Evaluation of the possibility of clinical application by surface dose measurement of ViewRay in MFW treatment (MFW 치료 시 ViewRay의 표면선량 측정에 따른 임상 적용 가능성 평가)

  • Sun, Geo Jun;Son, Sang Jun;Lee, Yang Hoon;Lee, Je Hee
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.169-176
    • /
    • 2018
  • Purpose : The purpose of this study is to evaluate clinical applicability of Co-60 ViewRay treatment plan to increase the skin dose in case of high skin dose is required such as Malignant Fungating Wound By measuring the presence / absence of Bolus application and skin dose by the treatment device and comparing it Materials and Methods : Nine inner measuring points of 2.5 cm lattice arrangement and all 13 measuring points including upper and lower left and right measuring points touching the chest and skin were marked. After CT was taken, each treatment plan was formulated through Eclipse and ViewRay-TPS, and a Fixed beam-IMRT treatment plan was formulated so that the left chest V2Gy=95 % is delivered. Before measurement QED detector was calibrated and the QED detector was positioned at the 13 measurement points displayed on Phantom and surface dose of each treatment planner was measured using 5 mm Bolus application using True-beam and View-ray before and after, measure three times and compare each before applying 5 mm Bolus. Results : The surface dose of the Co-60 ViewRay and the linear accelerator appeared at $76.8%{\pm}5.2%$ vs. $67.3{\pm}%7.5%$ and the surface dose after application of 5 mm Bolus was $87.6%{\pm}8.9%$ vs. $80.3%{\pm}10.2%$ It was measured at 10.2 % (p<0.001). Conclusion : As a result of the surface dose measurement of each treatment instrument, Co-60 ViewRay confirmed that the surface dose reached 95.6 % of 6 MV Linac with conventional 5 mm bolus, despite not using Bolus (p<0.001). Also, by utilizing magnetic resonance images for each treatment, it is possible to observe the change in the treatment site without the problem of exposure, it is easy to formulate an adaptive treatment plan and it is easy to secure the skin dose, so the size In the case of Malignant Fungating Wound patients who need fast skin changes and need high skin doses, Co-60 ViewRay is considered to be more useful than linear accelerators.

  • PDF

A Study on Developing Customized Bolus using 3D Printers (3D 프린터를 이용한 Customized Bolus 제작에 관한 연구)

  • Jung, Sang Min;Yang, Jin Ho;Lee, Seung Hyun;Kim, Jin Uk;Yeom, Du Seok
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.27 no.1
    • /
    • pp.61-71
    • /
    • 2015
  • Purpose : 3D Printers are used to create three-dimensional models based on blueprints. Based on this characteristic, it is feasible to develop a bolus that can minimize the air gap between skin and bolus in radiotherapy. This study aims to compare and analyze air gap and target dose at the branded 1 cm bolus with the developed customized bolus using 3D printers. Materials and Methods : RANDO phantom with a protruded tumor was used to procure images using CT simulator. CT DICOM file was transferred into the STL file, equivalent to 3D printers. Using this, customized bolus molding box (maintaining the 1 cm width) was created by processing 3D printers, and paraffin was melted to develop the customized bolus. The air gap of customized bolus and the branded 1 cm bolus was checked, and the differences in air gap was used to compare $D_{max}$, $D_{min}$, $D_{mean}$, $D_{95%}$ and $V_{95%}$ in treatment plan through Eclipse. Results : Customized bolus production period took about 3 days. The total volume of air gap was average $3.9cm^3$ at the customized bolus. And it was average $29.6cm^3$ at the branded 1 cm bolus. The customized bolus developed by the 3D printer was more useful in minimizing the air gap than the branded 1 cm bolus. In the 6 MV photon, at the customized bolus, $D_{max}$, $D_{min}$, $D_{mean}$, $D_{95%}$, $V_{95%}$ of GTV were 102.8%, 88.1%, 99.1%, 95.0%, 94.4% and the $D_{max}$, $D_{min}$, $D_{mean}$, $D_{95%}$, $V_{95%}$ of branded 1cm bolus were 101.4%, 92.0%, 98.2%, 95.2%, 95.7%, respectively. In the proton, at the customized bolus, $D_{max}$, $D_{min}$, $D_{mean}$, $D_{95%}$, $V_{95%}$ of GTV were 104.1%, 84.0%, 101.2%, 95.1%, 99.8% and the $D_{max}$, $D_{min}$, $D_{mean}$, $D_{95%}$, $V_{95%}$ of branded 1cm bolus were 104.8%, 87.9%, 101.5%, 94.9%, 99.9%, respectively. Thus, in treatment plan, there was no significant difference between the customized bolus and 1 cm bolus. However, the normal tissue nearby the GTV showed relatively lower radiation dose. Conclusion : The customized bolus developed by 3D printers was effective in minimizing the air gap, especially when it is used against the treatment area with irregular surface. However, the air gap between branded bolus and skin was not enough to cause a change in target dose. On the other hand, in the chest wall could confirm that dose decrease for small the air gap. Customized bolus production period took about 3 days and the development cost was quite expensive. Therefore, the commercialization of customized bolus developed by 3D printers requires low-cost 3D printer materials, adequate for the use of bolus.

  • PDF

The Application of 3D Bolus with Neck in the Treatment of Hypopharynx Cancer in VMAT (Hypopharynx Cancer의 VMAT 치료 시 Neck 3D Bolus 적용에 대한 유용성 평가)

  • An, Ye Chan;Kim, Jin Man;Kim, Chan Yang;Kim, Jong Sik;Park, Yong Chul
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.32
    • /
    • pp.41-52
    • /
    • 2020
  • Purpose: To find out the dosimetric usefulness, setup reproducibility and efficiency of applying 3D Bolus by comparing two treatment plans in which Commercial Bolus and 3D Bolus produced by 3D Printing Technology were applied to the neck during VMAT treatment of Hypopahrynx Cancer to evaluate the clinical applicability. Materials and Methods: Based on the CT image of the RANDO phantom to which CB was applied, 3D Bolus were fabricated in the same form. 3D Bolus was printed with a polyurethane acrylate resin with a density of 1.2g/㎤ through the SLA technique using OMG SLA 660 Printer and MaterializeMagics software. Based on two CT images using CB and 3D Bolus, a treatment plan was established assuming VMAT treatment of Hypopharynx Cancer. CBCT images were obtained for each of the two established treatment plans 18 times, and the treatment efficiency was evaluated by measuring the setup time each time. Based on the obtained CBCT image, the adaptive plan was performed through Pinnacle, a computerized treatment planning system, to evaluate target, normal organ dose evaluation, and changes in bolus volume. Results: The setup time for each treatment plan was reduced by an average of 28 sec in the 3D Bolus treatment plan compared to the CB treatment plan. The Bolus Volume change during the pretreatment period was 86.1±2.70㎤ in 83.9㎤ of CB Initial Plan and 99.8±0.46㎤ in 92.2㎤ of 3D Bolus Initial Plan. The change in CTV Min Value was 167.4±19.38cGy in CB Initial Plan 191.6cGy and 149.5±18.27cGy in 3D Bolus Initial Plan 167.3cGy. The change in CTV Mean Value was 228.3±0.38cGy in CB Initial Plan 227.1cGy and 227.7±0.30cGy in 3D Bolus Initial Plan 225.9cGy. The change in PTV Min Value was 74.9±19.47cGy in CB Initial Plan 128.5cGy and 83.2±12.92cGy in 3D Bolus Initial Plan 139.9cGy. The change in PTV Mean Value was 226.2±0.83cGy in CB Initial Plan 225.4cGy and 225.8±0.33cGy in 3D Bolus Initial Plan 224.1cGy. The maximum value for the normal organ spinal cord was the same as 135.6cGy on average each time. Conclusion: From the experimental results of this paper, it was found that the application of 3D Bolus to the irregular body surface is more dosimetrically useful than the application of Commercial Bolus, and the setup reproducibility and efficiency are excellent. If further case studies along with research on the diversity of 3D printing materials are conducted in the future, the application of 3D Bolus in the field of radiation therapy is expected to proceed more actively.

Scalp Dose Evaluation According Radiation Therapy Technique of Whole Brain Radiation Therapy (전뇌 방사선치료 시 치료방법에 따른 두피선량평가)

  • Jang, Joon-Yung;Park, Soo-Yun;Kim, Jong-Sik;Choi, Byeong-Gi;Song, Gi-Won
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.23 no.2
    • /
    • pp.103-108
    • /
    • 2011
  • Purpose: Opposing portal irradiation with helmet field shape that has been given to a patient with brain metastasis can cause excess dose in patient's scalp, resulting in hair loss. For this reason, this study is to quantitatively analyze scalp dose for effective prevention of hair loss by comparing opposing portal irradiation with scalp-shielding shape and tomotherapy designed to protect patient's scalp with conventional radiation therapy. Materials and Methods: Scalp dose was measured by using three therapies (HELMET, MLC, TOMO) after five thermo-luminescence dosimeters were positioned along center line of frontal lobe by using RANDO Phantom. Scalp dose and change in dose distribution were compared and analyzed with DVH after radiation therapy plan was made by using Radiation Treatment Planning System (Pinnacle3, Philips Medical System, USA) and 6 MV X-ray (Clinac 6EX, VARIAN, USA). Results: When surface dose of scalp by using thermo-luminescence dosimeters was measured, it was revealed that scalp dose decreased by average 87.44% at each point in MLC technique and that scalp dose decreased by average 88.03% at each point in TOMO compared with HELMET field therapy. In addition, when percentage of volume (V95%, V100%, V105% of prescribed dose) was calculated by using Dose Volume Histogram (DVH) in order to evaluate the existence or nonexistence of hotspot in scalp as to three therapies (HELMET, MLC, TOMO), it was revealed that MLC technique and TOMO plan had good dose coverage and did not have hot spot. Conclusion: Reducing hair loss of a patient who receives whole brain radiotherapy treatment can make a contribution to improve life quality of the patient. It is expected that making good use of opposing portal irradiation with scalp-shielding shape and tomotherapy to protect scalp of a patient based on this study will reduce hair loss of a patient.

  • PDF

Comparison of Three Different Helmet Bolus Device for Total Scalp Irradiation (Total Scalp의 방사선 치료 시 Helmet Bolus 제작방법에 관한 연구)

  • Song, Yong-Min;Kim, Jong-Sik;Hong, Chae-Seon;Ju, Sang-Gyu;Park, Ju-Young;Park, Su-Yeon
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.24 no.1
    • /
    • pp.31-37
    • /
    • 2012
  • Purpose: This study evaluated the usefulness of Helmet bolus device using Bolx-II, paraffin wax, solid thermoplastic material in total scalp irradiation. Materials and Methods: Using Rando phantom, we applied Bolx-II (Action Products, USA), paraffin wax (Densply, USA), solid thermoplastic material (Med-Tec, USA) on the whole scalp to make helmet bolus device. Computed tomography (GE, Ultra Light Speed16) images were acquired at 5 mm thickness. Then, we set up the optimum treatment plan and analyzed the variation in density of each bolus (Philips, Pinnacle). To evaluate the dose distribution, Dose-homogeneity index (DHI, $D_{90}/D_{10}$) and Conformity index (CI, $V_{95}/TV$) of Clinical Target Volume (CTV) using Dose-Volume Histogram (DVH) and $V_{20}$, $V_{30}$ of normal brain tissues. we assessed the efficiency of production process by measuring total time taken to produce. Thermoluminescent dosimeters (TLD) were used to verify the accuracy. Results: Density variation value of Bolx-II, paraffin wax, solid thermoplastic material turned out to be $0.952{\pm}0.13g/cm^3$, $0.842{\pm}0.17g/cm^3$, $0.908{\pm}0.24g/cm^3$, respectively. The DHI and CI of each helmet bolus device which used Bolx-II, paraffin wax, solid thermoplastic material were 0.89, 0.85, 0.77 and 0.86, 0.78, 0.74, respectively. The result of Bolx-II was the best. $V_{20}$ and $V_{30}$ of brain tissues were 11.50%, 10.80%, 10.07% and 7.62%, 7.40%, 7.31%, respectively. It took 30, 120, 90 minutes to produce. The measured TLD results were within ${\pm}7%$ of the planned values. Conclusion: The application of helmet bolus which used Bolx-II during total scalp irradiation not only improves homogeneity and conformity of Clinical Target Volume but also takes short time and the production method is simple. Thus, the helmet bolus which used Bolx-II is considered to be useful for the clinical trials.

  • PDF

Consideration on Measured Patients Dose of Three-Dimensional and Four-Dimensional Computer Tomography when CT-Simulation to Radiation Therapy (방사선치료를 위한 CT 검사 시 3DCT와 4DCT에 대한 피폭선량 고찰)

  • Park, Ryeong-Hwang;Kim, Min-Jung;Lee, Sang-Kyu;Park, Kwang-Woo;Jeon, Byeong-Cheol;Cho, Jeong-Hee;Yoo, Beong-Gyu;Lee, Jong-Seok
    • Journal of radiological science and technology
    • /
    • v.34 no.4
    • /
    • pp.341-349
    • /
    • 2011
  • This study was to measure the patient dose difference between 3D treatment planning CT and 4D respiratory gating CT. Study was performed with each 10 patients who have lung and liver cancer for measured patient exposure dose by using SOMATON SENSATION OPEN(SIMENS, GERMANY). CTDIvol and DLP value was used to analyze patient dose, and actual dose was measured in the location of liver and kidney for abdominal examination and lung, heart and spinal cord for chest examination. Rando phantom were used for the experiment. OSLD was used for in-vitro and in-vivo dosimetry. Increasing overall actual dose in 4D respiratory gated CT-simulation using OSLD increase the dose by 5.5 times for liver cancer patients and 6 times for lung cancer patients. In CT simulation of 10 lung cancer patients, CTDIvol value was increased by 5.7 times and DLP 2.4 times. For liver cancer patients, CTDIvol was risen by 3.8 times and DLP 1.6 times. The accuracy of treatment volume could be increased in 4D CT planning for position change due to the breaths of patient in the radiation therapy. However, patients dose was increased in 4D CT than 3D CT. In conclusion, constant efforts is required to reduce patients dose by reducing scan time and scan range.

Evaluation on Organ Dose and Image Quality of Lumbar Spine Radiography Using Glass Dosimeter (유리선량계를 이용한 요추검사의 장기선량 및 영상의 평가)

  • Kim, Jae-Kyeom;Kim, Jeong-Koo
    • Journal of radiological science and technology
    • /
    • v.39 no.1
    • /
    • pp.1-11
    • /
    • 2016
  • The purpose of this study was to provide resources for medical exposure reduction through evaluation of organ dose and image resolution for lumbar spine around according to the size of the collimator in DR system. The size of the collimator were varied from $8^{\prime\prime}{\times}17^{\prime\prime}$ to $14^{\prime\prime}{\times}17^{\prime\prime}$ by 1" in AP and lateral projection for the lumbar spine radiography with RANDO phantom. The organ dose measured for liver, stomach, pancreas, kidney and gonad by the glass dosimeter. The image resolution was analyzed using the Image J program. The organ dose of around lumbar spine were reduced as the size of the collimator is decreased in AP projection. There were no significant changes decreasing rate whenever the size of the collimator were reduced 1" in the gonad. The organ dose showed higher on liver and kidney near the surface in lateral projection. There were decreasing rate of less than 5% in liver and kidney, but decreasing rate was 24.34% in the gonad whenever the size of the collimator were reduced 1". Organ dose difference for internal and external of collimator measured $549.8{\mu}Gy$ in the liver and $264.6{\mu}Gy$ in the stomach. There were no significant changes organ dose difference that measured $1,135.1{\mu}Gy$ in the gonad. Image Quality made no difference because SNR and PSNR were over than 30 dB when the collimator size is less than $9^{\prime\prime}{\times}17^{\prime\prime}$ on AP projection and $10^{\prime\prime}{\times}17^{\prime\prime}$ on lateral projection. Therefore, we are considered that the recommendations criterion for control of collimator were suggested in order to reduce unnecessary X-ray exposure and to obtain good image quality because lumbar spine radiography contains a lot of peripheral organs rather than other area radiography.

Fabrication of a Patient-Customized Helmet with a Three-Dimensional Printer for Radiation Therapy of Scalp

  • Oh, Se An;Lee, Chang Min;Lee, Min Woo;Lee, Yeong Seok;Lee, Gyu Hwan;Kim, Seong Hoon;Kim, Sung Kyu;Park, Jae Won;Yea, Ji Woon
    • Progress in Medical Physics
    • /
    • v.28 no.3
    • /
    • pp.100-105
    • /
    • 2017
  • The purpose of the present study was to develop and evaluate patient-customized helmets with a three-dimensional (3D) printer for radiation therapy of malignant scalp tumors. Computed tomography was performed in a case an Alderson RANDO phantom without bolus (Non_Bolus), in a case with a dental wax bolus on the scalp (Wax_Bolus), and in a case with a patient-customized helmet fabricated using a 3D printer (3D Printing_Bolus); treatment plans for each of the 3 cases were compared. When wax bolus was used to fabricate a bolus, a drier was used to apply heat to the bolus to make the helmet. $3-matic^{(R)}$ (Materialise) was used for modeling and polyamide 12 (PA-12) was used as a material, 3D Printing bolus was fabricated using a HP JET Fusion 3D 4200. The average Hounsfield Unit (HU) for the Wax_Bolus was -100, and that of the 3D Printing_Bolus was -10. The average radiation doses to the normal brain with the Non_Bolus, Wax_Bolus, and 3D Printing_Bolus methods were 36.3%, 40.2%, and 36.9%, and the minimum radiation dose were 0.9%, 1.6%, 1.4%, respectively. The organs at risk dose were not significantly difference. However, the 95% radiation doses into the planning target volume (PTV) were 61.85%, 94.53%, and 97.82%, and the minimum doses were 0%, 77.1%, and 82.8%, respectively. The technique used to fabricate patient-customized helmets with a 3D printer for radiation therapy of malignant scalp tumors is highly useful, and is expected to accurately deliver doses by reducing the air gap between the patient and bolus.

Computed Tomographic Simulation of Craniospinal Irradiation (전산화 단층 촬영 장치를 이용한 뇌척수 조사의 치료 계획)

  • Lee CI;Kim HN;Oh TY;Hwang DS;Park NS;Kye CS;Kim YS
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.11 no.1
    • /
    • pp.53-59
    • /
    • 1999
  • The aim of this study is to improve the accuracy of field placement and junction between adjacent fields and block shielding through the use of a computed tomography(CT) simulator and virtual simulation. The information was acquired by assessment of Alderson Rando phantom image using CT simulator (I.Q. Xtra - Picker), determination of each field by virtual fluoroscopy of voxel IQ workstation AcQsim and colored critical structures that were obtained by contouring in virtual simulation. And also using a coronal, sagittal and axial view can determine the field and adjacent field gap correctly without calculation during the procedure. With the treatment planning by using the Helax TMS 4.0, the dose in the junction among the adjacent fields and the spinal cord and cribriform plate of the critical structure was evaluated by the dose volume histogram. The pilot image of coronal and sagittal view took about 2minutes and 26minutes to get 100 images. Image translation to the virtual simulation workstation took about 6minutes. Contouring a critical structure such as cribriform plate, spinal cord using a virtual fluoroscopy were eligible to determine a correct field and shielding. The process took about 20 minutes. As the result of the Helax planning, the dose distribution in adjacent field junction was ideal, and the dose level shows almost 100 percentage in the dose volume histogram of the spinal cord and cribriform plate CT simulation can get a correct therapy area due to enhancement of critical structures such as spinal cord and cribriform plate. In addition, using a Spiral CT scanner can be saved a lot of time to plan a simulation therefore this function can reduce difficulties to keep the patient position without any movements to the patient, physician and radiotherapy technician.

  • PDF

Radiation Dose and Image Evaluation for Position Change in Low Extremity Teleography (하지전신계측검사에서 자세의 변화에 따른 방사선량 및 영상평가)

  • Kim, Yeongcheon;Song, Jongnam;Choi, Namgil;Jeong, Yeon;Han, Jaebok
    • Journal of the Korean Society of Radiology
    • /
    • v.8 no.7
    • /
    • pp.409-415
    • /
    • 2014
  • In this study, radiation doses of major organs in various postures in lower extremity teleography were measured and compared to investigate the utility of the test methods. Ten adult males who underwent lower extremity teleography at a tube voltage of 73 kVp, tube current of 32 mAs, and SID of 180 cm. Using rando phantom, glass dosimeter was attached to the eye lens, thyroid gland, and genital gland to measure the radiation dose in each area 5 times in each anteroposterior posture and posteroanterior posture. The results were compared and analyzed through Paired T-test. The images from the anteroposterior posture and posteroanterior posture were evaluated through the blind test on a scale of 5. As a result, the posteroanterior method could reduce the dose than the anteroposterior posture method: less dose for the eye lens by 6%, thyroid gland by 6%, and genital gland by 26%. Since there was no significant difference in image evaluation, the posteroanterior posture is considered better than the anteroposterior posture in lower extremity teleography.