• Title/Summary/Keyword: Neutron Dose

Search Result 203, Processing Time 0.025 seconds

A Study on the Technology of Measuring and Analyzing Neutrons and Gamma-Rays Using a CZT Semiconductor Detector (CZT 반도체 검출기를 활용한 중성자 및 감마선 측정과 분석 기술에 관한 연구)

  • Jin, Dong-Sik;Hong, Yong-Ho;Kim, Hui-Gyeong;Kwak, Sang-Soo;Lee, Jae-Geun
    • Journal of radiological science and technology
    • /
    • v.45 no.1
    • /
    • pp.57-67
    • /
    • 2022
  • CZT detectors, which are compound semiconductors that have been widely used recently for gamma-ray detection purposes, are difficult to detect neutrons because direct interaction with them does not occur unlike gamma-rays. In this paper, a method of detecting and determining energy levels (fast neutrons and thermal neutrons) of neutrons, in addition of identifying energy and nuclide of gamma-rays, and evaluating gamma dose rates using a CZT semiconductor detector is described. Neutrons may be detected by a secondary photoelectric effect or compton scattering process with a characteristic gamma-ray of 558.6 keV generated by a capture reaction (113Cd + 1n → 114Cd + 𝛾) with cadmium (Cd) in the CZT detector. However, in the case of fast neutrons, the probability of capture reaction with cadmium (Cd) is very low, so it must be moderated to thermal neutrons using a moderator and the material and thickness of moderator should be determined in consideration of the portability and detection efficiency of the equipment. Conversely, in the case of thermal neutrons, the detection efficiency decreases due to shielding effect of moderator itself, so additional CZT detector that do not contain moderator must be configured. The CZT detector that does not contain moderator can be used to evaluate energy, nuclide, and gamma dose-rate for gamma-rays. The technology proposed in this paper provides a method for detecting both neutrons and gamma-rays using a CZT detector.

The Relative Biological Effectiveness of KCCH-Cyclotron neutron (원자력병원(原子力病院) 싸이클로트론 속중성자선(速中性子線)의 방사선생물학적효과비(RBE)에 관(關)한 연구(硏究))

  • Yoo, Seong-Yul;Koh, Kyoung-Hwan;Cho, Chul-Koo;Kim, Sung-Ho
    • Journal of Radiation Protection and Research
    • /
    • v.12 no.2
    • /
    • pp.1-8
    • /
    • 1987
  • The relative biological effectiveness has been measured for the biological characterization of $p^+(50.5\;MeV)$ Be neutron of KCCH-Cyclotron prior to clinical application. Measured RBE of mouse jejunal crypt cell in single whole body irradiation was 2.8. This RBE value is changed differently in different biologic systems such as mouse jejunal crypt cells, intestine and bone marrow in different irradiation method, so that in fractionated irradiation RBE is variable to the different fraction size and total dose, and also variable to the number of fractions.

  • PDF

A Study on the Radioactive Products of Components in Proton Accelerator on Short Term Usage Using Computed Simulation (몬테칼로 시뮬레이션을 활용한 양성자가속기 단기사용 시 구성품의 방사화 평가)

  • Bae, Sang-Il;Kim, Jung-Hoon
    • Journal of radiological science and technology
    • /
    • v.43 no.5
    • /
    • pp.389-395
    • /
    • 2020
  • The evaluation of radioactivated components of heavy-ion accelerator facilities affects the safety of radiation management and the exposure dose for workers. and this is an important issue when predicting the disposal cost of waste during maintenance and dismantling of accelerator facilities. In this study, the FLUKA code was used to simulate the proton treatment device nozzle and classify the radio-nuclides and total radioactivity generated by each component over a short period of time. The source term was evaluated using NIST reference beam data, and the neutron flux generated for each component was calculated using the evaluated beam data. Radioactive isotopes caused by generated neutrons were compared and evaluated using nuclide information from the International Radiation Protection Association and the Korea Radioisotope association. Most of the nuclides produced form of beta rays and electron capture, and short-lived nuclides dominated. However, In the case of 54Mn, which is a radioactive product of iron, the effect of gamma rays should be considered. In the case of tritium generated from a material with a low atomic number, it is considered that handling care should be taken due to its long half-life.

PROLONGATION OF THE BOR-60 REACTOR OPERATION

  • IZHUTOV, ALEXEY L.;KRASHENINNIKOV, YURI M.;ZHEMKOV, IGOR Y.;VARIVTSEV, ARTEM V.;NABOISHCHIKOV, YURI V.;NEUSTROEV, VICTOR S.;SHAMARDIN, VALENTIN K.
    • Nuclear Engineering and Technology
    • /
    • v.47 no.3
    • /
    • pp.253-259
    • /
    • 2015
  • The fast neutron reactor BOR-60 is one of the key experimental facilities worldwide to perform large-scale tests of fuel, absorbing, and structural materials for advanced reactors. The BOR-60 reactor was put into operation in December 1969, and by the end of 2014 it had been operating on power for ~265,000 hours. BOR-60 still demonstrates potential capabilities to extend the lifetime of sodium-cooled fast reactors. The BOR-60 lifetime should have expired at the end of 2014. Over the past few years, a great scope of work has been performed to justify the possibility of extending its lifetime. The work included inspection of the equipment conditions, calculations and experimental research on operating parameters and the conditions of nonremovable components, investigation of the structural material samples after their long-term operation under irradiation, etc. Based on the results of the work performed, the residual lifetime was evaluated and the reactor operator made a decision to extend the lifetime period of the BOR-60 reactor. After considering both a set of documents about the reactor conditions and the positive decision of independent experts, the Regulatory Authority of the Russian Federation extended the BOR-60 operating license up to 2020.

SHIELDING PERFORMANCE OF A NEWLY DESIGNED TRANSPORT CASK IN THE ADVANCED CONDITIONING SPENT FUEL PYROPROCESS FACILITIY

  • Park, Chang-Je;Jeong, Chang-Joon;Min, Deok-Ki;Kang, Hee-Young;Choi, Woo-Seok;Lee, Joo-Chan;Bang, Gyeoung-Sik;Seo, Ki-Seog
    • Nuclear Engineering and Technology
    • /
    • v.40 no.4
    • /
    • pp.319-326
    • /
    • 2008
  • To transport process wastes efficiently from the Advanced Spent Fuel Conditioning Pyro-process Facility (ACPF) at the Korea Atomic Energy Research Institute (KAERI), a new hot cell cask has been designed based on an existing hot cell padirac transport cask, with not only a neutron absorber for improved shielding capability, but also a docking facility for an easy docking system. In the new hot cell cask, two kinds of materials have been considered as shielding materials, polyethylene and resin. To verify the transport compatibility of the waste and spent fuel for the ACPF, neutron and photon shielding calculations were performed using the MCNPX code. The source term was evaluated by the ORIGEN-ARP code system based on spent PWR fuel. From the calculation, it was found that the maximum surface dose rates of the hot cell cask with the two candidates were estimated within the limit (2 mSv/hr).

Attenuation curves of neutrons from 400 to 550 Mev/u for Ca, Kr, Sn, and U ions in concrete on a graphite target for the design of shielding for the RAON in-flight fragment facility in Korea

  • Lee, Eunjoong;Kim, Junhyeok;Kim, Giyoon;Kim, Jinhwan;Park, Kyeongjin;Cho, Gyuseong
    • Nuclear Engineering and Technology
    • /
    • v.51 no.1
    • /
    • pp.275-283
    • /
    • 2019
  • Rare isotope beam facilities require shielding data in early stage of their design. There is much less shielding data on neutrons from the reactions between heavy ion beams and matter than the data on neutrons produced by protons. The purpose of the present work is to produce and thus increase the amount of shielding data on neutrons generated by high-energy heavy ion beams based on the RAON in-flight fragment facility. Calculations were performed with the computational Monte Carlo codes PHITS and MCNPX. The secondary neutron source terms were evaluated at 550 MeV/u for Ca, Kr, and Sn and at 400 MeV/u for U ions on a graphite target. Source terms and attenuation lengths were obtained by fitting the ambient dose equivalent inside an ordinary concrete shield.

X-Ray Diffraction line profile analysis of defects and precipitates in high displacement damage neutron-irradiated austenitic stainless steels

  • Shreevalli M.;Ran Vijay Kumar;Divakar R.;Ashish K.;Padmaprabu C.;Karthik V.;Archna Sagdeo
    • Nuclear Engineering and Technology
    • /
    • v.56 no.1
    • /
    • pp.114-122
    • /
    • 2024
  • Irradiation-induced defects and the precipitates in the wrapper material of the Indian Fast Breeder Test Reactor (FBTR), SS 316 are analyzed using the synchrotron source-based Angle Dispersive X-Ray Diffraction (ADXRD) technique with X-rays of energy 17.185 keV (wavelength ~0.72146 Å). The differences and similarities in the high displacement damage samples as a function of dpa (displacement per atom) and dpa rate in the range of 2.9 × 10-7- 9 × 10-7 dpa/s are studied. Ferrite and M23C6 are commonly observed in the present set of high displacement damage 40-74 dpa SS 316 samples irradiated at temperatures in the range of 400-483 ℃. Also, the dislocation density has increased as a function of the irradiation dose. The X-ray diffraction peak profile parameters quantified such as peak shift and asymmetry show that the irradiation-induced defects are sensitive to the dpa rate-irradiation temperature combinations. The increase in yield strength as a function of displacement damage is also correlated to the dislocation density.

Dosimetric Quantities for 300 keV Neutrons (300 keV 중성자(中性子)에 대한 방사선량(放射線量) 관계량(關係量)의 산정(算定))

  • Lee, Soo-Yong
    • Journal of Radiation Protection and Research
    • /
    • v.11 no.1
    • /
    • pp.37-43
    • /
    • 1986
  • Dosimetric quantities for 300 keV neutrons in the ICRU standard tissue sphere were evaluated. The Monte Carlo code NEDEP which performs neutron-photon-charged particles coupled transport was used in the direct estimation of absorbed dose and dose equivalent. Some important quantities calculated are as follows; Deep dose equivalent index $H_{I,d}:1.78{\times}10^{11}\;Sv-cm^2$ Shallow dose equivalent index $H_{I,s}:2.08{\times}10^{-11}\;Sv-cm^2$ Ambient dose equivalent $H^*(0.07):1.7{\times}10^{-11}\;Sv-cm^2$ Ambient dose equivalent $H^*(10):1.78{\times}10^{-11}\;Sv-cm^2$ Effective quality factor $\bar{Q}^*(10):12.4$

  • PDF

Development Treatment Planning System Based on Monte-Carlo Simulation for Boron Neutron Capture Therapy

  • Kim, Moo-Sub;Kubo, Kazuki;Monzen, Hajime;Yoon, Do-Kun;Shin, Han-Back;Kim, Sunmi;Suh, Tae Suk
    • Progress in Medical Physics
    • /
    • v.27 no.4
    • /
    • pp.232-235
    • /
    • 2016
  • The purpose of this study is to develop the treatment planning system (TPS) based on Monte-Carlo simulation for BNCT. In this paper, we will propose a method for dose estimation by Monte-Carlo simulation using the CT image, and will evaluate the accuracy of dose estimation of this TPS. The complicated geometry like a human body allows defining using the lattice function in MCNPX. The results of simulation such as flux or energy deposition averaged over a cell, can be obtained using the features of the tally provided by MCNPX. To assess the dose distribution and therapeutic effect, dose distribution was displayed on the CT image, and dose volume histogram (DVH) was employed in our developed system. The therapeutic effect can be efficiently evaluated by these evaluation tool. Our developed TPS could be effectively performed creating the voxel model from CT image, the estimation of each dose component, and evaluation of the BNCT plan.

Multi-slit prompt-gamma camera for locating of distal dose falloff in proton therapy

  • Park, Jong Hoon;Kim, Sung Hun;Ku, Youngmo;Kim, Chan Hyeong;Lee, Han Rim;Jeong, Jong Hwi;Lee, Se Byeong;Shin, Dong Ho
    • Nuclear Engineering and Technology
    • /
    • v.51 no.5
    • /
    • pp.1406-1416
    • /
    • 2019
  • In this research, a multi-slit prompt-gamma camera was developed to locate the distal dose falloff of the proton beam spots in spot scanning proton therapy. To see the performance of the developed camera, therapeutic proton beams were delivered to a solid plate phantom and then the prompt gammas from the phantom were measured using the camera. Our results show that the camera locates the 90% distal dose falloff (= d90%), within about 2-3 mm of error for the spots which are composed $3.8{\times}10^8$ protons or more. The measured location of d90% is not very sensitive to the irradiation depth of the proton beam (i.e., the depth of proton beam from the phantom surface toward which the camera is located). Considering the number of protons per spot for the most distal spots in typical treatment cases (i.e., 2 Gy dose divided in 2 fields), the camera can locate d90% only for a fraction of the spots depending on the treatment cases. However, the information of those spots is still valuable in that, in the multi-slit prompt-gamma camera, the distal dose falloff of the spots is located solely based on prompt gamma measurement, i.e., not referring to Monte Carlo simulation.