• Title/Summary/Keyword: Implant-abutment design

Search Result 108, Processing Time 0.024 seconds

Development of a Sealing-Type Abutment for Implant and the Performance Evaluation via Structural Analysis (임플란트용 실링 어버트먼트의 개발 및 구조해석을 통한 성능분석)

  • Kim, Jeong Min;Hong, Dae Sun
    • Journal of the Korean Society for Precision Engineering
    • /
    • v.33 no.9
    • /
    • pp.769-775
    • /
    • 2016
  • Currently, dental implants are widely used as artificial teeth due to their good chewing performance and long life cycle. Generally, a dental implant consists of an abutment as the upper part and a fixture as the lower part. When chewing forces are repeatedly applied to a dental implant, a gap is often generated at the interfacial surface between the abutment and the fixture, and it results in some deterioration such as loosening of the fastening screw, dental retraction and fixture fracture. To enhance the sealing performance for coping with such problems, this study proposes a new sealing-type abutment having a number of grooves along the conical surface circumference, and it carries out finite element analysis in consideration of the external chewing force and pretension between the abutment and the fixture. The result shows that the proposed sealing-type abutment shows an enhanced sealing performance in terms of the contact pressure in comparison with a conventional abutment.

STRESS ANALYSIS OF SUPPORTING TISSUES AND IMPLANTS ACCORDING TO IMPLANT FIXTURE SHAPES AND IMPLANT-ABUTMENT CONNECTIONS (임플랜트 고정체의 형태와 연결방식에 따른 임플랜트 및 지지조직의 응력분포)

  • Han Sang-Un;Park Ha-Ok;Yang Hong-So
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.42 no.2
    • /
    • pp.226-237
    • /
    • 2004
  • Purpose: Four finite element models were constructed in the mandible having a single implant fixture connected to the first premolar-shaped superstructure, in order to evaluate how the shape of the fixture and the implant-abutment connection would influence the stress level of the supporting tissues fixtures, and prosthethic components. Material and methods : The superstructures were constructed using UCLA type abutment, ADA type III gold alloy was used to fabricate a crown and then connected to the fixture with an abutment screw. The models BRA, END , FRI, ITI were constructed from the mandible implanted with Branemark, Endopore, Frialit-2, I.T.I. systems respectively. In each model, 150 N of vertical load was placed on the central pit of an occlusal plane and 150 N of $40^{\circ}$ oblique load was placed on the buccal cusp. The displacement and stress distribution in the supporting tissues and the other components were analysed using a 2-dimensional finite element analysis . The maximum stress in each reference area was compared. Results : 1. Under $40^{\circ}$ oblique loading, the maximum stress was larger in the implant, superstructure and supporting tissue, compared to the stress pattern under vertical loading. 2. In the implant, prosthesis and supporting tissue, the maximum stress was smaller with the internal connection type (FRI) and the morse taper type (ITI) when compared to that of the external connection type (BRA & END). 3. In the superstructure and implant/abutment interface, the maximum stress was smaller with the internal connection type (FRI) and the morse taper type (ITI) when compared to that of the external connection type (BRA & END). 4. In the implant fixture, the maximum stress was smaller with the internal connection type (FRI) and the morse taper type (ITI) when compared to that of the external connection type (BRA & END). 5 The stress was more evenly distributed in the bone/implant interface through the FRI of trapezoidal step design. Especially Under $40^{\circ}$ oblique loading, The maximum stress was smallest in the bone/implant interface. 6. In the implant and superstructure and supporting tissue, the maximum stress occured at the crown loading point through the ITI. Conclusion: The stress distribution of the supporting tissue was affected by shape of a fixture and implant-abutment connection. The magnitude of maximum stress was reduced with the internal connection type (FRI) and the morse taper type (ITI) in the implant, prosthesis and supporting tissue. Trapezoidal step design of FRI showed evenly distributed the stress at the bone/implant interface.

Influence of Implant Fixture-Abutment Connection and Abutment Design on Mechanical Strength (임플란트 고정체-지대주 연결부 및 지대주 디자인이 기계적 강도에 미치는 영향)

  • Chun, Mi-Hyun;Jeong, Chang-Mo;Jeon, Young-Chan;Eom, Tae-Gwan;Yoon, Ji-Hoon
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.24 no.3
    • /
    • pp.269-281
    • /
    • 2008
  • Fatigue or overload can result in mechanical problems of implant components. The mechanical strength in the implant system is dependent on several factors, such as screw and fixture diameters, material, and design of the fixture-abutment connection and abutment. In these factors, the last rules the strength and stability of the fixture-abutment assembly. There have been some previous reports on the mechanical strength of the fixture-abutment assembly with the compressive bending test or short-term cyclic loading test. However, it is restrictive to predict the long-term stability of the implant system with them. The purpose of this study was to evaluate the influence of the design of the fixture-abutment connection and abutment on the mechanical strength and failure mode by conducting the endurance limit test as well as the compressive bending strength test. Tests were performed according to a specified test(ISO/FDIS 14801) in 4 fixture-abutment assemblies of the Osstem implant system: an external butt joint with Cemented abutment (group BJT), an external butt joint with Safe abutment (group BJS), an internal conical joint with Solid abutment (group CJO), and an internal conical joint with ComOcta abutment (group CJT). The following conclusions were drawn within the limitation of this study. Compressive bending strengths were decreased in order of group BJS(1392.0N), group CJO(1261.8N), group BJT(1153.2N), and group CJT(1110.2N). There were no significant differences in compressive bending strengths between group BJT and group CJT(P>.05). Endurance limits were decreased in order of group CJO(600N), group CJT(453N), group BJS(360N), and group BJT(300N). 3. Compressive bending strengths were influenced by the connection and abutment design of the implant system, however endurance limits were affected more considerably by the connection design.

Three-dimensional finite element analysis of platform switched implant

  • Moon, Se-Young;Lim, Young-Jun;Kim, Myung-Joo;Kwon, Ho-Beom
    • The Journal of Advanced Prosthodontics
    • /
    • v.9 no.1
    • /
    • pp.31-37
    • /
    • 2017
  • PURPOSE. The purpose of this study was to analyze the influence of the platform switching concept on an implant system and peri-implant bone using three-dimensional finite element analysis. MATERIALS AND METHODS. Two three-dimensional finite element models for wide platform and platform switching were created. In the wide platform model, a wide platform abutment was connected to a wide platform implant. In the platform switching model, the wide platform abutment of the wide platform model was replaced by a regular platform abutment. A contact condition was set between the implant components. A vertical load of 300 N was applied to the crown. The maximum von Mises stress values and displacements of the two models were compared to analyze the biomechanical behavior of the models. RESULTS. In the two models, the stress was mainly concentrated at the bottom of the abutment and the top surface of the implant in both models. However, the von Mises stress values were much higher in the platform switching model in most of the components, except for the bone. The highest von Mises values and stress distribution pattern of the bone were similar in the two models. The components of the platform switching model showed greater displacement than those of the wide platform model. CONCLUSION. Due to the stress concentration generated in the implant and the prosthodontic components of the platform switched implant, the mechanical complications might occur when platform switching concept is used.

Long-term effect of implant-abutment connection type on marginal bone loss and survival of dental implants

  • Young-Min Kim;Jong-Bin Lee;Heung-Sik Um;Beom-Seok Chang;Jae-Kwan Lee
    • Journal of Periodontal and Implant Science
    • /
    • v.52 no.6
    • /
    • pp.496-508
    • /
    • 2022
  • Purpose: This study aimed to compare the long-term survival rate and peri-implant marginal bone loss between different types of dental implant-abutment connections. Methods: Implants with external or internal abutment connections, which were fitted at Gangneung-Wonju National University Dental Hospital from November 2011 to December 2015 and followed up for >5 years, were retrospectively investigated. Cumulative survival rates were evaluated for >5 years, and peri-implant marginal bone loss was evaluated at 1- and 5-year follow-up examinations after functional loading. Results: The 8-year cumulative survival rates were 93.3% and 90.7% in the external and internal connection types, respectively (P=0.353). The mean values of marginal bone loss were 1.23 mm (external) and 0.72 mm (internal) (P<0.001) after 1 year of loading, and 1.20 mm and 1.00 mm for external and internal abutment connections, respectively (P=0.137) after 5 years. Implant length (longer, P=0.018), smoking status (heavy, P=0.001), and prosthetic type (bridge, P=0.004) were associated with significantly greater marginal bone loss, and the use of screw-cement-retained prosthesis was significantly associated (P=0.027) with less marginal bone loss. Conclusions: There was no significant difference in the cumulative survival rate between implants with external and internal abutment connections. After 1 year of loading, marginal bone loss was greater around the implants with an external abutment connection. However, no significant difference between the external and internal connection groups was found after 5 years. Both types of abutment connections are viable treatment options for the reconstruction of partially edentulous ridges.

THREE-DIMENSIONAL FINITE ELEMENT STRESS ANALYSIS OF SINGLE IMPLANT RESTORATION USING DIFFERENT FIXTURE AND ABUTMENT SCREW DIAMETERS (단일치 임플랜트 지지 보철물에서 고정체와 지대주 나사 직경의 차이에 따른 삼차원 유한요소법적 응력 분석)

  • Kwon Joo-Hong;Choi Min-Ho;Kim Yu-Lee;Cho Hye-Won
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.43 no.1
    • /
    • pp.105-119
    • /
    • 2005
  • Statement of problem. As the effects of the various diameters of fixture and abutment screw on stress distribution was not yet examined, this study focused on the different design of single implant restoration using three dimensional finite element analysis. Purpose. This study was to compare five different fixture-abutment combinations for single implant supported restorations with different fixture and abutment screw diameters. Material of methods. The five kinds of finite element models were designed by 3 diameter fixtures ($\oslash$3.3, 3.75, 5.0 mm) with 3 different abutment screws $\oslash$1.5, 1.7, 2.0 mm). The crown for mandibular first molar was made using UCLA abutment according to Wheeler's anatomy. 244 N was applied at the central fossa with two different loading directions, vertically and obliquely (30$^{\circ}$) and at the buccal cusp vertically. Maximum von Mises stresses were recorded and compared in the supporting bone, crowns, fixtures, and abutment screws. Results. 1. The stresses in supporting bone and implant-abutment structure under oblique loading were greater than those under vertical or offset loading. The stresses under vertical loading were the least among 3 loading conditions regardless of the implant and abutment screw diameters. 2. The stresses in the narrow implants were greater than the wider implants. The narrow implant with narrow abutment screw showed highest stresses in the lingual crest, but the narrow implant with standard abutment screw showed highest stress in abutment screw. 3. The stresses of abutment screws were influenced by the diameter of fixtures and loading conditions. The wide implants showed least difference between two different abutment screw diameters. Conclusions. The wide implants showed lesser stresses than the narrow implants and affected least by the different abutment screw diameters. The narrow implants with standard abutment screw showed highest stresses in the lingual bony crest under oblique loading.

A STUDY ON THE FIT OF THE IMPLANT-ABUTMENT-SCREW INTERFACE (임플란트-지대주-나사의 적합에 관한 연구)

  • Kim Nak-Hyung;Chung Chae-Heon;Son Mee-Kyoung;Back Dae-Hwa
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.41 no.4
    • /
    • pp.503-518
    • /
    • 2003
  • Statement of problem : There have been previous studies about considerable variations in machining accuracy and consistency in the implant-abutment-screw interfaces. Purpose : The purpose of this study was to evaluate the machining accuracy and consistency of implant/abutment/screw combinations on two randomly selected implants from each of four manufactures. Material and methods : In this study, screws were respectively used to secure a cemented abutment, to a hexlock implant fixture ; teflon coated titanium alloy screw(Torq-Tite) and titanium alloy screw in Steri--Oss system, gold-plated gold-palladium alloy screw(Gold-Tite) and titanium alloy screw in 3i system gild screw ana titanium screw in AVANA Dental Implant system, and titanium screws in Paragon System. The implants were perpendicularly mounted in polymethyl methacrylate autopolymerizing acrylic resin block(Orthodontic resin, Densply International Inc. USA) by use of dental surveyer. Each abutment screw was secured to the implant with recommended torque value using a digital torque controller. Each screw was again tightened after 10 minutes. All samples were cross sectioned with grinder-polisher unit(Omnilap 2000 SBT Inc) after embeded in liquid unsaturated polyester (Epovia, Cray Valley Inc) Results : There were the largest gaps in the neck areas of screws in hexagonal extension implants which were examined in this study. The leading edge of the abutment screw thread (superior surface) was in contact with the implant body thread, and the majority of the contacting surfaces were localized to the middle portion of the mating threads. Considerable variation in the contacting surfaces was noted in the samples evaluated. Amounts of contact in the abutment screw thread were larger for assemblies with Gold-Tite screw, gold alloy screw. Torq-Tite screw than those with titanium screws. The findings of intimate contact between the screw and screw seat were seen in all samples, regardless of manufacturers. However, microgap between the head and lateral neck surface of the screw and the abutment could be dectected in all samples. The findings of intimate contact between the platform of the implant and the bottom of the abutment were consistent in all samples, regardless of manufacturers. However, microgaps between the lateral surface of external hex of the fixture and the abutment could be dectected in all samples. Conclusion : Considerable variations in machining accuracy and consistency were noted in the samples and the implant-abutment-screw interfaces were incomplete. From the results of this study, further development of the system will be required, including improvements in pattern design.

THE INFLUENCE OF IMPLANT FIXTURE-ABUTMENT CONNECTION DESIGN ON SCREW LOOSENING (임플랜트 지대주에 따른 나사 풀림의 연구)

  • Mun Yang-Suk;Park Sang-Won;Vang Mong-Sook;Yang Hong-So;Park Ha-Ok
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.44 no.2
    • /
    • pp.174-184
    • /
    • 2006
  • Purpose: Current trend in implant dentistry is changing from external connection to internal connection. To evaluate the splinting of external and internal connection implant on screw loosening, 2-units prosthesis was fabricated with BioPlant $System^(R)$ of external connection type and Lifecore STAGE-1 Single Stage Implant $System^(R)$ of internal connection type. Material and Method: Experimental group is classified into three groups. 1) $G_1-EE$: 2-units prosthesis was fabricated with two Bioplant $System^(R)$ of external connection type. 2) $G_1-EI$: 2-units prosthesis was fabricated with one BioPlant $System^(R)$ of external connection type and one Lifecore STAGE-1 Single Stage Implant $System^(R)$ of internal connection type. 3) $G_1-II$: 2-units prosthesis was fabricated with two Lifecore STAGE-1 Single Stage Implant $System^(R)$ of internal connection type. In fabricating 2-units prosthesis, two hexed abutments are recommended when two implants are installed parallel, otherwise one hexed abutment is used on major occlusal force area and one nonhexed abutment is used on the other area. Since it is rare to find two implants being parallel, it is hard to fabricate prosthesis with passive adaptation using two hexed abutments. It is much more difficult to acquire passive adaptation when using hex abutment compared to nonhex abutment. To evaluate the influence of hexed and nonhexed abutment on screw loosening, 2-units prosthesis was fabricated with hexed and nonhexed abutment. Experimental group is classified into three groups. 1) $G_2-HH$: 2-units prosthesis was fabricated with two hexed abutments. 2) $G_2-HN$: 2-units prosthesis was fabricated with one hexed abutment and one nonhexed abutment. 3) $G_2-NN$: 2-units prosthesis was fabricated with two nonhexed abutments. Result: The results of comparing the detorque value after loading on a each prosthesis periodically are as follows. 1. In splinting group of external and internal connection implant, $G_1-II$ group demonstrated the biggest detorque value, followed by $G_1-EI$ group and $G_1-EE$ group. 2. There is no notable significance between external connection implant of $G_1-EI$ group and $G_1-EE$ group and also no significance between internal connection implant of $G_1-EI$ group and $G_1-II$ group. 3. $G_2-HH$ group showed higher detorque value than $G_2-HN\;and\;G_2-NN$ group. From the results, we can concluded that using both external connection and internal connection implant together is clinically acceptable and in order to acquire a good passive adaptation in fabricating 2-units implant prosthesis we can use two nonhexed abutments.

Influence of the implant abutment types and the dynamic loading on initial screw loosening

  • Kim, Eun-Sook;Shin, Soo-Yeon
    • The Journal of Advanced Prosthodontics
    • /
    • v.5 no.1
    • /
    • pp.21-28
    • /
    • 2013
  • PURPOSE. This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS. Three groups of abutments were produced using different types of fabrication methods; stock abutment, gold cast abutment, and CAD/CAM custom abutment. A customized jig was fabricated to apply the load at $30^{\circ}$ to the long axis. The implant fixtures were fixed to the jig, and connected to the abutments with a 30 Ncm tightening torque. A sine curved dynamic load was applied for $10^5$ cycles between 25 and 250 N at 14 Hz. Removal torque before loading and after loading were evaluated. The SPSS was used for statistical analysis of the results. A Kruskal-Wallis test was performed to compare screw loosening between the abutment systems. A Wilcoxon signed-rank test was performed to compare screw loosening between before and after loading in each group (${\alpha}$=0.05). RESULTS. Removal torque value before loading and after loading was the highest in stock abutment, which was then followed by gold cast abutment and CAD/CAM custom abutment, but there were no significant differences. CONCLUSION. The abutment types did not have a significant influence on short term screw loosening. On the other hand, after $10^5$ cycles dynamic loading, CAD/CAM custom abutment affected the initial screw loosening, but stock abutment and gold cast abutment did not.

PHOTOELASTIC STRESS ANALYSIS OF LOAD TRANSFER TO SATELLITE ABUTMENT AS AN IMMEDIATE ABUTMENT (인공치아의 즉시부하를 위해 새로이 개발된 인공치아 지대주(Satellite Abutment)의 광탄성 응력 분석)

  • Park, Sang-Kyu;Lee, Baek-Soo;Engelke, W;Kim, Boo-Dong
    • Journal of the Korean Association of Oral and Maxillofacial Surgeons
    • /
    • v.28 no.6
    • /
    • pp.472-479
    • /
    • 2002
  • Since $Br^{\circ}anemark$ introduced the osseointegrated implants, they have been granted for useful methods for the restoration of oral function. The original $Br^{\circ}anemark$ protocol recommended long stress-free healing periods to achieve the osseointegration of dental implants. However, many clinical and experimental studies have shown that the osseointegration is no wonder in almost cases and that early and immediate loading may lead to predictable osseointegration. So we are willing to introduce the Satellite Abutment newly invented for immediate loading. We think that it will make the occlusal forces dispersed to surrounding bone and that we can restore the oral function immediately after implant installation not disturbing osseointegration. In case of using Satellite abutment, stress concentrated to bone contact area of implant was distributed not only fixation plate and screws but also superior, middle portion of implant and cortical layer of jaw bone. It was clearly decreased on the bone contact surfaces around dental implants. 1. Stress was decreased more than 76.5% when satellite straight abutment was used. 2. Stress was decreased more than 50% when satellite angled abutment was used. 3. The stress around dental implant was well distributed along the cortical bone surface and the fixation plate and screw. This study concludes that satellite abutment can be used as all immediate loading implant prothesis because it was possible to distribute periimplant occlusal stress through implant contact bone surface and cortical layer of jaw bone.