted(RBM), 20 anodized, and 20 anodized+hydroxyapatite sol-gel coated] were prepared, and one of each group was implanted in the tibia of a New Zealand white rabbit that total 20 of them were used. In order to test the implant stability and implant-tissue interface contact changing in the bone bed, each 10 rabbit were sacrificed 1 week and 6 week later while resonance frequency and removal torque were measured. One-way analysis of variance and the Tukey test were used for statistical analysis. Results : The results were as follows. 1. There was no statistically significant difference of implant stability quotients(ISQ) value in RFA between individual groups after 1 week of implantation and 6 weeks(p>0.05). But, there was statistically significant increase of ISQ value in 6 weeks group compared to 1 week group(p<0.05). 2. There was no statistically significant difference in removal torque analysis between individual groups after 1 week of implantation and 6 weeks(p>0.05). but there was statistically significant increase in all 4 groups after 6 weeks compared to 1 week later(p<0.05). 3. There was no statistically significant difference in removal torque analysis between anodized group and HA coating after anodic oxidation 6 weeks later(p>0.05), but significant difference was appeared in both groups compared to RBM group and smooth-machined group(p<0.05). Conclusions : It can be suggested that changes in surface characteristics affect bone reactions. Anodized and anodized+hydroxyapatite sol-gel coating showed significantly improved bone tissue response to implants, but further study on the effect of hydroxyapatite dissolution is needed.

Bone Response to Anodized Titanium Implants in Rabbits

  • Lee, Jae-Hyun;Lee, Cheol-Won;Kim, Chang-Hyen;Pyo, Sung-Woon
    • Journal of Korean Dental Science
    • /
    • v.4 no.1
    • /
    • pp.26-32
    • /
    • 2011
  • Purpose: The quality of implant surface is one of the factors that influence wound healing of implant site and subsequently affect osseointegration. The objective of modification of the surface properties of an implant is to affect the biological consequence. The purpose of this study is to evaluate the biologic response of osseous tissue to anodized implants. Materials and Methods: Two machined titanium implants for control group were installed in a tibia of each rabbit and two anodized implants for test group were installed in the other tibia of each rabbit. At the moment the implants were installed, resonance frequency analysis (RFA) values were measured. After healing periods of 1, 2, 3, and 7 weeks, the implants were uncovered and RFA values were measured again. Removal torque was measured for one implant in the test group and one implant in the control group. Histological evaluation was executed in the other implants. Results: Both of test group and control group have the tendency of greater RFA change rate and removal torque value as healing periods became longer, but were statistically insignificant (P>0.05). However, in the case of the same healing period, the test group tended to have greater RFA change rate and removal torque than the control group (P<0.05). More active new bone formation from endosteal surface was noted on the anodized surface than machined surface in specimen after 1 week. There were no significant differences between the test group and control group in histological evaluations. Conclusion: In summary, the anodized surface showed slightly favorable results and it is postulated that it may facilitate improved stability in bone.

The influence of systemically administered oxytocin on the implant-bone interface area: an experimental study in the rabbit

  • Cho, Sung-Am;Park, Sang-Hun;Cho, Jin-Hyun
    • The Journal of Advanced Prosthodontics
    • /
    • v.6 no.6
    • /
    • pp.505-511
    • /
    • 2014
  • PURPOSE. The purpose of this study was to assess the effect of systemically administered oxytocin (OT) on the implant-bone interface by using histomorphometric analysis and the removal torque test. MATERIALS AND METHODS. A total of 10 adult, New Zealand white, female rabbits were used in this experiment. We placed 2 implants (CSM; CSM Implant, Daegu, South Korea) in each distal femoral metaphysis on both the right and left sides; the implants on both sides were placed 10 mm apart. In each rabbit, 1 implant was prepared for histomorphometric analysis and the other 3 were prepared for the removal torque test (RT). The animals received intramuscular injections of either saline (control group; 0.15 M NaCl) or OT (experimental group; $200{\mu}g/rabbit$). The injections were initiated on Day 3 following the implant surgery and were continued for 4 subsequent weeks; the injections were administered twice per day (at a 12-h interval), for 2 days per week. RESULTS. While no statistically significant difference was observed between the two groups (P=.787), the control group had stronger removal torque values. The serum OT concentration (ELISA value) was higher in the OT-treated group, although no statistically significant difference was found. Further, the histomorphometric parameter (bone-to-implant contact [BIC], inter-thread bone, and peri-implant bone) values were higher in the experimental group, but the differences were not significant. CONCLUSION. We postulate that OT supplementation via intramuscular injection weakly contributes to the bone response at the implant-bone interface in rabbits. Therefore, higher concentrations or more frequent administration of OT may be required for a greater bone response to the implant. Further studies analyzing these aspects are needed.

An investigation of reosseointegration according to time course after mechanical loosening of the osseointegrated implant fixtures (표면처리 임플란트 고정체의 의원성 동요 후 시간 경과에 따른 재골유착에 관한 연구)

  • Ye, Sun-Hae;Cho, Jin-Hyun;Lee, Cheong-Hee
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.31 no.3
    • /
    • pp.203-211
    • /
    • 2015
  • Purpose: The purpose of this study was to investigate the reosseointegration periods when the rough surface implants, which had complete bone-implant ankylosis, suddenly losed the osseointegration. Materials and Methods: The implants with RBM surface treatment were inserted into both tibias of 23 rabbits. Two implants were submerged into each side. After six weeks, the primary removal torque was measured by Digital torque gauge, and then the implants were replaced and submerged to estimate the level of reosseointegration. After assigned healing periods for each group, the removal torque was measured again. BIC (Bone-Implant contact, %) ratio was measured through histomorphometric analysis.Paired t-test was processed by SPSS 14.0. One-way ANOVA and Tukey's post-hoc test was processed to analyze statistically significant differences among the groups. Results: In comparison with the primary removal torque, the secondary removal torque was increased after 11 days and significantly increased from 2 weeks. In fluorochrome labeling, the origin of mineralization was observed after 7 days, which showed as fluorescent bands around the bone-implant interfaces. After 11 days, the bone formation was apparent, and it is increased continuously with the passage of the time. Conclusion: In 11 days after the implant replacement, the secondary removal torque was almost as same as the primary value, and was significantly higher from 2 weeks. The mineralized shapes were observed in 7 days after the implant replacement, and then the bone formation appeared visibly in 11 days.

Influence of the implant abutment types and the dynamic loading on initial screw loosening

  • Kim, Eun-Sook;Shin, Soo-Yeon
    • The Journal of Advanced Prosthodontics
    • /
    • v.5 no.1
    • /
    • pp.21-28
    • /
    • 2013
  • PURPOSE. This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS. Three groups of abutments were produced using different types of fabrication methods; stock abutment, gold cast abutment, and CAD/CAM custom abutment. A customized jig was fabricated to apply the load at $30^{\circ}$ to the long axis. The implant fixtures were fixed to the jig, and connected to the abutments with a 30 Ncm tightening torque. A sine curved dynamic load was applied for $10^5$ cycles between 25 and 250 N at 14 Hz. Removal torque before loading and after loading were evaluated. The SPSS was used for statistical analysis of the results. A Kruskal-Wallis test was performed to compare screw loosening between the abutment systems. A Wilcoxon signed-rank test was performed to compare screw loosening between before and after loading in each group (${\alpha}$=0.05). RESULTS. Removal torque value before loading and after loading was the highest in stock abutment, which was then followed by gold cast abutment and CAD/CAM custom abutment, but there were no significant differences. CONCLUSION. The abutment types did not have a significant influence on short term screw loosening. On the other hand, after $10^5$ cycles dynamic loading, CAD/CAM custom abutment affected the initial screw loosening, but stock abutment and gold cast abutment did not.

Comparison of removal torque of dual-acid etched and single-acid etched implants in rabbit tibias (단일, 이중 산처리 임플란트의 회전제거력 비교)

  • Kim, Jong-Jin;Cho, Sung-Am
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.57 no.4
    • /
    • pp.335-341
    • /
    • 2019
  • Purpose: Chemically strong-acids (HF and $HCl/H_2SO_4$) dual etching implant surfaces have higher strengths of osseointegration than machined implant surfaces. However, the dual acid treatment deteriorates the physical properties of the titanium by weakening the fatigue resistance of the implant and causing microcracks. The removal torque comparison between the dual-acid etched (hydrochloric acid, sulfuric acid, HS) and single-acid etched implants (hydrochloric acid, H) could reveal the efficiency of implant surface acid treatment. Materials and methods: Nine $3.75{\times}4mm$ dual-acid etched SLA implants and nine single-acid etched SLA implants were inserted into New Zealand rabbit tibias. After 10 days, removal torque, roughness, and wetting angle were measured. Results: Mean removal torque values were as follows: Mean removal torque were 9.94 Ncm for HS group and 9.96 Ncm for H group (P=.995). Mean surface roughness value were $0.93{\mu}m$ for HS group and $0.84{\mu}m$ for H group (P=.170). Root mean square roughness (RSq) values were $1.21{\mu}m$ for HS group and $1.08{\mu}m$ for H group (P=.294), and mean wetting angle values were $99^{\circ}$ for HS group and $98^{\circ}$ for H group (P=.829). Statistical analysis showed no significant difference between the removal torques, roughness, or wetting angles of the two groups. Conclusion: In this experiment, we found no significant difference in removal torque, roughness, or wetting angle between dual-acid etched and single-acid etched implants.

Removal torque of sandblasted large grit, acid etched treated mini-implant (Sandblasted large grit, acid etched 표면처리에 따른 교정용 미니 임플랜트의 제거회전력에 관한 연구)

  • Oh, Nam-Hee;Kim, Seong-Hun;Kook, Yoon-Ah;Lee, Keun-Hye;Kang, Yoon-Goo;Mo, Sung-Seo
    • The korean journal of orthodontics
    • /
    • v.36 no.5
    • /
    • pp.324-330
    • /
    • 2006
  • Objective: The purposes of this study were to evaluate the differences between sand blasted, large grit and acid-etched (SLA) treated mini-implants and smooth surface orthodontic mini-implants in relation to the removal torque as well as the histologic analysis. Methods: Custom-made, screw-shaped, titanium implants with a length of 9.5 mm and an outer diameter of 1.8 mm were divided into 2 groups; the SLA group (20 SLA treated orthodontic mini-implants) and the smooth surface group (20 smooth surface mini-implants), and placed In the tibia metaphysis of 10 rabbits. Each rabbit had 4 mini-implants placed, 2 in each tibia. The right tibia were implanted with the SLA group mini-implants and the left tibia had the smooth group mini-implants placed. Each mini-implant group were immediately applied with a continuous traction force of 150 g using a Ni-Ti coil spring. The rabbits were sacrificed 6 weeks post-surgically. Subsequently, the legs were stabilized, the Ni-Ti coil springs were removed and the mini-implants were removed under reverse torque rotation with a digital torque gauge. Results: 6 weeks after placement, the SIA group presented a higher mean removal torque value (8.29 Ncm) than the smooth group (3.34 Ncm) and histologic analysis revealed a higher new bone formation aspect along the screw in the SLA group. Conclusion: Results of this study indicates that SLA treated mini-implants may endure higher orthodontic forces without loosening.

Comparative analysis of accuracy of how to use the implant torque controllers (사용 방법에 따른 임플란트 토크 조절기의 정확성 비교 분석 연구)

  • Cho, Lee-Ra;Park, Chan-Jin;Huh, Yoon-Hyuk;Lee, Ji-Hye;Kim, Dae-Gon
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.51 no.4
    • /
    • pp.300-306
    • /
    • 2013
  • Purpose: This study was to evaluate the accuracy of how to use the Mechanical Torque-Limiting Devices. Materials and methods: Three different implant manufacturers (Dentium, Osstem, Shinhung) were prepared for each two implant torque controllers. Divided into two groups depending on the method used for removal torque was measured. Repeated measures of ANOVA test (${\alpha}=.05$) was used as statistics to evaluate the effect of repeated loading number on the removal torque. Independent t-test was used to evaluate the difference in removal torque of two groups. Results: The removal torque significantly decreased as the number of loading repetition increased (P<.05). There was significant difference between two groups. Conclusion: Loosen the handle part of the implant torque controllers spring resilience to recover one group compared to the group that did not, showed a little more closed to the reference value.