• 제목/요약/키워드: Radiological measurement

검색결과 616건 처리시간 0.019초

Measurement uncertainty analysis of radiophotoluminescent glass dosimeter reader system based on GD-352M for estimation of protection quantity

  • Kim, Jae Seok;Park, Byeong Ryong;Yoo, Jaeryong;Ha, Wi-Ho;Jang, Seongjae;Jang, Won Il;Cho, Gyu Seok;Kim, Hyun;Chang, Insu;Kim, Yong Kyun
    • Nuclear Engineering and Technology
    • /
    • 제54권2호
    • /
    • pp.479-485
    • /
    • 2022
  • At the Korea Institute of Radiological and Medical Sciences, physical human phantoms were developed to evaluate various radiation protection quantities, based on the mesh-type reference computational phantoms of the International Commission on Radiological Protection. The physical human phantoms were fabricated such that a radiophotoluminescent glass dosimeter (RPLGD) with a Tin filter, namely GD-352M, could be inserted into them. A Tin filter is used to eliminate the overestimated signals in low-energy photons below 100 keV. The measurement uncertainty of the RPLGD reader system based on GD-352M should be analyzed for obtaining reliable protection quantities before using it for practical applications. Generally, the measurement uncertainty of RPLGD systems without Tin filters is analyzed for quality assurance of radiotherapy units using a high-energy photon beam. However, in this study, the measurement uncertainty of GD-352M was analyzed for evaluating the protection quantities. The measurement uncertainty factors in the RPLGD include the reference irradiation, regression curve, reproducibility, uniformity, energy dependence, and angular dependence, as described by the International Organization for Standardization (ISO). These factors were calculated using the Guide to the Expression of Uncertainty in Measurement method, applying ISO/ASTM standards 51261(2013), 51707(2015), and SS-ISO 22127(2019). The measurement uncertainties of the RPLGD reader system with a coverage factor of k = 2 were calculated to be 9.26% from 0.005 to 1 Gy and 8.16% from 1 to 10 Gy. A blind test was conducted to validate the RPLGD reader system, which demonstrated that the readout doses included blind doses of 0.1, 1, 2, and 5 Gy. Overall, the En values were considered satisfactory.

치과용 이동형 방사선장치의 선량평가 (Evaluation of Radiation Doses of Dental Portable Equipment)

  • 박훈희;강병삼
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제41권5호
    • /
    • pp.445-450
    • /
    • 2018
  • We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR.

In-Situ 측정법을 이용한 고리 원자로 방사선원항 평가 (Assessment of the Radiological Inventory for the Reactor at Kori NPP Using In-Situ Measurement Technology)

  • 정현철;정성엽
    • 방사성폐기물학회지
    • /
    • 제12권2호
    • /
    • pp.171-178
    • /
    • 2014
  • 원전 해체 시 원자력설비는 안전하게 해체되어야 한다. 고리 1호기나 월성 1호기와 같은 노후화된 원전의 경우 곧 원전 해체를 계획하고 있는 대상 원전이지만, 이 원전들의 가동 중단 후 해체 시 선원항 평가 기준, 제염 및 해체 기술 등의 독자적인 국내 기술 확보는 미흡한 실정이다. 본 연구의 목적은 원전 선원항 평가 기술 중 하나로 In-Situ 기법을 이용하여 대형 원전 기기를 직접 측정하여 측정대상체에 대한 선원항 평가방법을 개발하는 것이다. 원자로 헤드를 별도의 해체 없이 이동형 감마핵종분석기를 이용하여 직접 측정법으로 분석하고 간접 측정을 병행하여 측정 결과를 보완하였다. 그리고, 표면오염시료는 방사화학분석을 수행하였다. 분석 결과를 확장하여 원자로의 핵종 재고량을 계산하였다. 본 연구 결과를 토대로 각 핵종별 방사능량 변화에 따라 해체 시점을 결정할 수 있으며, 원전 해체 시 작업자의 피폭 저감에 도움이 될 것으로 기대한다.

유리선량계 특성평가 및 판독 보정인자에 대한 불확도 평가 (Evaluation of RPL Glass Dosimeter Characteristics and Uncertainty Evaluation of Reading Correction Factors)

  • 목성윤;강영록;김효진;계용욱;안현
    • 대한방사선기술학회지:방사선기술과학
    • /
    • 제46권3호
    • /
    • pp.219-229
    • /
    • 2023
  • In this study, basic characteristics such as reproducibility, linearity, and directionality of RPL glass dosimeters were evaluated to improve the reliability of dose evaluation through RPL glass dosimeters, and uncertainty elements such as sensitivity by glass element and magazine slot sensitivity were evaluated. Using a mathematical model to calibrate the measured values of the RPL glass dosimeter, the measurement uncertainty was calculated assuming an example. As a result of the characteristic evaluation, the RPL glass dosimeter showed excellent performance with a standard deviation of ±1% (1 SD) for the reproducibility of the reading process, a coefficient of determination for linearity of 0.99997. And the read-out of the RPL glass dosimeter are affected by the circular rotation direction of the glass dosimeter during irradiation, fading according to the period after irradiation, the number of laser pulses of the reader, and response degradation due to repeated reading, it is judged that measurement uncertainty can be reduced by irradiation and reading in consideration of these factors. In addition, it was confirmed that the dose should be determined by calculating the correction factors for the sensitivity of each element and, the sensitivity of each reading magazine slot. It is believed that the reliability of dosimetry using glass dosimeters can be improved by using a mathematical model for correction of glass dosimeter readings and calculating measurement uncertainty.

방사성핵종 이용 치료에서 핵의학영상을 이용한 흡수선량평가 (Internal Radiation Dosimetry using Nuclear Medicine Imaging in Radionuclide Therapy)

  • 김경민;변병현;천기정;임상무
    • Nuclear Medicine and Molecular Imaging
    • /
    • 제41권4호
    • /
    • pp.265-271
    • /
    • 2007
  • Radionuclide therapy has been an important field in nuclear medicine. In radionuclide therapy, relevant evaluation of Internally absorbed dose is essential for the achievement of efficient and sufficient treatment of incurable disease, and can be accomplish by means of accurate measurement of radioactivity in body and its changes with time. Recently, the advances of nuclear medicine imaging and multi modality imaging processing techniques can provide change of more accurate and easier measurement of the measures commented above, in cooperation of conventional imaging based approaches. in this review, basic concept for internal dosimetry using nuclear medicine imaging is summarized with several check points which should be considered In real practice.

Postal Dosimetry Audits for the Domestic Medical Linear Accelerator

  • Kim, Kum Bae;Choi, Sang Hyoun
    • 한국의학물리학회지:의학물리
    • /
    • 제31권2호
    • /
    • pp.20-28
    • /
    • 2020
  • Purpose: The objective of this study is to perform Postal dosimetry audits for medical linear accelerators in radiation therapy institutions using glass dosimeters and Gafchromic film reading systems and postal dosimetry audit procedures, and to evaluate radiation therapy doses and mechanical accuracy in medical institutions. Methods: Photon output measured and analyzed using a standard phantom for measuring photon output dose using a glass dosimeter for medical linear accelerators. Mechanical accuracy was measured and analyzed using software for film measurement. Results: Measurement and analysis of photon beam output dose using a standard phantom glass dosimeter for photon beam output dose measurement was completed. All tolerance doses were within 5%. Mechanical accuracy measurement and analysis using a standard phantom for verifying the mechanical accuracy of linear accelerator (LINAC) using a Gafchromic film were completed, and all results were shown within tolerances (2 mm or less). Conclusions: In this study, Postal dosimetry audits were performed on the output dose and mechanical accuracy of photon beams (207 beams) for 106 LINACs from 48 institutions. As a result of corrective action and re-execution, it was confirmed that all engines met the acceptable standard within 2 mm in the linear accelerator.