• Title/Summary/Keyword: Micro-fluidic chip

Search Result 22, Processing Time 0.026 seconds

Study on Effect of the printing direction and layer thickness for micro-fluidic chip fabrication via SLA 3D printing (적층 방식 3차원 프린팅에 의한 미세유로 칩 제작 공정에서 프린팅 방향 및 적층 두께의 영향에 관한 연구)

  • Jin, Jae-Ho;Kwon, Da-in;Oh, Jae-Hwan;Kang, Do-Hyun;Kim, Kwanoh;Yoon, Jae-Sung;Yoo, Yeong-Eun
    • Design & Manufacturing
    • /
    • v.16 no.3
    • /
    • pp.58-65
    • /
    • 2022
  • Micro-fluidic chip has been fabricated by lithography process on silicon or glass wafer, casting using PDMS, injection molding of thermoplastics or 3D printing, etc. Among these processes, 3D printing can fabricate micro-fluidic chip directly from the design without master or template for fluidic channel fabricated previously. Due to this direct printing, 3D printing provides very fast and economical method for prototyping micro-fluidic chip comparing to conventional fabrication process such as lithography, PDMS casting or injection molding. Although 3D printing is now used more extensively due to this fast and cheap process done automatically by single printing machine, there are some issues on accuracy or surface characteristics, etc. The accuracy of the shape and size of the micro-channel is limited by the resolution of the printing and printing direction or layering direction in case of SLM type of 3D printing using UV curable resin. In this study, the printing direction and thickness of each printing layer are investigated to see the effect on the size, shape and surface of the micro-channel. A set of micro-channels with different size was designed and arrayed orthogonal. Micro-fluidic chips are 3D printed in different directions to the micro-channel, orthogonal, parallel, or skewed. The shape of the cross-section of the micro-channel and the surface of the micro-channel are photographed using optical microscopy. From a series of experiments, an optimal printing direction and process conditions are investigated for 3D printing of micro-fluidic chip.

A Study on Fabrication of Fluidic Devices using Stereolithography Technology (Stereolithography 기술을 이용한 유체소자 제작에 관한 연구)

  • Lee Young Tae;Bae Yong Hwan
    • Journal of the Korean Society for Precision Engineering
    • /
    • v.21 no.10
    • /
    • pp.188-195
    • /
    • 2004
  • In this paper, we fabricated fluidic devices like micro-channel, pump, mixer and particular gas separator with the technology of stereolithouaphy using RP(rapid-prototyping). The fabricated fluidic devices are expected to be applied to develop Lab-on-a chip type liquid analyzer. Stereolithography technology seems effective for fabricating MEMS(Micro Electro Mechanical System) with complicated structure because it makes three dimensional fabrication possible but, exclusive devices are needed to be developed fur fabricating even more microscopic MEMS structure.

Microfluidic Device for Bio Analytical Systems

  • Junhong Min;Kim, Joon-Ho;Kim, Sanghyo
    • Biotechnology and Bioprocess Engineering:BBE
    • /
    • v.9 no.2
    • /
    • pp.100-106
    • /
    • 2004
  • Micro-fluidics is one of the major technologies used in developing micro-total analytical systems (${\mu}$-TAS), also known as “lab-on-a-chip”. With this technology, the analytical capabilities of room-size laboratories can be put on one small chip. In this paper, we will briefly introduce materials that can be used in micro-fluidic systems and a few modules (mixer, chamber, and sample prep. modules) for lab-on-a-chip to analyze biological samples. This is because a variety of fields have to be combined with micro-fluidic technologies in order to realize lab-on-a-chip.

A Multi-chip Microelectrofluidic Bench for Modular Fluidic and Electrical Interconnections (전기 및 유체 동시접속이 가능한 멀티칩 미소전기유체통합벤치의 설계, 제작 및 성능시험)

  • Chang Sung-Hwan;Suk Sang-Do;Cho Young-Ho
    • Transactions of the Korean Society of Mechanical Engineers A
    • /
    • v.30 no.4 s.247
    • /
    • pp.373-378
    • /
    • 2006
  • We present the design, fabrication, and characterization of a multi-chip microelectrofluidic bench, achieving both electrical and fluidic interconnections with a simple, low-loss and low-temperature electrofluidic interconnection method. We design 4-chip microelectrofluidic bench, having three electrical pads and two fluidic I/O ports. Each device chip, having three electrical interconnections and a pair of two fluidic I/O interconnections, can be assembled to the microelectofluidic bench with electrical and fluidic interconnections. In the fluidic and electrical characterization, we measure the average pressure drop of $13.6{\sim}125.4$ Pa/mm with the nonlinearity of 3.1 % for the flow-rates of $10{\sim}100{\mu}l/min$ in the fluidic line. The pressure drop per fluidic interconnection is measured as 0.19kPa. Experimentally, there are no significant differences in pressure drops between straight channels and elbow channels. The measured average electrical resistance is $0.26{\Omega}/mm$ in the electrical line. The electrical resistance per each electrical interconnection is measured as $0.64{\Omega}$. Mechanically, the maximum pressure, where the microelectrofluidic bench endures, reaches up to $115{\pm}11kPa$.

Numerical Simulation of Micro-Fluidic Flows of the Inkjet Printing Deposition Process for Microfabrication

  • Chau S.W.;Chen S.C.;Liou T.M.;Hsu K.L.;Shih K.C.;Lin Y.J.
    • 한국전산유체공학회:학술대회논문집
    • /
    • 2003.10a
    • /
    • pp.113-115
    • /
    • 2003
  • Droplet impinging into a cavity at micro-scale is one of important fluidic issues for microfabrications, e.g. bio-chip applications and inkjet deposition processes in the PLED panel manufacturing. The droplets generally dispensing from an inkjet head, which contains an array of nozzles, have a volume in several picoliters, while each nozzle jets the droplets into cavities with micron-meter size located on substrates. Due to measurement difficulties at micro-scale, the numerical simulation could serve as an efficient and preliminary way to evaluate the micro-sized droplet impinging behavior into a cavity. The micro-fluidic flow is computed by solving the three-dimensional Navier-Stokes equations through a finite volume discretization. The droplet front is predicted by a volume-of-fluid approach, in which the surface tension is modeled as a function of the fluid concentration. This paper discusses the influence of fluid properties, such as surface tension and fluid viscosity, on micro-fluidic characteristics at different jetting speeds in the deposition process via the proposed numerical approach.

  • PDF

Characterization of Dielectrophoretic Force for the Structural Shapes of Window in Microfluidic Dielectrophoretic Chip (미세유체칩내 electrode의 opening window형태에 따른 유전전기영동력 특성 규명)

  • Lee, Jaewoo;Kwak, Tae Joon;Yoon, Dae Sung;Lee, Sang Woo
    • Journal of Biomedical Engineering Research
    • /
    • v.34 no.4
    • /
    • pp.189-196
    • /
    • 2013
  • Dielectrophoresis(DEP) is useful in manipulation and separation of micro-sized particles including biological samples such as bacteria, blood cells, and cancer cells in a micro-fluidic device. Especially, those separation and manipulation techniques using DEP in combination of micro fabrication technique have been researched more and more. Recently, it is revealed that a window structure of insulating layer in microfluidic DEP chip is key role in trap of micro-particles around the window structure. However, the trap phenomenon-driven by DEP force gradient did not fully understand and is still illusive. In this study, we characterize the trap mechanism and efficiency with different shapes of window in a microfluidic DEP chip. To do this characterization, we fabricated 4 different windows shapes such as rhombus, circle, squares, and hexagon inside a micro-fluidic chip, and performed micro-sized particles manipulation experiments as varying the frequency and voltage of AC signal. Moreover, the numerical simulation with the same parameters that were used in the experiment was also performed in order to compare the simulation results and the experimental results. Those comparison shows that both results are closely matched. This study may be helpful in design and development of microfluidic DEP chip for trapping micro-scaled biological particle.

Highly Integrated DNA Chip Microarrays by Hydrophobic Interaction

  • Park, Yong-Sung;Kim, Do-Kyin;Kwon, Young-Soo
    • KIEE International Transactions on Electrophysics and Applications
    • /
    • v.11C no.2
    • /
    • pp.23-27
    • /
    • 2001
  • Microarray-based DNA chips provide an architecture for multi-analyte sensing. In this paper, we report a new approach for DNA chip microarray fabrication. Multifunctional DNA chip microarrays were made by immobilizing many kinds if DNAs on transducers (particles). DNA chip microarrays were prepared by randomly distributing a mixture of the particles on a chip pattern containing thousands of micro meter-scale sites. The particles occupied different sites from array to array. Each particle cam be distinguished by a tag that is established on the particle. The particles were arranged on the chip pattern by the random fluidic self-assembly (RFSA) method, using hydrophobic interaction.

Micro Fluidic Component for a Blood Analysis System (혈액분석기용 유체소자의 설계기술 개발)

  • Kim, Jae Yun;Kim, Duckjong;Heo, Pil Woo;Park, Sang-Jin;Yoon, Eui Soo
    • 유체기계공업학회:학술대회논문집
    • /
    • 2004.12a
    • /
    • pp.754-760
    • /
    • 2004
  • The miniaturization and integration are trend of modern blood analyses. Micro-Bio-Fluidics plays an important role in a micro blood analysis system. In this paper, analysis and design technology for blood analysis system is presented. Numerical simulations of a blood flow in micro separator and reservoir are conducted. As a result, we suggest on-chip micro separator, which performed plasma separation from whole blood in micro channels.

  • PDF

Effects of Packing Pressure and Time on Injection Molding of Plastic Micro-channel Plates (플라스틱 마이크로 채널 기판 사출성형 시 보압의 영향)

  • Woo, Sang-Won;Park, Si-Hwan
    • Journal of the Korean Society of Manufacturing Technology Engineers
    • /
    • v.25 no.3
    • /
    • pp.224-229
    • /
    • 2016
  • Recently, polymeric micro-fluidic biochips with numerous micro patterns on the surface were fabricated by injection molding for realizing low-cost mass production of devices. To evaluate the effects of process parameters on large-scale micro-structure replication, a $50{\times}50mm^2$ tool insert with surface structures having a patterns of trapezoidal shapes (height: $30{\mu}m$) was employed. During injection molding, PMMA was used; packing phase parameters and mold temperature were investigated. The replicated surface textures were quantitatively characterized by confocal laser microscopy with 10-nm resolution. The degree of replication at low mold temperatures was found to be higher than that at high mold temperature at the beginning of the packing stage. Thereafter, the degree of replication increased to a greater extent at higher mold temperatures; application of higher mold temperatures improved the degree of replication.

Fabrication and Simulation of Fluid Wing Structure for Microfluidic Blood Plasma Separation

  • Choe, Jeongun;Park, Jiyun;Lee, Jihye;Yeo, Jong-Souk
    • Applied Science and Convergence Technology
    • /
    • v.24 no.5
    • /
    • pp.196-202
    • /
    • 2015
  • Human blood consists of 55% of plasma and 45% of blood cells such as white blood cell (WBC) and red blood cell (RBC). In plasma, there are many kinds of promising biomarkers, which can be used for the diagnosis of various diseases and biological analysis. For diagnostic tools such as a lab-on-a-chip (LOC), blood plasma separation is a fundamental step for accomplishing a high performance in the detection of a disease. Highly efficient separators can increase the sensitivity and selectivity of biosensors and reduce diagnostic time. In order to achieve a higher yield in blood plasma separation, we propose a novel fluid wing structure that is optimized by COMSOL simulations by varying the fluidic channel width and the angle of the bifurcation. The fluid wing structure is inspired by the inertial particle separator system in helicopters where sand particles are prevented from following the air flow to an engine. The structure is ameliorated in order to satisfy biological and fluidic requirements at the micro scale to achieve high plasma yield and separation efficiency. In this study, we fabricated the fluid wing structure for the efficient microfluidic blood plasma separation. The high plasma yield of 67% is achieved with a channel width of $20{\mu}m$ in the fabricated fluidic chip and the result was not affected by the angle of the bifurcation.