Determination of Relative Output Factors Using PinPoint Ion Chamber and Gafchromic Film in Gamma Knife

  • Seo, Weon-Seop;Shin, Dong-Oh;Lim, Young-Jin;Im, Yong-Seok
    • 한국의학물리학회:학술대회논문집
    • /
    • 한국의학물리학회 2002년도 Proceedings
    • /
    • pp.300-301
    • /
    • 2002
  • Of many modalities for measuring output factor, we measured the outputs of 18mm, 14mm, 8mm and 4mm helmet in Gamma knife using ion chambers and a Gafrcromic MD-55 film and normalized the outputs of four helmets the one of a 18mm helmet. Ion chambers used for this paper were a PR-05P and a PinPoint having a volume of 0.07cc and 0.015cc respectively. The recommended output factors from a manufacture were 1, 0.984, 0.956, and 0.87. For PR-05P, the output factor of 14mm helmet showed a good agreement, but the ones of 8mm and 4mm helmet showed the difference of 4.6% and 47% respectively. For PinPoint, The output factors of 14mm and 8mm helmet showed a good agreement, but the one of a 4mm helmet showed a difference of 18%. The Gafcromic MD-55, however, showed a good agreement for all helmets.

  • PDF

DICOM 파일을 사용한 Geant4 시뮬레이션과 Gafchromic EBT2 필름에 의한 인체 내 흡수선량 비교 연구 (Comparative Studies on Absorbed Dose by Geant4-based Simulation Using DICOM File and Gafchromic EBT2 Film)

  • 모은희;이상호;안성환;김종일
    • 한국의학물리학회지:의학물리
    • /
    • 제24권1호
    • /
    • pp.48-53
    • /
    • 2013
  • 몬테카를로 방식은 지금까지 인체 내 흡수선량을 계산하는 가장 정확한 방법으로 알려져 왔고. 이러한 계산 방법을 이용하기 위한 인체 내부의 장기 묘사는 인체 모형 팬텀이 주로 사용되어 왔다. 그러나 최근 Geant4 코드를 사용한 몬테카를로 계산에서는 CT의 DICOM 파일에서 인체의 여러 장기에 대한 자료를 직접 추출하고 시뮬레이션에 필요한 geometry로 변환하여 사용하려는 다양한 노력이 시도되고 있다. 이와 같은 기능은 실제 인체의 해부학적 구조를 그대로 재현하면서 인체 내부의 흡수선량을 정확히 계산 할 수 있도록 한다. 따라서 본 연구에서는 DICOM 파일을 연동한 Geant4을 이용하여 인체 내 흡수선량을 계산하였고, 이를 Gafchromic EBT2 필름을 이용한 측정 선량과 비교함으로써 그 유용성을 확인하고자 하였다. 본 연구에서 시뮬레이션을 이용하여 계산한 선량과 EBT2 필름을 이용한 선속 중심축에서의 측정선량을 비교한 결과 피부표면에서부터 최대선량 깊이까지 선량이 급격하게 변화하는 build up 영역을 제외하고는 오차(difference) 범위가 평균 3.75% 임을 알 수 있었다. 또한 선량의 계산 값을 각 CT slice 별로 출력되도록 하였고, 또 각 slice에서도 복셀 하나하나의 선량 값이 출력되도록 하여 측정하고자 하는 장기별, 기관별 흡수선량을 쉽게 확인 할 수 있도록 하였다. 이처럼 인체 모형 팬텀이 아닌 실제 인체의 image data인 CT DICOM 파일을 이용한 선량계산을 각 slice, voxel 별로 선량 값을 출력하는 방식은 다양한 부위의 정확한 선량계산을 가능하게 하므로 향후 방사선 치료계획 시스템의 선량 계산에 유용할 것이라 생각한다. 또한 현재 사용 중인 여러 에너지 영역에도 적용이 가능하므로 인체 내 방사선의 흡수선량 확인을 위해 유용하게 활용되어질 수 있을 것으로 생각된다.

호흡동조방사선 치료 시 종양 치료의 정확도 평가 (Accuracy Evaluation of Tumor Therapy during Respiratory Gated Radiation Therapy)

  • 장은성;강수만;이철수;강세식
    • 대한방사선치료학회지
    • /
    • 제22권2호
    • /
    • pp.113-122
    • /
    • 2010
  • 목 적: 호흡동조방사선 치료 시 종양의 실제 움직임과 호흡추적장치로 측정한 피부 움직임의 차이를 자체 제작한 구동팬텀에서 가상의 종양을 이용하여 호흡과 유사하게 움직임에 따른 정적 상태, 동적 상태 및 호흡동조상태에서 표적 위치의 정확성의 측정치와 실측치를 평가하고 선량분포를 분석, 비교하고자 한다. 대상 및 방법: 호흡에 의해 움직이는 종양 측정을 위해 2차원적으로 움직이는 구동팬텀을 자체 제작하였다. 구동 팬톰의 움직임은 위. 아래 방향(SI) 각각 1.5 cm 왕복운동, 상 하 방향 2 cm으로 속도조절(1-5단계)이 되도록 하였다. 가로 4 cm, 세로 4 cm, 높이 0.5 cm의 아크릴 슬라이스에 직경 0.5 cm 종양을 납으로 표시하고, 위아래로 동일한 아크릴 슬라이스를 2장씩 쌓은 후 아크릴 슬라이스 세 번째와 네 번째 사이 Dmax 1.5 cm film을 삽입하였다. 가상의 타겟을 구동 팬텀 위에 위치시키고 6 MV X-선이 조사되는 정적인 상태, 호흡동조 및 동적인 상태에서 각각 5 Gy를 조사하였다. 구동팬텀 위에 표식자를 올린 후, 호흡추적장치를 이용하여 사전에 설정한 호흡시간의 변화에 따른 진폭과 위상변화를 분석하였다. 결 과: RPM respiratory gating system을 이용하여 호흡주기를 8단계로 나누어 각각을 12회씩 위상변화를 분석하여 평균과 표준편차를 구한결과 평균은 3.0 (1.5~1.5) sec에서 1.7 cm로 가장 크고, 3.0 (1.3~1.7) sec 5.0 (2.0~3.0) sec에서 0.2602 cm로 가장 크고 4.0 (2.0~2.0) sec에서 0.0866 cm로 가장 작았다. 또한 실측치에서 평균 및 표준편차를 구한 결과 t0에서 9.9 (6.6) mm $t_{10}$에서 10.6 mm (7.3), $t_{20}$ 16.5 mm (10.3), $t_{30}$ 10.2 mm (7.6)으로 나타났으며, 호기나 흡기 시간의 차이에 따른 규칙은 없고 대체로 균일한 평균과 표준편차의 분포를 나타내었다. 또한 정적 상태, 동적 상태 및 호흡동조상태에서 Gafchromic EBT film 의 방사선량을 분석한 결과, ICRU 62에서 권고한 90% 선량분포가 3 mm 이내에 포함되므로 정확성과 정도관리 측면에서 적합한 것으로 사료된다. 결 론: 구동팬톰을 이용하여 호흡움직임에 따른 정확성 및 선량분포차이를 Gafchromic film을 통하여 확인하였으며 결과를 바탕으로 호흡에 의해 변화가 생기는 장기에 대한 차이를 고려하여 치료계획을 한다면 종양과 정상조직에 적절한 선량계획을 세울 수 있어 치료효과 향상에 도움을 주게 될 것으로 생각한다.

  • PDF

Linear Energy Transfer Dependence Correction of Spread-Out Bragg Peak Measured by EBT3 Film for Dynamically Scanned Proton Beams

  • Lee, Moonhee;Ahn, Sunghwan;Cheon, Wonjoong;Han, Youngyih
    • 한국의학물리학회지:의학물리
    • /
    • 제31권4호
    • /
    • pp.135-144
    • /
    • 2020
  • Purpose: Gafchromic films for proton dosimetry are dependent on linear energy transfers (LETs), resulting in dose underestimation for high LETs. Despite efforts to resolve this problem for single-energy beams, there remains a need to do so for multi-energy beams. Here, a bimolecular reaction model was applied to correct the under-response of spread-out Bragg peaks (SOBPs). Methods: For depth-dose measurements, a Gafchromic EBT3 film was positioned in water perpendicular to the ground. The gantry was rotated at 15° to avoid disturbances in the beam path. A set of films was exposed to a uniformly scanned 112-MeV pristine proton beam with six different dose intensities, ranging from 0.373 to 4.865 Gy, at a 2-cm depth. Another set of films was irradiated with SOBPs with maximum energies of 110, 150, and 190 MeV having modulation widths of 5.39, 4.27, and 5.34 cm, respectively. The correction function was obtained using 150.8-MeV SOBP data. The LET of the SOBP was then analytically calculated. Finally, the model was validated for a uniform cubic dose distribution and compared with multilayered ionization chamber data. Results: The dose error in the plateau region was within 4% when normalized with the maximum dose. The discrepancy of the range was <1 mm for all measured energies. The highest errors occurred at 70 MeV owing to the steep gradient with the narrowest Bragg peak. Conclusions: With bimolecular model-based correction, an EBT3 film can be used to accurately verify the depth dose of scanned proton beams and could potentially be used to evaluate the depth-dose distribution for patient plans.

Radiochromic film dosimetry for linac-based stereotactic radiosurgery

  • Han, Seung-Hee;Park, Suk-Won;Oh, Do-Hoon
    • 한국의학물리학회:학술대회논문집
    • /
    • 한국의학물리학회 2002년도 Proceedings
    • /
    • pp.302-304
    • /
    • 2002
  • In linac-based stereotactic radiosurgery, assuring the quality of the planning and delivery of external photon beam requires accurate evaluation of beam parameters, usually including output factors, tissue-phantom ratio and off-axis ratios, and measurement of actual dose distributions from simulated treatment. We're going to test the use of calibrated radio chromic film (Gafchromic film; type MD-55, Nuclear associate) using a Lumiscan 75 digitizer to measure absolute dose and relative dose distributions for linac-based radiosurgery unit Relative dose distribution of a human-style spherical acryl phantom were measured using radiochromic film and calculated by treatment planning system. The absolute dose at the sphere center was measured by radiochromic film and micro chamber (Exradin A-14, 0.009cc). What we want to demonstrate in this work, the 'well selected' radiochromic films when external photon beam are used in linac-based stereotactic radiosurgery are very accurate detector for dosimetry.

  • PDF

Gafchromic EBT2필름과 다양한 검출기를 이용하여 $3cm^2$ 이하의 소조사면에서 출력비율의 선량검증 (Dosimetric Verifications of the Output Factors in the Small Field Less Than $3cm^2$ Using the Gafchromic EBT2 Films and the Various Detectors)

  • 오세안;예지원;이레나;박헌보;김성규
    • 한국의학물리학회지:의학물리
    • /
    • 제25권4호
    • /
    • pp.218-224
    • /
    • 2014
  • 소조사면의 선량검증은 고선량을 1회에 치료하는 정위적방사선수술(Stereotactic radiosurgery, SRS)과 고선량을 소분할 하여 치료하는 소분할방사선치료(hypo-fractionated radiotherapy)에서 작은 크기의 종양을 치료하기 위해서 자주 사용되기 때문에 현대의 방사선치료에서 있어서 매우 중요하다. 그러나, $3cm^2$ 이하의 소조사면에 대한 선량검증은 방사선치료에서 있어서 대단한 도전이다. 소조사면의 선량검증은 (a) 측방전자균형(lateral electronic equilibrium)의 부족, (b) 급격한 선량 기울기(steep dose gradient), (c) 선원의 부분적 차폐 때문에 어렵다. 이 연구의 목적은 6 MV 광자선의 $3cm^2$ 이하의 소조사면에서 출력비율을 다양한 검출기로 측정하고 검증하는 것이다. 출력비율은 CC13 이온함, CC01 이온함, EDGE 검출기, 열발광선량계(thermoluminescence dosimeters, TLD), Gafchromic EBT2 필름을 이용하여 $0.5{\times}0.5cm^2$, $1{\times}1cm^2$, $2{\times}2cm^2$, $3{\times}3cm^2$, $5{\times}5cm^2$, $10{\times}10cm^2$의 다양한 조사면에서 측정하였다. 출력비율의 차이는 조사면의 크기가 작아질수록 검출기간의 차이는 증가하였다. 본 연구의 결과는 $3cm^2$ 이하의 소조사면의 선량측정은 CC01 이온함, EDGE 검출기와 같은 작은 크기의 방사부부피(active volume)를 가지는 검출기를 사용해야 한다는 것을 입증하였다. 또한, $3cm^2$ 이하의 소조사면에서 EDGE 검출기의 출력비율은 Gafchromic EBT2 필름의 결과와 잘 일치하였다.

고에너지 광자선을 이용한 방사선 치료 시 소조사면에서의 흡수선량평가에 관한 연구 (Study on the Small Fields Dosimetry for High Energy Photon-based Radiation Therapy)

  • 정해선;한영이;금오연;김찬형
    • 한국의학물리학회지:의학물리
    • /
    • 제20권4호
    • /
    • pp.290-297
    • /
    • 2009
  • 고에너지 광자선 기반의 소조사면을 이용한 방사선 치료 시, 조사면의 가장자리에서의 급격한 선량 변화, 전자의 비평형상태, 검출기의 체적 효과 및 검출기와 팬텀 물질과의 불균질성 등으로 인하여 정확한 선량 측정이 어렵다. 따라서 본 연구에서는 선량 측정을 위해 널리 사용되는 전리함, 다이오드 검출기 및 물과 등가인 재질로 이루어져 측정 시 오차 유발 요인이 적은 것으로 알려진 $GAFCHROMIC^{(R)}$ EBT 필름을 이용하여 팬텀 내 소조사면에서의 흡수선량을 측정하고, 각 검출기들의 특성 및 EBT 필름의 유용성을 평가하였다. 각 검출기는 팬텀 표면으로부터 10 cm 깊이에 장착, 선원과의 거리(SAD)를 100 cm로 하였으며, 6 MV X-선 빔을 6개 조사면($5{\times}5\;cm^2$, $2{\times}2\;cm^2$, $1.5{\times}1.5\;cm^2$, $1{\times}1\;cm^2$, $0.7{\times}0.7\;cm^2$$0.5{\times}0.5\;cm^2$)으로 팬텀에 조사하였다. $5{\times}5\;cm^2{\sim}1.5{\times}1.5\;cm^2$ 조사면의 경우, 모든 검출기들의 선량값이 1% 이내로 정확하게 일치하였으나, $1{\times}1\;cm^2$ 이하 조사면에서는 전리함을 이용한 측정결과가 타 검출기들에 비해 선량값을 매우 낮게 평가하는 것으로 확인되었다. 이는 검출기 체적효과가 매우 큰 오차요인으로 작용한 것으로 예측되어, 이를 제거하기 위해 제적 효과를 보정하는 컨볼루션 이론을 적용하여 측정된 선량값을 보정하였다. 그 결과, 다이오드 검출기의 경우 $1{\times}1\;cm^2$의 조사면에서는 EBT 필름의 흡수선량보다 약 3%가 높게, 전리함은 약 1% 낮게 측정되었다. $0.5{\times}0.5\;cm^2$ 조사면에서 다이오드 검출기는 약 1% 높은 값을, 전리함은 7% 낮은 선량값을 나타내었다. 결론적으로 $GAFCHROMIC^{(R)}$ EBT 필름의 소조사면 선량측정기로서의 유용성을 확인하였으며, 몬테카를로 전산모사를 이용한 추가 검증이 수행될 예정이다.

  • PDF

선속 폭(Field Width) 변화에 따른 종축선량 분석 (Analysis on Longitudinal Dose according to Change of Field Width)

  • 정원석;백종걸;신령미;오병천;조준영;김기철;최태규
    • 대한방사선치료학회지
    • /
    • 제23권2호
    • /
    • pp.109-117
    • /
    • 2011
  • 목 적: 호흡에 의하여 영향을 받는 장기의 토모 치료 시 선속 폭(Field Width) 변화에 따른 종양 용적선량의 선량 전달 정확성을 분석하고 자체 제작한 Moving car를 사용하여 선량 변화의 차이를 확인하고 실질적인 종양 선량의 전달 선량을 평가해 보고자 한다. 대상 및 방법: 자체 제작한 Moving car를 사용하여 종축 움직임 차이(0.0 cm, 1.0 cm, 1.5 cm, 2.0 cm)를 적용하고 선속 폭(Field Width) 변화(1.05 cm, 2.50 cm, 5.02 cm)에 따른 측정된 선량의 값을 계산된 선량 값과 비교하여 오차범위를 파악하였다. Gafchromic EBT 필름을 이용하여 측정한 DQA (Delivery Quality Assurance) 필름의 감광정도를 선량윤곽(Dose Profile)과 Gamma Histogram을 이용하여 비교 분석하였다. 결 과: 선속 폭(Field Width) 1.05 cm, 2.50 cm, 5.02 cm일 때 오차범위가 각각 -2.00%, -0.39%, -2.55%의 선량 전달율을 나타내었다. Gafchromic EBT 필름 분석의 선량 윤곽(Dose profile)에서는 Moving car의 움직임이 있는 상태에서 종축방향으로의 움직임이 클수록, 선속 폭(Field Width)이 좁아질수록 고선량 부분에서 계산된 선량과의 오차가 크게 나타났다. Gamma histogram에서는 gamma index가 선속 폭(Field Width)이 좁을수록 움직임의 영향이 상당히 증가하는 것을 알 수 있었다. 결 론: 자체 제작한 Moving car를 이용하여 움직이는 장기의 종축선량 차이를 확인할 수 있었으며 토모 치료 시 선속 폭(Field Width)을 작게 하고 호흡에 의한 장기의 움직임을 최소화하기 위하여 호흡조절 장치와 고정기구의 개발이 필요하리라 사료된다.

  • PDF

Feasibility Study of a Custom-made Film for End-to-End Quality Assurance Test of Robotic Intensity Modulated Radiation Therapy System

  • Kim, Juhye;Park, Kwangwoo;Yoon, Jeongmin;Lee, Eungman;Cho, Samju;Ahn, Sohyun;Park, Jeongeun;Choi, Wonhoon;Lee, Ho
    • 한국의학물리학회지:의학물리
    • /
    • 제27권4호
    • /
    • pp.189-195
    • /
    • 2016
  • This paper aims to verify the clinical feasibility of a custom-made film created by a laser cutting tool for End-to-End (E2E) quality assurance in robotic intensity modulated radiation therapy system. The custom-made film was fabricated from the Gafchromic EBT3 film with the size of $8^{{\prime}{\prime}}{\times}10^{{\prime}{\prime}}$ using a drawing that is identical to the shape and scale of the original E2E film. The drawing was created by using a computer aided design program with the image file, which is obtained by scanning original E2E film. Beam delivery and evaluations were respectively performed with the original film and the custom-made film using fixed-cone collimator on three tracking modes: 6D skull (6DS), Xsight spine (XS), and Xsight lung (XL). The differences between total targeting errors of the original and custom-made films were recorded as 0.17 mm, 0.3 mm, and 0.17 mm at 6DS, XS, and XL tracking modes, respectively. This indicates that the custom-made film could yield nearly equivalent results to those of the original E2E film, given the uncertainties caused by distortions during film scanning and vibrations associated with film cutting. By confirming the clinical feasibility of a custom-made film for E2E testing, it can be expected that economic efficiency of the testing will increase accordingly.