Practical Considerations in Preparing an Institutional Procedure of Image Guided Radiation Therapy (방사선 치료용 영상 장치 지침서 작성을 위한 실용적인 고려사항)

  • Yi, Byong Yong
    • Progress in Medical Physics
    • /
    • v.24 no.4
    • /
    • pp.205-212
    • /
    • 2013
  • Recent developments of image guided radiation therapy (IGRT), especially the On Board Imaging (OBI) system and the cone beam CT (CBCT), enable the radiation treatment more accurate and reliable. IGRT is widely used in the radiation therapy as a standard of care. Use of IGRT is even expected to increase in the near future. IGRT is only beneficial to patients when it is used with proper considerations of safety and appropriateness of the techniques. Institutional procedure should be developed based on the clinical need and the deep understanding of the system before applying the new technique to the clinic. Comprehensive QA program should be established before to the clinic and imaging dose should be considered when preparing the departmental practice guidelines for IGRT.

Verify Image-Guided Shifts for 6DoF Couch using Yonsei Cancer Center QA Set (Yonsei Cancer Center QA Set을 이용한 6DoF Couch의 이동 정확성 검증)

  • Jung, Dongmin;Park, Hyokuk;Yoon, Jongwon;Lee, Sangkyu;Kim, Jooho;Cho, Jeonghee
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.29 no.1
    • /
    • pp.7-18
    • /
    • 2017
  • Purpose: A QA Set was established to verify the movement accuracy of image-guided 6DoF Couch and to evaluate its usefulness. Materials and Methods: Two sets of linear accelerators equipped with 6DoF Couch and CBCT were used. Using the established QA Set, each CBCT image was obtained over 15 times through the Penta-Guide Phantom installed with off-set shift values along six translational (Translation; TX, TY, TZ) and rotational (Rotation, Pitch; RX, Roll; RY, Yaw; RZ) directions. Using this method, we compared the reference image and the registration image, and we analyzed the error calculated by measuring the positional accuracy of the modified 6DoF Couch. Results: The Air Cavity corresponding to the Pixel of the reference image and the registration image were all contained between 30 and 66, and the revealing high registration accuracy. Error between the modified off-set value of 6DoF Couch and the measured value along translational directions were $0.25{\pm}0.18mm$ in the TX direction, $0.25{\pm}0.25mm$ in the TY direction, and $0.36{\pm}0.2mm$ in the TZ direction. Misalignments along the rotational axis were $0.18{\pm}0.08^{\circ}$ in the RX direction, $0.26{\pm}0.09^{\circ}$ in the RY direction, and $0.11{\pm}0.08^{\circ}$ in the RZ direction, it was corrected precisely for any value. Conclusion: Using the YCC QA Set, we were able to verify the error of 6DoF Couch along both the translational and rotational directions in a very simple method. This system would be useful in performing Daily IGRT QA of 6DoF Couch.

  • PDF

Accuracy and radiation exposure from image-guidance in Tomotherapy Hi-Art System (토모테라피 Hi-Art System의 영상유도 정확성 평가와 환자 피폭에 관한 연구)

  • Jang, Jae-Uk;Lim, Hyun-Soo;Han, Man-Seok;Kim, Yong-Kyun;Jeon, Min-Cheol
    • Journal of Digital Convergence
    • /
    • v.11 no.10
    • /
    • pp.577-584
    • /
    • 2013
  • IGRT(Image Guided Radiation Therapy) in radiation therapy is a very useful technique in order to increase setup of patient and position reproducibility. Tomotherapy can increase accuracy of setup to take IGRT by MVCT, but it be for verified accuracy of Image guided, and MVCT occurs the exposure of patient. Through this study, IGRT accuracy of Tomotherapy is very accurate within 1.0mm. When MVCT using Tomotherapy phantom for QA, QC be taken, exposure dose is Fine(2mm Slice thickness) 3cGy, Normal(4mm Slice thickness) 1.5cGy, Corse(6mmSlice thickness) 1.0cGy. Measurement value of spatial resolution using AAPM CT performance phantom did't cause a big difference. As a result, ability of IGRT in Tomotherapy is very accurate. While obtaining image for IGRT, we should minimize expose range because patient's be exposed to radiation. We should make an effort to do accurate radiation therapy to minimize exposure of patient by selecting the appropriate thickness of MVCT depending on patient's body and treat area.

The effective quality assurance for image guided device using the AMC G-Box (AMC G-Box를 이용한 영상유도장치의 효율적인 정도관리)

  • Kim, Chong Mi
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.199-206
    • /
    • 2014
  • Purpose : According to the rapid increase recently in image-guided radiation therapy, It is necessary to control of the image guidance system completely. In particular for the main subject to the accuracy of image guided radiation therapy device to be done essentially the quality assurance. We made efficient phantom in AMC for the management of the accurate and efficient. Materials and Methods : By setting up of five very important as a quality assurance inventory of the Image guidance system, we made (AMC G-Box) phantom for quality assurance efficient and accurate. Quality assurance list were the Iso-center align, the real measurement, the center align of four direction, the accuracy of table movement and the reproducibility of Hounsfield Unit. The rectangular phantom; acrylic with a thickness of 1 cm to $10cm{\time}10cm{\time}10cm$ was inserted the three materials with different densities respectively for measure the CBCT HU. The phantom was to perform a check of consistency centered by creating a marker that indicates the position of the center fixed. By performing the quality assurance using the phantom of existing, comparing the resulting value to the different resulting value using the AMC G-Box, experiment was analyzed time and problems. Therapy equipment was used Varian device. It was measured twice at 1-week intervals. Results : When implemented quality assurance of an image guidance system using AMC G-Box and a phantom existing has been completed, the quality assurance result is similar in $0.2mm{\pm}0.1$. In the case of the conventional method, it was 45 minutes at 30 minutes. When using AMC G-Box, it takes 20 minutes 15 minutes, and declined to 50% of the time. Conclusion : The consistency and accurate of image guidance system tend to decline using device. Therefore, We need to perform thoroughly on the quality assurance related. It needs to be checked daily to consistency check especially. When using the AMC G-Box, It is possible to enhance the accuracy of the patient care and equipment efficiently performing accurate quality assurance.

A Study on Mechanical Errors in Cone Beam Computed Tomography(CBCT) System (콘빔 전산화단층촬영(CBCT) 시스템에서 기계적 오류에 관한 연구)

  • Lee, Yi-Seong;Yoo, Eun-Jeong;Kim, Seung-Keun;Choi, Kyoung-Sik;Lee, Jeong-Woo;Suh, Tae-Suk;Kim, Joeng-Koo
    • Journal of radiological science and technology
    • /
    • v.36 no.2
    • /
    • pp.123-129
    • /
    • 2013
  • This study investigated the rate of setup variance by the rotating unbalance of gantry in image-guided radiation therapy. The equipments used linear accelerator(Elekta Synergy TM, UK) and a three-dimensional volume imaging mode(3D Volume View) in cone beam computed tomography(CBCT) system. 2D images obtained by rotating $360^{\circ}$and $180^{\circ}$ were reconstructed to 3D image. Catpan503 phantom and homogeneous phantom were used to measure the setup errors. Ball-bearing phantom was used to check the rotation axis of the CBCT. The volume image from CBCT using Catphan503 phantom and homogeneous phantom were analyzed and compared to images from conventional CT in the six dimensional view(X, Y, Z, Roll, Pitch, and Yaw). The variance ratio of setup error were difference in X 0.6 mm, Y 0.5 mm Z 0.5 mm when the gantry rotated $360^{\circ}$ in orthogonal coordinate. whereas rotated $180^{\circ}$, the error measured 0.9 mm, 0.2 mm, 0.3 mm in X, Y, Z respectively. In the rotating coordinates, the more increased the rotating unbalance, the more raised average ratio of setup errors. The resolution of CBCT images showed 2 level of difference in the table recommended. CBCT had a good agreement compared to each recommended values which is the mechanical safety, geometry accuracy and image quality. The rotating unbalance of gentry vary hardly in orthogonal coordinate. However, in rotating coordinate of gantry exceeded the ${\pm}1^{\circ}$ of recommended value. Therefore, when we do sophisticated radiation therapy six dimensional correction is needed.