Evaluation on Organ Dose and Image Quality of Lumbar Spine Radiography Using Glass Dosimeter  

Kim, Jae-Kyeom (Dept. of Radiological Science, Hanseo University)
Kim, Jeong-Koo (Dept. of Radiology, The Catholic University of Korea Incheon St. Mary's Hospital)
Publication Information
Journal of radiological science and technology / v.39, no.1, 2016 , pp. 1-11 More about this Journal
Abstract
The purpose of this study was to provide resources for medical exposure reduction through evaluation of organ dose and image resolution for lumbar spine around according to the size of the collimator in DR system. The size of the collimator were varied from $8^{\prime\prime}{\times}17^{\prime\prime}$ to $14^{\prime\prime}{\times}17^{\prime\prime}$ by 1" in AP and lateral projection for the lumbar spine radiography with RANDO phantom. The organ dose measured for liver, stomach, pancreas, kidney and gonad by the glass dosimeter. The image resolution was analyzed using the Image J program. The organ dose of around lumbar spine were reduced as the size of the collimator is decreased in AP projection. There were no significant changes decreasing rate whenever the size of the collimator were reduced 1" in the gonad. The organ dose showed higher on liver and kidney near the surface in lateral projection. There were decreasing rate of less than 5% in liver and kidney, but decreasing rate was 24.34% in the gonad whenever the size of the collimator were reduced 1". Organ dose difference for internal and external of collimator measured $549.8{\mu}Gy$ in the liver and $264.6{\mu}Gy$ in the stomach. There were no significant changes organ dose difference that measured $1,135.1{\mu}Gy$ in the gonad. Image Quality made no difference because SNR and PSNR were over than 30 dB when the collimator size is less than $9^{\prime\prime}{\times}17^{\prime\prime}$ on AP projection and $10^{\prime\prime}{\times}17^{\prime\prime}$ on lateral projection. Therefore, we are considered that the recommendations criterion for control of collimator were suggested in order to reduce unnecessary X-ray exposure and to obtain good image quality because lumbar spine radiography contains a lot of peripheral organs rather than other area radiography.
Keywords
Organ dose; Image Quality; Lumbar spine radiography; Glass Dosimeter;
Citations & Related Records
Times Cited By KSCI : 4  (Citation Analysis)
연도 인용수 순위
1 KFDA : Standard Radiographic Techniques in General Radiography Projection, 1-4, 2014.
2 KFDA : Guideline for Measurement of Patient Dose, 1-25, 2007
3 Lee Jong Woong : Implementation of a systematic radiation management program for reducing the patient dose on digital radiography examinations, Kwangwoon University, 1, 2013.
4 KFDA : Guideline for Reference Level of Patient Dose in General Radiography Projection, 2-42, 2012
5 Zetterberg LG, Espeland A : Lumbar spine radiography-poor collimation practices after implementation of digital technology, The British journal of radiology, 84(1002), 566-569, 2011.   DOI
6 JeeHye Kim, DongWook Sung, JeongWook Kim et. al. : Nationwise Survey of the X-ray Beam Collimator Utilization in General Diagnostic Radiograph, Korean journal of medical physics, 24(2), 119-126, 2013.
7 Sung Hyun Choi, Sun Kwang Hwang, Rae Gon Lee et. al. : A Study on Evaluating the Image Quality in DR a System with and without a Grid, Journal of the Korean Physical Society, 60(11), 1977-1982, 2012.   DOI
8 Chang Gyu Kim : Spatial dose distribution and exposure dose during lumbar lateral test, Journal of the Korea Convergence Society, 5(1), 17-22, 2014   DOI
9 Seong Kwan Choi : Change of the Scattered Dose by Field Size in X-ray Radiography, The Journal of the Korea Contents Association, 13(3), 198-203, 2013
10 Kim Kwang Hyun : A Study on Optimization of IPTV Video Service Quality in Next Generation IP Network, Pukyong National University, 100-124, 2009.
11 Boram Lee, Sunyoung Lee, Injeong Yang et. al. : Evaluation of the Stepwise Collimation Method for the Reduction of the Patient Dose in Full Spine Radiography, Journal of the Korean Physical Society, 64(9), 1380-1384, 2014   DOI
12 Jong Chang Park : An evaluation on Entrans Surface Dose and Effective Dose by the field size change of lumbar Anteroposterior test, Gachon University, 55-58, 2015.