Development of a Sealing-Type Abutment for Implant and the Performance Evaluation via Structural Analysis  

Kim, Jeong Min (Department of Mechanical Design and Manufacturing, Changwon National University)
Hong, Dae Sun (School of Mechanical Engineering, Changwon National University)
Publication Information
Abstract
Currently, dental implants are widely used as artificial teeth due to their good chewing performance and long life cycle. Generally, a dental implant consists of an abutment as the upper part and a fixture as the lower part. When chewing forces are repeatedly applied to a dental implant, a gap is often generated at the interfacial surface between the abutment and the fixture, and it results in some deterioration such as loosening of the fastening screw, dental retraction and fixture fracture. To enhance the sealing performance for coping with such problems, this study proposes a new sealing-type abutment having a number of grooves along the conical surface circumference, and it carries out finite element analysis in consideration of the external chewing force and pretension between the abutment and the fixture. The result shows that the proposed sealing-type abutment shows an enhanced sealing performance in terms of the contact pressure in comparison with a conventional abutment.
Keywords
Dental implant; Sealing-Type abutment; Finite element analysis; Fixture;
Citations & Related Records
Times Cited By KSCI : 2  (Citation Analysis)
연도 인용수 순위
1 Branemark, P.-I, Breine, U., Adell, R., Hansson, B. O., Lindstrom, J., et al., "Intra-Osseous Anchorage of Dental Prostheses: I. Experimental Studies," Scandinavian Journal of Plastic and Reconstructive Surgery, Vol. 3, No. 2, pp. 81-100, 1969.   DOI
2 Gaviria, L., Salcido, J. P., Guda, T., and Ong, J. L., "Current Trends in Dental Implants," Journal of the Korean Association of Oral and Maxillofacial Surgeons, Vol. 40, No. 2, pp. 50-60, 2014.   DOI
3 McGlumphy, E. A., Mendel, D. A., and Holloway, J. A., "Implant Schew Mechanics," Dental Clinics of North America, Vol. 42, No. 1, pp. 71-89, 1998.
4 Carlson, B. and Carlsson, G. E., "Prosthodontic Complications in Osseointegrated Dental Implant Treatment," International Journal of Oral and Maxillofacial Implants, Vol. 9, No. 1, p. 90, 1994.
5 Siamos, G., Winkler, S., and Boberick, K. G., "The Relationship between Implant Preload and Screw Loosening on Implant-Supported Prostheses," Journal of Oral Implantology, Vol. 28, No. 2, pp. 67-73, 2002.   DOI
6 Haack, J. E., Sakaguchi, R. L., Sun, T., and Coffey, J. P., "Elongation and Preload Stress in Dental Implant Abutment Screws," International Journal of Oral and Maxillofacial Implants, Vol. 10, No. 5, p. 529, 1995.
7 Martin, W. C., Woody, R. D., Miller, B. H., and Miller, A. W., "Implant Abutment Screw Rotations and Preloads for Four Different Screw Materials and Surfaces," The Journal of Prosthetic Dentistry, Vol. 86, No. 1, pp. 24-32, 2001.   DOI
8 Lee, M.-Y., Heo, S.-J., Park, E.-J., and Park, J.-M., "Comparative Study on Stress Distribution around Internal Tapered Connection Implants according to Fit of Cement-and-Screw-Retained Prosthesis," The Journal of Advanced Prosthodontics, Vol. 5, No. 3, pp. 312-318, 2013.   DOI
9 Kim, N., Choi, K. H., and Oh, H. S., "The Evaluation of Fitness on Implant Bolt Joints as Tightening Torques," Proc. of the Korean Society of Mechanical Engineers Autumn Conference, pp. 1786-1791, 2010.
10 ANSYS Inc., "ANSYS Mechanical APDL Structural Analysis Guide," http://congtrinhthep.vn/tailieuketcaucom/Umy/28-01-2012/ansys13_structural%20guide.pdf, (Accessed 13 December 2015)
11 Koak, J. Y., Kim, H. W., Lee, K. R., and Heo, S. J., "DLC-Coated Dental Titanium Implant and The Method of Processing Thereof," KOR Patent, No. 1020010033305, 2002.
12 Hong, J. P., "Mechanical Design," BooksHill, pp. 100-106, 2001.
13 ISO 14801, "Dentistry-Implants-Dynamics Fatigue Test for Endosseous Dental Implants," 2007.
14 ANSYS Inc., Workbench Training Manual, "WB Mechanical Mesh," pp. 11-21, 2015.